{"id":16585,"date":"2022-05-06T21:02:53","date_gmt":"2022-05-06T21:02:53","guid":{"rendered":"https:\/\/admin.acceleratingscience.com\/behindthebench\/?p=16585"},"modified":"2022-07-07T13:07:17","modified_gmt":"2022-07-07T13:07:17","slug":"what-are-ways-of-expanding-flow-cytometry-panels","status":"publish","type":"post","link":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/what-are-ways-of-expanding-flow-cytometry-panels\/","title":{"rendered":"What are Ways of Expanding Flow Cytometry Panels?"},"content":{"rendered":"<h1><strong>Pushing the Limits<\/strong><\/h1>\n<h2><strong>Ways of Expanding Immunophenotyping Panels<\/strong><\/h2>\n<p>Advances in cell biology have expanded applications of immunophenotyping in drug discovery, diagnosis, and treatment planning. This has led scientists to design flow cytometry experiments with bigger panels, allowing an ever-increasing complexity of biomarkers. Expanding panels is tricky as it increases the complexity of design, workflow, and data analysis. Creating a panel is often compared to solving sudoku puzzles as there are many variables to be considered.<\/p>\n<p>One way to expand your panel is to increase the number of lasers and detectors. By doing so, more fluorophores can be used in the panel. This requires instrument upgrades and increases costs. But what if you are already at the maximum possible limit of lasers and detectors and still want to increase panel size? Well, another option is to move from conventional to full spectrum flow cytometry. Thus, you may require instrument upgrade or an entirely new flow cytometer with spectral capabilities. This adds extra cost and significant resources for setup, training etc., so it may not be always practical.<\/p>\n<h2><strong>Can Flow Cytometry Panel Expansion be Simple?<\/strong><\/h2>\n<p>As the saying goes \u201cnecessity is the mother of invention&#8221;. To accommodate the panel expansion needs, scientists can add UV and Violet lasers and compatible conjugated antibodies to their set-up or go the spectral analysis route. <a href=\"https:\/\/www.thermofisher.com\/us\/en\/home\/life-science\/cellanalysis\/flow-cytometry\/brilliant-polymer-dyes.html?icid=bid_pca_aup_ro1_co_1434_pjt9045_ext1905_0s0_blg_op_awa_kt_s00_flopan1\">Brilliant Ultra Violet\u2122 (BUV) and Brilliant Violet\u2122 (BV)<\/a>\u00a0are brighter antibodies which can stain dim markers on cells and provide minimum spillover and maximum compatibility with UV and Violet lasers and detector channels. Thus, occupying UV and Violet detector channels with antibodies labeled with BUVs and BVs will reduce the compensation issues. Polymer based technology of BUVs and BVs make it ideal for both \u2018traditional\u2019 and \u2018spectral\u2019 flow cytometry.<\/p>\n<p>Looking outward to acquire more lasers or making major hardware upgrades to your flow cytometer is possible and a great tool. However, this may not always be an optimal or a feasible solution. But what if there was a way to get more out of your flow cytometer (with unchanged instrument hardware)? What if you could expand your panel and achieve your panel design goals by making a slight change in the panel design and fluorochrome selection? Sounds ideal, right? Good news, it is achievable though selection of appropriate dyes and conjugates.<\/p>\n<h2><strong>Ability to Control Conjugate Brightness Makes Panel Expansion Easier<\/strong><\/h2>\n<p>Even with violet and UV in the mix, often there are situations where overlaps in the signal exist using common dyes. A novel solution is available for this problem as well. You don\u2019t need to look any further than <a href=\"https:\/\/www.thermofisher.com\/us\/en\/home\/life-science\/cell-analysis\/flow-cytometry\/novafluor-dyes.html?icid=bid_pca_aup_ro1_co_1434_pjt9045_ext1905_0s0_blg_op_awa_kt_s00_flopan2\">Invitrogen<sup>TM<\/sup> NovaFluor<sup>TM<\/sup> Dyes<\/a> made with Phiton technology (macrostructure labeled with small-molecule fluorophores). NovaFluors unlocks previously unusedable channels in current instrumentation. They also provide high resolution data for cleaner single-cell analysis. These dyes can thus serve as a \u2018gap filler\u2019 along with BUVs and other dyes. Along with variable brightness for NovaFluor Blue 610 and NovaFluor Blue 660, NovaFluors have minimal cross-laser excitation, allowing the replacement of PE-tandem, dyes.\u00a0Yes, you read that right! Now you can increase your conventional panel size without a complete panel redesign.