{"id":19168,"date":"2025-10-14T08:00:00","date_gmt":"2025-10-14T08:00:00","guid":{"rendered":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/?p=19168"},"modified":"2025-10-07T14:11:19","modified_gmt":"2025-10-07T14:11:19","slug":"understanding-dna-melting-analysis-using-uv-visible-spectroscopy","status":"publish","type":"post","link":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/understanding-dna-melting-analysis-using-uv-visible-spectroscopy\/","title":{"rendered":"Understanding DNA Melting Analysis Using UV-Visible Spectroscopy"},"content":{"rendered":"\n<p>DNA melting analysis is a vital molecular biology technique used to study the stability and properties of DNA by monitoring its transition from double-stranded to single-stranded form with increasing temperature. UV-visible spectroscopy measures changes in DNA absorbance during this process, providing insights into DNA stability, sequence composition, and molecular interactions. This method supports key applications such as PCR optimization, mutation detection, genotyping, and drug discovery, making it an invaluable tool in genetic research.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\" id=\"h-the-principle-of-dna-melting\">The Principle of DNA Melting<\/h2>\n\n\n\n<p>DNA melting occurs when the hydrogen bonds between the complementary bases of the double-stranded DNA are disrupted due to increased temperature. As the temperature rises, the double helix structure unwinds, leading to single-stranded DNA. This transition can be monitored by measuring the absorbance at 260 nm, where single-stranded DNA absorbs more UV light than double-stranded DNA.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\" id=\"h-what-is-uv-visible-spectroscopy\">What is UV-Visible Spectroscopy<\/h2>\n\n\n\n<p><a href=\"https:\/\/www.thermofisher.com\/us\/en\/home\/industrial\/spectroscopy-elemental-isotope-analysis\/molecular-spectroscopy\/uv-vis-spectrophotometry\/resources.html?icid=CAD_blog_LSM_2025Oct#spectroscopy-principles\">UV-visible spectroscopy<\/a> (UV-Vis) is a powerful analytical technique used extensively in chemistry, biology, <a href=\"https:\/\/www.thermofisher.com\/us\/en\/home\/industrial\/spectroscopy-elemental-isotope-analysis\/molecular-spectroscopy\/uv-vis-spectrophotometry\/applications.html?icid=CAD_blog_LSM_2025Oct#life-science\">life science<\/a>, and material science to analyze the absorbance and transmittance of light in the ultraviolet and visible regions of the electromagnetic spectrum. When a sample is exposed to UV or visible light, molecules within the sample absorb specific wavelengths of light, causing electronic transitions between energy levels. The resulting absorbance spectrum provides valuable information such as analyte concentration.<\/p>\n\n\n\n<p>UV-Vis spectroscopy is widely applied in quantitative analysis, chemical kinetics, quality control, and environmental monitoring due to its non-destructive nature, high sensitivity, rapid analysis, and versatility. Key components of a UV-Vis spectrophotometer include a light source, monochromator, sample holder, and detector. Factors such as concentration, path length, solvent, and temperature can influence absorbance measurements. Understanding and utilizing UV-Vis spectroscopy can enhance analytical capabilities and drive advancements in various scientific disciplines.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\" id=\"h-uv-visible-spectroscopy-in-dna-melting-analysis\">UV-Visible Spectroscopy in DNA Melting Analysis<\/h2>\n\n\n\n<p>UV-visible spectroscopy is an effective tool for DNA melting analysis due to its sensitivity and simplicity. By recording the absorbance at various temperatures, a melting curve can be generated. The melting curve typically shows a sigmoidal shape, where the midpoint corresponds to the Tm. This data is valuable for understanding the DNA&#8217;s thermodynamic properties and for applications such as PCR optimization, genotyping, and studying DNA-protein interactions.<\/p>\n\n\n\n<h3 class=\"wp-block-heading\" id=\"h-factors-affecting-dna-melting-temperature\">Factors Affecting DNA Melting Temperature<\/h3>\n\n\n\n<p>Several factors influence the Tm of DNA, including:<\/p>\n\n\n\n<ol start=\"1\" class=\"wp-block-list\">\n<li><strong>Base Composition<\/strong>: GC-rich regions have a higher T<sub>m<\/sub> compared to AT-rich regions due to the stronger hydrogen bonding in GC pairs.<\/li>\n\n\n\n<li><strong>DNA Length<\/strong>: Longer DNA fragments tend to have higher T<sub>m<\/sub> values.<\/li>\n\n\n\n<li><strong>Base-Pair Mismatch:<\/strong> Compliment strands without the correct base-pairing leads to a less stable duplex, decreasing T<sub>m<\/sub>.<\/li>\n\n\n\n<li><strong>Salt Concentration<\/strong>: Higher ionic strength stabilizes the DNA duplex and increases the T<sub>m<\/sub>.<\/li>\n\n\n\n<li><strong>pH<\/strong>: Extreme pH levels can destabilize the DNA structure, affecting the T<sub>m<\/sub>.