{"id":6639,"date":"2020-03-27T00:00:00","date_gmt":"2020-03-27T00:00:00","guid":{"rendered":"https:\/\/www.thermofisher.com\/blog\/analyteguru\/what-do-you-know-about-high-temperature-static-headspace\/"},"modified":"2024-02-14T22:19:08","modified_gmt":"2024-02-14T22:19:08","slug":"what-do-you-know-about-high-temperature-static-headspace","status":"publish","type":"post","link":"https:\/\/www.thermofisher.com\/blog\/analyteguru\/what-do-you-know-about-high-temperature-static-headspace\/","title":{"rendered":"What Do You Know About High-Temperature Static Headspace?"},"content":{"rendered":"<div class=\"lia-message-body-content\">\n<p class=\"alignnone\" style=\"text-align: left\">The characterization of the volatile fraction of a liquid or solid sample is definitely a field of gas chromatography coupled to headspace sampling technique. Static or dynamic or through SPME enrichment, the headspace sampling offers a convenient solvent-less extraction approach, requiring almost no sample preparation, for the analysis of volatile compounds in a sample. Particularly suitable to be fully automated, the headspace sampling is very popular in testing laboratories doing routine volatiles determination.<\/p>\n<p><span class=\"lia-inline-image-display-wrapper lia-image-align-inline\"><img loading=\"lazy\" decoding=\"async\" class=\" wp-image-11908 alignright\" src=\"http:\/\/admin.acceleratingscience.com\/analyteguru\/wp-content\/uploads\/sites\/25\/2023\/06\/3673-og-1-300x246.jpg\" alt=\"\" width=\"360\" height=\"295\" srcset=\"https:\/\/admin.acceleratingscience.com\/analyteguru\/wp-content\/uploads\/sites\/25\/2023\/06\/3673-og-1-300x246.jpg 300w, https:\/\/admin.acceleratingscience.com\/analyteguru\/wp-content\/uploads\/sites\/25\/2023\/06\/3673-og-1-768x629.jpg 768w, https:\/\/admin.acceleratingscience.com\/analyteguru\/wp-content\/uploads\/sites\/25\/2023\/06\/3673-og-1.jpg 999w\" sizes=\"auto, (max-width: 360px) 100vw, 360px\" \/><\/span>The basic principle underlying all instrumentation for headspace-gas chromatography is that an aliquot of volatile compounds from the vapor phase above a liquid or solid sample in a sealed vial must be reproducibly and effectively transferred to the inlet of a <a href=\"https:\/\/www.thermofisher.com\/us\/en\/home\/industrial\/chromatography\/gas-chromatography-gc.html\" target=\"_blank\" rel=\"nofollow noopener\">gas chromatograph<\/a>. There are several means for accomplishing this, including gas-tight syringe, pressure-balanced time-base, valve-and-loop and collection on a\u00a0<a href=\"https:\/\/www.sciencedirect.com\/topics\/chemistry\/sorbent\" target=\"_blank\" rel=\"noopener nofollow noreferrer\">sorbent<\/a>. The most important challenges are to ensure that the sample composition that reaches the gas chromatograph is truly representative of the composition of the headspace vapor in the vial and that the headspace vapor is representative of the composition of the original sample<sup>1<\/sup>.<\/p>\n<p>Short and fully inert sample-path, precise and accurate pressure control during transfer, and efficient heating of the entire sample path are the key points to assure high data quality, robustness and system reliability. The valve and loop-based <a href=\"https:\/\/assets.thermofisher.com\/TFS-Assets\/CMD\/brochures\/br-10690-triplus-500-gc-headspace-autosampler-br10690-en.pdf\" target=\"_blank\" rel=\"nofollow noopener\">Thermo Scientific\u2122 TriPlus\u2122 500 Headspace autosampler<\/a>, with its innovative direct column connection design, offers all those benefits, including a superior recovery over a wide volatility range, compared to the standard static headspace autosamplers on the market.<\/p>\n<p>Capable to operate up to 300\u00b0C with extended recovery of &gt; 60% for C40, the TriPlus 500 HS autosampler can be successfully used to explore high-temperature applications.