{"id":19966,"date":"2026-09-08T17:15:17","date_gmt":"2026-09-08T17:15:17","guid":{"rendered":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/?p=19966"},"modified":"2026-09-10T20:09:48","modified_gmt":"2026-09-10T20:09:48","slug":"multiplex-assay-optimization-qpcr","status":"publish","type":"post","link":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/multiplex-assay-optimization-qpcr\/","title":{"rendered":"Building Better Multiplex Panels: Identifying Assay Interactions Early"},"content":{"rendered":"\n<figure class=\"wp-block-image aligncenter size-large\"><img loading=\"lazy\" decoding=\"async\" width=\"1024\" height=\"203\" src=\"https:\/\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/09\/Header-banner-1024x203.jpg\" alt=\"Multicolored overlapping fluorescence signals representing multiplex qPCR detection.\" class=\"wp-image-19967\" srcset=\"https:\/\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/09\/Header-banner-1024x203.jpg 1024w, https:\/\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/09\/Header-banner-300x59.jpg 300w, https:\/\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/09\/Header-banner-768x152.jpg 768w, https:\/\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/09\/Header-banner-1536x304.jpg 1536w, https:\/\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/09\/Header-banner-2048x405.jpg 2048w\" sizes=\"auto, (max-width: 1024px) 100vw, 1024px\" \/><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\">Multiplex qPCR has transformed molecular research by enabling scientists to detect multiple targets in a single reaction. The benefits are clear: reduced sample consumption, lower costs, faster workflows, and more data from every experiment.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">However, designing successful multiplex assays isn&#8217;t always straightforward. Traditionally, identifying problematic assay combinations required extensive wet-lab optimization and iterative testing. Today, in silico screening approaches can help researchers identify potential interaction risks before analytical validation begins.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\" id=\"h-what-is-multiplex-qpcr\"><strong>What is Multiplex qPCR?<\/strong><\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">Multiplex qPCR is a real-time PCR technique that simultaneously detects multiple genetic targets within a single reaction using distinct primer and probe sets. By consolidating multiple assays into one reaction, researchers can improve throughput, conserve valuable samples, and streamline molecular workflows.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">For researchers looking to implement multiplex workflows, explore Thermo Fisher Scientific&#8217;s <a href=\"https:\/\/www.thermofisher.com\/us\/en\/home\/life-science\/pcr\/real-time-pcr\/taqman-multiplex-qpcr-solution.html\" target=\"_blank\" rel=\"noreferrer noopener\">Multiplex qPCR Solutions<\/a> &nbsp;assays, master mixes, instruments, and workflow resources designed to support multiplex assay development.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\" id=\"h-what-causes-multiplex-qpcr-assay-interactions\"><strong>What causes multiplex qPCR assay interactions?<\/strong><\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">Multiplex assay interactions occur when primers, probes, or amplicons interact with unintended targets within the reaction mixture. These interactions compete with intended amplification events and can compromise assay performance.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Common interaction types include:<\/p>\n\n\n\n<h3 class=\"wp-block-heading\" id=\"h-primer-primer-interactions\"><strong>Primer\u2013Primer interactions<\/strong><\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">Primers may hybridize with one another rather than their intended targets, creating primer-dimer products that consume reagents and reduce amplification efficiency.<\/p>\n\n\n\n<h3 class=\"wp-block-heading\" id=\"h-probe-probe-or-probe-primer-interactions\"><strong>Probe\u2013Probe or Probe\u2013Primer interactions<\/strong><\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">Probes can form secondary structures or interact with neighboring oligonucleotides, increasing background fluorescence and reducing signal clarity.<\/p>\n\n\n\n<h3 class=\"wp-block-heading\" id=\"h-oligo-amplicon-interactions\"><strong>Oligo\u2013Amplicon interactions<\/strong><\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">Primers or probes may partially bind amplified products generated by other assays in the multiplex reaction, potentially affecting specificity and assay sensitivity<\/p>\n\n\n\n<h3 class=\"wp-block-heading\" id=\"h-competition-for-reagents\"><strong>Competition for reagents<\/strong><\/h3>\n\n\n\n<p class=\"wp-block-paragraph\" id=\"h-primer-primer-interactions\">Multiple assays amplify simultaneously and compete for polymerase activity, nucleotides, and reaction components. Poorly balanced assay combinations can reduce overall performance. As researchers move from duplex assays to higher-order multiplexing strategies involving three, four, or more targets, the likelihood of assay compatibility issues increases significantly.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\" id=\"h-introducing-the-taqman-multiplex-interaction-check-mic-tool\"><strong>Introducing the TaqMan Multiplex Interaction Check (MIC) Tool<\/strong><\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">To help simplify multiplex assay development, Thermo Fisher Scientific developed the <a href=\"https:\/\/www.thermofisher.com\/tools\/multiplex-interaction-check\" target=\"_blank\" rel=\"noreferrer noopener\"><strong>TaqMan Multiplex Interaction Check (MIC) tool<\/strong><\/a>.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">The MIC tool is an in silico screening solution that evaluates potential assay interactions before wet-lab testing begins. By analyzing possible oligo\u2013oligo and oligo\u2013amplicon interactions within a multiplex panel, the tool helps identify configurations that may be at higher risk for performance issues. Rather than replacing experimental validation, the MIC tool helps researchers make more informed design decisions earlier in the development process.<br><\/p>\n\n\n\n<h2 class=\"wp-block-heading\" id=\"h-planning-a-multiplex-panel\"><strong>Planning a multiplex panel?<\/strong><\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">Use the <a href=\"https:\/\/www.thermofisher.com\/tools\/multiplex-interaction-check\" target=\"_blank\" rel=\"noreferrer noopener\">TaqMan Multiplex Interaction Check (MIC) Tool<\/a> to evaluate potential assay interactions before analytical validation and identify high-confidence panel configurations earlier in development.<\/p>\n\n\n\n<div class=\"wp-block-limecuda-button lc-button lc-button--rg lc-button--outline is-style-outline\"><a href=\"https:\/\/www.thermofisher.com\/tools\/multiplex-interaction-check\" rel=\"noopener\" target=\"_blank\"><div class=\"lc-button__content\"><span class=\"lc-button__label\">Start your multiplex using the TaqMan Multiplex Interaction Check (MIC) tool<\/span><\/div><\/a><\/div>\n\n\n<style>.wp-block-kadence-spacer.kt-block-spacer-19966_a85ed3-c9 .kt-block-spacer{height:60px;}.wp-block-kadence-spacer.kt-block-spacer-19966_a85ed3-c9 .kt-divider{border-top-width:1px;height:1px;border-top-color:#eee;width:80%;border-top-style:solid;}<\/style>\n<div class=\"wp-block-kadence-spacer aligncenter kt-block-spacer-19966_a85ed3-c9\"><div class=\"kt-block-spacer kt-block-spacer-halign-center\"><hr class=\"kt-divider\" \/><\/div><\/div>\n\n\n\n<h2 class=\"wp-block-heading\" id=\"h-how-the-mic-tool-supports-multiplex-assay-development\"><strong>How the MIC Tool Supports Multiplex Assay Development<\/strong><\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">Built using data generated from approximately 200 multiplex panels across gene expression and pathogen detection applications, the MIC tool was designed to recognize interaction patterns associated with false positives and sensitivity loss.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">By flagging potential risks before laboratory testing, researchers can:<\/p>\n\n\n\n<ul class=\"wp-block-list\">\n<li>Reduce assay optimization cycles<\/li>\n\n\n\n<li>Focus testing on the most promising panel designs<\/li>\n\n\n\n<li>Improve confidence when transitioning from singleplex to multiplex assays<\/li>\n\n\n\n<li>Accelerate development timelines<\/li>\n<\/ul>\n\n\n\n<h2 class=\"wp-block-heading\" id=\"h-building-a-more-efficient-multiplex-workflow\"><strong>Building a More Efficient Multiplex Workflow<\/strong><\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">The MIC tool is part of Thermo Fisher Scientific&#8217;s broader multiplex qPCR ecosystem, which includes TaqMan assays, multiplex master mixes, QuantStudio real-time PCR systems, and assay design resources. Together, these solutions help researchers develop robust multiplex panels more efficiently.