{"id":8197,"date":"2016-04-04T18:28:53","date_gmt":"2016-04-04T22:28:53","guid":{"rendered":"http:\/\/admin.acceleratingscience.com\/?p=8197"},"modified":"2022-08-18T20:29:12","modified_gmt":"2022-08-18T20:29:12","slug":"hands-on-with-digital-pcr","status":"publish","type":"post","link":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/hands-on-with-digital-pcr\/","title":{"rendered":"Hands-on With Digital PCR"},"content":{"rendered":"<p><a href=\"http:\/\/www.thermofisher.com\/uk\/en\/home\/life-science\/pcr\/digital-pcr.html?socid=social_btb_genequant\" target=\"_blank\" rel=\"noopener noreferrer\">Digital PCR<\/a>\u00a0is a relatively new approach to nucleic acid detection and quantification that offers an alternate method to conventional real-time quantitative PCR for absolute quantification and rare allele detection. It all works by partitioning a sample of DNA or cDNA into many individual, parallel PCR reactions; some of these reactions contain the target molecule (positive) while others do not (negative). A single molecule can be amplified a million-fold or more. During amplification,\u00a0<a href=\"http:\/\/www.thermofisher.com\/uk\/en\/home\/life-science\/pcr\/real-time-pcr\/real-time-pcr-assays.html?socid=social_btb_genequant\" target=\"_blank\" rel=\"noopener noreferrer\">TaqMan chemistry<\/a>\u00a0with dye-labeled probes is used to detect sequence-specific targets. When no target sequence is present, no signal is emitted. Following PCR analysis, the fraction of negative reactions is used to generate an absolute count of the number of target molecules in the sample, without the need for standards or endogenous controls.<\/p>\n<p><iframe loading=\"lazy\" width=\"760\" height=\"428\" src=\"https:\/\/www.youtube.com\/embed\/PjgwDhN63Zc?feature=oembed\" frameborder=\"0\" allow=\"accelerometer; autoplay; encrypted-media; gyroscope; picture-in-picture\" allowfullscreen><\/iframe><\/p>\n<p>That&#8217;s how it all works on paper, but seeing it in person is a\u00a0different game\u00a0altogether. The workflow for the<a href=\"http:\/\/www.thermofisher.com\/uk\/en\/home\/life-science\/pcr\/digital-pcr\/quantstudio-3d-digital-pcr-system.html?socid=social_btb_genequant\" target=\"_blank\" rel=\"noopener noreferrer\">\u00a0QuantStudio 3D digital PCR\u00a0system<\/a>\u00a0was\u00a0recently updated and\u00a0posted on Thermo Fisher Scientific&#8217;s <a href=\"https:\/\/www.youtube.com\/user\/LifeTechnologiesCorp?socid=social_btb\" target=\"_blank\" rel=\"noopener noreferrer\">YouTube channel<\/a>. Take a look at the workflow in the video below, and let us know if you have any questions in the comments section.<\/p>\n","protected":false},"excerpt":{"rendered":"<p>Digital PCR\u00a0is a relatively new approach to nucleic acid detection and quantification that offers an alternate method to conventional real-time quantitative PCR for absolute quantification and rare allele detection. It all works by partitioning a sample of DNA or cDNA into many individual, parallel PCR reactions; some of these reactions contain the target molecule (positive)<\/p>\n","protected":false},"author":120,"featured_media":8198,"comment_status":"open","ping_status":"closed","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_kad_blocks_custom_css":"","_kad_blocks_head_custom_js":"","_kad_blocks_body_custom_js":"","_kad_blocks_footer_custom_js":"","_monsterinsights_skip_tracking":false,"_genesis_hide_title":false,"_genesis_hide_breadcrumbs":false,"_genesis_hide_singular_image":false,"_genesis_hide_footer_widgets":false,"_genesis_custom_body_class":"","_genesis_custom_post_class":"","_genesis_layout":"","_jetpack_newsletter_access":"","_jetpack_dont_email_post_to_subs":false,"_jetpack_newsletter_tier_id":0,"_jetpack_memberships_contains_paywalled_content":false,"_jetpack_memberships_contains_paid_content":false,"footnotes":""},"categories":[101],"tags":[17,57,104,43],"division":[],"class_list":{"0":"post-8197","1":"post","2":"type-post","3":"status-publish","4":"format-standard","5":"has-post-thumbnail","7":"category-product-news","8":"tag-applied-biosystems","9":"tag-digital-pcr","10":"tag-gene-expression","11":"tag-pcr","12":"entry"},"_selected_authors":"","_selected_reviewers":"","acf":[],"yoast_head":"<!