<\/p>\n<p>This is exactly what Dr. Matthew Noestheden from <a href=\"http:\/\/suprarnd.ca\/\" target=\"_blank\" rel=\"noopener nofollow noreferrer\">Supra RnD<\/a> realized when using on a daily basis his TriPlus 500 HS coupled to the <a href=\"https:\/\/www.thermofisher.com\/order\/catalog\/product\/ISQ7K-VPI#\/ISQ7K-VPI\" target=\"_blank\" rel=\"nofollow noopener\">ISQ\u2122 7000<\/a> GCMS. Matt\u2019s idea was to use the headspace autosampler as a surrogate vaporization environment to study the decomposition processes in e-cigarettes, working at the temperatures used in vaporizers, in order to characterize possible health risks associated to a given ingredient. <a href=\"https:\/\/www.cdc.gov\/mmwr\/volumes\/68\/wr\/mm6845e2.htm?s_cid=mm6845e2_w\" target=\"_blank\" rel=\"noopener nofollow noreferrer\">The recent set of severe illnesses and 47<\/a> deaths appeared as a cluster after a few years of widespread use of vaporizers. Many of these adverse health effects have been associated with vitamin E acetate (\u03b1-tocopheryl acetate), a product that is not toxic by ingestion and has never been identified as a problematic ingredient<sup>2<\/sup>. Read more on Matt\u2019s study on <a href=\"https:\/\/assets.thermofisher.com\/TFS-Assets\/CMD\/Application-Notes\/can-73327-gc-ms-thermal-decomposition-vitamin-e-acetate-can73327-en.pdf\" target=\"_blank\" rel=\"nofollow noopener\">Vitamin E degradation by high-temperature HS-GC-MS.<\/a><\/p>\n<p>Extending the applicability of the static headspace sampling to high temperatures operation opens up new testing capabilities covering less-volatile compounds. We have seen that the TriPlus 500 HS working at high temperature has been used to investigate the possible use of static headspace for the detection of volatile and semivolatile <a href=\"https:\/\/event.on24.com\/wcc\/r\/2217560\/AA71615A5A0DC0600428153DBF4639C9\" target=\"_blank\" rel=\"noopener nofollow noreferrer\">residual organics from oil contamination in aluminum foils<\/a> during manufacturing, a test required to ensure the delivery of high-quality and safe products. Several industries make use of these materials including the automotive, HVAC, transformer and food packaging industries.<\/p>\n<p>For this application the TriPlus 500 headspace autosampler was used at temperatures between 150-300\u00b0C, permitting the analysis of less volatile compounds. Important to say that working at a high temperature requires also the use of suitable high-temperature consumables. In particular, the caps and septa for the headspace vials need to be suitable for high-temperature operation, assuring the correct tightness and less bleeding, like the 20 mm magnetic tin plate crimp cap, 20 mm red silicone\/natural PTFE high temperature, <a href=\"https:\/\/www.thermofisher.com\/order\/catalog\/product\/60180-504#\/60180-504\" target=\"_blank\" rel=\"nofollow noopener\">P\/N 20-MCB-ST3HT<\/a> (highly recommended if going up to 300\u00b0C, to be used with an <a href=\"https:\/\/www.thermofisher.com\/order\/catalog\/product\/60180-ECRH11KIT#\/60180-ECRH11KIT\" target=\"_blank\" rel=\"nofollow noopener\">automatic crimper<\/a>).<\/p>\n<p>The HT operation along with the direct column connection design of the TriPlus 500 HS, allowed to recover &gt; 800 different hydrocarbons within a lubricant oil sample, including several fatty acid esters. The whole analysis could be carried out in automated, completely solvent-free fashion, making it a simple and fast testing approach that can be easily implemented in a quality control lab to monitor the manufacturing process of aluminum foils.