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><a href=\"https:\/\/www.thermofisher.com\/us\/en\/home\/life-science\/pcr\/real-time-pcr\/real-time-pcr-assays.html\" target=\"_blank\" rel=\"noreferrer noopener\"><strong>TaqMan Assays<\/strong><\/a><strong>: <\/strong>Highly specific probe-based detection for gene expression, pathogen detection, SNP genotyping, and other qPCR applications. Researchers can explore the complete <a href=\"https:\/\/www.thermofisher.com\/us\/en\/home\/life-science\/pcr\/real-time-pcr\/real-time-pcr-assays.html\" target=\"_blank\" rel=\"noreferrer noopener\">TaqMan Assay portfolio<\/a> to identify analytically validated assays for multiplex panel development:<\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><a href=\"https:\/\/www.thermofisher.com\/us\/en\/home\/life-science\/pcr\/real-time-pcr\/real-time-pcr-assays.html\" target=\"_blank\" rel=\"noreferrer noopener\">Learn more about TaqMan Assays<\/a><\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>TaqMan\u2122 Multiplex Solutions: <\/strong>A collection of multiplex-compatible assays, master mixes, instruments, software, and workflow resources designed to simplify multiplex assay development.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><a href=\"https:\/\/www.thermofisher.com\/us\/en\/home\/life-science\/pcr\/real-time-pcr\/taqman-multiplex-qpcr-solution.html\" target=\"_blank\" rel=\"noreferrer noopener\">Explore TaqMan multiplex solutions<\/a><\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>QuantStudio\u2122 Real-Time PCR Systems: <\/strong>Instruments supporting multiplex detection through advanced optical configurations and multiple fluorescence channels designed for high-order multiplexing.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><a href=\"https:\/\/www.thermofisher.com\/us\/en\/home\/life-science\/pcr\/real-time-pcr.html\" target=\"_blank\" rel=\"noreferrer noopener\">View Quantstudio real-time PCR instruments<\/a><br><br><strong>Multiplex Assay Design Support: <\/strong>Researchers developing custom multiplex panels can leverage Thermo Fisher Scientific assay design expertise and bioinformatics support to improve compatibility assessment and panel performance:<\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><a href=\"https:\/\/www.thermofisher.com\/us\/en\/home\/life-science\/pcr\/real-time-pcr\/real-time-pcr-assays\/assay-design.html\" target=\"_blank\" rel=\"noreferrer noopener\">View multiplex assay design resources<\/a><br><\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>Key Takeaway<\/strong><\/p>\n\n\n\n<ul class=\"wp-block-list\">\n<li>Multiplex qPCR enables simultaneous detection of multiple targets within a single reaction.<\/li>\n\n\n\n<li>Assay interactions can negatively affect specificity, sensitivity, and reproducibility.<\/li>\n\n\n\n<li>Oligo\u2013oligo and oligo\u2013amplicon interactions become more common as multiplex panel complexity increases.<\/li>\n\n\n\n<li>The <a href=\"https:\/\/www.thermofisher.com\/tools\/multiplex-interaction-check\" target=\"_blank\" rel=\"noreferrer noopener\">TaqMan\u2122 Multiplex Interaction Check (MIC) Tool<\/a> helps identify potential compatibility risks before wet-lab testing.<\/li>\n\n\n\n<li>Early interaction screening can reduce optimization cycles and accelerate assay development.<\/li>\n<\/ul>\n\n\n\n<h2 class=\"wp-block-heading\" id=\"h-looking-ahead\"><strong>Looking Ahead<\/strong><\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">As multiplex qPCR applications continue to expand, predictive tools that help identify potential assay interactions before experimentation are becoming increasingly valuable.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">The <a href=\"https:\/\/www.thermofisher.com\/tools\/multiplex-interaction-check\">TaqMan\u2122 Multipl<\/a><a href=\"https:\/\/www.thermofisher.com\/tools\/multiplex-interaction-check\" target=\"_blank\" rel=\"noreferrer noopener\">e<\/a><a href=\"https:\/\/www.thermofisher.com\/tools\/multiplex-interaction-check\">x Interaction Check tool<\/a> helps provide researchers with an additional layer of confidence during assay design, helping reduce avoidable iterations and supporting more efficient multiplex assay development. Combined with optimized assays, multiplex reagents, and QuantStudio instrumentation, early interaction screening can help reduce avoidable iterations and accelerate the development of robust multiplex panels.<br><br>While empirical analytical validation remains essential, starting with a smarter design strategy can help researchers get reliable results faster.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\" id=\"h-frequently-asked-questions\">Frequently Asked Questions<\/h2>\n\n\n<style>.kt-accordion-id19966_a462f5-ed .kt-accordion-inner-wrap{column-gap:var(--global-kb-gap-md, 2rem);row-gap:10px;}.kt-accordion-id19966_a462f5-ed .kt-accordion-panel-inner{border-top:0px solid transparent;border-right:1px solid transparent;border-bottom:1px solid transparent;border-left:1px solid transparent;background:#ffffff;padding-top:var(--global-kb-spacing-sm, 1.5rem);padding-right:var(--global-kb-spacing-sm, 1.5rem);padding-bottom:var(--global-kb-spacing-sm, 1.5rem);padding-left:var(--global-kb-spacing-sm, 1.5rem);}.kt-accordion-id19966_a462f5-ed > .kt-accordion-inner-wrap > .wp-block-kadence-pane > .kt-accordion-header-wrap > .kt-blocks-accordion-header{border-top:1px solid #eeeeee;border-right:1px solid #eeeeee;border-bottom:1px solid #eeeeee;border-left:2px solid #eeeeee;border-top-left-radius:0px;border-top-right-radius:0px;border-bottom-right-radius:0px;border-bottom-left-radius:0px;background:#ffffff;color:#444444;padding-top:14px;padding-right:16px;padding-bottom:14px;padding-left:16px;}.kt-accordion-id19966_a462f5-ed:not( .kt-accodion-icon-style-basiccircle ):not( .kt-accodion-icon-style-xclosecircle ):not( .kt-accodion-icon-style-arrowcircle )  > .kt-accordion-inner-wrap > .wp-block-kadence-pane > .kt-accordion-header-wrap .kt-blocks-accordion-icon-trigger:after, .kt-accordion-id19966_a462f5-ed:not( .kt-accodion-icon-style-basiccircle ):not( .kt-accodion-icon-style-xclosecircle ):not( .kt-accodion-icon-style-arrowcircle )  > .kt-accordion-inner-wrap > .wp-block-kadence-pane > .kt-accordion-header-wrap .kt-blocks-accordion-icon-trigger:before{background:#444444;}.kt-accordion-id19966_a462f5-ed:not( .kt-accodion-icon-style-basic ):not( .kt-accodion-icon-style-xclose ):not( .kt-accodion-icon-style-arrow ) .kt-blocks-accordion-icon-trigger{background:#444444;}.kt-accordion-id19966_a462f5-ed:not( .kt-accodion-icon-style-basic ):not( .kt-accodion-icon-style-xclose ):not( .kt-accodion-icon-style-arrow ) .kt-blocks-accordion-icon-trigger:after, .kt-accordion-id19966_a462f5-ed:not( .kt-accodion-icon-style-basic ):not( .kt-accodion-icon-style-xclose ):not( .kt-accodion-icon-style-arrow ) .kt-blocks-accordion-icon-trigger:before{background:#ffffff;}.kt-accordion-id19966_a462f5-ed > .kt-accordion-inner-wrap > .wp-block-kadence-pane > .kt-accordion-header-wrap > .kt-blocks-accordion-header:hover, \n\t\t\t\tbody:not(.hide-focus-outline) .kt-accordion-id19966_a462f5-ed .kt-blocks-accordion-header:focus-visible{color:#444444;background:#ffffff;border-top-color:#878787;border-right-color:#878787;border-bottom-color:#878787;border-left-color:#878787;}.kt-accordion-id19966_a462f5-ed:not( .kt-accodion-icon-style-basiccircle ):not( .kt-accodion-icon-style-xclosecircle ):not( .kt-accodion-icon-style-arrowcircle ) .kt-accordion-header-wrap .kt-blocks-accordion-header:hover .kt-blocks-accordion-icon-trigger:after, .kt-accordion-id19966_a462f5-ed:not( .kt-accodion-icon-style-basiccircle ):not( .kt-accodion-icon-style-xclosecircle ):not( .kt-accodion-icon-style-arrowcircle ) .kt-accordion-header-wrap .kt-blocks-accordion-header:hover .kt-blocks-accordion-icon-trigger:before, body:not(.hide-focus-outline) .kt-accordion-id19966_a462f5-ed:not( .kt-accodion-icon-style-basiccircle ):not( .kt-accodion-icon-style-xclosecircle ):not( .kt-accodion-icon-style-arrowcircle ) .kt-blocks-accordion--visible .kt-blocks-accordion-icon-trigger:after, body:not(.hide-focus-outline) .kt-accordion-id19966_a462f5-ed:not( .kt-accodion-icon-style-basiccircle ):not( .kt-accodion-icon-style-xclosecircle ):not( .kt-accodion-icon-style-arrowcircle ) .kt-blocks-accordion-header:focus-visible .kt-blocks-accordion-icon-trigger:before{background:#444444;}.kt-accordion-id19966_a462f5-ed:not( .kt-accodion-icon-style-basic ):not( .kt-accodion-icon-style-xclose ):not( .kt-accodion-icon-style-arrow ) .kt-accordion-header-wrap .kt-blocks-accordion-header:hover .kt-blocks-accordion-icon-trigger, body:not(.hide-focus-outline) .kt-accordion-id19966_a462f5-ed:not( .kt-accodion-icon-style-basic ):not( .kt-accodion-icon-style-xclose ):not( .kt-accodion-icon-style-arrow ) .kt-accordion-header-wrap .kt-blocks-accordion-header:focus-visible .kt-blocks-accordion-icon-trigger{background:#444444;}.kt-accordion-id19966_a462f5-ed:not( .kt-accodion-icon-style-basic ):not( .kt-accodion-icon-style-xclose ):not( .kt-accodion-icon-style-arrow ) .kt-accordion-header-wrap .kt-blocks-accordion-header:hover .kt-blocks-accordion-icon-trigger:after, .kt-accordion-id19966_a462f5-ed:not( .kt-accodion-icon-style-basic ):not( .kt-accodion-icon-style-xclose ):not( .kt-accodion-icon-style-arrow ) .kt-accordion-header-wrap .kt-blocks-accordion-header:hover .kt-blocks-accordion-icon-trigger:before, body:not(.hide-focus-outline) .kt-accordion-id19966_a462f5-ed:not( .kt-accodion-icon-style-basic ):not( .kt-accodion-icon-style-xclose ):not( .kt-accodion-icon-style-arrow ) .kt-accordion-header-wrap .kt-blocks-accordion-header:focus-visible .kt-blocks-accordion-icon-trigger:after, body:not(.hide-focus-outline) .kt-accordion-id19966_a462f5-ed:not( .kt-accodion-icon-style-basic ):not( .kt-accodion-icon-style-xclose ):not( .kt-accodion-icon-style-arrow ) .kt-accordion-header-wrap .kt-blocks-accordion-header:focus-visible .kt-blocks-accordion-icon-trigger:before{background:#ffffff;}.kt-accordion-id19966_a462f5-ed .kt-accordion-header-wrap .kt-blocks-accordion-header:focus-visible,\n\t\t\t\t.kt-accordion-id19966_a462f5-ed > .kt-accordion-inner-wrap > .wp-block-kadence-pane > .kt-accordion-header-wrap > .kt-blocks-accordion-header.kt-accordion-panel-active{color:#444444;background:#ffffff;border-top-color:#eeeeee;border-right-color:#eeeeee;border-bottom-color:#eeeeee;border-left-color:#0e9cd1;}.kt-accordion-id19966_a462f5-ed:not( .kt-accodion-icon-style-basiccircle ):not( .kt-accodion-icon-style-xclosecircle ):not( .kt-accodion-icon-style-arrowcircle )  > .kt-accordion-inner-wrap > .wp-block-kadence-pane > .kt-accordion-header-wrap > .kt-blocks-accordion-header.kt-accordion-panel-active .kt-blocks-accordion-icon-trigger:after, .kt-accordion-id19966_a462f5-ed:not( .kt-accodion-icon-style-basiccircle ):not( .kt-accodion-icon-style-xclosecircle ):not( .kt-accodion-icon-style-arrowcircle )  > .kt-accordion-inner-wrap > .wp-block-kadence-pane > .kt-accordion-header-wrap > .kt-blocks-accordion-header.kt-accordion-panel-active .kt-blocks-accordion-icon-trigger:before{background:#444444;}.kt-accordion-id19966_a462f5-ed:not( .kt-accodion-icon-style-basic ):not( .kt-accodion-icon-style-xclose ):not( .kt-accodion-icon-style-arrow ) .kt-blocks-accordion-header.kt-accordion-panel-active .kt-blocks-accordion-icon-trigger{background:#444444;}.kt-accordion-id19966_a462f5-ed:not( .kt-accodion-icon-style-basic ):not( .kt-accodion-icon-style-xclose ):not( .kt-accodion-icon-style-arrow ) .kt-blocks-accordion-header.kt-accordion-panel-active .kt-blocks-accordion-icon-trigger:after, .kt-accordion-id19966_a462f5-ed:not( .kt-accodion-icon-style-basic ):not( .kt-accodion-icon-style-xclose ):not( .kt-accodion-icon-style-arrow ) .kt-blocks-accordion-header.kt-accordion-panel-active .kt-blocks-accordion-icon-trigger:before{background:#ffffff;}@media all and (max-width: 1024px){.kt-accordion-id19966_a462f5-ed .kt-accordion-panel-inner{border-top:0px solid transparent;border-right:1px solid transparent;border-bottom:1px solid transparent;border-left:1px solid transparent;}}@media all and (max-width: 1024px){.kt-accordion-id19966_a462f5-ed > .kt-accordion-inner-wrap > .wp-block-kadence-pane > .kt-accordion-header-wrap > .kt-blocks-accordion-header{border-top:1px solid #eeeeee;border-right:1px solid #eeeeee;border-bottom:1px solid #eeeeee;border-left:2px solid #eeeeee;}}@media all and (max-width: 1024px){.kt-accordion-id19966_a462f5-ed > .kt-accordion-inner-wrap > .wp-block-kadence-pane > .kt-accordion-header-wrap > .kt-blocks-accordion-header:hover, \n\t\t\t\tbody:not(.hide-focus-outline) .kt-accordion-id19966_a462f5-ed .kt-blocks-accordion-header:focus-visible{border-top-color:#878787;border-right-color:#878787;border-bottom-color:#878787;border-left-color:#878787;}}@media all and (max-width: 1024px){.kt-accordion-id19966_a462f5-ed .kt-accordion-header-wrap .kt-blocks-accordion-header:focus-visible,\n\t\t\t\t.kt-accordion-id19966_a462f5-ed > .kt-accordion-inner-wrap > .wp-block-kadence-pane > .kt-accordion-header-wrap > .kt-blocks-accordion-header.kt-accordion-panel-active{border-top-color:#eeeeee;border-right-color:#eeeeee;border-bottom-color:#eeeeee;border-left-color:#0e9cd1;}}@media all and (max-width: 767px){.kt-accordion-id19966_a462f5-ed .kt-accordion-inner-wrap{display:block;}.kt-accordion-id19966_a462f5-ed .kt-accordion-inner-wrap .kt-accordion-pane:not(:first-child){margin-top:10px;}.kt-accordion-id19966_a462f5-ed .kt-accordion-panel-inner{border-top:0px solid transparent;border-right:1px solid transparent;border-bottom:1px solid transparent;border-left:1px solid transparent;}.kt-accordion-id19966_a462f5-ed > .kt-accordion-inner-wrap > .wp-block-kadence-pane > .kt-accordion-header-wrap > .kt-blocks-accordion-header{border-top:1px solid #eeeeee;border-right:1px solid #eeeeee;border-bottom:1px solid #eeeeee;border-left:2px solid #eeeeee;}.kt-accordion-id19966_a462f5-ed > .kt-accordion-inner-wrap > .wp-block-kadence-pane > .kt-accordion-header-wrap > .kt-blocks-accordion-header:hover, \n\t\t\t\tbody:not(.hide-focus-outline) .kt-accordion-id19966_a462f5-ed .kt-blocks-accordion-header:focus-visible{border-top-color:#878787;border-right-color:#878787;border-bottom-color:#878787;border-left-color:#878787;}.kt-accordion-id19966_a462f5-ed .kt-accordion-header-wrap .kt-blocks-accordion-header:focus-visible,\n\t\t\t\t.kt-accordion-id19966_a462f5-ed > .kt-accordion-inner-wrap > .wp-block-kadence-pane > .kt-accordion-header-wrap > .kt-blocks-accordion-header.kt-accordion-panel-active{border-top-color:#eeeeee;border-right-color:#eeeeee;border-bottom-color:#eeeeee;border-left-color:#0e9cd1;}}<\/style>\n<div class=\"wp-block-kadence-accordion alignnone\"><div class=\"kt-accordion-wrap kt-accordion-id19966_a462f5-ed kt-accordion-has-2-panes kt-active-pane-0 kt-accordion-block kt-pane-header-alignment-left kt-accodion-icon-style-arrow kt-accodion-icon-side-right\" style=\"max-width:none\"><div class=\"kt-accordion-inner-wrap\" data-allow-multiple-open=\"false\" data-start-open=\"0\">\n<div class=\"wp-block-kadence-pane kt-accordion-pane kt-accordion-pane-1 kt-pane19966_539586-38\"><h2 class=\"kt-accordion-header-wrap\"><button class=\"kt-blocks-accordion-header kt-acccordion-button-label-show\" type=\"button\"><span class=\"kt-blocks-accordion-title-wrap\"><span class=\"kt-blocks-accordion-title\"><strong>What causes multiplex qPCR assay interactions?<\/strong><\/span><\/span><span class=\"kt-blocks-accordion-icon-trigger\"><\/span><\/button><\/h2><div class=\"kt-accordion-panel kt-accordion-panel-hidden\"><div class=\"kt-accordion-panel-inner\">\n<p class=\"wp-block-paragraph\">Multiplex assay interactions occur when primers, probes, or amplified products bind unintended targets within the reaction. These interactions can reduce amplification efficiency, increase background signal, generate false positives, and decrease assay analytical sensitivity. As multiplex panels grow larger, evaluating compatibility becomes increasingly important. Researchers developing complex panels can learn more about Thermo Fisher&#8217;s multiplex workflow solutions and compatibility assessment tools on the <a href=\"https:\/\/www.thermofisher.com\/us\/en\/home\/life-science\/pcr\/real-time-pcr\/taqman-multiplex-qpcr-solution.html\" target=\"_blank\" rel=\"noreferrer noopener\"><strong>TaqMan\u2122 Multiplex qPCR Solutions<\/strong><\/a> page.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><a href=\"https:\/\/www.thermofisher.com\/us\/en\/home\/life-science\/pcr\/real-time-pcr\/taqman-multiplex-qpcr-solution.html\" target=\"_blank\" rel=\"noreferrer noopener\">Explore multiplex assay development solutions<\/a><\/p>\n<\/div><\/div><\/div>\n\n\n\n<div class=\"wp-block-kadence-pane kt-accordion-pane kt-accordion-pane-2 kt-pane19966_38ba9c-ca\"><h2 class=\"kt-accordion-header-wrap\"><button class=\"kt-blocks-accordion-header kt-acccordion-button-label-show\" type=\"button\"><span class=\"kt-blocks-accordion-title-wrap\"><span class=\"kt-blocks-accordion-title\"><strong>How can researchers identify multiplex assay interactions before analytical validation?<\/strong><\/span><\/span><span class=\"kt-blocks-accordion-icon-trigger\"><\/span><\/button><\/h2><div class=\"kt-accordion-panel kt-accordion-panel-hidden\"><div class=\"kt-accordion-panel-inner\">\n<p class=\"wp-block-paragraph\">Researchers can use in silico screening approaches such as the <a href=\"https:\/\/www.thermofisher.com\/tools\/multiplex-interaction-check\" target=\"_blank\" rel=\"noreferrer noopener\">TaqMan\u2122 Multiplex Interaction Check (MIC) Tool<\/a> to evaluate potential oligo\u2013oligo and oligo\u2013amplicon interactions before laboratory testing begins. Early screening can help prioritize high-confidence assay combinations and reduce costly optimization cycles. To learn more about Thermo Fisher&#8217;s multiplex qPCR ecosystem, including assays, master mixes, and assay compatibility resources, visit the <a href=\"https:\/\/www.thermofisher.com\/us\/en\/home\/life-science\/pcr\/real-time-pcr\/taqman-multiplex-qpcr-solution.html\" target=\"_blank\" rel=\"noreferrer noopener\">TaqMan\u2122 Multiplex qPCR Solutions hub<\/a>.