<\/li>\n<\/ol>\n\n\n\n<h3 class=\"wp-block-heading\" id=\"h-applications-of-dna-melting-analysis\">Applications of DNA Melting Analysis<\/h3>\n\n\n\n<p>DNA melting analysis has a wide range of applications in molecular biology and genetics:<\/p>\n\n\n\n<ol start=\"1\" class=\"wp-block-list\">\n<li><strong>PCR Optimization<\/strong>: Determining the optimal annealing temperature for primers.<\/li>\n\n\n\n<li><strong>Mutation Detection<\/strong>: Identifying single nucleotide polymorphisms (SNPs) and other mutations.<\/li>\n\n\n\n<li><strong>Genotyping<\/strong>: Differentiating between genetic variants.<\/li>\n\n\n\n<li><strong>Drug Discovery<\/strong>: Studying the binding affinity of drugs to DNA.<\/li>\n<\/ol>\n\n\n\n<p>We studied the effect of base-pair mismatch and solvent environment with analysis of DNA melting through UV-Visible absorption spectroscopy.&nbsp; You can read about the experiment and see the references and results, including diagrams, melt curves, sequences, instruments used and more in our application note <a href=\"https:\/\/documents.thermofisher.com\/TFS-Assets\/MSD\/Application-Notes\/analysis-dna-melting-uv-visible-an56369-en.pdf?icid=CAD_blog_LSM_2025Oct\"><em>Analysis of DNA melting through UV-Visible absorption spectroscopy<\/em><\/a><em>.<\/em><\/p>\n\n\n\n<h2 class=\"wp-block-heading\" id=\"h-summary\">Summary<\/h2>\n\n\n\n<p>DNA melting analysis using UV-visible spectroscopy is a powerful molecular biology technique that monitors the transition of DNA from double-stranded to single-stranded form as temperature increases. Measuring DNA absorbance changes at 260 nm provides critical insights into DNA stability, sequence composition, and molecular interactions. This analytical method supports diverse applications including PCR optimization, mutation detection, genotyping, and drug discovery. Its sensitivity, simplicity, and versatility make UV-Vis spectroscopy an essential tool for advancing genetic research and diagnostics. Optimizing experimental conditions and understanding factors influencing melting temperature enhance the accuracy and reliability of DNA melting analysis for various scientific studies.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\" id=\"h-additional-resources\">Additional Resources<\/h2>\n\n\n\n<ul class=\"wp-block-list\">\n<li>Application Note: <a href=\"https:\/\/documents.thermofisher.com\/TFS-Assets\/MSD\/Application-Notes\/analysis-dna-melting-uv-visible-an56369-en.pdf?icid=CAD_blog_LSM_2025Oct\">Analysis of DNA melting through UV-Visible absorption spectroscopy<\/a><\/li>\n\n\n\n<li><a href=\"https:\/\/www.thermofisher.com\/us\/en\/home\/industrial\/spectroscopy-elemental-isotope-analysis\/molecular-spectroscopy\/uv-vis-spectrophotometry\/resources.html?icid=CAD_blog_LSM_2025Oct#spectroscopy-principles\">Uv-Vis spectrometers resources<\/a><\/li>\n<\/ul>\n\n\n\n<p><\/p>\n","protected":false},"excerpt":{"rendered":"<p>DNA melting analysis is a vital molecular biology technique used to study the stability and properties of DNA by monitoring its transition from double-stranded to single-stranded form with increasing temperature. UV-visible spectroscopy measures changes in DNA absorbance during this process, providing insights into DNA stability, sequence composition, and molecular interactions. This method supports key applications<\/p>\n","protected":false},"author":38,"featured_media":19169,"comment_status":"open","ping_status":"closed","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_kad_blocks_custom_css":"","_kad_blocks_head_custom_js":"","_kad_blocks_body_custom_js":"","_kad_blocks_footer_custom_js":"","_monsterinsights_skip_tracking":false,"_genesis_hide_title":false,"_genesis_hide_breadcrumbs":false,"_genesis_hide_singular_image":false,"_genesis_hide_footer_widgets":false,"_genesis_custom_body_class":"","_genesis_custom_post_class":"","_genesis_layout":"","_jetpack_newsletter_access":"","_jetpack_dont_email_post_to_subs":false,"_jetpack_newsletter_tier_id":0,"_jetpack_memberships_contains_paywalled_content":false,"_jetpack_memberships_contains_paid_content":false,"footnotes":""},"categories":[1087],"tags":[2303,2527,2532,2526,2536,2530,2497,2535,260,2528,2534,2531,2529,2533,2525],"division":[2350],"class_list":{"0":"post-19168","1":"post","2":"type-post","3":"status-publish","4":"format-standard","5":"has-post-thumbnail","7":"category-pharmamfg","8":"tag-dna","9":"tag-dna-analysis","10":"tag-dna-base-composition","11":"tag-dna-melting-analysis","12":"tag-dna-melting-temperature","13":"tag-dna-stability","14":"tag-drug-discovery","15":"tag-genetic-variation-analysis","16":"tag-genotyping","17":"tag-molecular-biology","18":"tag-molecular-biology-techniques","19":"tag-mutation-detection","20":"tag-pcr-optimization","21":"tag-uv-vis-absorption-spectroscopy","22":"tag-uv-visible-spectroscopy","23":"division-cad","24":"entry"},"_selected_authors":[773],"_selected_reviewers":[],"acf":[],"yoast_head":"<!-- This site is optimized with the Yoast SEO Premium plugin v27.8 (Yoast SEO v27.8) - https:\/\/yoast.com\/product\/yoast-seo-premium-wordpress\/ -->\n<title>Understanding DNA Melting Analysis Using UV-Visible Spectroscopy - Behind the Bench<\/title>\n<meta name=\"description\" content=\"DNA melting analysis using UV-visible spectroscopy reveals critical insights into DNA stability and interactions, supporting applications such as PCR optimization, mutation detection, genotyping, and drug discovery.