-- This site is optimized with the Yoast SEO Premium plugin v27.8 (Yoast SEO v27.8) - https:\/\/yoast.com\/product\/yoast-seo-premium-wordpress\/ -->\n<title>Hands-on With Digital PCR - Behind the Bench<\/title>\n<meta name=\"robots\" content=\"index, follow, max-snippet:-1, max-image-preview:large, max-video-preview:-1\" \/>\n<link rel=\"canonical\" href=\"https:\/\/www.thermofisher.com\/blog\/behindthebench\/hands-on-with-digital-pcr\/\" \/>\n<meta property=\"og:locale\" content=\"en_US\" \/>\n<meta property=\"og:type\" content=\"article\" \/>\n<meta property=\"og:title\" content=\"Hands-on With Digital PCR\" \/>\n<meta property=\"og:description\" content=\"Digital PCR\u00a0is a relatively new approach to nucleic acid detection and quantification that offers an alternate method to conventional real-time quantitative PCR for absolute quantification and rare allele detection. It all works by partitioning a sample of DNA or cDNA into many individual, parallel PCR reactions; some of these reactions contain the target molecule (positive)\" \/>\n<meta property=\"og:url\" content=\"https:\/\/www.thermofisher.com\/blog\/behindthebench\/hands-on-with-digital-pcr\/\" \/>\n<meta property=\"og:site_name\" content=\"Behind the Bench\" \/>\n<meta property=\"article:publisher\" content=\"https:\/\/www.facebook.com\/thermofisher\" \/>\n<meta property=\"article:published_time\" content=\"2016-04-04T22:28:53+00:00\" \/>\n<meta property=\"article:modified_time\" content=\"2022-08-18T20:29:12+00:00\" \/>\n<meta property=\"og:image\" content=\"https:\/\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2016\/04\/Digital-PCR-Workflow-6-1.png\" \/>\n\t<meta property=\"og:image:width\" content=\"966\" \/>\n\t<meta property=\"og:image:height\" content=\"542\" \/>\n\t<meta property=\"og:image:type\" content=\"image\/png\" \/>\n<meta name=\"author\" content=\"Behind The Bench Staff\" \/>\n<meta name=\"twitter:card\" content=\"summary_large_image\" \/>\n<meta name=\"twitter:creator\" content=\"@thermofisher\" \/>\n<meta name=\"twitter:site\" content=\"@thermofisher\" \/>\n<meta name=\"twitter:label1\" content=\"Written by\" \/>\n\t<meta name=\"twitter:data1\" content=\"Behind The Bench Staff\" \/>\n\t<meta name=\"twitter:label2\" content=\"Est. reading time\" \/>\n\t<meta name=\"twitter:data2\" content=\"1 minute\" \/>\n<script type=\"application\/ld+json\" class=\"yoast-schema-graph\">{\"@context\":\"https:\\\/\\\/schema.org\",\"@graph\":[{\"@type\":\"Article\",\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/behindthebench\\\/hands-on-with-digital-pcr\\\/#article\",\"isPartOf\":{\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/behindthebench\\\/hands-on-with-digital-pcr\\\/\"},\"author\":{\"name\":\"Behind The Bench Staff\",\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/behindthebench\\\/#\\\/schema\\\/person\\\/9bbb2d67616274d3d4a21a584f7f4a0c\"},\"headline\":\"Hands-on With Digital