<\/p>\n<p>Now you know a bit more about the high-temperature headspace capability of the TriPlus 500 HS, but should you still have additional questions, discover more features and benefits on <a href=\"http:\/\/www.thermoscientific.com\/triplus500\" target=\"_blank\" rel=\"noopener nofollow noreferrer\">www.thermoscientific.com\/triplus500<\/a><\/p>\n<ol>\n<li><a href=\"https:\/\/www.sciencedirect.com\/science\/article\/pii\/B9780123855404000092\" target=\"_blank\" rel=\"noopener nofollow noreferrer\">Headspace-Gas Chromatography<\/a>, Michael J.Sithersingh, Nicholas H.Snow, in <a href=\"https:\/\/www.sciencedirect.com\/book\/9780123855404\" target=\"_blank\" rel=\"noopener nofollow noreferrer\">Gas Chromatography<\/a>, 2012<\/li>\n<li>Blundell, M. S.; Dargan, P. I.; Wood, D. M. The dark cloud of recreational drugs and vaping. QJM 2018, 111, 145\u2013148.<\/li>\n<\/ol>\n<\/div>\n","protected":false},"excerpt":{"rendered":"<p>The characterization of the volatile fraction of a liquid or solid sample is definitely a field of gas chromatography coupled to headspace sampling<\/p>\n","protected":false},"author":1327,"featured_media":6641,"comment_status":"open","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_kad_blocks_custom_css":"","_kad_blocks_head_custom_js":"","_kad_blocks_body_custom_js":"","_kad_blocks_footer_custom_js":"","_monsterinsights_skip_tracking":false,"_genesis_hide_title":false,"_genesis_hide_breadcrumbs":false,"_genesis_hide_singular_image":false,"_genesis_hide_footer_widgets":false,"_genesis_custom_body_class":"","_genesis_custom_post_class":"","_genesis_layout":"","_jetpack_newsletter_access":"","_jetpack_dont_email_post_to_subs":false,"_jetpack_newsletter_tier_id":0,"_jetpack_memberships_contains_paywalled_content":false,"_jetpack_memberships_contains_paid_content":false,"footnotes":""},"categories":[57],"tags":[59],"division":[],"class_list":{"0":"post-6639","1":"post","2":"type-post","3":"status-publish","4":"format-standard","5":"has-post-thumbnail","7":"category-gas-chromatography","8":"tag-gas-chromatography","9":"entry"},"_selected_authors":"","_selected_reviewers":"","acf":[],"yoast_head":"<!-- This site is optimized with the Yoast SEO Premium plugin v27.8 (Yoast SEO v27.8) - https:\/\/yoast.com\/product\/yoast-seo-premium-wordpress\/ -->\n<title>What do you know about High-Temperature Static Headspace?<\/title>\n<meta name=\"description\" content=\"The characterization of the volatile fraction of a liquid or solid sample is a field of gas chromatography coupled to headspace sampling.\" \/>\n<meta name=\"robots\" content=\"index, follow, max-snippet:-1, max-image-preview:large, max-video-preview:-1\" \/>\n<link rel=\"canonical\" href=\"https:\/\/www.thermofisher.com\/blog\/analyteguru\/what-do-you-know-about-high-temperature-static-headspace\/\" \/>\n<meta property=\"og:locale\" content=\"en_US\" \/>\n<meta property=\"og:type\" content=\"article\" \/>\n<meta property=\"og:title\" content=\"What Do You Know About High-Temperature Static Headspace?\" \/>\n<meta property=\"og:description\" content=\"The characterization of the volatile fraction of a liquid or solid sample is a field of gas chromatography coupled to headspace sampling.\" \/>\n<meta property=\"og:url\" content=\"https:\/\/www.thermofisher.com\/blog\/analyteguru\/what-do-you-know-about-high-temperature-static-headspace\/\" \/>\n<meta property=\"og:site_name\" content=\"AnalyteGuru\" \/>\n<meta property=\"article:publisher\" content=\"https:\/\/www.facebook.com\/ChromatographySolutions\" \/>\n<meta property=\"article:published_time\" content=\"2020-03-27T00:00:00+00:00\" \/>\n<meta property=\"article:modified_time\" content=\"2024-02-14T22:19:08+00:00\" \/>\n<meta property=\"og:image\" content=\"https:\/\/admin.acceleratingscience.com\/analyteguru\/wp-content\/uploads\/sites\/25\/2023\/06\/3673-og.jpg\" \/>\n\t<meta property=\"og:image:width\" content=\"999\" \/>\n\t<meta property=\"og:image:height\" content=\"818\" \/>\n\t<meta