\" \/>\n<meta name=\"robots\" content=\"index, follow, max-snippet:-1, max-image-preview:large, max-video-preview:-1\" \/>\n<link rel=\"canonical\" href=\"https:\/\/www.thermofisher.com\/blog\/behindthebench\/understanding-dna-melting-analysis-using-uv-visible-spectroscopy\/\" \/>\n<meta property=\"og:locale\" content=\"en_US\" \/>\n<meta property=\"og:type\" content=\"article\" \/>\n<meta property=\"og:title\" content=\"Understanding DNA Melting Analysis Using UV-Visible Spectroscopy\" \/>\n<meta property=\"og:description\" content=\"DNA melting analysis using UV-visible spectroscopy reveals critical insights into DNA stability and interactions, supporting applications such as PCR optimization, mutation detection, genotyping, and drug discovery.\" \/>\n<meta property=\"og:url\" content=\"https:\/\/www.thermofisher.com\/blog\/behindthebench\/understanding-dna-melting-analysis-using-uv-visible-spectroscopy\/\" \/>\n<meta property=\"og:site_name\" content=\"Behind the Bench\" \/>\n<meta property=\"article:publisher\" content=\"https:\/\/www.facebook.com\/thermofisher\" \/>\n<meta property=\"article:published_time\" content=\"2025-10-14T08:00:00+00:00\" \/>\n<meta property=\"og:image\" content=\"https:\/\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2025\/10\/iStock-2200215349.jpg\" \/>\n\t<meta property=\"og:image:width\" content=\"2121\" \/>\n\t<meta property=\"og:image:height\" content=\"1414\" \/>\n\t<meta property=\"og:image:type\" content=\"image\/jpeg\" \/>\n<meta name=\"author\" content=\"Marlene Gasdia-Cochrane\" \/>\n<meta name=\"twitter:card\" content=\"summary_large_image\" \/>\n<meta name=\"twitter:creator\" content=\"@thermofisher\" \/>\n<meta name=\"twitter:site\" content=\"@thermofisher\" \/>\n<meta name=\"twitter:label1\" content=\"Written by\" \/>\n\t<meta name=\"twitter:data1\" content=\"Marlene Gasdia-Cochrane\" \/>\n\t<meta name=\"twitter:label2\" content=\"Est. reading time\" \/>\n\t<meta name=\"twitter:data2\" content=\"4 minutes\" \/>\n<script type=\"application\/ld+json\" class=\"yoast-schema-graph\">{\"@context\":\"https:\\\/\\\/schema.org\",\"@graph\":[{\"@type\":\"Article\",\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/behindthebench\\\/understanding-dna-melting-analysis-using-uv-visible-spectroscopy\\\/#article\",\"isPartOf\":{\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/behindthebench\\\/understanding-dna-melting-analysis-using-uv-visible-spectroscopy\\\/\"},\"author\":{\"name\":\"Marlene Gasdia-Cochrane\",\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/behindthebench\\\/#\\\/schema\\\/person\\\/11bbcaf39c3abda3e0648991cb0f5ea8\"},\"headline\":\"Understanding DNA Melting Analysis Using UV-Visible Spectroscopy\",\"datePublished\":\"2025-10-14T08:00:00+00:00\",\"mainEntityOfPage\":{\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/behindthebench\\\/understanding-dna-melting-analysis-using-uv-visible-spectroscopy\\\/\"},\"wordCount\":664,\"commentCount\":0,\"image\":{\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/behindthebench\\\/understanding-dna-melting-analysis-using-uv-visible-spectroscopy\\\/#primaryimage\"},\"thumbnailUrl\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/behindthebench\\\/wp-content\\\/uploads\\\/sites\\\/9\\\/2025\\\/10\\\/iStock-2200215349.jpg\",\"keywords\":[\"dna\",\"DNA analysis\",\"DNA base composition\",\"DNA melting analysis\",\"DNA melting temperature\",\"DNA stability\",\"drug discovery\",\"Genetic variation analysis\",\"Genotyping\",\"molecular biology\",\"Molecular biology techniques\",\"Mutation detection\",\"PCR optimization\",\"UV-Vis absorption spectroscopy\",\"UV-Visible Spectroscopy\"],\"articleSection\":[\"Pharma &amp; Biotech Manufacturing\"],\"inLanguage\":\"en-US\",\"potentialAction\":[{\"@type\":\"CommentAction\",\"name\":\"Comment\",\"target\":[\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/behindthebench\\\/understanding-dna-melting-analysis-using-uv-visible-spectroscopy\\\/#respond\"]}]},{\"@type\":\"WebPage\",\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/behindthebench\\\/understanding-dna-melting-analysis-using-uv-visible-spectroscopy\\\/\",\"url\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/behindthebench\\\/understanding-dna-melting-analysis-using-uv-visible-spectroscopy\\\/\",\"name\":\"Understanding DNA Melting Analysis Using UV-Visible Spectroscopy - Behind the Bench\",\"isPartOf\":{\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/behindthebench\\\/#website\"},\"primaryImageOfPage\":{\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/behindthebench\\\/understanding-dna-melting-analysis-using-uv-visible-spectroscopy\\\/#primaryimage\"},\"image\":{\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/behindthebench\\\/understanding-dna-melting-analysis-using-uv-visible-spectroscopy\\\/#primaryimage\"},\"thumbnailUrl\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/behindthebench\\\/wp-content\\\/uploads\\\/sites\\\/9\\\/2025\\\/10\\\/iStock-2200215349.jpg\",\"datePublished\":\"2025-10-14T08:00:00+00:00\",\"author\":{\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/behindthebench\\\/#\\\/schema\\\/person\\\/11bbcaf39c3abda3e0648991cb0f5ea8\"},\"description\":\"DNA melting analysis using UV-visible spectroscopy reveals critical insights into DNA stability and interactions, supporting applications such as PCR optimization, mutation detection, genotyping, and drug discovery.