<\/p>\n<p>Novafluor dyes are also more stable than other tandem dyes, which increases their shelf life and drastically reduces lot-to-lot variability. In addition to BUVs and NovaFluors, SuperBright Polymer Dyes and eFluor Organic Dyes have unique spectral signatures and are compatible with both \u2018traditional\u2019 and \u2018spectral\u2019 workflows. Thus, providing you all the flavors needed for your perfect <a href=\"https:\/\/www.thermofisher.com\/us\/en\/home\/life-science\/cell-analysis\/flow-cytometry\/antibodies-for-flow-cytometry\/flow-cytometry-panel-builder.html?cid=fl-bid-flowpanel?cid=bid_pca_aup_ro1_co_1434_pjt9045_ext1905_0s0_blg_op_awa_kt_s00_flopan3\">panel design<\/a>.<\/p>\n<h2><strong>Take Home (or to the Lab) Lesson <\/strong><\/h2>\n<p>Using specialized dyes such as BUVs and NovaFluors in combination can bring versatility to \u2018traditional\u2019 and \u2018spectral\u2019 flow cytometry studies. You can spread your panels across a broader range with these dyes without significant changes to your panel design or instrument hardware. In short, BUVs and NovaFluors together can help you uncover the hidden potential of your flow cytometer.<\/p>\n<p><img loading=\"lazy\" decoding=\"async\" class=\"alignnone wp-image-16586 size-large\" src=\"http:\/\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2022\/05\/PG2631-PJT9045-COL019505-Resize-Spectral-Story-Imagery-Blog-article-Global-F-1-1024x224.jpg\" alt=\"Brilliant UltraViolet\" width=\"760\" height=\"166\" srcset=\"https:\/\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2022\/05\/PG2631-PJT9045-COL019505-Resize-Spectral-Story-Imagery-Blog-article-Global-F-1-1024x224.jpg 1024w, https:\/\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2022\/05\/PG2631-PJT9045-COL019505-Resize-Spectral-Story-Imagery-Blog-article-Global-F-1-300x66.jpg 300w, https:\/\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2022\/05\/PG2631-PJT9045-COL019505-Resize-Spectral-Story-Imagery-Blog-article-Global-F-1-768x168.jpg 768w, https:\/\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2022\/05\/PG2631-PJT9045-COL019505-Resize-Spectral-Story-Imagery-Blog-article-Global-F-1-1536x336.jpg 1536w, https:\/\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2022\/05\/PG2631-PJT9045-COL019505-Resize-Spectral-Story-Imagery-Blog-article-Global-F-1.jpg 1600w\" sizes=\"auto, (max-width: 760px) 100vw, 760px\" \/><\/p>\n<p><strong>Learn more about the product in this blog: <\/strong><a href=\"https:\/\/www.thermofisher.com\/us\/en\/home\/life-science\/cell-analysis\/flow-cytometry\/brilliant-polymer-dyes.html?cid=bid_pca_aup_ro1_co_1434_pjt9045_ext1905_0s0_blg_op_awa_kt_s00_flopan4\">Brilliant Ultra Violet\u2122 (BUVs) and Brilliant Violet\u2122 (BVs)<\/a><strong>, <\/strong><a href=\"https:\/\/www.thermofisher.com\/us\/en\/home\/life-science\/cell-analysis\/flow-cytometry\/novafluor-dyes.html?icid=bid_pca_aup_ro1_co_1434_pjt9045_ext1905_0s0_blg_op_awa_kt_s00_flopan5\">Invitrogen<sup>TM<\/sup> NovaFluor<sup>TM<\/sup> Dyes<\/a><\/p>\n<h2><strong>Resources and Solutions for Flow Cytometry<\/strong><\/h2>\n<ul>\n<li><a href=\"https:\/\/www.thermofisher.com\/us\/en\/home\/life-science\/cell-analysis\/flow-cytometry\/antibodies-for-flow-cytometry\/flow-cytometry-panel-builder.html?icid=bid_pca_aup_ro1_co_1434_pjt9045_ext1905_0s0_blg_op_awa_kt_s00_flopan7\">Flow Cytometry Panel Builder<\/a> \u2013 this online tool helps you design an effective reagent panel for your specific application on your conventional or spectral flow cytometer<\/li>\n<li><a href=\"https:\/\/www.thermofisher.com\/us\/en\/home\/global\/forms\/flow-cytometry-protocols-handbook.html?icid=bid_pca_aup_ro1_co_1434_pjt9045_ext1905_0s0_blg_op_awa_kt_s00_flopan8\">Flow Cytometry Protocols Handbook<\/a> \u2013 protocols that fit your needs in flow cytometry ranging from sample preparation to numerous cell stimulation conditions, staining, immunophenotyping and data analysis strategies<\/li>\n<li><a href=\"https:\/\/www.thermofisher.com\/us\/en\/home\/life-science\/cell-analysis\/cell-analysis-learning-center\/immunology-at-work.html?icid=bid_pca_aup_ro1_co_1434_pjt9045_ext1905_0s0_blg_op_awa_kt_s00_flopan9\">Immunology at Work Resource Center<\/a> \u2013 practical guidance and protocols to help you get started quickly and confidently<\/li>\n<\/ul>\n<p>Brilliant Violet\u2122 and Brilliant Ultra Violet\u2122 are trademarks of Becton, Dickinson and Company or its affiliates, and are used under license.