{"id":249,"date":"2020-04-02T16:11:58","date_gmt":"2020-04-02T23:11:58","guid":{"rendered":"https:\/\/www.thermofisher.com\/blog\/ask-a-scientist\/?p=249"},"modified":"2022-10-06T16:21:05","modified_gmt":"2022-10-06T23:21:05","slug":"how-does-a-covid-19-test-work","status":"publish","type":"post","link":"https:\/\/www.thermofisher.com\/blog\/clinical-conversations\/how-does-a-covid-19-test-work\/","title":{"rendered":"How does a SARS-CoV-2 (COVID-19) test work?"},"content":{"rendered":"<p>Hundreds of thousands of people are being tested for SARS-CoV-2 (COVID-19) around the world and the tests themselves are a crucial starting point in efforts to flatten the spread of the virus.<\/p>\n<p>In a recent episode of Thermo Fisher Scientific\u2019s Science with a Twist podcast, we talk with Senior Director of Genetic Sciences, Joshua Tratta. Tratta explains how tests are created, how they\u2019re performed and how they\u2019re analyzed.<\/p>\n<p><strong>QUESTION: Prior to the CDC announcement that they have developed a test, there was this kind of race to develop a test. Talk about what it is that encompasses the time that it takes to develop a test. Why does it take so long? What is the process you have to go through to develop a test?<\/strong><\/p>\n<p>ANSWER: Most outbreaks, and this one is no different, you start with asking the question, \u2018Why are people getting sick? We don\u2019t understand.\u2019 So something happens. In this case, someone was getting sick and people didn\u2019t understand why. The first thing that most outbreaks do is use Next Generation Sequencing technology to globally look at the sample and say, \u2018What is it? Is it bacterial? Is it viral? Is it fungal? Is it something we have no idea what it is?\u2019 And it allows you to start narrowing down what it is.<\/p>\n<p>Once we\u2019ve done that, as is the case of the Chinese government, they actually published their NextGen sequencing results, which allowed people to actually see that it was in the coronavirus family.<\/p>\n<p>Not only was it a coronavirus, compared to like SARS, which is also a coronavirus \u2013 it\u2019s a novel coronavirus. So they kind of knew what it was. They\u2019ve narrowed it down, but there\u2019s something new about this particular virus.<\/p>\n<p>So what you have to do once you identify that, instead of running that big broad screen over and over again, which can be cost prohibitive for trying to ramp up to millions or hundreds of thousands of \u00a0people. People move into designing specific qPCR assays to detect just that area which is now identified.<\/p>\n<p>So to do that we use something called bioinformatics tools. It\u2019s designing specific genetic sequences and assays that cover that region. Then, unlike some of the other technologies, qPCR can give you the results in a two-to-2.5 hour timeframe at a reduced cost. And because it\u2019s been a technology that has been around over 25 years, a lot of people are using it for other things. For other viruses. For other bacteria. For research applications. So the prevalence of instrumentation is out there. It\u2019s something you can now push more broadly to people that already have these instruments and can use them to specifically look at something like the coronavirus.<\/p>\n<p><strong>QUESTION: What exactly is an assay?<\/strong><\/p>\n<p>ANSWER: An assay is a genetic tool that actually helps you identify a virus or a bacteria. Being able to match up that genetic material in one piece versus another that says you actually have this genetic information. \u2018I have compared you against this test.\u2019 Or you don\u2019t.<\/p>\n<p><strong>QUESTION: You provided me with an example using Lucky Charms cereal. Can you, using the Lucky Charms analogy talk about how the coronavirus may or may not be present in my Lucky Charms. <\/strong><\/p>\n<p>Hopefully you don\u2019t have coronavirus in your Lucky Charms. But using it as the analogy\u2026 within Lucky Charms you have multiple types of marshmallows. So imagine that the coronavirus is the rainbow marshmallow. Your cereal bowl is your genetic makeup. An assay would try to detect the rainbow. So when you apply that, you would take rainbow \u2013 which is the qPCR assay \u2013 and you would look over your bowl and say, \u2018Are there any rainbows? Are there any rainbows in my cereal?