PCR\",\"datePublished\":\"2016-04-04T22:28:53+00:00\",\"dateModified\":\"2022-08-18T20:29:12+00:00\",\"mainEntityOfPage\":{\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/behindthebench\\\/hands-on-with-digital-pcr\\\/\"},\"wordCount\":200,\"commentCount\":0,\"image\":{\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/behindthebench\\\/hands-on-with-digital-pcr\\\/#primaryimage\"},\"thumbnailUrl\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/behindthebench\\\/wp-content\\\/uploads\\\/sites\\\/9\\\/2016\\\/04\\\/Digital-PCR-Workflow-6-1.png\",\"keywords\":[\"Applied Biosystems\",\"digital PCR\",\"gene expression\",\"PCR\"],\"articleSection\":[\"Product News\"],\"inLanguage\":\"en-US\",\"potentialAction\":[{\"@type\":\"CommentAction\",\"name\":\"Comment\",\"target\":[\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/behindthebench\\\/hands-on-with-digital-pcr\\\/#respond\"]}]},{\"@type\":\"WebPage\",\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/behindthebench\\\/hands-on-with-digital-pcr\\\/\",\"url\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/behindthebench\\\/hands-on-with-digital-pcr\\\/\",\"name\":\"Hands-on With Digital PCR - Behind the Bench\",\"isPartOf\":{\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/behindthebench\\\/#website\"},\"primaryImageOfPage\":{\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/behindthebench\\\/hands-on-with-digital-pcr\\\/#primaryimage\"},\"image\":{\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/behindthebench\\\/hands-on-with-digital-pcr\\\/#primaryimage\"},\"thumbnailUrl\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/behindthebench\\\/wp-content\\\/uploads\\\/sites\\\/9\\\/2016\\\/04\\\/Digital-PCR-Workflow-6-1.png\",\"datePublished\":\"2016-04-04T22:28:53+00:00\",\"dateModified\":\"2022-08-18T20:29:12+00:00\",\"author\":{\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/behindthebench\\\/#\\\/schema\\\/person\\\/9bbb2d67616274d3d4a21a584f7f4a0c\"},\"breadcrumb\":{\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/behindthebench\\\/hands-on-with-digital-pcr\\\/#breadcrumb\"},\"inLanguage\":\"en-US\",\"potentialAction\":[{\"@type\":\"ReadAction\",\"target\":[\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/behindthebench\\\/hands-on-with-digital-pcr\\\/\"]}]},{\"@type\":\"ImageObject\",\"inLanguage\":\"en-US\",\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/behindthebench\\\/hands-on-with-digital-pcr\\\/#primaryimage\",\"url\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/behindthebench\\\/wp-content\\\/uploads\\\/sites\\\/9\\\/2016\\\/04\\\/Digital-PCR-Workflow-6-1.png\",\"contentUrl\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/behindthebench\\\/wp-content\\\/uploads\\\/sites\\\/9\\\/2016\\\/04\\\/Digital-PCR-Workflow-6-1.png\",\"width\":966,\"height\":542},{\"@type\":\"BreadcrumbList\",\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/behindthebench\\\/hands-on-with-digital-pcr\\\/#breadcrumb\",\"itemListElement\":[{\"@type\":\"ListItem\",\"position\":1,\"name\":\"Home\",\"item\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/behindthebench\\\/\"},{\"@type\":\"ListItem\",\"position\":2,\"name\":\"Hands-on With Digital PCR\"}]},{\"@type\":\"WebSite\",\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/behindthebench\\\/#website\",\"url\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/behindthebench\\\/\",\"name\":\"Behind the Bench\",\"description\":\"\",\"potentialAction\":[{\"@type\":\"SearchAction\",\"target\":{\"@type\":\"EntryPoint\",\"urlTemplate\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/behindthebench\\\/?s={search_term_string}\"},\"query-input\":{\"@type\":\"PropertyValueSpecification\",\"valueRequired\":true,\"valueName\":\"search_term_string\"}}],\"inLanguage\":\"en-US\"},{\"@type\":\"Person\",\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/behindthebench\\\/#\\\/schema\\\/person\\\/9bbb2d67616274d3d4a21a584f7f4a0c\",\"name\":\"Behind The