property=\"og:image:type\" content=\"image\/jpeg\" \/>\n<meta name=\"author\" content=\"AnalyteGuru Staff\" \/>\n<meta name=\"twitter:card\" content=\"summary_large_image\" \/>\n<meta name=\"twitter:creator\" content=\"@ChromSolutions\" \/>\n<meta name=\"twitter:site\" content=\"@ChromSolutions\" \/>\n<meta name=\"twitter:label1\" content=\"Written by\" \/>\n\t<meta name=\"twitter:data1\" content=\"AnalyteGuru Staff\" \/>\n\t<meta name=\"twitter:label2\" content=\"Est. reading time\" \/>\n\t<meta name=\"twitter:data2\" content=\"4 minutes\" \/>\n<script type=\"application\/ld+json\" class=\"yoast-schema-graph\">{\"@context\":\"https:\\\/\\\/schema.org\",\"@graph\":[{\"@type\":\"Article\",\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/analyteguru\\\/what-do-you-know-about-high-temperature-static-headspace\\\/#article\",\"isPartOf\":{\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/analyteguru\\\/what-do-you-know-about-high-temperature-static-headspace\\\/\"},\"author\":{\"name\":\"AnalyteGuru Staff\",\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/analyteguru\\\/#\\\/schema\\\/person\\\/47226155467d66d692f24162ee320270\"},\"headline\":\"What Do You Know About High-Temperature Static Headspace?\",\"datePublished\":\"2020-03-27T00:00:00+00:00\",\"dateModified\":\"2024-02-14T22:19:08+00:00\",\"mainEntityOfPage\":{\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/analyteguru\\\/what-do-you-know-about-high-temperature-static-headspace\\\/\"},\"wordCount\":724,\"commentCount\":0,\"image\":{\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/analyteguru\\\/what-do-you-know-about-high-temperature-static-headspace\\\/#primaryimage\"},\"thumbnailUrl\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/analyteguru\\\/wp-content\\\/uploads\\\/sites\\\/25\\\/2023\\\/06\\\/3673-og.jpg\",\"keywords\":[\"Gas Chromatography\"],\"articleSection\":[\"Gas Chromatography\"],\"inLanguage\":\"en-US\",\"potentialAction\":[{\"@type\":\"CommentAction\",\"name\":\"Comment\",\"target\":[\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/analyteguru\\\/what-do-you-know-about-high-temperature-static-headspace\\\/#respond\"]}]},{\"@type\":\"WebPage\",\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/analyteguru\\\/what-do-you-know-about-high-temperature-static-headspace\\\/\",\"url\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/analyteguru\\\/what-do-you-know-about-high-temperature-static-headspace\\\/\",\"name\":\"What do you know about High-Temperature Static Headspace?\",\"isPartOf\":{\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/analyteguru\\\/#website\"},\"primaryImageOfPage\":{\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/analyteguru\\\/what-do-you-know-about-high-temperature-static-headspace\\\/#primaryimage\"},\"image\":{\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/analyteguru\\\/what-do-you-know-about-high-temperature-static-headspace\\\/#primaryimage\"},\"thumbnailUrl\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/analyteguru\\\/wp-content\\\/uploads\\\/sites\\\/25\\\/2023\\\/06\\\/3673-og.jpg\",\"datePublished\":\"2020-03-27T00:00:00+00:00\",\"dateModified\":\"2024-02-14T22:19:08+00:00\",\"author\":{\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/analyteguru\\\/#\\\/schema\\\/person\\\/47226155467d66d692f24162ee320270\"},\"description\":\"The characterization of the volatile fraction of a liquid or solid sample is a field of gas chromatography coupled to headspace sampling.