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><a href=\"https:\/\/www.thermofisher.com\/us\/en\/home\/life-science\/pcr\/real-time-pcr\/taqman-multiplex-qpcr-solution.html\" target=\"_blank\" rel=\"noreferrer noopener\">Discover multiplex compatibility tools<\/a><\/p>\n<\/div><\/div><\/div>\n<\/div><\/div><\/div>\n\n\n<style>.kt-accordion-id19966_c42905-c3 .kt-accordion-inner-wrap{column-gap:var(--global-kb-gap-md, 2rem);row-gap:10px;}.kt-accordion-id19966_c42905-c3 .kt-accordion-panel-inner{border-top:0px solid transparent;border-right:1px solid transparent;border-bottom:1px solid transparent;border-left:1px solid transparent;background:#ffffff;padding-top:var(--global-kb-spacing-sm, 1.5rem);padding-right:var(--global-kb-spacing-sm, 1.5rem);padding-bottom:var(--global-kb-spacing-sm, 1.5rem);padding-left:var(--global-kb-spacing-sm, 1.5rem);}.kt-accordion-id19966_c42905-c3 > .kt-accordion-inner-wrap > .wp-block-kadence-pane > .kt-accordion-header-wrap > .kt-blocks-accordion-header{border-top:1px solid #eeeeee;border-right:1px solid #eeeeee;border-bottom:1px solid #eeeeee;border-left:2px solid #eeeeee;border-top-left-radius:0px;border-top-right-radius:0px;border-bottom-right-radius:0px;border-bottom-left-radius:0px;background:#ffffff;color:#444444;padding-top:14px;padding-right:16px;padding-bottom:14px;padding-left:16px;}.kt-accordion-id19966_c42905-c3:not( .kt-accodion-icon-style-basiccircle ):not( .kt-accodion-icon-style-xclosecircle ):not( .kt-accodion-icon-style-arrowcircle )  > .kt-accordion-inner-wrap > .wp-block-kadence-pane > .kt-accordion-header-wrap .kt-blocks-accordion-icon-trigger:after, .kt-accordion-id19966_c42905-c3:not( .kt-accodion-icon-style-basiccircle ):not( .kt-accodion-icon-style-xclosecircle ):not( .kt-accodion-icon-style-arrowcircle )  > .kt-accordion-inner-wrap > .wp-block-kadence-pane > .kt-accordion-header-wrap .kt-blocks-accordion-icon-trigger:before{background:#444444;}.kt-accordion-id19966_c42905-c3:not( .kt-accodion-icon-style-basic ):not( .kt-accodion-icon-style-xclose ):not( .kt-accodion-icon-style-arrow ) .kt-blocks-accordion-icon-trigger{background:#444444;}.kt-accordion-id19966_c42905-c3:not( .kt-accodion-icon-style-basic ):not( .kt-accodion-icon-style-xclose ):not( .kt-accodion-icon-style-arrow ) .kt-blocks-accordion-icon-trigger:after, .kt-accordion-id19966_c42905-c3:not( .kt-accodion-icon-style-basic ):not( .kt-accodion-icon-style-xclose ):not( .kt-accodion-icon-style-arrow ) .kt-blocks-accordion-icon-trigger:before{background:#ffffff;}.kt-accordion-id19966_c42905-c3 > .kt-accordion-inner-wrap > .wp-block-kadence-pane > .kt-accordion-header-wrap > .kt-blocks-accordion-header:hover, \n\t\t\t\tbody:not(.hide-focus-outline) .kt-accordion-id19966_c42905-c3 .kt-blocks-accordion-header:focus-visible{color:#444444;background:#ffffff;border-top-color:#878787;border-right-color:#878787;border-bottom-color:#878787;border-left-color:#878787;}.kt-accordion-id19966_c42905-c3:not( .kt-accodion-icon-style-basiccircle ):not( .kt-accodion-icon-style-xclosecircle ):not( .kt-accodion-icon-style-arrowcircle ) .kt-accordion-header-wrap .kt-blocks-accordion-header:hover .kt-blocks-accordion-icon-trigger:after, .kt-accordion-id19966_c42905-c3:not( .kt-accodion-icon-style-basiccircle ):not( .kt-accodion-icon-style-xclosecircle ):not( .kt-accodion-icon-style-arrowcircle ) .kt-accordion-header-wrap .kt-blocks-accordion-header:hover .kt-blocks-accordion-icon-trigger:before, body:not(.hide-focus-outline) .kt-accordion-id19966_c42905-c3:not( .kt-accodion-icon-style-basiccircle ):not( .kt-accodion-icon-style-xclosecircle ):not( .kt-accodion-icon-style-arrowcircle ) .kt-blocks-accordion--visible .kt-blocks-accordion-icon-trigger:after, body:not(.hide-focus-outline) .kt-accordion-id19966_c42905-c3:not( .kt-accodion-icon-style-basiccircle ):not( .kt-accodion-icon-style-xclosecircle ):not( .kt-accodion-icon-style-arrowcircle ) .kt-blocks-accordion-header:focus-visible .kt-blocks-accordion-icon-trigger:before{background:#444444;}.kt-accordion-id19966_c42905-c3:not( .kt-accodion-icon-style-basic ):not( .kt-accodion-icon-style-xclose ):not( .kt-accodion-icon-style-arrow ) .kt-accordion-header-wrap .kt-blocks-accordion-header:hover .kt-blocks-accordion-icon-trigger, body:not(.hide-focus-outline) .kt-accordion-id19966_c42905-c3:not( .kt-accodion-icon-style-basic ):not( .kt-accodion-icon-style-xclose ):not( .kt-accodion-icon-style-arrow ) .kt-accordion-header-wrap .kt-blocks-accordion-header:focus-visible .kt-blocks-accordion-icon-trigger{background:#444444;}.kt-accordion-id19966_c42905-c3:not( .kt-accodion-icon-style-basic ):not( .kt-accodion-icon-style-xclose ):not( .kt-accodion-icon-style-arrow ) .kt-accordion-header-wrap .kt-blocks-accordion-header:hover .kt-blocks-accordion-icon-trigger:after, .kt-accordion-id19966_c42905-c3:not( .kt-accodion-icon-style-basic ):not( .kt-accodion-icon-style-xclose ):not( .kt-accodion-icon-style-arrow ) .kt-accordion-header-wrap .kt-blocks-accordion-header:hover .kt-blocks-accordion-icon-trigger:before, body:not(.hide-focus-outline) .kt-accordion-id19966_c42905-c3:not( .kt-accodion-icon-style-basic ):not( .kt-accodion-icon-style-xclose ):not( .kt-accodion-icon-style-arrow ) .kt-accordion-header-wrap .kt-blocks-accordion-header:focus-visible .kt-blocks-accordion-icon-trigger:after, body:not(.hide-focus-outline) .kt-accordion-id19966_c42905-c3:not( .kt-accodion-icon-style-basic ):not( .kt-accodion-icon-style-xclose ):not( .kt-accodion-icon-style-arrow ) .kt-accordion-header-wrap .kt-blocks-accordion-header:focus-visible .kt-blocks-accordion-icon-trigger:before{background:#ffffff;}.kt-accordion-id19966_c42905-c3 .kt-accordion-header-wrap .kt-blocks-accordion-header:focus-visible,\n\t\t\t\t.kt-accordion-id19966_c42905-c3 > .kt-accordion-inner-wrap > .wp-block-kadence-pane > .kt-accordion-header-wrap > .kt-blocks-accordion-header.kt-accordion-panel-active{color:#444444;background:#ffffff;border-top-color:#eeeeee;border-right-color:#eeeeee;border-bottom-color:#eeeeee;border-left-color:#0e9cd1;}.kt-accordion-id19966_c42905-c3:not( .kt-accodion-icon-style-basiccircle ):not( .kt-accodion-icon-style-xclosecircle ):not( .kt-accodion-icon-style-arrowcircle )  > .kt-accordion-inner-wrap > .wp-block-kadence-pane > .kt-accordion-header-wrap > .kt-blocks-accordion-header.kt-accordion-panel-active .kt-blocks-accordion-icon-trigger:after, .kt-accordion-id19966_c42905-c3:not( .kt-accodion-icon-style-basiccircle ):not( .kt-accodion-icon-style-xclosecircle ):not( .kt-accodion-icon-style-arrowcircle )  > .kt-accordion-inner-wrap > .wp-block-kadence-pane > .kt-accordion-header-wrap > .kt-blocks-accordion-header.kt-accordion-panel-active .kt-blocks-accordion-icon-trigger:before{background:#444444;}.kt-accordion-id19966_c42905-c3:not( .kt-accodion-icon-style-basic ):not( .kt-accodion-icon-style-xclose ):not( .kt-accodion-icon-style-arrow ) .kt-blocks-accordion-header.kt-accordion-panel-active .kt-blocks-accordion-icon-trigger{background:#444444;}.kt-accordion-id19966_c42905-c3:not( .kt-accodion-icon-style-basic ):not( .kt-accodion-icon-style-xclose ):not( .kt-accodion-icon-style-arrow ) .kt-blocks-accordion-header.kt-accordion-panel-active .kt-blocks-accordion-icon-trigger:after, .kt-accordion-id19966_c42905-c3:not( .kt-accodion-icon-style-basic ):not( .kt-accodion-icon-style-xclose ):not( .kt-accodion-icon-style-arrow ) .kt-blocks-accordion-header.kt-accordion-panel-active .kt-blocks-accordion-icon-trigger:before{background:#ffffff;}@media all and (max-width: 1024px){.kt-accordion-id19966_c42905-c3 .kt-accordion-panel-inner{border-top:0px solid transparent;border-right:1px solid transparent;border-bottom:1px solid transparent;border-left:1px solid transparent;}}@media all and (max-width: 1024px){.kt-accordion-id19966_c42905-c3 > .kt-accordion-inner-wrap > .wp-block-kadence-pane > .kt-accordion-header-wrap > .kt-blocks-accordion-header{border-top:1px solid #eeeeee;border-right:1px solid #eeeeee;border-bottom:1px solid #eeeeee;border-left:2px solid #eeeeee;}}@media all and (max-width: 1024px){.kt-accordion-id19966_c42905-c3 > .kt-accordion-inner-wrap > .wp-block-kadence-pane > .kt-accordion-header-wrap > .kt-blocks-accordion-header:hover, \n\t\t\t\tbody:not(.hide-focus-outline) .kt-accordion-id19966_c42905-c3 .kt-blocks-accordion-header:focus-visible{border-top-color:#878787;border-right-color:#878787;border-bottom-color:#878787;border-left-color:#878787;}}@media all and (max-width: 1024px){.kt-accordion-id19966_c42905-c3 .kt-accordion-header-wrap .kt-blocks-accordion-header:focus-visible,\n\t\t\t\t.kt-accordion-id19966_c42905-c3 > .kt-accordion-inner-wrap > .wp-block-kadence-pane > .kt-accordion-header-wrap > .kt-blocks-accordion-header.kt-accordion-panel-active{border-top-color:#eeeeee;border-right-color:#eeeeee;border-bottom-color:#eeeeee;border-left-color:#0e9cd1;}}@media all and (max-width: 767px){.kt-accordion-id19966_c42905-c3 .kt-accordion-inner-wrap{display:block;}.kt-accordion-id19966_c42905-c3 .kt-accordion-inner-wrap .kt-accordion-pane:not(:first-child){margin-top:10px;}.kt-accordion-id19966_c42905-c3 .kt-accordion-panel-inner{border-top:0px solid transparent;border-right:1px solid transparent;border-bottom:1px solid transparent;border-left:1px solid transparent;}.kt-accordion-id19966_c42905-c3 > .kt-accordion-inner-wrap > .wp-block-kadence-pane > .kt-accordion-header-wrap > .kt-blocks-accordion-header{border-top:1px solid #eeeeee;border-right:1px solid #eeeeee;border-bottom:1px solid #eeeeee;border-left:2px solid #eeeeee;}.kt-accordion-id19966_c42905-c3 > .kt-accordion-inner-wrap > .wp-block-kadence-pane > .kt-accordion-header-wrap > .kt-blocks-accordion-header:hover, \n\t\t\t\tbody:not(.hide-focus-outline) .kt-accordion-id19966_c42905-c3 .kt-blocks-accordion-header:focus-visible{border-top-color:#878787;border-right-color:#878787;border-bottom-color:#878787;border-left-color:#878787;}.kt-accordion-id19966_c42905-c3 .kt-accordion-header-wrap .kt-blocks-accordion-header:focus-visible,\n\t\t\t\t.kt-accordion-id19966_c42905-c3 > .kt-accordion-inner-wrap > .wp-block-kadence-pane > .kt-accordion-header-wrap > .kt-blocks-accordion-header.kt-accordion-panel-active{border-top-color:#eeeeee;border-right-color:#eeeeee;border-bottom-color:#eeeeee;border-left-color:#0e9cd1;}}<\/style>\n<div class=\"wp-block-kadence-accordion alignnone\"><div class=\"kt-accordion-wrap kt-accordion-id19966_c42905-c3 kt-accordion-has-2-panes kt-active-pane-0 kt-accordion-block kt-pane-header-alignment-left kt-accodion-icon-style-arrow kt-accodion-icon-side-right\" style=\"max-width:none\"><div class=\"kt-accordion-inner-wrap\" data-allow-multiple-open=\"false\" data-start-open=\"0\">\n<div class=\"wp-block-kadence-pane kt-accordion-pane kt-accordion-pane-1 kt-pane19966_8d0a18-41\"><h2 class=\"kt-accordion-header-wrap\"><button class=\"kt-blocks-accordion-header kt-acccordion-button-label-show\" type=\"button\"><span class=\"kt-blocks-accordion-title-wrap\"><span class=\"kt-blocks-accordion-title\"><strong>Can multiplex assays reduce sample consumption?<\/strong><\/span><\/span><span class=\"kt-blocks-accordion-icon-trigger\"><\/span><\/button><\/h2><div class=\"kt-accordion-panel kt-accordion-panel-hidden\"><div class=\"kt-accordion-panel-inner\">\n<p class=\"wp-block-paragraph\">Yes. Multiplex qPCR allows multiple targets to be analyzed within a single reaction, reducing sample requirements and conserving valuable biological material. This is especially beneficial when sample quantity is limited or when multiple biomarkers must be evaluated simultaneously.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><a href=\"https:\/\/www.thermofisher.com\/us\/en\/home\/life-science\/pcr\/real-time-pcr\/taqman-multiplex-qpcr-solution.html\" target=\"_blank\" rel=\"noreferrer noopener\">Learn more about multiplex workflow advantages<\/a><\/p>\n<\/div><\/div><\/div>\n\n\n\n<div class=\"wp-block-kadence-pane kt-accordion-pane kt-accordion-pane-2 kt-pane19966_29a60a-90\"><h2 class=\"kt-accordion-header-wrap\"><button class=\"kt-blocks-accordion-header kt-acccordion-button-label-show\" type=\"button\"><span class=\"kt-blocks-accordion-title-wrap\"><span class=\"kt-blocks-accordion-title\"><strong>What are the advantages of multiplex qPCR compared with singleplex qPCR?<\/strong><\/span><\/span><span class=\"kt-blocks-accordion-icon-trigger\"><\/span><\/button><\/h2><div class=\"kt-accordion-panel kt-accordion-panel-hidden\"><div class=\"kt-accordion-panel-inner\">\n<p class=\"wp-block-paragraph\">Multiplex qPCR offers several advantages over singleplex workflows, including higher throughput, reduced sample consumption, lower reagent usage, improved workflow efficiency, and the ability to incorporate internal controls within the same reaction. These benefits make multiplex qPCR particularly valuable when sample quantity is limited or when researchers need to analyze multiple targets simultaneously. For practical examples and multiplex design considerations, see <a href=\"https:\/\/www.thermofisher.com\/blog\/behindthebench\/multiplex-qpcr-basics-guide\/\" target=\"_blank\" rel=\"noreferrer noopener\">Thermo Fisher&#8217;s Multiplex qPCR Basics Guide<\/a>.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><a href=\"https:\/\/www.thermofisher.com\/blog\/behindthebench\/multiplex-qpcr-basics-guide\/\" target=\"_blank\" rel=\"noreferrer noopener\">See multiplex qPCR best practices<\/a><\/p>\n<\/div><\/div><\/div>\n<\/div><\/div><\/div>\n\n\n\n<p class=\"has-small-font-size wp-block-paragraph\"><br>For Research Use Only. Not for use in diagnostic procedures.<br><br>\u00a9 2026 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries unless otherwise specified.<\/p>\n","protected":false},"excerpt":{"rendered":"<p>Multiplex qPCR has transformed molecular research by enabling scientists to detect multiple targets in a single reaction. The benefits are clear: reduced sample consumption, lower costs, faster workflows, and more data from every experiment. However, designing successful multiplex assays isn&#8217;t always straightforward. Traditionally, identifying problematic assay combinations required extensive wet-lab optimization and iterative testing. Today,<\/p>\n","protected":false},"author":1793,"featured_media":19967,"comment_status":"open","ping_status":"closed","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_kad_blocks_custom_css":"","_kad_blocks_head_custom_js":"","_kad_blocks_body_custom_js":"","_kad_blocks_footer_custom_js":"","_monsterinsights_skip_tracking":false,"_genesis_hide_title":false,"_genesis_hide_breadcrumbs":false,"_genesis_hide_singular_image":false,"_genesis_hide_footer_widgets":false,"_genesis_custom_body_class":"","_genesis_custom_post_class":"","_genesis_layout":"","_jetpack_newsletter_access":"","_jetpack_dont_email_post_to_subs":false,"_jetpack_newsletter_tier_id":0,"_jetpack_memberships_contains_paywalled_content":false,"_jetpack_memberships_contains_paid_content":false,"footnotes":""},"categories":[108,2259,101],"tags":[2729,2723,2726,2722,2728,2610,2724,2725,2721,2727],"division":[2575],"class_list":["post-19966","post","type-post","status-publish","format-standard","has-post-thumbnail","category-lab-tips-and-tricks","category-molecular-diagnostics","category-product-news","tag-assay-compatibility","tag-assay-interactions","tag-in-silico-screening","tag-multiplex-assay-optimization","tag-multiplex-panel-development","tag-multiplex-qpcr","tag-primer-dimer","tag-primer-probe-interactions","tag-qpcr-assay-design","tag-qpcr-optimization","division-gsd","entry"],"_selected_authors":[1955],"_selected_reviewers":[],"acf":[],"yoast_head":"<!-- This site is optimized with the Yoast SEO Premium plugin v27.8 (Yoast SEO v27.8) - https:\/\/yoast.com\/product\/yoast-seo-premium-wordpress\/ -->\n<title>Multiplex Assay Optimization for qPCR Panels<\/title>\n<meta name=\"description\" content=\"Learn how to identify multiplex qPCR assay interactions early using in silico screening to improve sensitivity and streamline panel development.\" \/>\n<meta name=\"robots\" content=\"index, follow, max-snippet:-1, max-image-preview:large, max-video-preview:-1\" \/>\n<link rel=\"canonical\" href=\"https:\/\/www.thermofisher.com\/blog\/behindthebench\/multiplex-assay-optimization-qpcr\/\" \/>\n<meta property=\"og:locale\" content=\"en_US\" \/>\n<meta property=\"og:type\" content=\"article\" \/>\n<meta property=\"og:title\" content=\"Building Better Multiplex Panels: Identifying Assay Interactions Early\" \/>\n<meta property=\"og:description\" content=\"Learn how to identify multiplex qPCR assay interactions early using in silico screening to improve sensitivity and streamline panel development.\" \/>\n<meta property=\"og:url\" content=\"https:\/\/www.thermofisher.com\/blog\/behindthebench\/multiplex-assay-optimization-qpcr\/\" \/>\n<meta property=\"og:site_name\" content=\"Behind the Bench\" \/>\n<meta property=\"article:publisher\" content=\"https:\/\/www.facebook.com\/thermofisher\" \/>\n<meta property=\"article:published_time\" content=\"2026-09-08T17:15:17+00:00\" \/>\n<meta property=\"article:modified_time\" content=\"2026-09-10T20:09:48+00:00\" \/>\n<meta property=\"og:image\" content=\"http:\/\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/09\/Header-banner-scaled.jpg\" \/>\n\t<meta property=\"og:image:width\" content=\"2560\" \/>\n\t<meta property=\"og:image:height\" content=\"507\" \/>\n\t<meta property=\"og:image:type\" content=\"image\/jpeg\" \/>\n<meta name=\"author\" content=\"Ashleigh Barlow\" \/>\n<meta name=\"twitter:card\" content=\"summary_large_image\" \/>\n<meta name=\"twitter:creator\" content=\"@thermofisher\" \/>\n<meta name=\"twitter:site\" content=\"@thermofisher\" \/>\n<meta name=\"twitter:label1\" content=\"Written by\" \/>\n\t<meta name=\"twitter:data1\" content=\"Ashleigh Barlow\" \/>\n\t<meta name=\"twitter:label2\" content=\"Est. reading time\" \/>\n\t<meta name=\"twitter:data2\" content=\"6 minutes\" \/>\n<script