Bench Staff\",\"image\":{\"@type\":\"ImageObject\",\"inLanguage\":\"en-US\",\"@id\":\"https:\\\/\\\/secure.gravatar.com\\\/avatar\\\/52d5c16ece4bd1c0a49a63521c2b6a25c54a3f0c33f7afed215ba7e0cdff69e3?s=96&d=mm&r=g\",\"url\":\"https:\\\/\\\/secure.gravatar.com\\\/avatar\\\/52d5c16ece4bd1c0a49a63521c2b6a25c54a3f0c33f7afed215ba7e0cdff69e3?s=96&d=mm&r=g\",\"contentUrl\":\"https:\\\/\\\/secure.gravatar.com\\\/avatar\\\/52d5c16ece4bd1c0a49a63521c2b6a25c54a3f0c33f7afed215ba7e0cdff69e3?s=96&d=mm&r=g\",\"caption\":\"Behind The Bench Staff\"},\"url\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/author\\\/behind-the-bench-staff\\\/\"}]}<\/script>\n<!-- \/ Yoast SEO Premium plugin. -->","yoast_head_json":{"title":"Hands-on With Digital PCR - Behind the Bench","robots":{"index":"index","follow":"follow","max-snippet":"max-snippet:-1","max-image-preview":"max-image-preview:large","max-video-preview":"max-video-preview:-1"},"canonical":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/hands-on-with-digital-pcr\/","og_locale":"en_US","og_type":"article","og_title":"Hands-on With Digital PCR","og_description":"Digital PCR\u00a0is a relatively new approach to nucleic acid detection and quantification that offers an alternate method to conventional real-time quantitative PCR for absolute quantification and rare allele detection. It all works by partitioning a sample of DNA or cDNA into many individual, parallel PCR reactions; some of these reactions contain the target molecule (positive)","og_url":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/hands-on-with-digital-pcr\/","og_site_name":"Behind the Bench","article_publisher":"https:\/\/www.facebook.com\/thermofisher","article_published_time":"2016-04-04T22:28:53+00:00","article_modified_time":"2022-08-18T20:29:12+00:00","og_image":[{"width":966,"height":542,"url":"https:\/\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2016\/04\/Digital-PCR-Workflow-6-1.png","type":"image\/png"}],"author":"Behind The Bench Staff","twitter_card":"summary_large_image","twitter_creator":"@thermofisher","twitter_site":"@thermofisher","twitter_misc":{"Written by":"Behind The Bench Staff","Est. reading time":"1 minute"},"schema":{"@context":"https:\/\/schema.org","@graph":[{"@type":"Article","@id":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/hands-on-with-digital-pcr\/#article","isPartOf":{"@id":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/hands-on-with-digital-pcr\/"},"author":{"name":"Behind The Bench Staff","@id":"https:\/\/admin.acceleratingscience.com\/behindthebench\/#\/schema\/person\/9bbb2d67616274d3d4a21a584f7f4a0c"},"headline":"Hands-on With Digital PCR","datePublished":"2016-04-04T22:28:53+00:00","dateModified":"2022-08-18T20:29:12+00:00","mainEntityOfPage":{"@id":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/hands-on-with-digital-pcr\/"},"wordCount":200,"commentCount":0,"image":{"@id":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/hands-on-with-digital-pcr\/#primaryimage"},"thumbnailUrl":"https:\/\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2016\/04\/Digital-PCR-Workflow-6-1.png","keywords":["Applied Biosystems","digital PCR","gene expression","PCR"],"articleSection":["Product News"],"inLanguage":"en-US","potentialAction":[{"@type":"CommentAction","name":"Comment","target":["https:\/\/www.thermofisher.com\/blog\/behindthebench\/hands-on-with-digital-pcr\/#respond"]}]},{"@type":"WebPage","@id":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/hands-on-with-digital-pcr\/","url":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/hands-on-with-digital-pcr\/","name":"Hands-on With Digital PCR - Behind the