{"id":16503,"date":"2022-04-27T14:20:50","date_gmt":"2022-04-27T14:20:50","guid":{"rendered":"https:\/\/admin.acceleratingscience.com\/behindthebench\/?p=16503"},"modified":"2022-05-09T20:12:48","modified_gmt":"2022-05-09T20:12:48","slug":"standardization-challenges-how-to-compare-flow-cytometry-data-across-experiments","status":"publish","type":"post","link":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/standardization-challenges-how-to-compare-flow-cytometry-data-across-experiments\/","title":{"rendered":"Standardization Challenges: How to Compare Your Flow Cytometry Data Across Experiments?"},"content":{"rendered":"<h3><strong>Need for Standardization<\/strong><\/h3>\n<p>Applications of flow cytometry in multi-parametric analysis of single cells or cell populations are widely known. But standardization of flow cytometric assays, not as much. It is extremely important for comparison of data across studies, instruments, and locations. There are many variables in flow cytometry such as instrument setup, reagents, controls, sample handling, data analysis, etc. It\u2019s like making a great burger (only flow cytometric assay standardization is numerous times more difficult) is nice, but for universal acceptance, standardization is necessary in the cooking process, i.e. quality of ingredients, etc., across all locations.<\/p>\n<h3><strong>Why is it Difficult to Standardize Assays? <\/strong><\/h3>\n<p>Standardization is critical for comparing data among samples within an experiment or across experiments, to ascertain the differences observed are biological and not due to variables. Standardization is also important for comparing data between different instruments, scientists, laboratories, and sites. It is relatively less difficult to standardize steps for flow cytometry analysis in a single laboratory (yes, \u201crelatively\u201d as scientists are aware of the challenges with standardization processes). It is much more challenging and daunting to implement standardization across sites and laboratories. Process of standardization is time consuming and takes technical expertise in error management and variability. Bead standards are useful for instruments controls, as standards in studies and compensation. Consistent data from instruments with minimal variation for alignment, size calibration and sort set-up; sample conservation; and comparable data between experiments and instruments are some of the many advantages offered by bead standards.<\/p>\n<h3><strong>Bead Standards to The Rescue!<\/strong><\/h3>\n<p>We now know the challenges involved with standardization of flow cytometry assays, but where to start with standardization? Well, the instrument is the first place to look. Standardize optical and fluidics systems, to ensure that cells pass the lasers at a specific interrogation point. At this point, light scatter and fluorescent light are generated, ensuring no loss of sensitivity or errors in measurements.<\/p>\n<p>The <a href=\"https:\/\/www.thermofisher.com\/us\/en\/home\/life-science\/cell-analysis\/flow-cytometry\/flow-cytometry-calibration\/flow-cytometer-alignment-and-linearity-beads.html?icid=bid_pca_aup_ro1_co_1434_pjt9045_ext1905_0s0_blg_op_awa_kt_s00_comflo1\">Invitrogen<sup>TM<\/sup> AccuCheck<sup>TM <\/sup>ERF Reference Particles<\/a> are used for standardization and calibration of an intensity unit of fluorescence for flow cytometry analysis. For a given fluorophore, determine instrument sensitivity and accuracy through the application of the assigned NIST-traceable intensity standard. This permits data comparison between flow cytometry instruments. In combination with a biological standard, you can use these particles to quantify the level of biomarker expression. This solution offers confidence, reliability, and compatibility especially where comparison of results over time is important; because ERF values of <a href=\"https:\/\/www.thermofisher.com\/us\/en\/home\/life-science\/cell-analysis\/flow-cytometry\/flow-cytometry-calibration\/flow-cytometer-alignment-and-linearity-beads.html?icid=bid_pca_aup_ro1_co_1434_pjt9045_ext1905_0s0_blg_op_awa_kt_s00_comflo2\">AccuCheck<sup>TM <\/sup>ERF Reference Particles<\/a> are NIST-traceable and comparable between manufacturers unlike MESF beads available through other manufacturers.<\/p>\n<p>Now going deep into the assays, fluorochrome-conjugated antibodies are important components of flow cytometry studies; their performance is an area of focus in the context of large, multi-parameter panels, in terms of standardization. Antibodies tend to underperform in large panels due to compensation, unmixing in the detectors or unexpected interactions between dyes, cells, or buffers. Using appropriate controls can be helpful, such as The <a href=\"https:\/\/www.thermofisher.com\/us\/en\/home\/life-science\/cell-analysis\/flow-cytometry\/flow-cytometry-calibration\/flow-cytometry-compensation-tools.html?icid=bid_pca_aup_ro1_co_1434_pjt9045_ext1905_0s0_blg_op_awa_kt_s00_comflo3\">Invitrogen<sup>TM<\/sup> Flow Cytometry Compensation Beads<\/a>. These beads help set voltages and gating parameters for obtaining accurate fluorescence signals. Compensation beads are particularly effective in cases where multiple fluorophore emissions overlap, limited sample is available (which is often the case with immunology studies), large multicolor immunophenotyping panels, or large experiments.