type=\"application\/ld+json\" class=\"yoast-schema-graph\">{\"@context\":\"https:\\\/\\\/schema.org\",\"@graph\":[{\"@type\":\"Article\",\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/behindthebench\\\/multiplex-assay-optimization-qpcr\\\/#article\",\"isPartOf\":{\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/behindthebench\\\/multiplex-assay-optimization-qpcr\\\/\"},\"author\":{\"name\":\"Ashleigh Barlow\",\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/behindthebench\\\/#\\\/schema\\\/person\\\/9cd6fa1f2eca52a1cab4a12601fc43ee\"},\"headline\":\"Building Better Multiplex Panels: Identifying Assay Interactions Early\",\"datePublished\":\"2026-09-08T17:15:17+00:00\",\"dateModified\":\"2026-09-10T20:09:48+00:00\",\"mainEntityOfPage\":{\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/behindthebench\\\/multiplex-assay-optimization-qpcr\\\/\"},\"wordCount\":1169,\"commentCount\":0,\"image\":{\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/behindthebench\\\/multiplex-assay-optimization-qpcr\\\/#primaryimage\"},\"thumbnailUrl\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/behindthebench\\\/wp-content\\\/uploads\\\/sites\\\/9\\\/2026\\\/09\\\/Header-banner-scaled.jpg\",\"keywords\":[\"assay compatibility\",\"assay interactions\",\"in silico screening\",\"multiplex assay optimization\",\"multiplex panel development\",\"Multiplex qPCR\",\"primer dimer\",\"primer probe interactions\",\"qPCR assay design\",\"qPCR optimization\"],\"articleSection\":[\"Lab Tips and Tricks\",\"Molecular Diagnostics\",\"Product News\"],\"inLanguage\":\"en-US\",\"potentialAction\":[{\"@type\":\"CommentAction\",\"name\":\"Comment\",\"target\":[\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/behindthebench\\\/multiplex-assay-optimization-qpcr\\\/#respond\"]}]},{\"@type\":\"WebPage\",\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/behindthebench\\\/multiplex-assay-optimization-qpcr\\\/\",\"url\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/behindthebench\\\/multiplex-assay-optimization-qpcr\\\/\",\"name\":\"Multiplex Assay Optimization for qPCR Panels\",\"isPartOf\":{\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/behindthebench\\\/#website\"},\"primaryImageOfPage\":{\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/behindthebench\\\/multiplex-assay-optimization-qpcr\\\/#primaryimage\"},\"image\":{\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/behindthebench\\\/multiplex-assay-optimization-qpcr\\\/#primaryimage\"},\"thumbnailUrl\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/behindthebench\\\/wp-content\\\/uploads\\\/sites\\\/9\\\/2026\\\/09\\\/Header-banner-scaled.jpg\",\"datePublished\":\"2026-09-08T17:15:17+00:00\",\"dateModified\":\"2026-09-10T20:09:48+00:00\",\"author\":{\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/behindthebench\\\/#\\\/schema\\\/person\\\/9cd6fa1f2eca52a1cab4a12601fc43ee\"},\"description\":\"Learn how to identify multiplex qPCR assay interactions early using in silico screening to improve sensitivity and streamline panel development.\",\"breadcrumb\":{\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/behindthebench\\\/multiplex-assay-optimization-qpcr\\\/#breadcrumb\"},\"inLanguage\":\"en-US\",\"potentialAction\":[{\"@type\":\"ReadAction\",\"target\":[\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/behindthebench\\\/multiplex-assay-optimization-qpcr\\\/\"]}]},{\"@type\":\"ImageObject\",\"inLanguage\":\"en-US\",\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/behindthebench\\\/multiplex-assay-optimization-qpcr\\\/#primaryimage\",\"url\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/behindthebench\\\/wp-content\\\/uploads\\\/sites\\\/9\\\/2026\\\/09\\\/Header-banner-scaled.jpg\",\"contentUrl\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/behindthebench\\\/wp-content\\\/uploads\\\/sites\\\/9\\\/2026\\\/09\\\/Header-banner-scaled.jpg\",\"width\":2560,\"height\":507,\"caption\":\"Successful multiplex qPCR depends on compatible assays that maintain sensitivity, specificity, and reproducibility as panel complexity increases.\"},{\"@type\":\"BreadcrumbList\",\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/behindthebench\\\/multiplex-assay-optimization-qpcr\\\/#breadcrumb\",\"itemListElement\":[{\"@type\":\"ListItem\",\"position\":1,\"name\":\"Home\",\"item\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/behindthebench\\\/\"},{\"@type\":\"ListItem\",\"position\":2,\"name\":\"Building Better Multiplex Panels: Identifying Assay Interactions Early\"}]},{\"@type\":\"WebSite\",\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/behindthebench\\\/#website\",\"url\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/behindthebench\\\/\",\"name\":\"Behind the Bench\",\"description\":\"\",\"potentialAction\":[{\"@type\":\"SearchAction\",\"target\":{\"@type\":\"EntryPoint\",\"urlTemplate\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/behindthebench\\\/?s={search_term_string}\"},\"query-input\":{\"@type\":\"PropertyValueSpecification\",\"valueRequired\":true,\"valueName\":\"search_term_string\"}}],\"inLanguage\":\"en-US\"},{\"@type\":\"Person\",\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/behindthebench\\\/#\\\/schema\\\/person\\\/9cd6fa1f2eca52a1cab4a12601fc43ee\",\"name\":\"Ashleigh Barlow\",\"image\":{\"@type\":\"ImageObject\",\"inLanguage\":\"en-US\",\"@id\":\"https:\\\/\\\/secure.gravatar.com\\\/avatar\\\/ffa9b05d0f35cf4c6041578d6c5a047634604e604efa2286b297cae04e97512a?s=96&d=mm&r=g\",\"url\":\"https:\\\/\\\/secure.gravatar.com\\\/avatar\\\/ffa9b05d0f35cf4c6041578d6c5a047634604e604efa2286b297cae04e97512a?s=96&d=mm&r=g\",\"contentUrl\":\"https:\\\/\\\/secure.gravatar.com\\\/avatar\\\/ffa9b05d0f35cf4c6041578d6c5a047634604e604efa2286b297cae04e97512a?s=96&d=mm&r=g\",\"caption\":\"Ashleigh Barlow\"},\"url\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/author\\\/ashleighbarlow\\\/\"}]}<\/script>\n<!-- \/ Yoast SEO Premium plugin. -->","yoast_head_json":{"title":"Multiplex Assay Optimization for qPCR Panels","description":"Learn how to identify multiplex qPCR assay interactions early using in silico screening to improve sensitivity and streamline panel development.","robots":{"index":"index","follow":"follow","max-snippet":"max-snippet:-1","max-image-preview":"max-image-preview:large","max-video-preview":"max-video-preview:-1"},"canonical":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/multiplex-assay-optimization-qpcr\/","og_locale":"en_US","og_type":"article","og_title":"Building Better Multiplex Panels: Identifying Assay Interactions Early","og_description":"Learn how to identify multiplex qPCR assay interactions early using in silico screening to improve sensitivity and streamline panel development.","og_url":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/multiplex-assay-optimization-qpcr\/","og_site_name":"Behind the Bench","article_publisher":"https:\/\/www.facebook.com\/thermofisher","article_published_time":"2026-09-08T17:15:17+00:00","article_modified_time":"2026-09-10T20:09:48+00:00","og_image":[{"width":2560,"height":507,"url":"http:\/\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/09\/Header-banner-scaled.jpg","type":"image\/jpeg"}],"author":"Ashleigh Barlow","twitter_card":"summary_large_image","twitter_creator":"@thermofisher","twitter_site":"@thermofisher","twitter_misc":{"Written by":"Ashleigh Barlow","Est. reading time":"6 minutes"},"schema":{"@context":"https:\/\/schema.org","@graph":[{"@type":"Article","@id":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/multiplex-assay-optimization-qpcr\/#article","isPartOf":{"@id":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/multiplex-assay-optimization-qpcr\/"},"author":{"name":"Ashleigh Barlow","@id":"https:\/\/admin.acceleratingscience.com\/behindthebench\/#\/schema\/person\/9cd6fa1f2eca52a1cab4a12601fc43ee"},"headline":"Building Better Multiplex Panels: Identifying Assay Interactions Early","datePublished":"2026-09-08T17:15:17+00:00","dateModified":"2026-09-10T20:09:48+00:00","mainEntityOfPage":{"@id":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/multiplex-assay-optimization-qpcr\/"},"wordCount":1169,"commentCount":0,"image":{"@id":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/multiplex-assay-optimization-qpcr\/#primaryimage"},"thumbnailUrl":"https:\/\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/09\/Header-banner-scaled.jpg","keywords":["assay compatibility","assay interactions","in silico screening","multiplex assay optimization","multiplex panel development","Multiplex qPCR","primer dimer","primer probe interactions","qPCR assay design","qPCR optimization"],"articleSection":["Lab Tips and Tricks","Molecular Diagnostics","Product News"],"inLanguage":"en-US","potentialAction":[{"@type":"CommentAction","name":"Comment","target":["https:\/\/www.thermofisher.com\/blog\/behindthebench\/multiplex-assay-optimization-qpcr\/#respond"]}]},{"@type":"WebPage","@id":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/multiplex-assay-optimization-qpcr\/","url":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/multiplex-assay-optimization-qpcr\/","name":"Multiplex Assay Optimization for qPCR