Bench","isPartOf":{"@id":"https:\/\/admin.acceleratingscience.com\/behindthebench\/#website"},"primaryImageOfPage":{"@id":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/hands-on-with-digital-pcr\/#primaryimage"},"image":{"@id":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/hands-on-with-digital-pcr\/#primaryimage"},"thumbnailUrl":"https:\/\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2016\/04\/Digital-PCR-Workflow-6-1.png","datePublished":"2016-04-04T22:28:53+00:00","dateModified":"2022-08-18T20:29:12+00:00","author":{"@id":"https:\/\/admin.acceleratingscience.com\/behindthebench\/#\/schema\/person\/9bbb2d67616274d3d4a21a584f7f4a0c"},"breadcrumb":{"@id":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/hands-on-with-digital-pcr\/#breadcrumb"},"inLanguage":"en-US","potentialAction":[{"@type":"ReadAction","target":["https:\/\/www.thermofisher.com\/blog\/behindthebench\/hands-on-with-digital-pcr\/"]}]},{"@type":"ImageObject","inLanguage":"en-US","@id":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/hands-on-with-digital-pcr\/#primaryimage","url":"https:\/\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2016\/04\/Digital-PCR-Workflow-6-1.png","contentUrl":"https:\/\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2016\/04\/Digital-PCR-Workflow-6-1.png","width":966,"height":542},{"@type":"BreadcrumbList","@id":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/hands-on-with-digital-pcr\/#breadcrumb","itemListElement":[{"@type":"ListItem","position":1,"name":"Home","item":"https:\/\/admin.acceleratingscience.com\/behindthebench\/"},{"@type":"ListItem","position":2,"name":"Hands-on With Digital PCR"}]},{"@type":"WebSite","@id":"https:\/\/admin.acceleratingscience.com\/behindthebench\/#website","url":"https:\/\/admin.acceleratingscience.com\/behindthebench\/","name":"Behind the Bench","description":"","potentialAction":[{"@type":"SearchAction","target":{"@type":"EntryPoint","urlTemplate":"https:\/\/admin.acceleratingscience.com\/behindthebench\/?s={search_term_string}"},"query-input":{"@type":"PropertyValueSpecification","valueRequired":true,"valueName":"search_term_string"}}],"inLanguage":"en-US"},{"@type":"Person","@id":"https:\/\/admin.acceleratingscience.com\/behindthebench\/#\/schema\/person\/9bbb2d67616274d3d4a21a584f7f4a0c","name":"Behind The Bench Staff","image":{"@type":"ImageObject","inLanguage":"en-US","@id":"https:\/\/secure.gravatar.com\/avatar\/52d5c16ece4bd1c0a49a63521c2b6a25c54a3f0c33f7afed215ba7e0cdff69e3?s=96&d=mm&r=g","url":"https:\/\/secure.gravatar.com\/avatar\/52d5c16ece4bd1c0a49a63521c2b6a25c54a3f0c33f7afed215ba7e0cdff69e3?s=96&d=mm&r=g","contentUrl":"https:\/\/secure.gravatar.com\/avatar\/52d5c16ece4bd1c0a49a63521c2b6a25c54a3f0c33f7afed215ba7e0cdff69e3?s=96&d=mm&r=g","caption":"Behind The Bench Staff"},"url":"https:\/\/admin.acceleratingscience.com\/author\/behind-the-bench-staff\/"}]}},"taxonomy_info":{"category":[{"value":101,"label":"Product News"}],"post_tag":[{"value":17,"label":"Applied Biosystems"},{"value":57,"label":"digital PCR"},{"value":104,"label":"gene expression"},{"value":43,"label":"PCR"}]},"featured_image_src_large":["https:\/\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2016\/04\/Digital-PCR-Workflow-6-1.png",760,426,false],"author_info":{"display_name":"Behind The Bench Staff","author_link":"https:\/\/admin.acceleratingscience.com\/author\/behind-the-bench-staff\/"},"comment_info":0,"category_info":[{"term_id":101,"name":"Product News","slug":"product-news","term_group":0,"term_taxonomy_id":101,"taxonomy":"category","description":"","parent":0,"count":105,"filter":"raw","meta":[],"cat_ID":101,"category_count":105,"category_description":"","cat_name":"Product News","category_nicename":"product-news","category_parent":0}],"tag_info":[{"term_id":17,"name":"Applied