<\/p>\n","protected":false},"excerpt":{"rendered":"<p>Pushing the Limits Ways of Expanding Immunophenotyping Panels Advances in cell biology have expanded applications of immunophenotyping in drug discovery, diagnosis, and treatment planning. This has led scientists to design flow cytometry experiments with bigger panels, allowing an ever-increasing complexity of biomarkers. Expanding panels is tricky as it increases the complexity of design, workflow, and<\/p>\n","protected":false},"author":120,"featured_media":16586,"comment_status":"open","ping_status":"closed","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_kad_blocks_custom_css":"","_kad_blocks_head_custom_js":"","_kad_blocks_body_custom_js":"","_kad_blocks_footer_custom_js":"","_monsterinsights_skip_tracking":false,"_genesis_hide_title":false,"_genesis_hide_breadcrumbs":false,"_genesis_hide_singular_image":false,"_genesis_hide_footer_widgets":false,"_genesis_custom_body_class":"","_genesis_custom_post_class":"","_genesis_layout":"","_jetpack_newsletter_access":"","_jetpack_dont_email_post_to_subs":false,"_jetpack_newsletter_tier_id":0,"_jetpack_memberships_contains_paywalled_content":false,"_jetpack_memberships_contains_paid_content":false,"footnotes":""},"categories":[135],"tags":[210,159],"division":[],"class_list":{"0":"post-16585","1":"post","2":"type-post","3":"status-publish","4":"format-standard","5":"has-post-thumbnail","7":"category-immunology","8":"tag-flow-cytometry","9":"tag-immunology","10":"entry"},"_selected_authors":"","_selected_reviewers":"","acf":[],"yoast_head":"<!-- This site is optimized with the Yoast SEO Premium plugin v27.8 (Yoast SEO v27.8) - https:\/\/yoast.com\/product\/yoast-seo-premium-wordpress\/ -->\n<title>What are Ways of Expanding Flow Cytometry Panels? - Behind the Bench<\/title>\n<meta name=\"description\" content=\"Maximize the potential of your flow cytometry panels with Phiton based NovaFluors dyes, polymer dyes and many such fluorescent conjugates.\" \/>\n<meta name=\"robots\" content=\"index, follow, max-snippet:-1, max-image-preview:large, max-video-preview:-1\" \/>\n<link rel=\"canonical\" href=\"https:\/\/www.thermofisher.com\/blog\/behindthebench\/what-are-ways-of-expanding-flow-cytometry-panels\/\" \/>\n<meta property=\"og:locale\" content=\"en_US\" \/>\n<meta property=\"og:type\" content=\"article\" \/>\n<meta property=\"og:title\" content=\"What are Ways of Expanding Flow Cytometry Panels?\" \/>\n<meta property=\"og:description\" content=\"Maximize the potential of your flow cytometry panels with Phiton based NovaFluors dyes, polymer dyes and many such fluorescent conjugates.\" \/>\n<meta property=\"og:url\" content=\"https:\/\/www.thermofisher.com\/blog\/behindthebench\/what-are-ways-of-expanding-flow-cytometry-panels\/\" \/>\n<meta property=\"og:site_name\" content=\"Behind the Bench\" \/>\n<meta property=\"article:publisher\" content=\"https:\/\/www.facebook.com\/thermofisher\" \/>\n<meta property=\"article:published_time\" content=\"2022-05-06T21:02:53+00:00\" \/>\n<meta property=\"article:modified_time\" content=\"2022-07-07T13:07:17+00:00\" \/>\n<meta property=\"og:image\" content=\"https:\/\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2022\/05\/PG2631-PJT9045-COL019505-Resize-Spectral-Story-Imagery-Blog-article-Global-F-1.jpg\" \/>\n\t<meta property=\"og:image:width\" content=\"1600\" \/>\n\t<meta property=\"og:image:height\" content=\"350\" \/>\n\t<meta property=\"og:image:type\" content=\"image\/jpeg\" \/>\n<meta name=\"author\" content=\"Behind The Bench Staff\" \/>\n<meta name=\"twitter:card\" content=\"summary_large_image\" \/>\n<meta name=\"twitter:creator\" content=\"@thermofisher\" \/>\n<meta name=\"twitter:site\" content=\"@thermofisher\" \/>\n<meta name=\"twitter:label1\" content=\"Written by\" \/>\n\t<meta name=\"twitter:data1\" content=\"Behind The Bench Staff\" \/>\n\t<meta name=\"twitter:label2\" content=\"Est. reading time\" \/>\n\t<meta name=\"twitter:data2\" content=\"4 minutes\" \/>\n<script type=\"application\/ld+json\" class=\"yoast-schema-graph\">{\"@context\":\"https:\\\/\\\/schema.org\",\"@graph\":[{\"@type\":\"Article\",\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/behindthebench\\\/what-are-ways-of-expanding-flow-cytometry-panels\\\/#article\",\"isPartOf\":{\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/behindthebench\\\/what-are-ways-of-expanding-flow-cytometry-panels\\\/\"},\"author\":{\"name\":\"Behind The Bench Staff\",\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/behindthebench\\\/#\\\/schema\\\/person\\\/9bbb2d67616274d3d4a21a584f7f4a0c\"},\"headline\":\"What are Ways of Expanding Flow Cytometry Panels?