Panels","isPartOf":{"@id":"https:\/\/admin.acceleratingscience.com\/behindthebench\/#website"},"primaryImageOfPage":{"@id":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/multiplex-assay-optimization-qpcr\/#primaryimage"},"image":{"@id":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/multiplex-assay-optimization-qpcr\/#primaryimage"},"thumbnailUrl":"https:\/\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/09\/Header-banner-scaled.jpg","datePublished":"2026-09-08T17:15:17+00:00","dateModified":"2026-09-10T20:09:48+00:00","author":{"@id":"https:\/\/admin.acceleratingscience.com\/behindthebench\/#\/schema\/person\/9cd6fa1f2eca52a1cab4a12601fc43ee"},"description":"Learn how to identify multiplex qPCR assay interactions early using in silico screening to improve sensitivity and streamline panel development.","breadcrumb":{"@id":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/multiplex-assay-optimization-qpcr\/#breadcrumb"},"inLanguage":"en-US","potentialAction":[{"@type":"ReadAction","target":["https:\/\/www.thermofisher.com\/blog\/behindthebench\/multiplex-assay-optimization-qpcr\/"]}]},{"@type":"ImageObject","inLanguage":"en-US","@id":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/multiplex-assay-optimization-qpcr\/#primaryimage","url":"https:\/\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/09\/Header-banner-scaled.jpg","contentUrl":"https:\/\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/09\/Header-banner-scaled.jpg","width":2560,"height":507,"caption":"Successful multiplex qPCR depends on compatible assays that maintain sensitivity, specificity, and reproducibility as panel complexity increases."},{"@type":"BreadcrumbList","@id":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/multiplex-assay-optimization-qpcr\/#breadcrumb","itemListElement":[{"@type":"ListItem","position":1,"name":"Home","item":"https:\/\/admin.acceleratingscience.com\/behindthebench\/"},{"@type":"ListItem","position":2,"name":"Building Better Multiplex Panels: Identifying Assay Interactions Early"}]},{"@type":"WebSite","@id":"https:\/\/admin.acceleratingscience.com\/behindthebench\/#website","url":"https:\/\/admin.acceleratingscience.com\/behindthebench\/","name":"Behind the Bench","description":"","potentialAction":[{"@type":"SearchAction","target":{"@type":"EntryPoint","urlTemplate":"https:\/\/admin.acceleratingscience.com\/behindthebench\/?s={search_term_string}"},"query-input":{"@type":"PropertyValueSpecification","valueRequired":true,"valueName":"search_term_string"}}],"inLanguage":"en-US"},{"@type":"Person","@id":"https:\/\/admin.acceleratingscience.com\/behindthebench\/#\/schema\/person\/9cd6fa1f2eca52a1cab4a12601fc43ee","name":"Ashleigh Barlow","image":{"@type":"ImageObject","inLanguage":"en-US","@id":"https:\/\/secure.gravatar.com\/avatar\/ffa9b05d0f35cf4c6041578d6c5a047634604e604efa2286b297cae04e97512a?s=96&d=mm&r=g","url":"https:\/\/secure.gravatar.com\/avatar\/ffa9b05d0f35cf4c6041578d6c5a047634604e604efa2286b297cae04e97512a?s=96&d=mm&r=g","contentUrl":"https:\/\/secure.gravatar.com\/avatar\/ffa9b05d0f35cf4c6041578d6c5a047634604e604efa2286b297cae04e97512a?s=96&d=mm&r=g","caption":"Ashleigh Barlow"},"url":"https:\/\/admin.acceleratingscience.com\/author\/ashleighbarlow\/"}]}},"taxonomy_info":{"category":[{"value":108,"label":"Lab Tips and Tricks"},{"value":2259,"label":"Molecular Diagnostics"},{"value":101,"label":"Product News"}],"post_tag":[{"value":2729,"label":"assay compatibility"},{"value":2723,"label":"assay interactions"},{"value":2726,"label":"in silico screening"},{"value":2722,"label":"multiplex assay optimization"},{"value":2728,"label":"multiplex panel development"},{"value":2610,"label":"Multiplex qPCR"},{"value":2724,"label":"primer dimer"},{"value":2725,"label":"primer probe interactions"},{"value":2721,"label":"qPCR assay design"},{"value":2727,"label":"qPCR optimization"}],"division":[{"value":2575,"label":"GSD"}]},"featured_image_src_large":["https:\/\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/09\/Header-banner-1024x203.jpg",760,151,true],"author_info":{"display_name":"Ashleigh Barlow","author_link":"https:\/\/admin.acceleratingscience.com\/author\/ashleighbarlow\/"},"comment_info":3,"category_info":[{"term_id":108,"name":"Lab Tips and Tricks","slug":"lab-tips-and-tricks","term_group":0,"term_taxonomy_id":108,"taxonomy":"category","description":"","parent":0,"count":102,"filter":"raw","meta":[],"cat_ID":108,"category_count":102,"category_description":"","cat_name":"Lab Tips and Tricks","category_nicename":"lab-tips-and-tricks","category_parent":0},{"term_id":2259,"name":"Molecular Diagnostics","slug":"molecular-diagnostics","term_group":0,"term_taxonomy_id":2259,"taxonomy":"category","description":"","parent":0,"count":19,"filter":"raw","meta":[],"cat_ID":2259,"category_count":19,"category_description":"","cat_name":"Molecular Diagnostics","category_nicename":"molecular-diagnostics","category_parent":0},{"term_id":101,"name":"Product News","slug":"product-news","term_group":0,"term_taxonomy_id":101,"taxonomy":"category","description":"","parent":0,"count":105,"filter":"raw","meta":[],"cat_ID":101,"category_count":105,"category_description":"","cat_name":"Product News","category_nicename":"product-news","category_parent":0}],"tag_info":[{"term_id":2729,"name":"assay compatibility","slug":"assay-compatibility","term_group":0,"term_taxonomy_id":2729,"taxonomy":"post_tag","description":"","parent":0,"count":1,"filter":"raw","meta":[]},{"term_id":2723,"name":"assay interactions","slug":"assay-interactions","term_group":0,"term_taxonomy_id":2723,"taxonomy":"post_tag","description":"","parent":0,"count":1,"filter":"raw","meta":[]},{"term_id":2726,"name":"in silico screening","slug":"in-silico-screening","term_group":0,"term_taxonomy_id":2726,"taxonomy":"post_tag","description":"","parent":0,"count":1,"filter":"raw","meta":[]},{"term_id":2722,"name":"multiplex assay optimization","slug":"multiplex-assay-optimization","term_group":0,"term_taxonomy_id":2722,"taxonomy":"post_tag","description":"","parent":0,"count":1,"filter":"raw","meta":[]},{"term_id":2728,"name":"multiplex panel development","slug":"multiplex-panel-development","term_group":0,"term_taxonomy_id":2728,"taxonomy":"post_tag","description":"","parent":0,"count":1,"filter":"raw","meta":[]},{"term_id":2610,"name":"Multiplex qPCR","slug":"multiplex-qpcr","term_group":0,"term_taxonomy_id":2610,"taxonomy":"post_tag","description":"","parent":0,"count":2,"filter":"raw","meta":[]},{"term_id":2724,"name":"primer dimer","slug":"primer-dimer","term_group":0,"term_taxonomy_id":2724,"taxonomy":"post_tag","description":"","parent":0,"count":1,"filter":"raw","meta":[]},{"term_id":2725,"name":"primer probe interactions","slug":"primer-probe-interactions","term_group":0,"term_taxonomy_id":2725,"taxonomy":"post_tag","description":"","parent":0,"count":1,"filter":"raw","meta":[]},{"term_id":2721,"name":"qPCR assay design","slug":"qpcr-assay-design","term_group":0,"term_taxonomy_id":2721,"taxonomy":"post_tag","description":"","parent":0,"count":2,"filter":"raw","meta":[]},{"term_id":2727,"name":"qPCR optimization","slug":"qpcr-optimization","term_group":0,"term_taxonomy_id":2727,"taxonomy":"post_tag","description":"","parent":0,"count":1,"filter":"raw","meta":[]}],"jetpack-related-posts":[{"id":19685,"url":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/multiplex-qpcr-primer-optimization\/","url_meta":{"origin":19966,"position":0},"title":"Multiplex qPCR primer optimization: How to balance amplification across targets","author":"Ashleigh Barlow","date":"May 1, 2026","format":false,"excerpt":"What happens when one target amplifies more efficiently than another in a multiplex qPCR reaction? In many experiments, targets are not present at equal abundance. Some amplify earlier, generate stronger signals, and begin to dominate the reaction. Others may appear delayed, or not at all. These effects are not always\u2026","rel":"","context":"In &quot;Lab Tips and Tricks&quot;","block_context":{"text":"Lab Tips and Tricks","link":"https:\/\/admin.acceleratingscience.com\/behindthebench\/lab-tips-and-tricks\/"},"img":{"alt_text":"DNA strands with multicolored signals representing simultaneous detection of multiple targets in multiplex qPCR.","src":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/05\/Blog_qPCR_Reagent_multiplex_qPCR_primerlimitations-1-Multiplex-qPCR_hero.jpeg?resize=350%2C200&ssl=1","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/05\/Blog_qPCR_Reagent_multiplex_qPCR_primerlimitations-1-Multiplex-qPCR_hero.jpeg?resize=350%2C200&ssl=1 1x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/05\/Blog_qPCR_Reagent_multiplex_qPCR_primerlimitations-1-Multiplex-qPCR_hero.jpeg?resize=525%2C300&ssl=1 1.5x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/05\/Blog_qPCR_Reagent_multiplex_qPCR_primerlimitations-1-Multiplex-qPCR_hero.jpeg?resize=700%2C400&ssl=1 2x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/05\/Blog_qPCR_Reagent_multiplex_qPCR_primerlimitations-1-Multiplex-qPCR_hero.jpeg?resize=1050%2C600&ssl=1 