<\/p>\n<h3><strong>Count the Benefits of Counting Beads<\/strong><\/h3>\n<p>Flow Cytometry offers a rapid method to quantify cell characteristics, but most flow cytometers cannot provide cell concentration or absolute cell counts in a sample. There are two methods to obtaining an absolute cell count. One method is to use a hematology analyzer by combining separate cell concentration numbers. Another method involves adding an internal microsphere counting standard to the samples, also known as single platform testing. The single platform testing method is preferential due to lack of complications and it avoids interlaboratory variability and underestimations. Various <a href=\"https:\/\/www.thermofisher.com\/us\/en\/home\/life-science\/cell-analysis\/flow-cytometry\/flow-cytometry-calibration\/flow-cytometer-cell-counting-beads.html?icid=bid_pca_aup_ro1_co_1434_pjt9045_ext1905_0s0_blg_op_awa_kt_s00_comflo4\">Invitrogen Flow Cytometry Cell Counting Beads<\/a> are available for cell as well as bacterial counting.<\/p>\n<p>Standardization in flow cytometry assays is a complex process with multiple variables; however, a suite of bead standards can make your studies more manageable while providing high quality data.<\/p>\n<p><strong>Learn more about the product in this blog: <\/strong><a href=\"https:\/\/www.thermofisher.com\/us\/en\/home\/life-science\/cell-analysis\/flow-cytometry\/flow-cytometry-calibration\/flow-cytometer-alignment-and-linearity-beads.html?icid=bid_pca_aup_ro1_co_1434_pjt9045_ext1905_0s0_blg_op_awa_kt_s00_comflo5\">Invitrogen<sup>TM<\/sup> AccuCheck<sup>TM <\/sup>ERF Reference Particles<\/a><strong>, <\/strong><a href=\"https:\/\/www.thermofisher.com\/us\/en\/home\/life-science\/cell-analysis\/flow-cytometry\/flow-cytometry-calibration\/flow-cytometry-compensation-tools.html?icid=bid_pca_aup_ro1_co_1434_pjt9045_ext1905_0s0_blg_op_awa_kt_s00_comflo6\">Invitrogen<sup>TM<\/sup> Flow Cytometry Compensation Beads<\/a><strong>, <\/strong><a href=\"https:\/\/www.thermofisher.com\/us\/en\/home\/life-science\/cell-analysis\/flow-cytometry\/flow-cytometry-calibration\/flow-cytometer-cell-counting-beads.html?icid=bid_pca_aup_ro1_co_1434_pjt9045_ext1905_0s0_blg_op_awa_kt_s00_comflo7\">Invitrogen Flow Cytometry Cell Counting Beads<\/a><\/p>\n<p><img loading=\"lazy\" decoding=\"async\" class=\"aligncenter wp-image-16504 size-large\" title=\"Attune Cytpix and Compensation Beads\" src=\"http:\/\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2022\/04\/PG2631-PJT9045-COL019324-Resize-Attune-CytPix-and-Compensation-Beads-Images-Global-F-v2-compressed-1024x224.jpg\" alt=\"standardization of flow cytometry assay\" width=\"760\" height=\"166\" srcset=\"https:\/\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2022\/04\/PG2631-PJT9045-COL019324-Resize-Attune-CytPix-and-Compensation-Beads-Images-Global-F-v2-compressed-1024x224.jpg 1024w, https:\/\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2022\/04\/PG2631-PJT9045-COL019324-Resize-Attune-CytPix-and-Compensation-Beads-Images-Global-F-v2-compressed-300x66.jpg 300w, https:\/\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2022\/04\/PG2631-PJT9045-COL019324-Resize-Attune-CytPix-and-Compensation-Beads-Images-Global-F-v2-compressed-768x168.jpg 768w, https:\/\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2022\/04\/PG2631-PJT9045-COL019324-Resize-Attune-CytPix-and-Compensation-Beads-Images-Global-F-v2-compressed-1536x336.jpg 1536w, https:\/\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2022\/04\/PG2631-PJT9045-COL019324-Resize-Attune-CytPix-and-Compensation-Beads-Images-Global-F-v2-compressed.jpg 1600w\" sizes=\"auto, (max-width: 760px) 100vw, 760px\" \/><\/p>\n<h3><strong>F<\/strong><strong>low Cytometry Resources and Solutions<\/strong><\/h3>\n<ul>\n<li><a href=\"https:\/\/www.thermofisher.com\/us\/en\/home\/life-science\/cell-analysis\/flow-cytometry\/antibodies-for-flow-cytometry\/flow-cytometry-panel-builder.html?icid=bid_pca_aup_ro1_co_1434_pjt9045_ext1905_0s0_blg_op_awa_kt_s00_comflo8\">Flow Cytometry Panel Builder<\/a> \u2013 this online tool helps you design an effective reagent panel for your specific application on your conventional or spectral flow cytometer<\/li>\n<li><a href=\"https:\/\/www.thermofisher.com\/us\/en\/home\/global\/forms\/flow-cytometry-protocols-handbook.html?icid=bid_pca_aup_ro1_co_1434_pjt9045_ext1905_0s0_blg_op_awa_kt_s00_comflo9\">Flow Cytometry Protocols Handbook<\/a> \u2013 protocols that fit your needs in flow cytometry ranging from sample preparation to numerous cell stimulation conditions, staining, immunophenotyping and data analysis strategies<\/li>\n<li><a href=\"https:\/\/www.thermofisher.com\/us\/en\/home\/life-science\/cell-analysis\/cell-analysis-learning-center\/immunology-at-work.html?icid=bid_pca_aup_ro1_co_1434_pjt9045_ext1905_0s0_blg_op_awa_kt_s00_comflo10\">Immunology at Work Resource