\u2019. And if there was a match, you\u2019d have coronavirus. If not, you\u2019d say \u2018I don\u2019t have the coronavirus.\u2019 Doesn\u2019t mean there\u2019s not something else that makes you sick but you can say you don\u2019t have the coronavirus.<\/p>\n<p><strong>QUESTION: I know that qPCR has been around for a long time. What is qPCR? How would I describe qPCR or explain it to someone who doesn\u2019t know what it was? <\/strong><\/p>\n<p>qPCR has been around for 25 years. It\u2019s a detection methodology that is akin to identifying using genetic information to identify bacteria, viruses, fungal targets. Historically people have used culture for many years. Which is still a robustly and routinely used technology. But in this particular case it allows you to do something faster. It\u2019s being able to use genetic information to identify a pathogen in a very short amount of time.<\/p>\n<p><strong>QUESTION: How does the virus\u2019s genetic code factor into the development of the test?<\/strong><\/p>\n<p>The genetic code plays into the test. You want to be very specific about identifying just that virus.<\/p>\n<p>So when you develop the test you need to really understand the genetic makeup of the virus. So that you can create the most specific test. Just like for you or for me, I don\u2019t want you to tell me a test that says, \u00a0\u2018Oh I\u2019m only 30% sure you\u2019ve got the coronavirus.\u2019 I want you to give me an answer so I can feel confident behind it.<\/p>\n<p>That\u2019s what we call specificity.<\/p>\n<p>So you want something to be specifically targeting something you\u2019re concerned I may have. And if you\u2019re not sure, and you have a lower specificity, it means that you could be entering a quarantine with somebody when you shouldn\u2019t have been in quarantine.<\/p>\n<p>I want to make sure we are understating the virus, knowing the details will help you design very specific assays for detection.<\/p>\n<p><strong>QUESTION: Everyone that has a respiratory infection doesn\u2019t necessarily have the coronavirus. Talk about how technology enables you to differentiate between the different sources of infection using, for example, a multi-panel test. \u00a0<\/strong><\/p>\n<p>So if you look at the symptoms associated with the coronavirus, how do you differentiate between someone who has the flu \u2013 that\u2019s pretty prevalent and spreading across the globe right now \u2013 or if you have the coronavirus? Or if you have some bacteria? The beauty of technologies like qPCR, we have something that\u2019s called a syndromic approach. Which is having multiple targets that you can design for and test for them all at once. So in a case like for what Thermo Fisher has designed, we have a test that can test for 42 different targets. And they will look at the most common bacterial, viral and fungal targets that cause respiratory illness. Imagine you walk in and somebody suspects that you have the coronavirus, on that test is something that will say \u2018yes or no \u2013 you have the coronavirus.\u2019 If the answer is no, wouldn\u2019t it be great to say that you actually have a bacteria, and there\u2019s a specific antibiotic that will be used to treat you. Or that you have flu, and it\u2019s just flu A. They\u2019re going to tell you to go home, get some rest and drink some fluids.<\/p>\n<p><strong>QUESTION: Talk about the importance of detection and the importance of testing. <\/strong><\/p>\n<p>I\u2019d say detection is the beginning of ensuring that we don\u2019t spread the thing. Detection is the starting point. Knowing how to make good decisions from that point.