Biosystems","slug":"applied-biosystems","term_group":0,"term_taxonomy_id":17,"taxonomy":"post_tag","description":"","parent":0,"count":193,"filter":"raw","meta":[]},{"term_id":57,"name":"digital PCR","slug":"digital-pcr","term_group":0,"term_taxonomy_id":57,"taxonomy":"post_tag","description":"","parent":0,"count":51,"filter":"raw","meta":[]},{"term_id":104,"name":"gene expression","slug":"gene-expression","term_group":0,"term_taxonomy_id":104,"taxonomy":"post_tag","description":"","parent":0,"count":21,"filter":"raw","meta":[]},{"term_id":43,"name":"PCR","slug":"pcr","term_group":0,"term_taxonomy_id":43,"taxonomy":"post_tag","description":"","parent":0,"count":142,"filter":"raw","meta":[]}],"jetpack-related-posts":[{"id":19566,"url":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/pcr-inhibitors-digital-pcr\/","url_meta":{"origin":8197,"position":0},"title":"When PCR Gets Dirty &#8211; Tackling PCR Inhibitors","author":"Behind The Bench Staff","date":"March 24, 2026","format":false,"excerpt":"In this first Absolute Gene-ius: Science Snapshot episode, hosts Jordan Ruggieri and Lisa Crawford revisit expert insights on one persistent molecular biology challenge: PCR inhibitors. Across wastewater surveillance, environmental microbiology, and gene therapy manufacturing, inhibitors can distort PCR data and complicate quantification. Featuring perspectives from Ray Ketchum, PhD; Sarah Philo,\u2026","rel":"","context":"In &quot;Lab Tips and Tricks&quot;","block_context":{"text":"Lab Tips and Tricks","link":"https:\/\/admin.acceleratingscience.com\/behindthebench\/lab-tips-and-tricks\/"},"img":{"alt_text":"","src":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/03\/Ketchum-headshot.jpg?resize=350%2C200&ssl=1","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/03\/Ketchum-headshot.jpg?resize=350%2C200&ssl=1 1x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/03\/Ketchum-headshot.jpg?resize=525%2C300&ssl=1 1.5x"},"classes":[]},{"id":19888,"url":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/streptococcus-pneumoniae-serotype-identification-qpcr-nested-pcr\/","url_meta":{"origin":8197,"position":1},"title":"qPCR, Nested PCR, and the Search for Bacterial Serotypes","author":"Ashleigh Barlow","date":"August 5, 2026","format":false,"excerpt":"Article Summary In this episode of Absolute Gene-ius, Donna Paznar, former Head of Molecular Infection Diagnostics at the Medical University of Vienna, discusses her research on Streptococcus pneumoniae and its impact across the Austrian population. She explains why children can act as asymptomatic reservoirs, why tracking bacterial serotypes matters, and\u2026","rel":"","context":"In &quot;Infectious Diseases&quot;","block_context":{"text":"Infectious Diseases","link":"https:\/\/admin.acceleratingscience.com\/behindthebench\/infectious-diseases\/"},"img":{"alt_text":"Donna Paznar discussing molecular methods for identifying Streptococcus pneumoniae serotypes.","src":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/08\/Paznar_Headshot_Jordan-Ruggieri-scaled.jpg?resize=350%2C200&ssl=1","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/08\/Paznar_Headshot_Jordan-Ruggieri-scaled.jpg?resize=350%2C200&ssl=1 1x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/08\/Paznar_Headshot_Jordan-Ruggieri-scaled.jpg?resize=525%2C300&ssl=1 1.5x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/08\/Paznar_Headshot_Jordan-Ruggieri-scaled.jpg?resize=700%2C400&ssl=1 2x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/08\/Paznar_Headshot_Jordan-Ruggieri-scaled.jpg?resize=1050%2C600&ssl=1 3x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/08\/Paznar_Headshot_Jordan-Ruggieri-scaled.jpg?resize=1400%2C800&ssl=1 