Center<\/a> \u2013 practical guidance and protocols to help you get started quickly and confidently<\/li>\n<\/ul>\n<p>&nbsp;<\/p>\n<h2><strong>Related Blogs<\/strong><\/h2>\n<ul>\n<li><a href=\"https:\/\/www.thermofisher.com\/blog\/behindthebench\/clog-resistance-of-non-pressure-based-flow-cytometers\/\">https:\/\/www.thermofisher.com\/blog\/behindthebench\/clog-resistance-of-non-pressure-based-flow-cytometers\/<\/a><\/li>\n<li><a href=\"https:\/\/www.thermofisher.com\/blog\/behindthebench\/is-it-a-t-cell-or-b-cell-antibodies-for-immunophenotyping\/\">https:\/\/www.thermofisher.com\/blog\/behindthebench\/is-it-a-t-cell-or-b-cell-antibodies-for-immunophenotyping\/<\/a><\/li>\n<\/ul>\n","protected":false},"excerpt":{"rendered":"<p>Need for Standardization Applications of flow cytometry in multi-parametric analysis of single cells or cell populations are widely known. But standardization of flow cytometric assays, not as much. It is extremely important for comparison of data across studies, instruments, and locations. There are many variables in flow cytometry such as instrument setup, reagents, controls, sample<\/p>\n","protected":false},"author":120,"featured_media":16504,"comment_status":"open","ping_status":"closed","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_kad_blocks_custom_css":"","_kad_blocks_head_custom_js":"","_kad_blocks_body_custom_js":"","_kad_blocks_footer_custom_js":"","_monsterinsights_skip_tracking":false,"_genesis_hide_title":false,"_genesis_hide_breadcrumbs":false,"_genesis_hide_singular_image":false,"_genesis_hide_footer_widgets":false,"_genesis_custom_body_class":"","_genesis_custom_post_class":"","_genesis_layout":"","_jetpack_newsletter_access":"","_jetpack_dont_email_post_to_subs":false,"_jetpack_newsletter_tier_id":0,"_jetpack_memberships_contains_paywalled_content":false,"_jetpack_hide_featured_image":false,"_jetpack_featured_image_duplicate_dismissed":0,"_jetpack_memberships_contains_paid_content":false,"footnotes":""},"categories":[135],"tags":[210,159],"division":[],"class_list":["post-16503","post","type-post","status-publish","format-standard","has-post-thumbnail","category-immunology","tag-flow-cytometry","tag-immunology","entry"],"_selected_authors":"","_selected_reviewers":"","acf":[],"yoast_head":"<!-- This site is optimized with the Yoast SEO Premium plugin v28.9 (Yoast SEO v28.9) - https:\/\/yoast.com\/product\/yoast-seo-premium-wordpress\/ -->\n<title>Standardization Challenges: How to Compare Your Flow Cytometry Data Across Experiments? Behind the Bench<\/title>\n<meta name=\"description\" content=\"Standardize Flow Cytometry research and data with the easy to use Bead Standards - Invitrogen AccuCheck ERF Reference Particles.\" \/>\n<meta name=\"robots\" content=\"index, follow, max-snippet:-1, max-image-preview:large, max-video-preview:-1\" \/>\n<link rel=\"canonical\" href=\"https:\/\/www.thermofisher.com\/blog\/behindthebench\/standardization-challenges-how-to-compare-flow-cytometry-data-across-experiments\/\" \/>\n<meta property=\"og:locale\" content=\"en_US\" \/>\n<meta property=\"og:type\" content=\"article\" \/>\n<meta property=\"og:title\" content=\"Standardization Challenges: How to Compare Your Flow Cytometry Data Across Experiments?\" \/>\n<meta property=\"og:description\" content=\"Standardize Flow Cytometry research and data with the easy to use Bead Standards - Invitrogen AccuCheck ERF Reference Particles.\" \/>\n<meta property=\"og:url\" content=\"https:\/\/www.thermofisher.com\/blog\/behindthebench\/standardization-challenges-how-to-compare-flow-cytometry-data-across-experiments\/\" \/>\n<meta property=\"og:site_name\" content=\"Behind the Bench\" \/>\n<meta property=\"article:publisher\" content=\"https:\/\/www.facebook.com\/thermofisher\" \/>\n<meta property=\"article:published_time\" content=\"2022-04-27T14:20:50+00:00\" \/>\n<meta property=\"article:modified_time\" content=\"2022-05-09T20:12:48+00:00\" \/>\n<meta property=\"og:image\" content=\"https:\/\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2022\/04\/PG2631-PJT9045-COL019324-Resize-Attune-CytPix-and-Compensation-Beads-Images-Global-F-v2-compressed.jpg\" \/>\n\t<meta property=\"og:image:width\" content=\"1600\" \/>\n\t<meta property=\"og:image:height\" content=\"350\" \/>\n\t<meta property=\"og:image:type\" content=\"image\/jpeg\" \/>\n<meta name=\"author\" content=\"Behind The Bench Staff\" \/>\n<meta name=\"twitter:card\" content=\"summary_large_image\" \/>\n<meta name=\"twitter:creator\" content=\"@thermofisher\" \/>\n<meta name=\"twitter:site\" content=\"@thermofisher\" \/>\n<meta name=\"twitter:label1\" content=\"Written by\" \/>\n\t<meta name=\"twitter:data1\" content=\"Behind The Bench Staff\" \/>\n\t<meta name=\"twitter:label2\" content=\"Est. reading time\" \/>\n\t<meta name=\"twitter:data2\" content=\"4 minutes\" \/>\n<script type=\"application\/ld+json\" class=\"yoast-schema-graph\">{\"@context\":\"https:\\\/\\\/schema.org\",\"@graph\":[{\"@type\":\"Article\",\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/behindthebench\\\/standardization-challenges-how-to-compare-flow-cytometry-data-across-experiments\\\/#article\",\"isPartOf\":{\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/behindthebench\\\/standardization-challenges-how-to-compare-flow-cytometry-data-across-experiments\\\/\"},\"author\":{\"name\":\"Behind The Bench Staff\",\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/behindthebench\\\/#\\\/schema\\\/person\\\/7c9eade775fde6f2d8b4160ac5cc3bae\"},\"headline\":\"Standardization Challenges: How to Compare Your Flow Cytometry Data Across Experiments?