\",\"datePublished\":\"2022-05-06T21:02:53+00:00\",\"dateModified\":\"2022-07-07T13:07:17+00:00\",\"mainEntityOfPage\":{\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/behindthebench\\\/what-are-ways-of-expanding-flow-cytometry-panels\\\/\"},\"wordCount\":760,\"commentCount\":0,\"image\":{\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/behindthebench\\\/what-are-ways-of-expanding-flow-cytometry-panels\\\/#primaryimage\"},\"thumbnailUrl\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/behindthebench\\\/wp-content\\\/uploads\\\/sites\\\/9\\\/2022\\\/05\\\/PG2631-PJT9045-COL019505-Resize-Spectral-Story-Imagery-Blog-article-Global-F-1.jpg\",\"keywords\":[\"Flow Cytometry\",\"immunology\"],\"articleSection\":[\"Immunology\"],\"inLanguage\":\"en-US\",\"potentialAction\":[{\"@type\":\"CommentAction\",\"name\":\"Comment\",\"target\":[\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/behindthebench\\\/what-are-ways-of-expanding-flow-cytometry-panels\\\/#respond\"]}]},{\"@type\":\"WebPage\",\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/behindthebench\\\/what-are-ways-of-expanding-flow-cytometry-panels\\\/\",\"url\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/behindthebench\\\/what-are-ways-of-expanding-flow-cytometry-panels\\\/\",\"name\":\"What are Ways of Expanding Flow Cytometry Panels? - Behind the Bench\",\"isPartOf\":{\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/behindthebench\\\/#website\"},\"primaryImageOfPage\":{\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/behindthebench\\\/what-are-ways-of-expanding-flow-cytometry-panels\\\/#primaryimage\"},\"image\":{\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/behindthebench\\\/what-are-ways-of-expanding-flow-cytometry-panels\\\/#primaryimage\"},\"thumbnailUrl\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/behindthebench\\\/wp-content\\\/uploads\\\/sites\\\/9\\\/2022\\\/05\\\/PG2631-PJT9045-COL019505-Resize-Spectral-Story-Imagery-Blog-article-Global-F-1.jpg\",\"datePublished\":\"2022-05-06T21:02:53+00:00\",\"dateModified\":\"2022-07-07T13:07:17+00:00\",\"author\":{\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/behindthebench\\\/#\\\/schema\\\/person\\\/9bbb2d67616274d3d4a21a584f7f4a0c\"},\"description\":\"Maximize the potential of your flow cytometry panels with Phiton based NovaFluors dyes, polymer dyes and many such fluorescent conjugates.\",\"breadcrumb\":{\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/behindthebench\\\/what-are-ways-of-expanding-flow-cytometry-panels\\\/#breadcrumb\"},\"inLanguage\":\"en-US\",\"potentialAction\":[{\"@type\":\"ReadAction\",\"target\":[\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/behindthebench\\\/what-are-ways-of-expanding-flow-cytometry-panels\\\/\"]}]},{\"@type\":\"ImageObject\",\"inLanguage\":\"en-US\",\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/behindthebench\\\/what-are-ways-of-expanding-flow-cytometry-panels\\\/#primaryimage\",\"url\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/behindthebench\\\/wp-content\\\/uploads\\\/sites\\\/9\\\/2022\\\/05\\\/PG2631-PJT9045-COL019505-Resize-Spectral-Story-Imagery-Blog-article-Global-F-1.jpg\",\"contentUrl\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/behindthebench\\\/wp-content\\\/uploads\\\/sites\\\/9\\\/2022\\\/05\\\/PG2631-PJT9045-COL019505-Resize-Spectral-Story-Imagery-Blog-article-Global-F-1.jpg\",\"width\":1600,\"height\":350,\"caption\":\"Flow cytometry panels\"},{\"@type\":\"BreadcrumbList\",\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/behindthebench\\\/what-are-ways-of-expanding-flow-cytometry-panels\\\/#breadcrumb\",\"itemListElement\":[{\"@type\":\"ListItem\",\"position\":1,\"name\":\"Home\",\"item\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/behindthebench\\\/\"},{\"@type\":\"ListItem\",\"position\":2,\"name\":\"What are Ways of Expanding Flow Cytometry Panels?\"}]},{\"@type\":\"WebSite\",\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/behindthebench\\\/#website\",\"url\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/behindthebench\\\/\",\"name\":\"Behind the Bench\",\"description\":\"\",\"potentialAction\":[{\"@type\":\"SearchAction\",\"target\":{\"@type\":\"EntryPoint\",\"urlTemplate\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/behindthebench\\\/?s={search_term_string}\"},\"query-input\":{\"@type\":\"PropertyValueSpecification\",\"valueRequired\":true,\"valueName\":\"search_term_string\"}}],\"inLanguage\":\"en-US\"},{\"@type\":\"Person\",\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/behindthebench\\\/#\\\/schema\\\/person\\\/9bbb2d67616274d3d4a21a584f7f4a0c\",\"name\":\"Behind The Bench Staff\",\"image\":{\"@type\":\"ImageObject\",\"inLanguage\":\"en-US\",\"@id\":\"https:\\\/\\\/secure.gravatar.com\\\/avatar\\\/52d5c16ece4bd1c0a49a63521c2b6a25c54a3f0c33f7afed215ba7e0cdff69e3?s=96&d=mm&r=g\",\"url\":\"https:\\\/\\\/secure.gravatar.com\\\/avatar\\\/52d5c16ece4bd1c0a49a63521c2b6a25c54a3f0c33f7afed215ba7e0cdff69e3?s=96&d=mm&r=g\",\"contentUrl\":\"https:\\\/\\\/secure.gravatar.com\\\/avatar\\\/52d5c16ece4bd1c0a49a63521c2b6a25c54a3f0c33f7afed215ba7e0cdff69e3?s=96&d=mm&r=g\",\"caption\":\"Behind The Bench Staff\"},\"url\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/author\\\/behind-the-bench-staff\\\/\"}]}<\/script>\n<!