3x"},"classes":[]},{"id":19606,"url":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/multiplex-qpcr-optimization-6-plex\/","url_meta":{"origin":19966,"position":1},"title":"Optimizing multiplex qPCR: How to amplify up to 6 targets without compromising analytical sensitivity","author":"Behind The Bench Staff","date":"April 1, 2026","format":false,"excerpt":"What happens when a multiplex qPCR assay begins to lose performance as additional targets are introduced? An assay that performs reproducibly in singleplex can exhibit Ct shift, reduced sensitivity, or compressed dynamic range when scaled to 5- or 6-plex. As plex level increases, competition among primers, probes, reporter dyes, and\u2026","rel":"","context":"In &quot;Cancer Research&quot;","block_context":{"text":"Cancer Research","link":"https:\/\/admin.acceleratingscience.com\/behindthebench\/cancer-research\/"},"img":{"alt_text":"Graph comparing Ct values between singleplex and multiplex qPCR reactions showing \u0394Ct shift as targets increase.","src":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/04\/mulitplexing-image.png?resize=350%2C200&ssl=1","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/04\/mulitplexing-image.png?resize=350%2C200&ssl=1 1x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/04\/mulitplexing-image.png?resize=525%2C300&ssl=1 1.5x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/04\/mulitplexing-image.png?resize=700%2C400&ssl=1 2x"},"classes":[]},{"id":19271,"url":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/multiplex-qpcr-basics-guide\/","url_meta":{"origin":19966,"position":2},"title":"Multiplex vs. Singleplex qPCR: What do I need to know?","author":"Ashleigh Barlow","date":"December 3, 2025","format":false,"excerpt":"Real-time PCR (qPCR) has transformed how scientists detect and quantify nucleic acid targets, offering analytical sensitivity, speed, and scalability. Many real-time PCR reactions measure a single target per well \u2014 a setup called singleplex qPCR. However, the effectiveness of real-time PCR can be increased by an approach called multiplex qPCR.\u2026","rel":"","context":"In &quot;General&quot;","block_context":{"text":"General","link":"https:\/\/admin.acceleratingscience.com\/behindthebench\/general\/"},"img":{"alt_text":"","src":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2025\/12\/Web-header-900-x-600-custom-web-page-size-feature-pods.png?resize=350%2C200&ssl=1","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2025\/12\/Web-header-900-x-600-custom-web-page-size-feature-pods.png?resize=350%2C200&ssl=1 1x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2025\/12\/Web-header-900-x-600-custom-web-page-size-feature-pods.png?resize=525%2C300&ssl=1 1.5x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2025\/12\/Web-header-900-x-600-custom-web-page-size-feature-pods.png?resize=700%2C400&ssl=1 2x"},"classes":[]},{"id":19461,"url":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/grapevine-downy-mildew-fungicide-resistance-detection\/","url_meta":{"origin":19966,"position":3},"title":"Tracking grape pathogens from air to vine \u2013 a conversation with Lexi Heger","author":"Behind The Bench Staff","date":"February 28, 2026","format":false,"excerpt":"On the Absolute Gene-ius podcast, Lexi Heger, PhD candidate in Plant Pathology and Molecular Plant Sciences at Michigan State University, explores the molecular detection of grapevine pathogens and how qPCR and digital PCR (dPCR) are transforming plant disease research. Her research centers on downy mildew of grape (Plasmopara viticola) and\u2026","rel":"","context":"In &quot;Animal Health and AgriGenomics&quot;","block_context":{"text":"Animal Health and AgriGenomics","link":"https:\/\/admin.acceleratingscience.com\/behindthebench\/animal-health-and-agrigenomics\/"},"img":{"alt_text":"Portrait of a smiling plant pathology researcher standing outdoors beside grapevine leaves.","src":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/02\/Heger-headshot-1_Jordan-Ruggieri-scaled.jpeg?resize=350%2C200&ssl=1","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/02\/Heger-headshot-1_Jordan-Ruggieri-scaled.jpeg?resize=350%2C200&ssl=1 1x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/02\/Heger-headshot-1_Jordan-Ruggieri-scaled.jpeg?resize=525%2C300&ssl=1 1.5x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/02\/Heger-headshot-1_Jordan-Ruggieri-scaled.jpeg?resize=700%2C400&ssl=1 2x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/02\/Heger-headshot-1_Jordan-Ruggieri-scaled.jpeg?resize=1050%2C600&ssl=1 3x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/02\/Heger-headshot-1_Jordan-Ruggieri-scaled.jpeg?resize=1400%2C800&ssl=1 4x"},"classes":[]},{"id":19860,"url":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/choose-sybr-green-qpcr-master-mix\/","url_meta":{"origin":19966,"position":4},"title":"Choosing the Right SYBR Green Master Mix for Your Research and Budget","author":"Behind The Bench Staff","date":"July 28, 2026","format":false,"excerpt":"Every qPCR experiment begins with a series of decisions. How many targets should you analyze? How many samples and replicates are enough? Should you expand your sample cohort or reserve budget for downstream analytical validation studies? The answers depend not only on the scientific question, but also on the resources\u2026","rel":"","context":"In &quot;Lab Tips and Tricks&quot;","block_context":{"text":"Lab Tips and Tricks","link":"https:\/\/admin.acceleratingscience.com\/behindthebench\/lab-tips-and-tricks\/"},"img":{"alt_text":"Research scientist preparing samples for a real-time PCR experiment in a molecular biology laboratory.","src":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/07\/Blog-1280-x-720_Michealla-Scott.jpeg?resize=350%2C200&ssl=1","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/07\/Blog-1280-x-720_Michealla-Scott.jpeg?resize=350%2C200&ssl=1 1x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/07\/Blog-1280-x-720_Michealla-Scott.jpeg?resize=525%2C300&ssl=1 1.5x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/07\/Blog-1280-x-720_Michealla-Scott.jpeg?resize=700%2C400&ssl=1 2x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/07\/Blog-1280-x-720_Michealla-Scott.jpeg?resize=1050%2C600&ssl=1 3x"},"classes":[]},{"id":19748,"url":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/andes-virus-rt-qpcr-detection-workflow\/","url_meta":{"origin":19966,"position":5},"title":"Responding to Hantavirus: Advancing Rapid Molecular Detection","author":"Behind The Bench Staff","date":"May 14, 2026","format":false,"excerpt":"Article Summary Laboratories responding to the emerging 2026 Andes hantavirus incident need rapid, scalable molecular workflows that can integrate into existing RT-qPCR infrastructure. This workflow is best suited for public health laboratories, surveillance programs, academic research labs, and high-throughput testing facilities seeking rapid research-use-only (RUO) molecular detection capabilities. By combining\u2026","rel":"","context":"In &quot;Infectious Diseases&quot;","block_context":{"text":"Infectious Diseases","link":"https:\/\/admin.acceleratingscience.com\/behindthebench\/infectious-diseases\/"},"img":{"alt_text":"Illustration of hantavirus transmission showing rodent-derived viral particles entering the human respiratory system.","src":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/05\/Hantavirus-3-1_Geoffrey-Cassell.png?resize=350%2C200&ssl=1","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/05\/Hantavirus-3-1_Geoffrey-Cassell.png?resize=350%2C200&ssl=1 1x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/05\/Hantavirus-3-1_Geoffrey-Cassell.png?resize=525%2C300&ssl=1 1.5x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/05\/Hantavirus-3-1_Geoffrey-Cassell.png?resize=700%2C400&ssl=1 2x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/05\/Hantavirus-3-1_Geoffrey-Cassell.png?resize=1050%2C600&ssl=1 3x"},"classes":[]}],"jetpack_sharing_enabled":true,"jetpack_featured_media_url":"https:\/\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/09\/Header-banner-scaled.jpg","_links":{"self":[{"href":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/wp-json\/wp\/v2\/posts\/19966","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/wp-json\/wp\/v2\/users\/1793"}],"replies":[{"embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/wp-json\/wp\/v2\/comments?post=19966"}],"version-history":[{"count":0,"href":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/wp-json\/wp\/v2\/posts\/19966\/revisions"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/wp-json\/wp\/v2\/media\/19967"}],"wp:attachment":[{"href":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/wp-json\/wp\/v2\/media?parent=19966"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/wp-json\/wp\/v2\/categories?post=19966"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/wp-json\/wp\/v2\/tags?post=19966"},{"taxonomy":"division","embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/wp-json\/wp\/v2\/division?post=19966"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}