4x"},"classes":[]},{"id":19911,"url":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/multiplex-mirna-profiling-precision-oncology\/","url_meta":{"origin":8197,"position":2},"title":"A scalable multiplex miRNA profiling workflow for precision oncology research","author":"Ashleigh Barlow","date":"August 18, 2026","format":false,"excerpt":"The promise of precision oncology rests on our ability to reliably monitor tumor biology in real time. Circulating microRNAs (miRNAs) offer that opportunity, serving as potentially stable, non-invasive biomarkers linked to disease progression and therapeutic response; however, their adoption in routine practice remains constrained by technical hurdles. (1,2) Because cancer\u2026","rel":"","context":"In &quot;Cancer Research&quot;","block_context":{"text":"Cancer Research","link":"https:\/\/admin.acceleratingscience.com\/behindthebench\/cancer-research\/"},"img":{"alt_text":"Scientist pipetting a laboratory sample during molecular biomarker research.","src":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/08\/lisa-blog-picture.jpg?resize=350%2C200&ssl=1","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/08\/lisa-blog-picture.jpg?resize=350%2C200&ssl=1 1x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/08\/lisa-blog-picture.jpg?resize=525%2C300&ssl=1 1.5x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/08\/lisa-blog-picture.jpg?resize=700%2C400&ssl=1 2x"},"classes":[]},{"id":19776,"url":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/epigenetic-methylation-analysis-qpcr-dpcr\/","url_meta":{"origin":8197,"position":3},"title":"Epigenetics unlocked: How behavior shapes biology","author":"Ashleigh Barlow","date":"May 28, 2026","format":false,"excerpt":"Article Summary Epigenetics is helping researchers understand how environment, behavior, and molecular regulation intersect. Rather than changing the underlying DNA sequence, epigenetic changes influence how genes are expressed, including when they are turned on, turned off, or regulated in response to biological and environmental signals. In this episode of Absolute\u2026","rel":"","context":"In &quot;Epigenetics&quot;","block_context":{"text":"Epigenetics","link":"https:\/\/admin.acceleratingscience.com\/behindthebench\/epigenetics\/"},"img":{"alt_text":"Portrait of Dr. Robert Philibert discussing epigenetics, DNA methylation, and behavioral biomarkers.","src":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/05\/Philibert-Photo_Jordan-Ruggieri.jpg?resize=350%2C200&ssl=1","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/05\/Philibert-Photo_Jordan-Ruggieri.jpg?resize=350%2C200&ssl=1 1x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/05\/Philibert-Photo_Jordan-Ruggieri.jpg?resize=525%2C300&ssl=1 1.5x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/05\/Philibert-Photo_Jordan-Ruggieri.jpg?resize=700%2C400&ssl=1 2x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/05\/Philibert-Photo_Jordan-Ruggieri.jpg?resize=1050%2C600&ssl=1 3x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2026\/05\/Philibert-Photo_Jordan-Ruggieri.jpg?resize=1400%2C800&ssl=1 4x"},"classes":[]},{"id":19447,"url":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/philadelphia-chromosome-like-all-digital-pcr\/","url_meta":{"origin":8197,"position":4},"title":"Unlocking Cancer\u2019s Genetic Secrets with Digital PCR","author":"Behind The Bench Staff","date":"February 27, 2026","format":false,"excerpt":"In this episode of Absolute Gene-ius, we explore how digital PCR (dPCR) is transforming leukemia research with Valeria Rangel, a PhD candidate at the University of California, Irvine. Her work focuses on Philadelphia chromosome-like (Ph-like) B-cell