\",\"breadcrumb\":{\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/behindthebench\\\/understanding-dna-melting-analysis-using-uv-visible-spectroscopy\\\/#breadcrumb\"},\"inLanguage\":\"en-US\",\"potentialAction\":[{\"@type\":\"ReadAction\",\"target\":[\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/behindthebench\\\/understanding-dna-melting-analysis-using-uv-visible-spectroscopy\\\/\"]}]},{\"@type\":\"ImageObject\",\"inLanguage\":\"en-US\",\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/behindthebench\\\/understanding-dna-melting-analysis-using-uv-visible-spectroscopy\\\/#primaryimage\",\"url\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/behindthebench\\\/wp-content\\\/uploads\\\/sites\\\/9\\\/2025\\\/10\\\/iStock-2200215349.jpg\",\"contentUrl\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/behindthebench\\\/wp-content\\\/uploads\\\/sites\\\/9\\\/2025\\\/10\\\/iStock-2200215349.jpg\",\"width\":2121,\"height\":1414,\"caption\":\"Close-up Of Dna Molecule On Blue Background\"},{\"@type\":\"BreadcrumbList\",\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/behindthebench\\\/understanding-dna-melting-analysis-using-uv-visible-spectroscopy\\\/#breadcrumb\",\"itemListElement\":[{\"@type\":\"ListItem\",\"position\":1,\"name\":\"Home\",\"item\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/behindthebench\\\/\"},{\"@type\":\"ListItem\",\"position\":2,\"name\":\"Understanding DNA Melting Analysis Using UV-Visible Spectroscopy\"}]},{\"@type\":\"WebSite\",\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/behindthebench\\\/#website\",\"url\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/behindthebench\\\/\",\"name\":\"Behind the Bench\",\"description\":\"\",\"potentialAction\":[{\"@type\":\"SearchAction\",\"target\":{\"@type\":\"EntryPoint\",\"urlTemplate\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/behindthebench\\\/?s={search_term_string}\"},\"query-input\":{\"@type\":\"PropertyValueSpecification\",\"valueRequired\":true,\"valueName\":\"search_term_string\"}}],\"inLanguage\":\"en-US\"},{\"@type\":\"Person\",\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/behindthebench\\\/#\\\/schema\\\/person\\\/11bbcaf39c3abda3e0648991cb0f5ea8\",\"name\":\"Marlene Gasdia-Cochrane\",\"image\":{\"@type\":\"ImageObject\",\"inLanguage\":\"en-US\",\"@id\":\"https:\\\/\\\/secure.gravatar.com\\\/avatar\\\/d4ecccff16bb946462f4389b39255e116430923ec42d08f298b5633eba2d07dd?s=96&d=mm&r=g\",\"url\":\"https:\\\/\\\/secure.gravatar.com\\\/avatar\\\/d4ecccff16bb946462f4389b39255e116430923ec42d08f298b5633eba2d07dd?s=96&d=mm&r=g\",\"contentUrl\":\"https:\\\/\\\/secure.gravatar.com\\\/avatar\\\/d4ecccff16bb946462f4389b39255e116430923ec42d08f298b5633eba2d07dd?s=96&d=mm&r=g\",\"caption\":\"Marlene Gasdia-Cochrane\"},\"url\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/author\\\/mgasdiacochrane\\\/\"}]}<\/script>\n<!-- \/ Yoast SEO Premium plugin. -->","yoast_head_json":{"title":"Understanding DNA Melting Analysis Using UV-Visible Spectroscopy - Behind the Bench","description":"DNA melting analysis using UV-visible spectroscopy reveals critical insights into DNA stability and interactions, supporting applications such as PCR optimization, mutation detection, genotyping, and drug discovery.","robots":{"index":"index","follow":"follow","max-snippet":"max-snippet:-1","max-image-preview":"max-image-preview:large","max-video-preview":"max-video-preview:-1"},"canonical":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/understanding-dna-melting-analysis-using-uv-visible-spectroscopy\/","og_locale":"en_US","og_type":"article","og_title":"Understanding DNA Melting Analysis Using UV-Visible Spectroscopy","og_description":"DNA melting analysis using UV-visible spectroscopy reveals critical insights into DNA stability and interactions, supporting applications such as PCR optimization, mutation detection, genotyping, and drug discovery.","og_url":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/understanding-dna-melting-analysis-using-uv-visible-spectroscopy\/","og_site_name":"Behind the Bench","article_publisher":"https:\/\/www.facebook.com\/thermofisher","article_published_time":"2025-10-14T08:00:00+00:00","og_image":[{"width":2121,"height":1414,"url":"https:\/\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2025\/10\/iStock-2200215349.jpg","type":"image\/jpeg"}],"author":"Marlene Gasdia-Cochrane","twitter_card":"summary_large_image","twitter_creator":"@thermofisher","twitter_site":"@thermofisher","twitter_misc":{"Written by":"Marlene Gasdia-Cochrane","Est. reading time":"4 