-- \/ Yoast SEO Premium plugin. -->","yoast_head_json":{"title":"What are Ways of Expanding Flow Cytometry Panels? - Behind the Bench","description":"Maximize the potential of your flow cytometry panels with Phiton based NovaFluors dyes, polymer dyes and many such fluorescent conjugates.","robots":{"index":"index","follow":"follow","max-snippet":"max-snippet:-1","max-image-preview":"max-image-preview:large","max-video-preview":"max-video-preview:-1"},"canonical":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/what-are-ways-of-expanding-flow-cytometry-panels\/","og_locale":"en_US","og_type":"article","og_title":"What are Ways of Expanding Flow Cytometry Panels?","og_description":"Maximize the potential of your flow cytometry panels with Phiton based NovaFluors dyes, polymer dyes and many such fluorescent conjugates.","og_url":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/what-are-ways-of-expanding-flow-cytometry-panels\/","og_site_name":"Behind the Bench","article_publisher":"https:\/\/www.facebook.com\/thermofisher","article_published_time":"2022-05-06T21:02:53+00:00","article_modified_time":"2022-07-07T13:07:17+00:00","og_image":[{"width":1600,"height":350,"url":"https:\/\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2022\/05\/PG2631-PJT9045-COL019505-Resize-Spectral-Story-Imagery-Blog-article-Global-F-1.jpg","type":"image\/jpeg"}],"author":"Behind The Bench Staff","twitter_card":"summary_large_image","twitter_creator":"@thermofisher","twitter_site":"@thermofisher","twitter_misc":{"Written by":"Behind The Bench Staff","Est. reading time":"4 minutes"},"schema":{"@context":"https:\/\/schema.org","@graph":[{"@type":"Article","@id":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/what-are-ways-of-expanding-flow-cytometry-panels\/#article","isPartOf":{"@id":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/what-are-ways-of-expanding-flow-cytometry-panels\/"},"author":{"name":"Behind The Bench Staff","@id":"https:\/\/admin.acceleratingscience.com\/behindthebench\/#\/schema\/person\/9bbb2d67616274d3d4a21a584f7f4a0c"},"headline":"What are Ways of Expanding Flow Cytometry Panels?","datePublished":"2022-05-06T21:02:53+00:00","dateModified":"2022-07-07T13:07:17+00:00","mainEntityOfPage":{"@id":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/what-are-ways-of-expanding-flow-cytometry-panels\/"},"wordCount":760,"commentCount":0,"image":{"@id":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/what-are-ways-of-expanding-flow-cytometry-panels\/#primaryimage"},"thumbnailUrl":"https:\/\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2022\/05\/PG2631-PJT9045-COL019505-Resize-Spectral-Story-Imagery-Blog-article-Global-F-1.jpg","keywords":["Flow Cytometry","immunology"],"articleSection":["Immunology"],"inLanguage":"en-US","potentialAction":[{"@type":"CommentAction","name":"Comment","target":["https:\/\/www.thermofisher.com\/blog\/behindthebench\/what-are-ways-of-expanding-flow-cytometry-panels\/#respond"]}]},{"@type":"WebPage","@id":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/what-are-ways-of-expanding-flow-cytometry-panels\/","url":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/what-are-ways-of-expanding-flow-cytometry-panels\/","name":"What are Ways of Expanding Flow Cytometry Panels? - Behind the Bench","isPartOf":{"@id":"https:\/\/admin.acceleratingscience.com\/behindthebench\/#website"},"primaryImageOfPage":{"@id":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/what-are-ways-of-expanding-flow-cytometry-panels\/#primaryimage"},"image":{"@id":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/what-are-ways-of-expanding-flow-cytometry-panels\/#primaryimage"},"thumbnailUrl":"https:\/\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2022\/05\/PG2631-PJT9045-COL019505-Resize-Spectral-Story-Imagery-Blog-article-Global-F-1.jpg","datePublished":"2022-05-06T21:02:53+00:00","dateModified":"2022-07-07T13:07:17+00:00","author":{"@id":"https:\/\/admin.acceleratingscience.com\/behindthebench\/#\/schema\/person\/9bbb2d67616274d3d4a21a584f7f4a0c"},"description":"Maximize the potential of your flow cytometry panels with Phiton based NovaFluors dyes, polymer dyes and many such fluorescent conjugates.","breadcrumb":{"@id":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/what-are-ways-of-expanding-flow-cytometry-panels\/#breadcrumb"},"inLanguage":"en-US","potentialAction":[{"@type":"ReadAction","target":["https:\/\/www.thermofisher.com\/blog\/behindthebench\/what-are-ways-of-expanding-flow-cytometry-panels\/"]}]},{"@type":"ImageObject","inLanguage":"en-US","@id":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/what-are-ways-of-expanding-flow-cytometry-panels\/#primaryimage","url":"https:\/\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2022\/05\/PG2631-PJT9045-COL019505-Resize-Spectral-Story-Imagery-Blog-article-Global-F-1.jpg","contentUrl":"https:\/\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2022\/05\/PG2631-PJT9045-COL019505-Resize-Spectral-Story-Imagery-Blog-article-Global-F-1.jpg","width":1600,"height":350,"caption":"Flow cytometry panels"},{"@type":"BreadcrumbList","@id":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/what-are-ways-of-expanding-flow-cytometry-panels\/#breadcrumb","itemListElement":[{"@type":"ListItem","position":1,"name":"Home","item":"https:\/\/admin.acceleratingscience.com\/behindthebench\/"},{"@type":"ListItem","position":2,"name":"What are Ways of Expanding Flow Cytometry Panels?"