acute lymphoblastic leukemia (ALL), a rare, aggressive subtype that disproportionately affects Hispanic populations. By\u2026","rel":"","context":"In &quot;Cancer Research&quot;","block_context":{"text":"Cancer Research","link":"https:\/\/admin.acceleratingscience.com\/behindthebench\/cancer-research\/"},"img":{"alt_text":"","src":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2025\/11\/RANGEL-headshot-1.jpg?resize=350%2C200&ssl=1","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2025\/11\/RANGEL-headshot-1.jpg?resize=350%2C200&ssl=1 1x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2025\/11\/RANGEL-headshot-1.jpg?resize=525%2C300&ssl=1 1.5x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2025\/11\/RANGEL-headshot-1.jpg?resize=700%2C400&ssl=1 2x"},"classes":[]},{"id":19235,"url":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/crispr-digital-pcr-gene-editing-workflow\/","url_meta":{"origin":8197,"position":5},"title":"Revolutionizing Gene Editing with CRISPR and Digital PCR &#8211; A Conversation with Dr. Dustin Rubinstein","author":"Behind The Bench Staff","date":"November 11, 2025","format":false,"excerpt":"In a recent Absolute Gene-ius podcast episode titled \u201cCRISPR, dPCR, and pigs - a cut above the rest\u201d, Dr. Dustin Rubinstein, Director of the Advanced Genome Editing Laboratory at the University of Wisconsin-Madison, shared his insights on how CRISPR and digital PCR (dPCR) are revolutionizing gene editing workflows. This Q&A\u2026","rel":"","context":"In &quot;Cancer Research&quot;","block_context":{"text":"Cancer Research","link":"https:\/\/admin.acceleratingscience.com\/behindthebench\/cancer-research\/"},"img":{"alt_text":"","src":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2025\/11\/S3-E1-Header-Image-1_Jordan-Ruggieri-scaled.png?resize=350%2C200&ssl=1","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2025\/11\/S3-E1-Header-Image-1_Jordan-Ruggieri-scaled.png?resize=350%2C200&ssl=1 1x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2025\/11\/S3-E1-Header-Image-1_Jordan-Ruggieri-scaled.png?resize=525%2C300&ssl=1 1.5x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2025\/11\/S3-E1-Header-Image-1_Jordan-Ruggieri-scaled.png?resize=700%2C400&ssl=1 2x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2025\/11\/S3-E1-Header-Image-1_Jordan-Ruggieri-scaled.png?resize=1050%2C600&ssl=1 3x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2025\/11\/S3-E1-Header-Image-1_Jordan-Ruggieri-scaled.png?resize=1400%2C800&ssl=1 4x"},"classes":[]}],"jetpack_sharing_enabled":true,"jetpack_featured_media_url":"https:\/\/admin.acceleratingscience.com\/behindthebench\/wp-content\/uploads\/sites\/9\/2016\/04\/Digital-PCR-Workflow-6-1.png","_links":{"self":[{"href":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/wp-json\/wp\/v2\/posts\/8197","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/wp-json\/wp\/v2\/users\/120"}],"replies":[{"embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/wp-json\/wp\/v2\/comments?post=8197"}],"version-history":[{"count":0,"href":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/wp-json\/wp\/v2\/posts\/8197\/revisions"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/wp-json\/wp\/v2\/media\/8198"}],"wp:attachment":[{"href":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/wp-json\/wp\/v2\/media?parent=8197"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/wp-json\/wp\/v2\/categories?post=8197"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/wp-json\/wp\/v2\/tags?post=8197"},{"taxonomy":"division","embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/behindthebench\/wp-json\/wp\/v2\/division?post=8197"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}