\",\"datePublished\":\"2022-04-27T14:20:50+00:00\",\"dateModified\":\"2022-05-09T20:12:48+00:00\",\"mainEntityOfPage\":{\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/behindthebench\\\/standardization-challenges-how-to-compare-flow-cytometry-data-across-experiments\\\/\"},\"wordCount\":796,\"commentCount\":0,\"image\":{\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/behindthebench\\\/standardization-challenges-how-to-compare-flow-cytometry-data-across-experiments\\\/#primaryimage\"},\"thumbnailUrl\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/behindthebench\\\/wp-content\\\/uploads\\\/sites\\\/9\\\/2022\\\/04\\\/PG2631-PJT9045-COL019324-Resize-Attune-CytPix-and-Compensation-Beads-Images-Global-F-v2-compressed.jpg\",\"keywords\":[\"Flow Cytometry\",\"immunology\"],\"articleSection\":[\"Immunology\"],\"inLanguage\":\"en-US\",\"potentialAction\":[{\"@type\":\"CommentAction\",\"name\":\"Comment\",\"target\":[\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/behindthebench\\\/standardization-challenges-how-to-compare-flow-cytometry-data-across-experiments\\\/#respond\"]}]},{\"@type\":\"WebPage\",\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/behindthebench\\\/standardization-challenges-how-to-compare-flow-cytometry-data-across-experiments\\\/\",\"url\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/behindthebench\\\/standardization-challenges-how-to-compare-flow-cytometry-data-across-experiments\\\/\",\"name\":\"Standardization Challenges: How to Compare Your Flow Cytometry Data Across Experiments? Behind the Bench\",\"isPartOf\":{\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/behindthebench\\\/#website\"},\"primaryImageOfPage\":{\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/behindthebench\\\/standardization-challenges-how-to-compare-flow-cytometry-data-across-experiments\\\/#primaryimage\"},\"image\":{\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/behindthebench\\\/standardization-challenges-how-to-compare-flow-cytometry-data-across-experiments\\\/#primaryimage\"},\"thumbnailUrl\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/behindthebench\\\/wp-content\\\/uploads\\\/sites\\\/9\\\/2022\\\/04\\\/PG2631-PJT9045-COL019324-Resize-Attune-CytPix-and-Compensation-Beads-Images-Global-F-v2-compressed.jpg\",\"datePublished\":\"2022-04-27T14:20:50+00:00\",\"dateModified\":\"2022-05-09T20:12:48+00:00\",\"author\":{\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/behindthebench\\\/#\\\/schema\\\/person\\\/7c9eade775fde6f2d8b4160ac5cc3bae\"},\"description\":\"Standardize Flow Cytometry research and data with the easy to use Bead Standards - Invitrogen AccuCheck ERF Reference Particles.\",\"breadcrumb\":{\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/behindthebench\\\/standardization-challenges-how-to-compare-flow-cytometry-data-across-experiments\\\/#breadcrumb\"},\"inLanguage\":\"en-US\",\"potentialAction\":[{\"@type\":\"ReadAction\",\"target\":[\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/behindthebench\\\/standardization-challenges-how-to-compare-flow-cytometry-data-across-experiments\\\/\"]}]},{\"@type\":\"ImageObject\",\"inLanguage\":\"en-US\",\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/behindthebench\\\/standardization-challenges-how-to-compare-flow-cytometry-data-across-experiments\\\/#primaryimage\",\"url\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/behindthebench\\\/wp-content\\\/uploads\\\/sites\\\/9\\\/2022\\\/04\\\/PG2631-PJT9045-COL019324-Resize-Attune-CytPix-and-Compensation-Beads-Images-Global-F-v2-compressed.jpg\",\"contentUrl\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/behindthebench\\\/wp-content\\\/uploads\\\/sites\\\/9\\\/2022\\\/04\\\/PG2631-PJT9045-COL019324-Resize-Attune-CytPix-and-Compensation-Beads-Images-Global-F-v2-compressed.jpg\",\"width\":1600,\"height\":350,\"caption\":\"Attune CytPix and Compensation Beads\"},{\"@type\":\"BreadcrumbList\",\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/behindthebench\\\/standardization-challenges-how-to-compare-flow-cytometry-data-across-experiments\\\/#breadcrumb\",\"itemListElement\":[{\"@type\":\"ListItem\",\"position\":1,\"name\":\"Home\",\"item\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/behindthebench\\\/\"},{\"@type\":\"ListItem\",\"position\":2,\"name\":\"Standardization Challenges: How to Compare Your Flow Cytometry Data Across Experiments?