minutes"},"schema":{"@context":"https:\/\/schema.org","@graph":[{"@type":"Article","@id":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/understanding-dna-melting-analysis-using-uv-visible-spectroscopy\/#article","isPartOf":{"@id":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/understanding-dna-melting-analysis-using-uv-visible-spectroscopy\/"},"author":{"name":"Marlene Gasdia-Cochrane","@id":"https:\/\/admin.acceleratingscience.com\/behindthebench\/#\/schema\/person\/11bbcaf39c3abda3e0648991cb0f5ea8"},"headline":"Understanding DNA Melting Analysis Using UV-Visible Spectroscopy","datePublished":"2025-10-14T08:00:00+00:00","mainEntityOfPage":{"@id":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/understanding-dna-melting-analysis-using-uv-visible-spectroscopy\/"},"wordCount":664,"commentCount":0,"image":{"@id":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/understanding-dna-melting-analysis-using-uv-visible-spectroscopy\/#primaryimage"},"thumbnailUrl":"https:\/\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2025\/10\/iStock-2200215349.jpg","keywords":["dna","DNA analysis","DNA base composition","DNA melting analysis","DNA melting temperature","DNA stability","drug discovery","Genetic variation analysis","Genotyping","molecular biology","Molecular biology techniques","Mutation detection","PCR optimization","UV-Vis absorption spectroscopy","UV-Visible Spectroscopy"],"articleSection":["Pharma &amp; Biotech Manufacturing"],"inLanguage":"en-US","potentialAction":[{"@type":"CommentAction","name":"Comment","target":["https:\/\/www.thermofisher.com\/blog\/behindthebench\/understanding-dna-melting-analysis-using-uv-visible-spectroscopy\/#respond"]}]},{"@type":"WebPage","@id":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/understanding-dna-melting-analysis-using-uv-visible-spectroscopy\/","url":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/understanding-dna-melting-analysis-using-uv-visible-spectroscopy\/","name":"Understanding DNA Melting Analysis Using UV-Visible Spectroscopy - Behind the Bench","isPartOf":{"@id":"https:\/\/admin.acceleratingscience.com\/behindthebench\/#website"},"primaryImageOfPage":{"@id":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/understanding-dna-melting-analysis-using-uv-visible-spectroscopy\/#primaryimage"},"image":{"@id":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/understanding-dna-melting-analysis-using-uv-visible-spectroscopy\/#primaryimage"},"thumbnailUrl":"https:\/\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2025\/10\/iStock-2200215349.jpg","datePublished":"2025-10-14T08:00:00+00:00","author":{"@id":"https:\/\/admin.acceleratingscience.com\/behindthebench\/#\/schema\/person\/11bbcaf39c3abda3e0648991cb0f5ea8"},"description":"DNA melting analysis using UV-visible spectroscopy reveals critical insights into DNA stability and interactions, supporting applications such as PCR optimization, mutation detection, genotyping, and drug discovery.","breadcrumb":{"@id":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/understanding-dna-melting-analysis-using-uv-visible-spectroscopy\/#breadcrumb"},"inLanguage":"en-US","potentialAction":[{"@type":"ReadAction","target":["https:\/\/www.thermofisher.com\/blog\/behindthebench\/understanding-dna-melting-analysis-using-uv-visible-spectroscopy\/"]}]},{"@type":"ImageObject","inLanguage":"en-US","@id":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/understanding-dna-melting-analysis-using-uv-visible-spectroscopy\/#primaryimage","url":"https:\/\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2025\/10\/iStock-2200215349.jpg","contentUrl":"https:\/\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2025\/10\/iStock-2200215349.jpg","width":2121,"height":1414,"caption":"Close-up Of Dna Molecule On Blue Background"},{"@type":"BreadcrumbList","@id":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/understanding-dna-melting-analysis-using-uv-visible-spectroscopy\/#breadcrumb","itemListElement":[{"@type":"ListItem","position":1,"name":"Home","item":"https:\/\/admin.acceleratingscience.com\/behindthebench\/"},{"@type":"ListItem","position":2,"name":"Understanding DNA Melting Analysis Using UV-Visible Spectroscopy"}]},{"@type":"WebSite","@id":"https:\/\/admin.acceleratingscience.com\/behindthebench\/#website","url":"https:\/\/admin.acceleratingscience.com\/behindthebench\/","name":"Behind the Bench","description":"","potentialAction":[{"@type":"SearchAction","target":{"@type":"EntryPoint","urlTemplate":"https:\/\/admin.acceleratingscience.com\/behindthebench\/?s={search_term_string}"},"query-input":{"@type":"PropertyValueSpecification","valueRequired":true,"valueName":"search_term_string"}}],"inLanguage":"en-US"},{"@type":"Person","@id":"https:\/\/admin.acceleratingscience.com\/behindthebench\/#\/schema\/person\/11bbcaf39c3abda3e0648991cb0f5ea8","name":"Marlene Gasdia-Cochrane","image":{"@type":"ImageObject","inLanguage":"en-US","@id":"https:\/\/secure.gravatar.com\/avatar\/d4ecccff16bb946462f4389b39255e116430923ec42d08f298b5633eba2d07dd?s=96&d=mm&r=g","url":"https:\/\/secure.gravatar.com\/avatar\/d4ecccff16bb946462f4389b39255e116430923ec42d08f298b5633eba2d07dd?s=96&d=mm&r=g","contentUrl":"https:\/\/secure.gravatar.com\/avatar\/d4ecccff16bb946462f4389b39255e116430923ec42d08f298b5633eba2d07dd?s=96&d=mm&r=g","caption":"Marlene