\",\"breadcrumb\":{\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/analyteguru\\\/what-do-you-know-about-high-temperature-static-headspace\\\/#breadcrumb\"},\"inLanguage\":\"en-US\",\"potentialAction\":[{\"@type\":\"ReadAction\",\"target\":[\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/analyteguru\\\/what-do-you-know-about-high-temperature-static-headspace\\\/\"]}]},{\"@type\":\"ImageObject\",\"inLanguage\":\"en-US\",\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/analyteguru\\\/what-do-you-know-about-high-temperature-static-headspace\\\/#primaryimage\",\"url\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/analyteguru\\\/wp-content\\\/uploads\\\/sites\\\/25\\\/2023\\\/06\\\/3673-og.jpg\",\"contentUrl\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/analyteguru\\\/wp-content\\\/uploads\\\/sites\\\/25\\\/2023\\\/06\\\/3673-og.jpg\",\"width\":999,\"height\":818},{\"@type\":\"BreadcrumbList\",\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/analyteguru\\\/what-do-you-know-about-high-temperature-static-headspace\\\/#breadcrumb\",\"itemListElement\":[{\"@type\":\"ListItem\",\"position\":1,\"name\":\"Home\",\"item\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/analyteguru\\\/\"},{\"@type\":\"ListItem\",\"position\":2,\"name\":\"What Do You Know About High-Temperature Static Headspace?\"}]},{\"@type\":\"WebSite\",\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/analyteguru\\\/#website\",\"url\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/analyteguru\\\/\",\"name\":\"AnalyteGuru\",\"description\":\"Your educational resource for biopharma, pharma, environmental, food and agriculture, industrial, and clinical labs\",\"potentialAction\":[{\"@type\":\"SearchAction\",\"target\":{\"@type\":\"EntryPoint\",\"urlTemplate\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/analyteguru\\\/?s={search_term_string}\"},\"query-input\":{\"@type\":\"PropertyValueSpecification\",\"valueRequired\":true,\"valueName\":\"search_term_string\"}}],\"inLanguage\":\"en-US\"},{\"@type\":\"Person\",\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/analyteguru\\\/#\\\/schema\\\/person\\\/47226155467d66d692f24162ee320270\",\"name\":\"AnalyteGuru Staff\",\"image\":{\"@type\":\"ImageObject\",\"inLanguage\":\"en-US\",\"@id\":\"https:\\\/\\\/secure.gravatar.com\\\/avatar\\\/d716c4c68d79e076572f67bcc11df0ef8db404f6c3463e9cfd3d20514a881439?s=96&d=mm&r=g\",\"url\":\"https:\\\/\\\/secure.gravatar.com\\\/avatar\\\/d716c4c68d79e076572f67bcc11df0ef8db404f6c3463e9cfd3d20514a881439?s=96&d=mm&r=g\",\"contentUrl\":\"https:\\\/\\\/secure.gravatar.com\\\/avatar\\\/d716c4c68d79e076572f67bcc11df0ef8db404f6c3463e9cfd3d20514a881439?s=96&d=mm&r=g\",\"caption\":\"AnalyteGuru Staff\"},\"url\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/author\\\/kerrycolligan\\\/\"}]}<\/script>\n<!-- \/ Yoast SEO Premium plugin. -->","yoast_head_json":{"title":"What do you know about High-Temperature Static Headspace?","description":"The characterization of the volatile fraction of a liquid or solid sample is a field of gas chromatography coupled to headspace sampling.","robots":{"index":"index","follow":"follow","max-snippet":"max-snippet:-1","max-image-preview":"max-image-preview:large","max-video-preview":"max-video-preview:-1"},"canonical":"https:\/\/www.thermofisher.com\/blog\/analyteguru\/what-do-you-know-about-high-temperature-static-headspace\/","og_locale":"en_US","og_type":"article","og_title":"What Do You Know About High-Temperature Static Headspace?","og_description":"The characterization of the volatile fraction of a liquid or solid sample is a field of gas chromatography coupled to headspace sampling.","og_url":"https:\/\/www.thermofisher.com\/blog\/analyteguru\/what-do-you-know-about-high-temperature-static-headspace\/","og_site_name":"AnalyteGuru","article_publisher":"https:\/\/www.facebook.com\/ChromatographySolutions","article_published_time":"2020-03-27T00:00:00+00:00","article_modified_time":"2024-02-14T22:19:08+00:00","og_image":[{"width":999,"height":818,"url":"https:\/\/admin.acceleratingscience.com\/analyteguru\/wp-content\/uploads\/sites\/25\/2023\/06\/3673-og.jpg","type":"image\/jpeg"}],"author":"AnalyteGuru