Stellar MS Raises the Standard for DBS Steroid Ester Detection

Dried blood spot analysis has promised anti-doping laboratories a faster, simpler, and less invasive way to detect steroid ester use. The problem has never been the promise. It has been the analytical reality.

Close-up of runners racing on a track, emphasizing speed and high performance.

DBS samples are small. Whole blood is complex. Steroid esters can be present at trace levels. And when the signal gets that low, MS2 workflows can run out of specificity before the science runs out of need.

A recent publication from Andreas Thomas, Jasmin Thelen, Panagiotis Sakellariou, and Mario Thevis in Journal of Mass Spectrometry* demonstrates a path forward: native steroid ester detection in DBS by LC–MS3 using the Stellar mass spectrometer.

The method demonstrates sensitive, selective detection of 17 underivatized steroid esters from DBS without derivatization. More importantly, it shows how Stellar MS enables high-confidence trace-level results by generating multiple diagnostic MS³ product ions. Scientists can monitor three to five product ions, supporting clean identification and strong product-ion ratio assessment with great confidence at the lowest concentration levels.

MS2 alone is not enough

Steroid ester analysis in DBS is not a routine MS/MS problem. It is a low-abundance, high-background, high-confidence identification challenge.

At sub-ng/mL concentrations, conventional MS2-based workflows can struggle to deliver the specificity needed in complex DBS matrices. HRAM MS platforms provide accurate mass and resolution, but accurate mass alone does not automatically deliver highly diagnostic, sequential dissociation data at the required sensitivity. Triple quadrupole MS systems remain powerful for targeted quantitation, but they are not designed to deliver ion trap-based MS3 workflows that can generate additional diagnostic product ions from already-isolated fragments.

That is where Stellar MS changes the conversation.

In the published DBS steroid ester method, precursor ions were first activated by CID, selected product ions were isolated, and additional dissociation was used to generate cleaner, more diagnostic signals. The result was sensitive, selective detection of native steroid esters from DBS without derivatization.

The Stellar MS difference

Stellar MS is built for targeted MSn performance. Its hybrid quadrupole–dual-pressure linear ion trap architecture enables practical LC–MS3 workflows that are difficult to reproduce on conventional targeted platforms.

The published method used CID at the MS2 stage followed by HCD at the MS3 stage. This CID→HCD strategy helped cut through DBS matrix complexity while maintaining sensitivity at levels that matter for real-world doping control.

Case study: A decade of DBS steroid innovation in the German Sports University Cologne adds an important workflow point: ion trap CID at the MS2 stage allowed the same CID setting to be used across steroid esters, while the MS2 product ion selected for MS3 was conserved by base steroid class. That makes method expansion more straightforward as additional anabolic steroid esters are added.

That is a major advantage.

Other platforms may deliver accurate mass. Triple quadrupoles may deliver robust MRM. Some systems may offer MSn capability. But the differentiator is not MS3 capability alone. It is the ability to improve lower limits of quantitation while matching the limit of identification at the lowest relevant concentrations.

In DBS workflows, that distinction matters. When sensitivity outpaces identification confidence, laboratories may detect a low-level signal but still need a follow-up experiment from another DBS spot to satisfy identification criteria. Stellar MS helps close that gap by combining MS3 selectivity, high acquisition speed, and multiple diagnostic product ions. In this workflow, Stellar MS enables more than 10 data points across an approximately 3-second-wide chromatographic peak at full-width half-maximum (FWHM), supporting reproducible peak integration and confident product-ion ratio assessment at trace levels.

Native analysis, leaner workflow

Previous DBS steroid ester methods often relied on derivatization to improve ionization efficiency and sensitivity. That adds complexity, time, and workflow burden.

The LC–MS3 method eliminates that bottleneck and improves scalability.

By analyzing native, underivatized steroid esters, the workflow simplifies sample preparation and preserves the possibility of using the same extract for additional prohibited substances. For anti-doping laboratories, that is more than an efficiency gain. It is a more scalable way to extract greater analytical value from limited DBS sample volume.

The data that matters

The publication reports analysis of 17 steroid esters composed of testosterone, nandrolone, and boldenone species. Most analytes extracted from a single DBS spot had detection limits below 0.1 ng/mL. Recovery rates were reported at 40%–80%, and the method used four stable isotope-labeled internal standards.

The method was also practical across DBS formats, working with both cellulose-based DMPK cards and polymer-based TASSO devices.

Selectivity was demonstrated across blank samples from healthy volunteers, with no interfering signals at the relevant retention times. At the proposed 1 ng/mL level, all target steroid esters were detected in both TASSO and DMPK card samples.

That is the type of performance DBS workflows require: low-level detection, low-level identification confidence, and practical robustness across sampling devices.

Beyond steroid esters

The immediate application is direct detection of intact steroid esters in DBS for sports drug testing. But the broader message is bigger.

Scientists working in targeted quantitation and trace-level confirmation are often forced to choose between sensitivity, specificity, throughput, and workflow simplicity. Stellar MS changes that equation by making MS3 practical for small-molecule workflows where MS2 is not selective enough and HRAM MS alone does not solve the problem.

The German Sports University at Cologne case study also points to future opportunities beyond anabolic steroid esters, including broader small-molecule screening, targeted confirmation workflows, isomer differentiation, and analytes where additional diagnostic product ions can simplify fast, reliable testing.

The bottom line

DBS steroid ester analysis has spent more than a decade fighting the same obstacles: tiny sample volumes, complex matrices, low concentrations, and insufficient specificity from conventional approaches.

Stellar MS gives laboratories a practical way forward.

By combining native LC–MS3 analysis, ion trap CID, high acquisition speed, and multiple diagnostic product ions, Stellar MS helps improve lower limits of quantitation while aligning them with limits of identification. That means laboratories can push lower, identify with great confidence, and reduce the risk that a low-level result becomes only a suspicious finding requiring another DBS spot.

For laboratories still trying to solve tomorrow’s DBS challenges with yesterday’s MS2 assumptions, the message is direct:

Raise the standard.

Explore the Thermo Scientific Stellar mass spectrometer at Stellar Mass Spectrometer.

References

*Thomas, A.; Thelen, J.; Sakellariou, P.; Thevis, M. Simplified Analysis of Native Steroid Esters in Dried Blood Spots by LC–MS³. Journal of Mass Spectrometry 2025, 60, e5188.

This article was authored by Mario Thevis as a guest contributor. Views expressed are their own.