Gasdia-Cochrane"},"url":"https:\/\/admin.acceleratingscience.com\/author\/mgasdiacochrane\/"}]}},"taxonomy_info":{"category":[{"value":1087,"label":"Pharma &amp; Biotech Manufacturing"}],"post_tag":[{"value":2303,"label":"dna"},{"value":2527,"label":"DNA analysis"},{"value":2532,"label":"DNA base composition"},{"value":2526,"label":"DNA melting analysis"},{"value":2536,"label":"DNA melting temperature"},{"value":2530,"label":"DNA stability"},{"value":2497,"label":"drug discovery"},{"value":2535,"label":"Genetic variation analysis"},{"value":260,"label":"Genotyping"},{"value":2528,"label":"molecular biology"},{"value":2534,"label":"Molecular biology techniques"},{"value":2531,"label":"Mutation detection"},{"value":2529,"label":"PCR optimization"},{"value":2533,"label":"UV-Vis absorption spectroscopy"},{"value":2525,"label":"UV-Visible Spectroscopy"}],"division":[{"value":2350,"label":"CAD"}]},"featured_image_src_large":["https:\/\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2025\/10\/iStock-2200215349-1024x683.jpg",760,507,true],"author_info":{"display_name":"Marlene Gasdia-Cochrane","author_link":"https:\/\/admin.acceleratingscience.com\/author\/mgasdiacochrane\/"},"comment_info":1,"category_info":[{"term_id":1087,"name":"Pharma &amp; Biotech Manufacturing","slug":"pharmamfg","term_group":0,"term_taxonomy_id":1087,"taxonomy":"category","description":"Raw material identification through the pharmaceutical manufacturing process, to finished and packaged pharmaceutical product inspection","parent":0,"count":60,"filter":"raw","meta":[],"cat_ID":1087,"category_count":60,"category_description":"Raw material identification through the pharmaceutical manufacturing process, to finished and packaged pharmaceutical product inspection","cat_name":"Pharma &amp; Biotech Manufacturing","category_nicename":"pharmamfg","category_parent":0}],"tag_info":[{"term_id":2303,"name":"dna","slug":"dna","term_group":0,"term_taxonomy_id":2303,"taxonomy":"post_tag","description":"","parent":0,"count":3,"filter":"raw","meta":[]},{"term_id":2527,"name":"DNA analysis","slug":"dna-analysis","term_group":0,"term_taxonomy_id":2527,"taxonomy":"post_tag","description":"","parent":0,"count":1,"filter":"raw","meta":[]},{"term_id":2532,"name":"DNA base composition","slug":"dna-base-composition","term_group":0,"term_taxonomy_id":2532,"taxonomy":"post_tag","description":"","parent":0,"count":1,"filter":"raw","meta":[]},{"term_id":2526,"name":"DNA melting analysis","slug":"dna-melting-analysis","term_group":0,"term_taxonomy_id":2526,"taxonomy":"post_tag","description":"","parent":0,"count":1,"filter":"raw","meta":[]},{"term_id":2536,"name":"DNA melting temperature","slug":"dna-melting-temperature","term_group":0,"term_taxonomy_id":2536,"taxonomy":"post_tag","description":"","parent":0,"count":1,"filter":"raw","meta":[]},{"term_id":2530,"name":"DNA stability","slug":"dna-stability","term_group":0,"term_taxonomy_id":2530,"taxonomy":"post_tag","description":"","parent":0,"count":1,"filter":"raw","meta":[]},{"term_id":2497,"name":"drug discovery","slug":"drug-discovery","term_group":0,"term_taxonomy_id":2497,"taxonomy":"post_tag","description":"","parent":0,"count":3,"filter":"raw","meta":[]},{"term_id":2535,"name":"Genetic variation analysis","slug":"genetic-variation-analysis","term_group":0,"term_taxonomy_id":2535,"taxonomy":"post_tag","description":"","parent":0,"count":1,"filter":"raw","meta":[]},{"term_id":260,"name":"Genotyping","slug":"genotyping","term_group":0,"term_taxonomy_id":260,"taxonomy":"post_tag","description":"","parent":0,"count":14,"filter":"raw","meta":[]},{"term_id":2528,"name":"molecular biology","slug":"molecular-biology","term_group":0,"term_taxonomy_id":2528,"taxonomy":"post_tag","description":"","parent":0,"count":4,"filter":"raw","meta":[]},{"term_id":2534,"name":"Molecular biology techniques","slug":"molecular-biology-techniques","term_group":0,"term_taxonomy_id":2534,"taxonomy":"post_tag","description":"","parent":0,"count":1,"filter":"raw","meta":[]},{"term_id":2531,"name":"Mutation detection","slug":"mutation-detection","term_group":0,"term_taxonomy_id":2531,"taxonomy":"post_tag","description":"","parent":0,"count":1,"filter":"raw","meta":[]},{"term_id":2529,"name":"PCR optimization","slug":"pcr-optimization","term_group":0,"term_taxonomy_id":2529,"taxonomy":"post_tag","description":"","parent":0,"count":1,"filter":"raw","meta":[]},{"term_id":2533,"name":"UV-Vis absorption spectroscopy","slug":"uv-vis-absorption-spectroscopy","term_group":0,"term_taxonomy_id":2533,"taxonomy":"post_tag","description":"","parent":0,"count":1,"filter":"raw","meta":[]},{"term_id":2525,"name":"UV-Visible Spectroscopy","slug":"uv-visible-spectroscopy","term_group":0,"term_taxonomy_id":2525,"taxonomy":"post_tag","description":"","parent":0,"count":1,"filter":"raw","meta":[]}],"jetpack-related-posts":[{"id":19640,"url":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/direct-from-lysate-snp-genotyping-in-research-applications\/","url_meta":{"origin":19168,"position":0},"title":"Direct-from-Lysate SNP Genotyping in Research Applications","author":"Ashleigh Barlow","date":"April 6, 