\"}]},{\"@type\":\"WebSite\",\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/behindthebench\\\/#website\",\"url\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/behindthebench\\\/\",\"name\":\"Behind the Bench\",\"description\":\"\",\"potentialAction\":[{\"@type\":\"SearchAction\",\"target\":{\"@type\":\"EntryPoint\",\"urlTemplate\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/behindthebench\\\/?s={search_term_string}\"},\"query-input\":{\"@type\":\"PropertyValueSpecification\",\"valueRequired\":true,\"valueName\":\"search_term_string\"}}],\"inLanguage\":\"en-US\"},{\"@type\":\"Person\",\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/behindthebench\\\/#\\\/schema\\\/person\\\/7c9eade775fde6f2d8b4160ac5cc3bae\",\"name\":\"Behind The Bench Staff\",\"image\":{\"@type\":\"ImageObject\",\"inLanguage\":\"en-US\",\"@id\":\"https:\\\/\\\/secure.gravatar.com\\\/avatar\\\/52d5c16ece4bd1c0a49a63521c2b6a25c54a3f0c33f7afed215ba7e0cdff69e3?s=96&d=mm&r=g\",\"url\":\"https:\\\/\\\/secure.gravatar.com\\\/avatar\\\/52d5c16ece4bd1c0a49a63521c2b6a25c54a3f0c33f7afed215ba7e0cdff69e3?s=96&d=mm&r=g\",\"contentUrl\":\"https:\\\/\\\/secure.gravatar.com\\\/avatar\\\/52d5c16ece4bd1c0a49a63521c2b6a25c54a3f0c33f7afed215ba7e0cdff69e3?s=96&d=mm&r=g\",\"caption\":\"Behind The Bench Staff\"},\"url\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/author\\\/behind-the-bench-staff\\\/\"}]}<\/script>\n<!-- \/ Yoast SEO Premium plugin. -->","yoast_head_json":{"title":"Standardization Challenges: How to Compare Your Flow Cytometry Data Across Experiments? Behind the Bench","description":"Standardize Flow Cytometry research and data with the easy to use Bead Standards - Invitrogen AccuCheck ERF Reference Particles.","robots":{"index":"index","follow":"follow","max-snippet":"max-snippet:-1","max-image-preview":"max-image-preview:large","max-video-preview":"max-video-preview:-1"},"canonical":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/standardization-challenges-how-to-compare-flow-cytometry-data-across-experiments\/","og_locale":"en_US","og_type":"article","og_title":"Standardization Challenges: How to Compare Your Flow Cytometry Data Across Experiments?","og_description":"Standardize Flow Cytometry research and data with the easy to use Bead Standards - Invitrogen AccuCheck ERF Reference Particles.","og_url":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/standardization-challenges-how-to-compare-flow-cytometry-data-across-experiments\/","og_site_name":"Behind the Bench","article_publisher":"https:\/\/www.facebook.com\/thermofisher","article_published_time":"2022-04-27T14:20:50+00:00","article_modified_time":"2022-05-09T20:12:48+00:00","og_image":[{"width":1600,"height":350,"url":"https:\/\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2022\/04\/PG2631-PJT9045-COL019324-Resize-Attune-CytPix-and-Compensation-Beads-Images-Global-F-v2-compressed.jpg","type":"image\/jpeg"}],"author":"Behind The Bench Staff","twitter_card":"summary_large_image","twitter_creator":"@thermofisher","twitter_site":"@thermofisher","twitter_misc":{"Written by":"Behind The Bench Staff","Est. reading time":"4 minutes"},"schema":{"@context":"https:\/\/schema.org","@graph":[{"@type":"Article","@id":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/standardization-challenges-how-to-compare-flow-cytometry-data-across-experiments\/#article","isPartOf":{"@id":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/standardization-challenges-how-to-compare-flow-cytometry-data-across-experiments\/"},"author":{"name":"Behind The Bench Staff","@id":"https:\/\/admin.acceleratingscience.com\/behindthebench\/#\/schema\/person\/7c9eade775fde6f2d8b4160ac5cc3bae"},"headline":"Standardization Challenges: How to Compare Your Flow Cytometry Data Across Experiments?","datePublished":"2022-04-27T14:20:50+00:00","dateModified":"2022-05-09T20:12:48+00:00","mainEntityOfPage":{"@id":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/standardization-challenges-how-to-compare-flow-cytometry-data-across-experiments\/"},"wordCount":796,"commentCount":0,"image":{"@id":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/standardization-challenges-how-to-compare-flow-cytometry-data-across-experiments\/#primaryimage"},"thumbnailUrl":"https:\/\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2022\/04\/PG2631-PJT9045-COL019324-Resize-Attune-CytPix-and-Compensation-Beads-Images-Global-F-v2-compressed.jpg","keywords":["Flow Cytometry","immunology"],"articleSection":["Immunology"],"inLanguage":"en-US","potentialAction":[{"@type":"CommentAction","name":"Comment","target":["https:\/\/www.thermofisher.com\/blog\/behindthebench\/standardization-challenges-how-to-compare-flow-cytometry-data-across-experiments\/#respond"]}]},{"@type":"WebPage","@id":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/standardization-challenges-how-to-compare-flow-cytometry-data-across-experiments\/","url":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/standardization-challenges-how-to-compare-flow-cytometry-data-across-experiments\/","name":"Standardization Challenges: How to Compare Your Flow Cytometry Data Across Experiments? Behind the