}]},{"@type":"WebSite","@id":"https:\/\/admin.acceleratingscience.com\/behindthebench\/#website","url":"https:\/\/admin.acceleratingscience.com\/behindthebench\/","name":"Behind the Bench","description":"","potentialAction":[{"@type":"SearchAction","target":{"@type":"EntryPoint","urlTemplate":"https:\/\/admin.acceleratingscience.com\/behindthebench\/?s={search_term_string}"},"query-input":{"@type":"PropertyValueSpecification","valueRequired":true,"valueName":"search_term_string"}}],"inLanguage":"en-US"},{"@type":"Person","@id":"https:\/\/admin.acceleratingscience.com\/behindthebench\/#\/schema\/person\/9bbb2d67616274d3d4a21a584f7f4a0c","name":"Behind The Bench Staff","image":{"@type":"ImageObject","inLanguage":"en-US","@id":"https:\/\/secure.gravatar.com\/avatar\/52d5c16ece4bd1c0a49a63521c2b6a25c54a3f0c33f7afed215ba7e0cdff69e3?s=96&d=mm&r=g","url":"https:\/\/secure.gravatar.com\/avatar\/52d5c16ece4bd1c0a49a63521c2b6a25c54a3f0c33f7afed215ba7e0cdff69e3?s=96&d=mm&r=g","contentUrl":"https:\/\/secure.gravatar.com\/avatar\/52d5c16ece4bd1c0a49a63521c2b6a25c54a3f0c33f7afed215ba7e0cdff69e3?s=96&d=mm&r=g","caption":"Behind The Bench Staff"},"url":"https:\/\/admin.acceleratingscience.com\/author\/behind-the-bench-staff\/"}]}},"taxonomy_info":{"category":[{"value":135,"label":"Immunology"}],"post_tag":[{"value":210,"label":"Flow Cytometry"},{"value":159,"label":"immunology"}]},"featured_image_src_large":["https:\/\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2022\/05\/PG2631-PJT9045-COL019505-Resize-Spectral-Story-Imagery-Blog-article-Global-F-1-1024x224.jpg",760,166,true],"author_info":{"display_name":"Behind The Bench Staff","author_link":"https:\/\/admin.acceleratingscience.com\/author\/behind-the-bench-staff\/"},"comment_info":1,"category_info":[{"term_id":135,"name":"Immunology","slug":"immunology","term_group":0,"term_taxonomy_id":135,"taxonomy":"category","description":"","parent":0,"count":45,"filter":"raw","meta":[],"cat_ID":135,"category_count":45,"category_description":"","cat_name":"Immunology","category_nicename":"immunology","category_parent":0}],"tag_info":[{"term_id":210,"name":"Flow Cytometry","slug":"flow-cytometry","term_group":0,"term_taxonomy_id":210,"taxonomy":"post_tag","description":"","parent":0,"count":18,"filter":"raw","meta":[]},{"term_id":159,"name":"immunology","slug":"immunology","term_group":0,"term_taxonomy_id":159,"taxonomy":"post_tag","description":"","parent":0,"count":21,"filter":"raw","meta":[]}],"jetpack-related-posts":[{"id":19159,"url":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/scalable-molecular-testing-workflows\/","url_meta":{"origin":16585,"position":0},"title":"From Bottlenecks to Breakthroughs: Scaling Molecular Testing with OpenArray","author":"Ashleigh Barlow","date":"October 1, 2025","format":false,"excerpt":"At Livingston Med Lab in central Texas, the team has long prided itself on doing quality molecular medicine work including qPCR, array cards, and infectious disease panels, and always pushing to offer more to its community. But a roadblock emerged when they asked: how do we scale more efficiently without\u2026","rel":"","context":"In &quot;Infectious Diseases&quot;","block_context":{"text":"Infectious Diseases","link":"https:\/\/admin.acceleratingscience.com\/behindthebench\/infectious-diseases\/"},"img":{"alt_text":"","src":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2025\/10\/Sean-Villasana-is-molecular-lab-supervisor-technical-respresentative-at-Livingston-Med-Lab-1.png?resize=350%2C200&ssl=1","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2025\/10\/Sean-Villasana-is-molecular-lab-supervisor-technical-respresentative-at-Livingston-Med-Lab-1.png?resize=350%2C200&ssl=1 1x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2025\/10\/Sean-Villasana-is-molecular-lab-supervisor-technical-respresentative-at-Livingston-Med-Lab-1.png?resize=525%2C300&ssl=1 1.5x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2025\/10\/Sean-Villasana-is-molecular-lab-supervisor-technical-respresentative-at-Livingston-Med-Lab-1.png?resize=700%2C400&ssl=1 2x"},"classes":[]},{"id":19966,"url":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/multiplex-assay-optimization-qpcr\/","url_meta":{"origin":16585,"position":1},"title":"Building Better