<\/p>\n<p><strong>QUESTION: We talked about a little bit about the sample \u2013 it\u2019s a nasal swab\/throat swab. If you could just help us understand how you get from swab to test result, that would be really helpful. <\/strong><\/p>\n<p>Imagine a swab &#8211; for any of our listeners who don\u2019t know what that is \u2013 imagine a big Q-Tip that is long. For any of you who have had the unpleasant experience of going to your doctor because you thought you had strep throat. Same approach. They\u2019re going to take a sample from your throat or nose and pull out what seems like nothing. But it will have material on that which that material is what we want to extract to get your DNA. So being able to get that DNA extracted, Thermo Fisher Scientific also makes kits to be able to do that. We have one that is focused on viral pathogens and extracts the DNA. That\u2019s kind of the starting point. We got to get the DNA so we can go back to our Lucky Charms example and do some comparisons of understanding if the rainbow is there. DNA extraction works in that way. And then you will actually move over into using multiplex assay. So you combine those pieces together with a single tube sample that they\u2019ve taken and some other reagents and you put it on to an instrument. Let it run for a few hours and a result will come back that will say the lucky charm rainbow is there or it is not.<\/p>\n","protected":false},"excerpt":{"rendered":"<p>Hundreds of thousands of people are being tested for SARS-CoV-2 (COVID-19) around the world and the tests themselves are a crucial starting point in efforts to flatten the spread of the virus. In a recent episode of Thermo Fisher Scientific\u2019s Science with a Twist podcast, we talk with Senior Director of Genetic Sciences, Joshua Tratta.<\/p>\n","protected":false},"author":1556,"featured_media":15840,"comment_status":"open","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_kad_blocks_custom_css":"","_kad_blocks_head_custom_js":"","_kad_blocks_body_custom_js":"","_kad_blocks_footer_custom_js":"","_monsterinsights_skip_tracking":false,"_genesis_hide_title":false,"_genesis_hide_breadcrumbs":false,"_genesis_hide_singular_image":false,"_genesis_hide_footer_widgets":false,"_genesis_custom_body_class":"","_genesis_custom_post_class":"","_genesis_layout":"","footnotes":""},"categories":[10],"tags":[121,22,90,26],"division":[],"class_list":["post-249","post","type-post","status-publish","format-standard","has-post-thumbnail","category-science-and-education","tag-content-article","tag-covid-19","tag-qpcr","tag-sars-cov-2","entry"],"_selected_authors":"","_selected_reviewers":"","acf":[],"yoast_head":"<!-- This site is optimized with the Yoast SEO Premium plugin v28.9 (Yoast SEO v28.9) - https:\/\/yoast.com\/product\/yoast-seo-premium-wordpress\/ -->\n<title>How does a COVID-19 test work? | Ask a Scientist<\/title>\n<meta name=\"description\" content=\"Understand how the tests used to diagnose COVID-19 are developed, how testing is being performed and how results are being generated.\" \/>\n<meta name=\"robots\" content=\"index, follow, max-snippet:-1, max-image-preview:large, max-video-preview:-1\" \/>\n<link rel=\"canonical\" href=\"https:\/\/www.thermofisher.com\/blog\/clinical-conversations\/how-does-a-covid-19-test-work\/\" \/>\n<meta property=\"og:locale\" content=\"en_US\" \/>\n<meta property=\"og:type\" content=\"article\" \/>\n<meta property=\"og:title\" content=\"How does a SARS-CoV-2 (COVID-19) test work?\" \/>\n<meta property=\"og:description\" content=\"Understand how the tests used to diagnose COVID-19 are developed, how testing is being performed and how results are being generated.