Bench","isPartOf":{"@id":"https:\/\/admin.acceleratingscience.com\/behindthebench\/#website"},"primaryImageOfPage":{"@id":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/standardization-challenges-how-to-compare-flow-cytometry-data-across-experiments\/#primaryimage"},"image":{"@id":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/standardization-challenges-how-to-compare-flow-cytometry-data-across-experiments\/#primaryimage"},"thumbnailUrl":"https:\/\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2022\/04\/PG2631-PJT9045-COL019324-Resize-Attune-CytPix-and-Compensation-Beads-Images-Global-F-v2-compressed.jpg","datePublished":"2022-04-27T14:20:50+00:00","dateModified":"2022-05-09T20:12:48+00:00","author":{"@id":"https:\/\/admin.acceleratingscience.com\/behindthebench\/#\/schema\/person\/7c9eade775fde6f2d8b4160ac5cc3bae"},"description":"Standardize Flow Cytometry research and data with the easy to use Bead Standards - Invitrogen AccuCheck ERF Reference Particles.","breadcrumb":{"@id":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/standardization-challenges-how-to-compare-flow-cytometry-data-across-experiments\/#breadcrumb"},"inLanguage":"en-US","potentialAction":[{"@type":"ReadAction","target":["https:\/\/www.thermofisher.com\/blog\/behindthebench\/standardization-challenges-how-to-compare-flow-cytometry-data-across-experiments\/"]}]},{"@type":"ImageObject","inLanguage":"en-US","@id":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/standardization-challenges-how-to-compare-flow-cytometry-data-across-experiments\/#primaryimage","url":"https:\/\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2022\/04\/PG2631-PJT9045-COL019324-Resize-Attune-CytPix-and-Compensation-Beads-Images-Global-F-v2-compressed.jpg","contentUrl":"https:\/\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2022\/04\/PG2631-PJT9045-COL019324-Resize-Attune-CytPix-and-Compensation-Beads-Images-Global-F-v2-compressed.jpg","width":1600,"height":350,"caption":"Attune CytPix and Compensation Beads"},{"@type":"BreadcrumbList","@id":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/standardization-challenges-how-to-compare-flow-cytometry-data-across-experiments\/#breadcrumb","itemListElement":[{"@type":"ListItem","position":1,"name":"Home","item":"https:\/\/admin.acceleratingscience.com\/behindthebench\/"},{"@type":"ListItem","position":2,"name":"Standardization Challenges: How to Compare Your Flow Cytometry Data Across Experiments?"}]},{"@type":"WebSite","@id":"https:\/\/admin.acceleratingscience.com\/behindthebench\/#website","url":"https:\/\/admin.acceleratingscience.com\/behindthebench\/","name":"Behind the Bench","description":"","potentialAction":[{"@type":"SearchAction","target":{"@type":"EntryPoint","urlTemplate":"https:\/\/admin.acceleratingscience.com\/behindthebench\/?s={search_term_string}"},"query-input":{"@type":"PropertyValueSpecification","valueRequired":true,"valueName":"search_term_string"}}],"inLanguage":"en-US"},{"@type":"Person","@id":"https:\/\/admin.acceleratingscience.com\/behindthebench\/#\/schema\/person\/7c9eade775fde6f2d8b4160ac5cc3bae","name":"Behind The Bench Staff","image":{"@type":"ImageObject","inLanguage":"en-US","@id":"https:\/\/secure.gravatar.com\/avatar\/52d5c16ece4bd1c0a49a63521c2b6a25c54a3f0c33f7afed215ba7e0cdff69e3?s=96&d=mm&r=g","url":"https:\/\/secure.gravatar.com\/avatar\/52d5c16ece4bd1c0a49a63521c2b6a25c54a3f0c33f7afed215ba7e0cdff69e3?s=96&d=mm&r=g","contentUrl":"https:\/\/secure.gravatar.com\/avatar\/52d5c16ece4bd1c0a49a63521c2b6a25c54a3f0c33f7afed215ba7e0cdff69e3?s=96&d=mm&r=g","caption":"Behind The Bench Staff"},"url":"https:\/\/admin.acceleratingscience.com\/author\/behind-the-bench-staff\/"}]}},"taxonomy_info":{"category":[{"value":135,"label":"Immunology"}],"post_tag":[{"value":210,"label":"Flow Cytometry"},{"value":159,"label":"immunology"}]},"featured_image_src_large":["https:\/\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2022\/04\/PG2631-PJT9045-COL019324-Resize-Attune-CytPix-and-Compensation-Beads-Images-Global-F-v2-compressed-1024x224.jpg",760,166,true],"author_info":{"display_name":"Behind The Bench Staff","author_link":"https:\/\/admin.acceleratingscience.com\/author\/behind-the-bench-staff\/"},"comment_info":0,"category_info":[{"term_id":135,"name":"Immunology","slug":"immunology","term_group":0,"term_taxonomy_id":135,"taxonomy":"category","description":"","parent":0,"count":45,"filter":"raw","meta":[],"cat_ID":135,"category_count":45,"category_description":"","cat_name":"Immunology","category_nicename":"immunology","category_parent":0}],"tag_info":[{"term_id":210,"name":"Flow Cytometry","slug":"flow-cytometry","term_group":0,"term_taxonomy_id":210,"taxonomy":"post_tag","description":"","parent":0,"count":18,"filter":"raw","meta":[]},{"term_id":159,"name":"immunology","slug":"immunology","term_group":0,"term_taxonomy_id":159,"taxonomy":"post_tag","description":"","parent":0,"count":21,"filter":"raw","meta":[]}],"jetpack-related-posts":[{"id":19805,"url":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/liquid-biopsy-workflow-standardization\/","url_meta":{"origin":16503,"position":0},"title":"Improving the Reliability of Liquid Biopsy in Real-World Settings Through Workflow Standardization: A Discussion with Dr. Matteo Allegretti","author":"Ashleigh Barlow","date":"June 18, 2026","format":false,"excerpt":"Liquid biopsy offers a powerful, noninvasive way to detect and track cancer, but