Staff","twitter_card":"summary_large_image","twitter_creator":"@ChromSolutions","twitter_site":"@ChromSolutions","twitter_misc":{"Written by":"AnalyteGuru Staff","Est. reading time":"4 minutes"},"schema":{"@context":"https:\/\/schema.org","@graph":[{"@type":"Article","@id":"https:\/\/www.thermofisher.com\/blog\/analyteguru\/what-do-you-know-about-high-temperature-static-headspace\/#article","isPartOf":{"@id":"https:\/\/www.thermofisher.com\/blog\/analyteguru\/what-do-you-know-about-high-temperature-static-headspace\/"},"author":{"name":"AnalyteGuru Staff","@id":"https:\/\/admin.acceleratingscience.com\/analyteguru\/#\/schema\/person\/47226155467d66d692f24162ee320270"},"headline":"What Do You Know About High-Temperature Static Headspace?","datePublished":"2020-03-27T00:00:00+00:00","dateModified":"2024-02-14T22:19:08+00:00","mainEntityOfPage":{"@id":"https:\/\/www.thermofisher.com\/blog\/analyteguru\/what-do-you-know-about-high-temperature-static-headspace\/"},"wordCount":724,"commentCount":0,"image":{"@id":"https:\/\/www.thermofisher.com\/blog\/analyteguru\/what-do-you-know-about-high-temperature-static-headspace\/#primaryimage"},"thumbnailUrl":"https:\/\/admin.acceleratingscience.com\/analyteguru\/wp-content\/uploads\/sites\/25\/2023\/06\/3673-og.jpg","keywords":["Gas Chromatography"],"articleSection":["Gas Chromatography"],"inLanguage":"en-US","potentialAction":[{"@type":"CommentAction","name":"Comment","target":["https:\/\/www.thermofisher.com\/blog\/analyteguru\/what-do-you-know-about-high-temperature-static-headspace\/#respond"]}]},{"@type":"WebPage","@id":"https:\/\/www.thermofisher.com\/blog\/analyteguru\/what-do-you-know-about-high-temperature-static-headspace\/","url":"https:\/\/www.thermofisher.com\/blog\/analyteguru\/what-do-you-know-about-high-temperature-static-headspace\/","name":"What do you know about High-Temperature Static Headspace?","isPartOf":{"@id":"https:\/\/admin.acceleratingscience.com\/analyteguru\/#website"},"primaryImageOfPage":{"@id":"https:\/\/www.thermofisher.com\/blog\/analyteguru\/what-do-you-know-about-high-temperature-static-headspace\/#primaryimage"},"image":{"@id":"https:\/\/www.thermofisher.com\/blog\/analyteguru\/what-do-you-know-about-high-temperature-static-headspace\/#primaryimage"},"thumbnailUrl":"https:\/\/admin.acceleratingscience.com\/analyteguru\/wp-content\/uploads\/sites\/25\/2023\/06\/3673-og.jpg","datePublished":"2020-03-27T00:00:00+00:00","dateModified":"2024-02-14T22:19:08+00:00","author":{"@id":"https:\/\/admin.acceleratingscience.com\/analyteguru\/#\/schema\/person\/47226155467d66d692f24162ee320270"},"description":"The characterization of the volatile fraction of a liquid or solid sample is a field of gas chromatography coupled to headspace sampling.","breadcrumb":{"@id":"https:\/\/www.thermofisher.com\/blog\/analyteguru\/what-do-you-know-about-high-temperature-static-headspace\/#breadcrumb"},"inLanguage":"en-US","potentialAction":[{"@type":"ReadAction","target":["https:\/\/www.thermofisher.com\/blog\/analyteguru\/what-do-you-know-about-high-temperature-static-headspace\/"]}]},{"@type":"ImageObject","inLanguage":"en-US","@id":"https:\/\/www.thermofisher.com\/blog\/analyteguru\/what-do-you-know-about-high-temperature-static-headspace\/#primaryimage","url":"https:\/\/admin.acceleratingscience.com\/analyteguru\/wp-content\/uploads\/sites\/25\/2023\/06\/3673-og.jpg","contentUrl":"https:\/\/admin.acceleratingscience.com\/analyteguru\/wp-content\/uploads\/sites\/25\/2023\/06\/3673-og.jpg","width":999,"height":818},{"@type":"BreadcrumbList","@id":"https:\/\/www.thermofisher.com\/blog\/analyteguru\/what-do-you-know-about-high-temperature-static-headspace\/#breadcrumb","itemListElement":[{"@type":"ListItem","position":1,"name":"Home","item":"https:\/\/admin.acceleratingscience.com\/analyteguru\/"},{"@type":"ListItem","position":2,"name":"What Do You Know About High-Temperature Static Headspace?"