Nano Flow LC–MS Could Expand What’s Possible in Discovery Lipidomics

New research demonstrates how nano flow chromatography (nano flow LC) coupled to high resolution accurate-mass (HRAM) mass spectrometer can increase lipid coverage, improve sensitivity, reduce unwanted fragmentation, and lower sample requirements in untargeted lipidomics.

Illustration of a lipid bilayer membrane with highlighted lipid molecules interacting at its center.

Untargeted lipidomics depends on a difficult analytical balance: researchers want to detect as much of the lipidome as possible, including low-abundance lipids, without consuming large amounts of precious biological sample. A recent peer reviewed publication by Thermo Fisher Scientific scientists examines whether nano flow liquid chromatography–mass spectrometry (nano flow LC–MS) can help address that challenge. The study, “Implementation of Nano Flow Chromatography Coupled to Mass Spectrometry as a Reliable and Sensitive Discovery Lipidomics Platform,” compares a nano flow LC–MS workflow with an established high flow workflow using HRAM Thermo Scientific Orbitrap Excedion Hybrid Mass Spectrometer. The results suggest that nano flow LC–MS can provide meaningful advantages for sample-limited and sensitivity-driven discovery lipidomics.

Why nano flow LC–MS for lipidomics?

Conventional analytical-flow LC–MS is widely used in lipidomics because the workflows are established, robust, and compatible with many column chemistries. But detecting low-abundance lipids can remain challenging, particularly when researchers have limited sample available. Nano flow chromatography operates at substantially lower flow rates. In this study, the nano flow method operated at 700 nL/min, compared with 260 µL/min for the high flow method. This reduction in flow rate also impacts the total amount of solvent used in experiments. This provides key cost savings in addition to the improvements in analytical performance.  The nano flow workflow used a Thermo Scientific EASY-Spray PepMap Neo UHPLC Column (75 µm × 15 cm C18 column) and a 0.1 µL injection volume on a Thermo Scientific Vanquish Neo UHPLC System. Lower flow rates can improve electrospray ionization efficiency where smaller droplets can increase charge density and promote faster desolvation, while fewer analyte molecules per droplet can reduce ion competition and ion suppression. These effects can translate into improved analytical sensitivity.

More lipid annotations with less sample

The most striking result was the difference in lipid annotation depth. At 25 ng on-column, nano flow LC–MS generated 1,266 total lipid annotations and 835 high-quality annotations from Thermo Scientific LipidSearch Software processing. By comparison, high flow LC-MS at 100 ng on-column generated 919 total annotations and 518 high-quality annotations. Even at only 5 ng on-column, nano flow produced 579 high-quality annotations, approximately 12% more than the high flow workflow while using 20-fold less material. At 25 ng, nano flow generated approximately 61% more high-quality annotations while using fourfold less material. The annotation gains were also broadly distributed. At 25 ng, nano flow produced more annotations in 19 of the 23 lipid classes evaluated, with notable differences in phosphatidylcholine, fatty acids, phosphatidylethanolamine, diacylglycerol, and sphingomyelin.

Detecting lipids at lower on-column loads

Nano flow also extended analytical response toward substantially lower sample loads. Using a SPLASH Lipidomix dilution series, researchers observed lower ends of the selected response range extending by 2-fold for sphingomyelin, 10-fold for phosphatidylcholine, and 40-fold for diacylglycerol compared with high flow LC–MS. For researchers working with limited biological samples or low-abundance lipids, that improvement could be particularly valuable.

Reduced unintentional lipid fragmentation

Sensitivity was not the only benefit with nano flow LC-MS. Unintentional fragmentation can reduce intact lipid signal and potentially complicate lipid annotation. Across six representative lipid species examined in the study, nano flow produced consistently less fragmentation than high flow. Reductions ranged from 21% for the phosphatidylinositol species to 24%–31% for cholesterol esters and 23%–30% for phosphatidylglycerols.

Can nano flow lipidomics be reliable for routine analysis?

Sensitivity alone is insufficient if a workflow cannot remain stable across a large analytical sequence. To examine robustness, the researchers analyzed pooled mouse plasma QC samples across approximately 100 injections. Representative lipid standards maintained consistent retention times and peak-area responses, mass errors remained within ±5 ppm, and no emitter failure or clogging was observed.

Importantly, the study goes beyond performance metrics and addresses the practical details that matter for routine implementation: sample preparation, injection volume, wash procedures, carryover control, and column re-equilibration. A strong wash and an alternating low/high-organic “zebra wash” were incorporated to address lipid buildup and carryover.

What does this mean for discovery lipidomics?

The study provides evidence that nano flow LC–MS can serve as a sensitive, complementary option for untargeted lipidomics, particularly when sample quantity or analyte abundance is limited. Importantly, the authors caution against attributing every observed difference solely to flow rate. The high flow workflow used a C30 stationary phase while the nano flow method used C18, meaning differences in chromatographic selectivity, gradients, loading, and source configuration may also contribute to the results.

Taken together, the complete nano flow workflow evaluated here delivered a compelling combination: deeper lipid annotation, lower sample consumption, reduced unintentional fragmentation, response at lower on-column loads, and reduced solvent consumption. For laboratories pursuing deeper lipidome coverage from increasingly precious samples, nano flow LC–MS may therefore represent an important direction for the next generation of discovery lipidomics.

Visit us on LinkedIn: #lipidomics #nanoLC

Dauly-Claire-headshot

Written by:

Claire Dauly

Product Manager, Thermo Fisher Scientific

Claire Dauly is a Product Manager at Thermo Fisher Scientific, responsible for Orbitrap Excedion Mass Spectrometers and Orbitrap Excedion Pro Mass Spectrometers. Combining expertise in analytical chemistry, mass spectrometry and commercialization, she helps advance innovative technologies for researchers worldwide.

Read more Dauly, Claire