ensuring reliability in real-world settings remains a key hurdle. Variability in workflows and the need for robust verification steps limit widespread adoption. Dr. Matteo Allegretti, Senior Researcher at the Translational Oncology Research Unit at the IRCCS Regina\u2026","rel":"","context":"In &quot;Cancer Research&quot;","block_context":{"text":"Cancer Research","link":"https:\/\/admin.acceleratingscience.com\/behindthebench\/cancer-research\/"},"img":{"alt_text":"Research scientist operating laboratory equipment during liquid biopsy and biomarker research.","src":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/06\/Allegretti-Header.png?resize=350%2C200&ssl=1","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/06\/Allegretti-Header.png?resize=350%2C200&ssl=1 1x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/06\/Allegretti-Header.png?resize=525%2C300&ssl=1 1.5x"},"classes":[]},{"id":19997,"url":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/ai-laboratory-automation-qpcr\/","url_meta":{"origin":16503,"position":1},"title":"AI and Laboratory Automation: The experimental context behind qPCR data","author":"Behind The Bench Staff","date":"September 11, 2026","format":false,"excerpt":"AI is changing how researchers analyze biological data, while automation is changing how that data is performed at scale. As more of the research process becomes automated and data-driven, an important question emerges: What should researchers consider about how data is generated before downstream analysis begins? In qPCR, assays, reagents,\u2026","rel":"","context":"In &quot;Lab Tips and Tricks&quot;","block_context":{"text":"Lab Tips and Tricks","link":"https:\/\/admin.acceleratingscience.com\/behindthebench\/lab-tips-and-tricks\/"},"img":{"alt_text":"qPCR reaction tube with amplification curves representing experimental data used in AI and laboratory automation.","src":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/09\/qPCR_GEN_Blog_AI_Lab_Automation_Amafel-Calalay.png?resize=350%2C200&ssl=1","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/09\/qPCR_GEN_Blog_AI_Lab_Automation_Amafel-Calalay.png?resize=350%2C200&ssl=1 1x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/09\/qPCR_GEN_Blog_AI_Lab_Automation_Amafel-Calalay.png?resize=525%2C300&ssl=1 1.5x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/09\/qPCR_GEN_Blog_AI_Lab_Automation_Amafel-Calalay.png?resize=700%2C400&ssl=1 2x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/09\/qPCR_GEN_Blog_AI_Lab_Automation_Amafel-Calalay.png?resize=1050%2C600&ssl=1 3x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/09\/qPCR_GEN_Blog_AI_Lab_Automation_Amafel-Calalay.png?resize=1400%2C800&ssl=1 4x"},"classes":[]},{"id":19435,"url":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/scaling-human-relevant-oncology-testing-oncopro-tumoroids-on-screenin3ds-upscale-platform\/","url_meta":{"origin":16503,"position":2},"title":"Scaling Human-Relevant Oncology Testing: OncoPro Tumoroids on ScreenIn3D\u2019s UpScale Platform","author":"Behind The Bench Staff","date":"February 25, 2026","format":false,"excerpt":"The United States National Institutes of Health has prioritized human cell-based in vitro approaches to \u201cmodel human disease and capture human variability and patient-specific characteristics,\u201d complementing the U.S. Food and Drug Administration\u2019s goals to use advanced in vitro assays as a New Approach Methodology (NAM) that could reduce animal testing.\u2026","rel":"","context":"In &quot;Cancer Research&quot;","block_context":{"text":"Cancer Research","link":"https:\/\/admin.acceleratingscience.com\/behindthebench\/cancer-research\/"},"img":{"alt_text":"","src":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/02\/HuCo021320-screen-v2-1.png?resize=350%2C200&ssl=1","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/02\/HuCo021320-screen-v2-1.png?resize=350%2C200&ssl=1 1x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/02\/HuCo021320-screen-v2-1.png?resize=525%2C300&ssl=1 1.5x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/02\/HuCo021320-screen-v2-1.png?resize=700%2C400&ssl=1 2x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/02\/HuCo021320-screen-v2-1.png?resize=1050%2C600&ssl=1 3x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/02\/HuCo021320-screen-v2-1.png?resize=1400%2C800&ssl=1 4x"},"classes":[]},{"id":19676,"url":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/avian-influenza-qpcr-assay-failure\/","url_meta":{"origin":16503,"position":3},"title":"Avian Influenza qPCR Testing: How Viral Evolution Contributes to Assay Failure\u2014and How Labs Can Stay Ahead","author":"Behind The Bench Staff","date":"May 1, 2026","format":false,"excerpt":"Molecular veterinary diagnostic laboratories rely on reverse transcriptase real-time PCR (RT qPCR) assays to detect avian influenza virus (AIV) quickly and accurately. These assays are central to surveillance programs, outbreak response, and disease control\u2014particularly for highly pathogenic avian influenza (HPAI) strains such as H5Nx. As Influenza A viruses continue to\u2026","rel":"","context":"In &quot;Animal