}]},{"@type":"WebSite","@id":"https:\/\/admin.acceleratingscience.com\/analyteguru\/#website","url":"https:\/\/admin.acceleratingscience.com\/analyteguru\/","name":"AnalyteGuru","description":"Your educational resource for biopharma, pharma, environmental, food and agriculture, industrial, and clinical labs","potentialAction":[{"@type":"SearchAction","target":{"@type":"EntryPoint","urlTemplate":"https:\/\/admin.acceleratingscience.com\/analyteguru\/?s={search_term_string}"},"query-input":{"@type":"PropertyValueSpecification","valueRequired":true,"valueName":"search_term_string"}}],"inLanguage":"en-US"},{"@type":"Person","@id":"https:\/\/admin.acceleratingscience.com\/analyteguru\/#\/schema\/person\/47226155467d66d692f24162ee320270","name":"AnalyteGuru Staff","image":{"@type":"ImageObject","inLanguage":"en-US","@id":"https:\/\/secure.gravatar.com\/avatar\/d716c4c68d79e076572f67bcc11df0ef8db404f6c3463e9cfd3d20514a881439?s=96&d=mm&r=g","url":"https:\/\/secure.gravatar.com\/avatar\/d716c4c68d79e076572f67bcc11df0ef8db404f6c3463e9cfd3d20514a881439?s=96&d=mm&r=g","contentUrl":"https:\/\/secure.gravatar.com\/avatar\/d716c4c68d79e076572f67bcc11df0ef8db404f6c3463e9cfd3d20514a881439?s=96&d=mm&r=g","caption":"AnalyteGuru Staff"},"url":"https:\/\/admin.acceleratingscience.com\/author\/kerrycolligan\/"}]}},"taxonomy_info":{"category":[{"value":57,"label":"Gas Chromatography"}],"post_tag":[{"value":59,"label":"Gas Chromatography"}]},"featured_image_src_large":["https:\/\/admin.acceleratingscience.com\/analyteguru\/wp-content\/uploads\/sites\/25\/2023\/06\/3673-og.jpg",760,622,false],"author_info":{"display_name":"AnalyteGuru Staff","author_link":"https:\/\/admin.acceleratingscience.com\/author\/kerrycolligan\/"},"comment_info":0,"category_info":[{"term_id":57,"name":"Gas Chromatography","slug":"gas-chromatography","term_group":0,"term_taxonomy_id":57,"taxonomy":"category","description":"","parent":0,"count":48,"filter":"raw","meta":[],"cat_ID":57,"category_count":48,"category_description":"","cat_name":"Gas Chromatography","category_nicename":"gas-chromatography","category_parent":0}],"tag_info":[{"term_id":59,"name":"Gas Chromatography","slug":"gas-chromatography","term_group":0,"term_taxonomy_id":59,"taxonomy":"post_tag","description":"","parent":0,"count":39,"filter":"raw","meta":[]}],"jetpack_sharing_enabled":true,"jetpack-related-posts":[],"jetpack_featured_media_url":"https:\/\/admin.acceleratingscience.com\/analyteguru\/wp-content\/uploads\/sites\/25\/2023\/06\/3673-og.jpg","_links":{"self":[{"href":"https:\/\/www.thermofisher.com\/blog\/analyteguru\/wp-json\/wp\/v2\/posts\/6639","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/www.thermofisher.com\/blog\/analyteguru\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/www.thermofisher.com\/blog\/analyteguru\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/analyteguru\/wp-json\/wp\/v2\/users\/1327"}],"replies":[{"embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/analyteguru\/wp-json\/wp\/v2\/comments?post=6639"}],"version-history":[{"count":0,"href":"https:\/\/www.thermofisher.com\/blog\/analyteguru\/wp-json\/wp\/v2\/posts\/6639\/revisions"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/analyteguru\/wp-json\/wp\/v2\/media\/6641"}],"wp:attachment":[{"href":"https:\/\/www.thermofisher.com\/blog\/analyteguru\/wp-json\/wp\/v2\/media?parent=6639"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/analyteguru\/wp-json\/wp\/v2\/categories?post=6639"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/analyteguru\/wp-json\/wp\/v2\/tags?post=6639"},{"taxonomy":"division","embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/analyteguru\/wp-json\/wp\/v2\/division?post=6639"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}