Multiplex Panels: Identifying Assay Interactions Early","author":"Ashleigh Barlow","date":"September 8, 2026","format":false,"excerpt":"Multiplex qPCR has transformed molecular research by enabling scientists to detect multiple targets in a single reaction. The benefits are clear: reduced sample consumption, lower costs, faster workflows, and more data from every experiment. However, designing successful multiplex assays isn't always straightforward. Traditionally, identifying problematic assay combinations required extensive wet-lab\u2026","rel":"","context":"In &quot;Lab Tips and Tricks&quot;","block_context":{"text":"Lab Tips and Tricks","link":"https:\/\/admin.acceleratingscience.com\/behindthebench\/lab-tips-and-tricks\/"},"img":{"alt_text":"For Research Use Only. Not for use in diagnostic procedures. \u00a9 2026 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries unless otherwise specified.","src":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/09\/Header-banner-scaled.jpg?resize=350%2C200&ssl=1","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/09\/Header-banner-scaled.jpg?resize=350%2C200&ssl=1 1x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/09\/Header-banner-scaled.jpg?resize=525%2C300&ssl=1 1.5x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/09\/Header-banner-scaled.jpg?resize=700%2C400&ssl=1 2x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/09\/Header-banner-scaled.jpg?resize=1050%2C600&ssl=1 3x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/09\/Header-banner-scaled.jpg?resize=1400%2C800&ssl=1 4x"},"classes":[]},{"id":19184,"url":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/magnetic-cell-separation-how-to-automate-cell-isolation\/","url_meta":{"origin":16585,"position":2},"title":"Magnetic Cell Separation: How to Automate Cell Isolation","author":"Behind The Bench Staff","date":"October 29, 2025","format":false,"excerpt":"In the rapidly advancing field of medical diagnostics, liquid biopsy stands out as an innovative technique that is advancing the way we detect and monitor diseases. Unlike traditional biopsies, which require invasive procedures to extract tissue samples, liquid biopsy offers a minimally invasive alternative by isolating cells from complex samples\u2026","rel":"","context":"In &quot;General&quot;","block_context":{"text":"General","link":"https:\/\/admin.acceleratingscience.com\/behindthebench\/general\/"},"img":{"alt_text":"Automated cell isolation","src":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2025\/10\/Blog-32-Automated-Cell-Isolation_1600x350px.jpg?resize=350%2C200&ssl=1","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2025\/10\/Blog-32-Automated-Cell-Isolation_1600x350px.jpg?resize=350%2C200&ssl=1 1x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2025\/10\/Blog-32-Automated-Cell-Isolation_1600x350px.jpg?resize=525%2C300&ssl=1 1.5x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2025\/10\/Blog-32-Automated-Cell-Isolation_1600x350px.jpg?resize=700%2C400&ssl=1 2x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2025\/10\/Blog-32-Automated-Cell-Isolation_1600x350px.jpg?resize=1050%2C600&ssl=1 3x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2025\/10\/Blog-32-Automated-Cell-Isolation_1600x350px.jpg?resize=1400%2C800&ssl=1 4x"},"classes":[]},{"id":19640,"url":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/direct-from-lysate-snp-genotyping-in-research-applications\/","url_meta":{"origin":16585,"position":3},"title":"Direct-from-Lysate SNP Genotyping in Research Applications","author":"Ashleigh Barlow","date":"April 6, 2026","format":false,"excerpt":"Direct-from-lysate workflow enabled by DNA Extract All Reagents compared to traditional DNA purification for SNP genotyping applications. Can You Perform SNP Genotyping Without DNA Purification? The short answer: In many research workflows, direct-from-lysate SNP genotyping can be performed without traditional DNA purification\u2014when appropriate chemistries and verification procedures are used. Direct-from-lysate\u2026","rel":"","context":"In &quot;General&quot;","block_context":{"text":"General","link":"https:\/\/admin.acceleratingscience.com\/behindthebench\/general\/"},"img":{"alt_text":"Comparison of traditional DNA purification workflow and direct-from-lysate workflow using DNA Extract All Reagents for SNP genotyping prior to qPCR.","src":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/04\/blog-image-1.png?resize=350%2C200&ssl=1","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/04\/blog-image-1.png?resize=350%2C200&ssl=1 1x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/04\/blog-image-1.png?resize=525%2C300&ssl=1 1.5x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/04\/blog-image-1.png?resize=700%2C400&ssl=1 