2026","format":false,"excerpt":"Direct-from-lysate workflow enabled by DNA Extract All Reagents compared to traditional DNA purification for SNP genotyping applications. Can You Perform SNP Genotyping Without DNA Purification? The short answer: In many research workflows, direct-from-lysate SNP genotyping can be performed without traditional DNA purification\u2014when appropriate chemistries and verification procedures are used. Direct-from-lysate\u2026","rel":"","context":"In &quot;General&quot;","block_context":{"text":"General","link":"https:\/\/admin.acceleratingscience.com\/behindthebench\/general\/"},"img":{"alt_text":"Comparison of traditional DNA purification workflow and direct-from-lysate workflow using DNA Extract All Reagents for SNP genotyping prior to qPCR.","src":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/04\/blog-image-1.png?resize=350%2C200&ssl=1","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/04\/blog-image-1.png?resize=350%2C200&ssl=1 1x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/04\/blog-image-1.png?resize=525%2C300&ssl=1 1.5x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/04\/blog-image-1.png?resize=700%2C400&ssl=1 2x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/04\/blog-image-1.png?resize=1050%2C600&ssl=1 3x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/04\/blog-image-1.png?resize=1400%2C800&ssl=1 4x"},"classes":[]},{"id":20019,"url":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/qpcr-dpcr-ngs-cancer-research\/","url_meta":{"origin":19168,"position":1},"title":"From tumor heterogeneity to low-frequency variants: How qPCR, dPCR, and NGS work together in cancer research","author":"Ashleigh Barlow","date":"September 26, 2026","format":false,"excerpt":"Dr. Javier Hernandez Losa explains how combining broad genomic profiling with targeted, sensitive confirmation can help researchers address some of cancer biology\u2019s most difficult analytical challenges. A Q&A with Dr. Javier Hernandez Losa, Director of the Molecular Biology Laboratory, Department of Pathology, Hospital Universitario Vall d\u2019Hebron, Barcelona Cancer research increasingly\u2026","rel":"","context":"In &quot;Cancer Research&quot;","block_context":{"text":"Cancer Research","link":"https:\/\/admin.acceleratingscience.com\/behindthebench\/cancer-research\/"},"img":{"alt_text":"Dr. Javier Hernandez Losa, Director of the Molecular Biology Laboratory at Hospital Universitario Vall d'Hebron in Barcelona.","src":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/09\/Picture133.jpg?resize=350%2C200&ssl=1","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/09\/Picture133.jpg?resize=350%2C200&ssl=1 1x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/09\/Picture133.jpg?resize=525%2C300&ssl=1 1.5x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/09\/Picture133.jpg?resize=700%2C400&ssl=1 2x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/09\/Picture133.jpg?resize=1050%2C600&ssl=1 3x"},"classes":[]},{"id":19579,"url":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/liquid-biopsy-cancer-research-ngs-pcr\/","url_meta":{"origin":19168,"position":2},"title":"From Discovery to Monitoring: How Sequencing and PCR Work Together in Oncology Research","author":"Behind The Bench Staff","date":"March 24, 2026","format":false,"excerpt":"Scientific, joins the Absolute Gene-ius podcast to discuss how molecular technologies are advancing precision oncology research. With experience spanning clinical laboratory research, digital PCR research applications, and next-generation sequencing (NGS), Dr. Sollweck shares how scientists are addressing critical unmet needs in cancer research. From liquid biopsy research strategies to minimal\u2026","rel":"","context":"In &quot;Cancer Research&quot;","block_context":{"text":"Cancer Research","link":"https:\/\/admin.acceleratingscience.com\/behindthebench\/cancer-research\/"},"img":{"alt_text":"","src":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/03\/Sollweck-headshot-200x250_Jordan-Ruggieri.jpeg?resize=350%2C200&ssl=1","width":350,"height":200},"classes":[]},{"id":19585,"url":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/advanced-verification-in-pharmaceutical-manufacturing-using-raman-spectroscopy\/","url_meta":{"origin":19168,"position":3},"title":"Advanced Verification in Pharmaceutical Manufacturing Using Raman Spectroscopy","author":"Marlene Gasdia-Cochrane","date":"April 21, 2026","format":false,"excerpt":"Article Summary Pharmaceutical manufacturing relies on strict quality control to ensure that every product released to the market is safe, effective, and compliant with regulatory standards. Analytical testing plays a critical role in verifying product identity, composition, and consistency. Among modern analytical techniques, Raman spectroscopy stands out as a powerful\u2026","rel":"","context":"In &quot;General&quot;","block_context":{"text":"General","link":"https:\/\/admin.acceleratingscience.com\/behindthebench\/general\/"},"img":{"alt_text":"Syringes, injection bottles and tablets lying on the table. Close-up, indoors, view from above. Day light, studio photo. Healthcare