Health and AgriGenomics&quot;","block_context":{"text":"Animal Health and AgriGenomics","link":"https:\/\/admin.acceleratingscience.com\/behindthebench\/animal-health-and-agrigenomics\/"},"img":{"alt_text":"Chickens in a flock on a farm representing poultry populations monitored for avian influenza surveillance.","src":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/05\/Untitled-3.png?resize=350%2C200&ssl=1","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/05\/Untitled-3.png?resize=350%2C200&ssl=1 1x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/05\/Untitled-3.png?resize=525%2C300&ssl=1 1.5x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/05\/Untitled-3.png?resize=700%2C400&ssl=1 2x"},"classes":[]},{"id":19976,"url":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/taqman-assay-design-genomic-data\/","url_meta":{"origin":16503,"position":4},"title":"What does it take to keep 20 million Applied Biosystems TaqMan Assays current?","author":"Ashleigh Barlow","date":"September 30, 2026","format":false,"excerpt":"How do you keep more than 20 million TaqMan assays current as genomic knowledge evolves? Every year, our understanding of biology becomes more complete. New variants are discovered, transcript annotations improve, gene models evolve, and our assay-design algorithms advance. That does not mean an assay designed years ago is suddenly\u2026","rel":"","context":"In &quot;Bioinformatics&quot;","block_context":{"text":"Bioinformatics","link":"https:\/\/admin.acceleratingscience.com\/behindthebench\/bioinformatics\/"},"img":{"alt_text":"","src":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/09\/BAN-16090146_1528x860.jpg?resize=350%2C200&ssl=1","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/09\/BAN-16090146_1528x860.jpg?resize=350%2C200&ssl=1 1x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/09\/BAN-16090146_1528x860.jpg?resize=525%2C300&ssl=1 1.5x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/09\/BAN-16090146_1528x860.jpg?resize=700%2C400&ssl=1 2x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/09\/BAN-16090146_1528x860.jpg?resize=1050%2C600&ssl=1 3x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/09\/BAN-16090146_1528x860.jpg?resize=1400%2C800&ssl=1 4x"},"classes":[]},{"id":19714,"url":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/what-happens-to-your-raman-model-when-the-instrument-changes\/","url_meta":{"origin":16503,"position":5},"title":"What Happens to Your Raman Model When the Instrument Changes?","author":"Marlene Gasdia-Cochrane","date":"May 19, 2026","format":false,"excerpt":"Overview Raman spectroscopy is incredibly powerful for bioprocess monitoring. It gives you real-time visibility into what\u2019s happening inside a bioreactor\u2014no waiting, no guesswork, no relying solely on offline samples. But there\u2019s a practical question that doesn\u2019t get talked about enough: What happens to your analytical model when the instrument changes?\u2026","rel":"","context":"In &quot;Pharma &amp; Biotech Manufacturing&quot;","block_context":{"text":"Pharma &amp; Biotech Manufacturing","link":"https:\/\/admin.acceleratingscience.com\/behindthebench\/pharmamfg\/"},"img":{"alt_text":"Digital biotechnology concept. 3D render. - blue molecules","src":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/05\/iStock-2193981513.jpg?resize=350%2C200&ssl=1","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/05\/iStock-2193981513.jpg?resize=350%2C200&ssl=1 1x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/05\/iStock-2193981513.jpg?resize=525%2C300&ssl=1 1.5x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/05\/iStock-2193981513.jpg?resize=700%2C400&ssl=1 2x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/05\/iStock-2193981513.jpg?resize=1050%2C600&ssl=1 3x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/05\/iStock-2193981513.jpg?resize=1400%2C800&ssl=1 4x"},"classes":[]}],"jetpack_sharing_enabled":true,"jetpack_featured_media_url":"https:\/\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2022\/04\/PG2631-PJT9045-COL019324-Resize-Attune-CytPix-and-Compensation-Beads-Images-Global-F-v2-compressed.jpg","_links":{"self":[{"href":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/wp-json\/wp\/v2\/posts\/16503","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/wp-json\/wp\/v2\/users\/120"}],"replies":[{"embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/wp-json\/wp\/v2\/comments?post=16503"}],"version-history":[{"count":0,"href":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/wp-json\/wp\/v2\/posts\/16503\/revisions"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/wp-json\/wp\/v2\/media\/16504"}],"wp:attachment":[{"href":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/wp-json\/wp\/v2\/media?parent=16503"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/wp-json\/wp\/v2\/categories?post=16503"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/wp-json\/wp\/v2\/tags?post=16503"},{"taxonomy":"division","embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/wp-json\/wp\/v2\/division?post=16503"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}