2x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/04\/blog-image-1.png?resize=1050%2C600&ssl=1 3x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/04\/blog-image-1.png?resize=1400%2C800&ssl=1 4x"},"classes":[]},{"id":19828,"url":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/cyclospora-molecular-testing-gi-pathogens\/","url_meta":{"origin":16585,"position":4},"title":"Cyclospora Outbreak in the United States: Why Rapid Molecular Testing Matters for Gastrointestinal Pathogens","author":"Ashleigh Barlow","date":"July 17, 2026","format":false,"excerpt":"Fresh produce is one of the most common sources of Cyclospora outbreaks. Molecular testing helps laboratories rapidly identify Cyclospora cayetanensis alongside other gastrointestinal pathogens from a single specimen. Article Summary The United States is currently investigating a multistate outbreak of Cyclospora cayetanensis, a foodborne parasite that causes prolonged diarrheal illness.\u2026","rel":"","context":"In &quot;Food Safety&quot;","block_context":{"text":"Food Safety","link":"https:\/\/admin.acceleratingscience.com\/behindthebench\/food-safety\/"},"img":{"alt_text":"Magnified view of parasite contamination on fresh leafy greens, illustrating foodborne pathogen detection.","src":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/07\/Cyclospora-Image-6000x4480_Geoffrey-Cassell-scaled.jpeg?resize=350%2C200&ssl=1","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/07\/Cyclospora-Image-6000x4480_Geoffrey-Cassell-scaled.jpeg?resize=350%2C200&ssl=1 1x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/07\/Cyclospora-Image-6000x4480_Geoffrey-Cassell-scaled.jpeg?resize=525%2C300&ssl=1 1.5x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/07\/Cyclospora-Image-6000x4480_Geoffrey-Cassell-scaled.jpeg?resize=700%2C400&ssl=1 2x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/07\/Cyclospora-Image-6000x4480_Geoffrey-Cassell-scaled.jpeg?resize=1050%2C600&ssl=1 3x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/07\/Cyclospora-Image-6000x4480_Geoffrey-Cassell-scaled.jpeg?resize=1400%2C800&ssl=1 4x"},"classes":[]},{"id":19606,"url":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/multiplex-qpcr-optimization-6-plex\/","url_meta":{"origin":16585,"position":5},"title":"Optimizing multiplex qPCR: How to amplify up to 6 targets without compromising analytical sensitivity","author":"Behind The Bench Staff","date":"April 1, 2026","format":false,"excerpt":"What happens when a multiplex qPCR assay begins to lose performance as additional targets are introduced? An assay that performs reproducibly in singleplex can exhibit Ct shift, reduced sensitivity, or compressed dynamic range when scaled to 5- or 6-plex. As plex level increases, competition among primers, probes, reporter dyes, and\u2026","rel":"","context":"In &quot;Cancer Research&quot;","block_context":{"text":"Cancer Research","link":"https:\/\/admin.acceleratingscience.com\/behindthebench\/cancer-research\/"},"img":{"alt_text":"Graph comparing Ct values between singleplex and multiplex qPCR reactions showing \u0394Ct shift as targets increase.","src":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/04\/mulitplexing-image.png?resize=350%2C200&ssl=1","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/04\/mulitplexing-image.png?resize=350%2C200&ssl=1 1x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/04\/mulitplexing-image.png?resize=525%2C300&ssl=1 1.5x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/04\/mulitplexing-image.png?resize=700%2C400&ssl=1 2x"},"classes":[]}],"jetpack_sharing_enabled":true,"jetpack_featured_media_url":"https:\/\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2022\/05\/PG2631-PJT9045-COL019505-Resize-Spectral-Story-Imagery-Blog-article-Global-F-1.jpg","_links":{"self":[{"href":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/wp-json\/wp\/v2\/posts\/16585","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/wp-json\/wp\/v2\/users\/120"}],"replies":[{"embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/wp-json\/wp\/v2\/comments?post=16585"}],"version-history":[{"count":0,"href":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/wp-json\/wp\/v2\/posts\/16585\/revisions"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/wp-json\/wp\/v2\/media\/16586"}],"wp:attachment":[{"href":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/wp-json\/wp\/v2\/media?parent=16585"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/wp-json\/wp\/v2\/categories?post=16585"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/wp-json\/wp\/v2\/tags?post=16585"},{"taxonomy":"division","embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/wp-json\/wp\/v2\/division?post=16585"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}