concept","src":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/04\/iStock-1355152090.jpg?resize=350%2C200&ssl=1","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/04\/iStock-1355152090.jpg?resize=350%2C200&ssl=1 1x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/04\/iStock-1355152090.jpg?resize=525%2C300&ssl=1 1.5x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/04\/iStock-1355152090.jpg?resize=700%2C400&ssl=1 2x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/04\/iStock-1355152090.jpg?resize=1050%2C600&ssl=1 3x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/04\/iStock-1355152090.jpg?resize=1400%2C800&ssl=1 4x"},"classes":[]},{"id":19658,"url":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/is-raman-spectroscopy-a-better-verification-method-in-pharmaceutical-manufacturing\/","url_meta":{"origin":19168,"position":4},"title":"Is Raman Spectroscopy a Better Verification Method in Pharmaceutical Manufacturing?","author":"Marlene Gasdia-Cochrane","date":"May 5, 2026","format":false,"excerpt":"Article Summary Verification testing is a critical step in pharmaceutical manufacturing, helping to ensure that every product released meets strict safety, quality, and regulatory standards. Verification helps ensure that products contain the correct active ingredients, meet concentration specifications, are free from contamination, and maintain structural and chemical integrity. This process\u2026","rel":"","context":"In &quot;Pharma &amp; Biotech Manufacturing&quot;","block_context":{"text":"Pharma &amp; Biotech Manufacturing","link":"https:\/\/admin.acceleratingscience.com\/behindthebench\/pharmamfg\/"},"img":{"alt_text":"Yellow pill and red check mark sitting on brown wood surface in front of abstract defocused background. Horizontal composition with selective focus and copy space. Front view. Drug approval concept.","src":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/04\/iStock-1223991858.jpg?resize=350%2C200&ssl=1","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/04\/iStock-1223991858.jpg?resize=350%2C200&ssl=1 1x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/04\/iStock-1223991858.jpg?resize=525%2C300&ssl=1 1.5x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/04\/iStock-1223991858.jpg?resize=700%2C400&ssl=1 2x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/04\/iStock-1223991858.jpg?resize=1050%2C600&ssl=1 3x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/04\/iStock-1223991858.jpg?resize=1400%2C800&ssl=1 4x"},"classes":[]},{"id":19232,"url":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/early-molecular-relapse-detection-dpcr\/","url_meta":{"origin":19168,"position":5},"title":"Following the Molecular Fingerprints of Cancer with Digital PCR","author":"Behind The Bench Staff","date":"November 11, 2025","format":false,"excerpt":"https:\/\/www.youtube.com\/watch?v=vbta1TG617s For clinical researchers and scientists, the value of a biomarker depends on its accurate measurement and trackability over time. Digital PCR (dPCR) delivers both. At Tata Medical Center, Dr. Mayur Parihar, Senior Consultant in Hematology and Cytogenetics and Clinical Research Scientist at the Tata Translational Cancer Research Center, recognizes\u2026","rel":"","context":"In &quot;Cancer Research&quot;","block_context":{"text":"Cancer Research","link":"https:\/\/admin.acceleratingscience.com\/behindthebench\/cancer-research\/"},"img":{"alt_text":"Pathologist preparing a molecular diagnostic sample in a clinical laboratory to support precision oncology decision-making.","src":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/01\/LC-Blog_Preferred-1.jpg?resize=350%2C200&ssl=1","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/01\/LC-Blog_Preferred-1.jpg?resize=350%2C200&ssl=1 1x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/01\/LC-Blog_Preferred-1.jpg?resize=525%2C300&ssl=1 1.5x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/01\/LC-Blog_Preferred-1.jpg?resize=700%2C400&ssl=1 2x"},"classes":[]}],"jetpack_sharing_enabled":true,"jetpack_featured_media_url":"https:\/\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2025\/10\/iStock-2200215349.jpg","_links":{"self":[{"href":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/wp-json\/wp\/v2\/posts\/19168","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/wp-json\/wp\/v2\/users\/38"}],"replies":[{"embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/wp-json\/wp\/v2\/comments?post=19168"}],"version-history":[{"count":0,"href":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/wp-json\/wp\/v2\/posts\/19168\/revisions"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/wp-json\/wp\/v2\/media\/19169"}],"wp:attachment":[{"href":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/wp-json\/wp\/v2\/media?parent=19168"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/wp-json\/wp\/v2\/categories?post=19168"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/wp-json\/wp\/v2\/tags?post=19168"},{"taxonomy":"division","embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/wp-json\/wp\/v2\/division?post=19168"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}