What computer configurations can I use with the ViiA™ 7 Fast Real-Time PCR System?

Answer

For computer and system requirements, please refer to the Release Notes for the ViiA™ 7 Software, which can be found on this page (http://www.thermofisher.com/us/en/home/technical-resources/software-downloads/applied-biosystems-viia-7-real-time-pcr-system.html )(Read Me).

Answer Id: E2328

Was this answer helpful?

Yes
No
Thank you for your response

Product FAQ

How do I set up a custom dye calibration in the ViiA™ 7 or QuantStudio™ 12K Flex systems?

Answer

Follow the directions on page 10 of this User Bulletin to add a custom dye to your ViiA™ 7 or QuantStudio™ 12K Flex system.

Answer Id: E6317

Was this answer helpful?

Yes
No
Thank you for your response

Product FAQ

What does “Error - Fatal Error Number 0x8158” mean?

Answer

This is generally caused by incorrect placement of tube strips, resulting in a sample block that can’t properly be lowered. We recommend contacting Instrument Services to have an engineer walk you through a manual drawer-lowering process during instrument power-up. To prevent this from happening in the future, place tubes toward the middle of the block or use a placeholder strip on the opposing side to counterbalance block pressure.

Answer Id: E7216

Was this answer helpful?

Yes
No
Thank you for your response

Product FAQ

When do I need to calibrate the ViiA™ 7 Real-Time PCR System?

Answer

The recommended maintenance schedule and calibration instructions can be found in the User Guide here: http://tools.thermofisher.com/content/sfs/manuals/4442661C.pdf
In brief, the background calibration is recommended to be done every month. All other calibrations (pure dye, ROI, uniformity, and normalization) are recommended to be done every six months.

Answer Id: E7183

Was this answer helpful?

Yes
No
Thank you for your response

Product FAQ

What qPCR instruments are compatible with Applied Biosystems™ SymphoniqPCR™ Analysis Software?

Answer

Applied Biosystems™ QuantStudio™ 12K Flex, Applied Biosystems™ QuantStudio™ 6 and 7, Applied Biosystems™ ViiA™ 7, 7500 Fast (eds files only), 7500, 7900 HT, and Applied Biosystems™ StepOne™/ StepOnePlus ™ real-time PCR systems are compatible with SymphoniqPCR Analysis Software.

Answer Id: E10235

Was this answer helpful?

Yes
No
Thank you for your response

Product FAQ

What is the normalization calibration for in the ViiA™ 7 or QuantStudio™ 12K Flex systems?

Answer

The normalization calibration generates factors that the software uses when comparing data from multiple instruments.

Answer Id: E6318

Was this answer helpful?

Yes
No
Thank you for your response

Product FAQ

What does “Error - Cannot calculate Pure Dye Matrix” mean?

Answer

This issue can occur if the run is interrupted. To prevent this from happening in the future, avoid ending the PCR with an infinite hold stage or interrupting the run. Please send the file to techsupport@thermofisher.com so that our software group can attempt to recover the data.

Answer Id: E7217

Was this answer helpful?

Yes
No
Thank you for your response

Product FAQ

What are the uniformity and normalization calibrations for on the ViiA™ 7 Real-Time PCR System?

Answer

The uniformity calibration generates data that allow the software to compensate for the physical effects of the ViiA™ 7 system filters using the ROI plate. The normalization calibration generates factors that the software uses when comparing data from multiple ViiA™ 7 instruments within a study. Two normalization plates are provided in the calibration kit, one containing FAM™/ROX™ dyes and the other with VIC™/ROX™ dyes.

Answer Id: E7184

Was this answer helpful?

Yes
No
Thank you for your response

Product FAQ

What should I do if I need to transfer my ViiA™ 7 system software license to another computer?

Answer

If you need to move your ViiA™ 7 system software to another computer, please contact instrumentservices@lifetech.com with your instrument serial number.

Answer Id: E6319

Was this answer helpful?

Yes
No
Thank you for your response

Product FAQ

What should I do if the software will not connect to the ViiA™ 7 Real-Time PCR System?

Answer

After launching the ViiA™ 7 system software, go to the Maintenance Manager. Check if the ViiA™ 7 system icon is present in the “On the Network” window. If not, click on the “Refresh” button to the left of “Remove from My Instruments”.

If you still do not see the ViiA™ 7 icon for the following reasons:
- There is an icon but you cannot add it into the “My Instruments” area
- There is a red “X” on the instrument
- A message says “Disconnected” at the bottom of the window with a red “X” in it then you will need to add the instrument to the “My Instruments” area. Click on the wrench icon in the upper right-hand corner of the icon as pictured below. You may get a message that says that the firmware is older than the one you are trying to run. If appropriate, choose to download the new firmware and let it progress to finish. Do not turn off the instrument during this update. After the firmware update is completed, restart the instrument to check the ViiA™ 7 system icon in ‘On the Network’ area.

Answer Id: E7194

Was this answer helpful?

Yes
No
Thank you for your response

Product FAQ

How do I renew my software license?

Answer

User Bulletin v1.2.1 has specific directions for how to renew your software license. Please search the term '4474347' on our website. The renewal directions apply to the ViiA™ 7, QuantStudio™ 6, QuantStudio™ 7, and QuantStudio™ 12K Flex systems.

Answer Id: E6627

Was this answer helpful?

Yes
No
Thank you for your response

Product FAQ

What real-time instruments can be run independently from a computer?

Answer

The StepOne™, StepOnePlus™, ViiA™ 7, and QuantStudio™ 12K Flex systems can all be run directly from the instrument touchscreen and don't need a separate computer to operate them.

Answer Id: E6303

Was this answer helpful?

Yes
No
Thank you for your response

Product FAQ

What does “Error - Cannot open the file because file contents are not recognized” mean?

Answer

This issue can occur if duplicate samples are added to the experiment, if illegal characters are used in the experiment/file name, or if the file extension was changed to .eds from a different extension.
Please send the file to techsupport@thermofisher.com to attempt recovery.

Answer Id: E7214

Was this answer helpful?

Yes
No
Thank you for your response

Product FAQ

Can I reuse my calibration plates?

Answer

The calibration plates can be stored and reused three times for up to 1 year after you first open them, so make sure to return them to their original packaging and return them to -20 degrees C storage until the next use. (Note: Due to the small volumes, calibration array cards cannot be reused. Please discard them after use.) If needed, you can make your own background plate using deionized water. Please follow the directions in the ViiA™ 7 Real-Time PCR System Installation, Networking, and Maintenance Guide for more details at http://tools.thermofisher.com/content/sfs/manuals/4442661C.pdf.

Answer Id: E7185

Was this answer helpful?

Yes
No
Thank you for your response

Product FAQ

Where are my files saved to on the StepOne™ and StepOnePlus™ Real-Time PCR Systems?

Answer

Run files will be saved to both the instrument and the connected computer. On the connected computer, files will be saved to the default data folder, unless you change it.

To find or change the default folder, go to Tools then Preferences then Defaults. Here you will see a Data Folder and an Import Folder. The default location is shown. If you want files to be saved to (or open from) a different location, click ‘Browse’ and choose the new folder.

Answer Id: E7208

Was this answer helpful?

Yes
No
Thank you for your response