\" \/>\n<meta property=\"og:url\" content=\"https:\/\/admin.acceleratingscience.com\/clinical-conversations\/how-does-a-covid-19-test-work\/\" \/>\n<meta property=\"og:site_name\" content=\"Clinical Conversations\" \/>\n<meta property=\"article:publisher\" content=\"https:\/\/www.facebook.com\/thermofisher\" \/>\n<meta property=\"article:published_time\" content=\"2020-04-02T23:11:58+00:00\" \/>\n<meta property=\"article:modified_time\" content=\"2022-10-06T23:21:05+00:00\" \/>\n<meta property=\"og:image\" content=\"https:\/\/admin.acceleratingscience.com\/clinical-conversations\/wp-content\/uploads\/sites\/18\/2020\/12\/COL011389-Coronavirus-310x274-1.jpg\" \/>\n\t<meta property=\"og:image:width\" content=\"310\" \/>\n\t<meta property=\"og:image:height\" content=\"274\" \/>\n\t<meta property=\"og:image:type\" content=\"image\/jpeg\" \/>\n<meta name=\"author\" content=\"onbeat\" \/>\n<meta name=\"twitter:card\" content=\"summary_large_image\" \/>\n<meta name=\"twitter:creator\" content=\"@thermofisher\" \/>\n<meta name=\"twitter:site\" content=\"@thermofisher\" \/>\n<meta name=\"twitter:label1\" content=\"Written by\" \/>\n\t<meta name=\"twitter:data1\" content=\"onbeat\" \/>\n\t<meta name=\"twitter:label2\" content=\"Est. reading time\" \/>\n\t<meta name=\"twitter:data2\" content=\"7 minutes\" \/>\n<script type=\"application\/ld+json\" class=\"yoast-schema-graph\">{\"@context\":\"https:\\\/\\\/schema.org\",\"@graph\":[{\"@type\":\"Article\",\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/clinical-conversations\\\/how-does-a-covid-19-test-work\\\/#article\",\"isPartOf\":{\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/clinical-conversations\\\/how-does-a-covid-19-test-work\\\/\"},\"author\":{\"name\":\"onbeat\",\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/clinical-conversations\\\/#\\\/schema\\\/person\\\/acd0b707cacf9a1ce3873f1b01d8f540\"},\"headline\":\"How does a SARS-CoV-2 (COVID-19) test work?\",\"datePublished\":\"2020-04-02T23:11:58+00:00\",\"dateModified\":\"2022-10-06T23:21:05+00:00\",\"mainEntityOfPage\":{\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/clinical-conversations\\\/how-does-a-covid-19-test-work\\\/\"},\"wordCount\":1501,\"commentCount\":1,\"image\":{\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/clinical-conversations\\\/how-does-a-covid-19-test-work\\\/#primaryimage\"},\"thumbnailUrl\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/clinical-conversations\\\/wp-content\\\/uploads\\\/sites\\\/18\\\/2020\\\/12\\\/COL011389-Coronavirus-310x274-1.jpg\",\"keywords\":[\"content-Article\",\"COVID-19\",\"qPCR\",\"sars-cov-2\"],\"articleSection\":[\"Science and education\"],\"inLanguage\":\"en-US\",\"potentialAction\":[{\"@type\":\"CommentAction\",\"name\":\"Comment\",\"target\":[\"https:\\\/\\\/admin.acceleratingscience.com\\\/clinical-conversations\\\/how-does-a-covid-19-test-work\\\/#respond\"]}]},{\"@type\":\"WebPage\",\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/clinical-conversations\\\/how-does-a-covid-19-test-work\\\/\",\"url\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/clinical-conversations\\\/how-does-a-covid-19-test-work\\\/\",\"name\":\"How does a COVID-19 test work? | Ask a Scientist\",\"isPartOf\":{\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/clinical-conversations\\\/#website\"},\"primaryImageOfPage\":{\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/clinical-conversations\\\/how-does-a-covid-19-test-work\\\/#primaryimage\"},\"image\":{\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/clinical-conversations\\\/how-does-a-covid-19-test-work\\\/#primaryimage\"},\"thumbnailUrl\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/clinical-conversations\\\/wp-content\\\/uploads\\\/sites\\\/18\\\/2020\\\/12\\\/COL011389-Coronavirus-310x274-1.jpg\",\"datePublished\":\"2020-04-02T23:11:58+00:00\",\"dateModified\":\"2022-10-06T23:21:05+00:00\",\"author\":{\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/clinical-conversations\\\/#\\\/schema\\\/person\\\/acd0b707cacf9a1ce3873f1b01d8f540\"},\"description\":\"Understand how the tests used to diagnose COVID-19 are developed, how testing is being performed and how results are being generated.\",\"breadcrumb\":{\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/clinical-conversations\\\/how-does-a-covid-19-test-work\\\/#breadcrumb\"},\"inLanguage\":\"en-US\",\"potentialAction\":[{\"@type\":\"ReadAction\",\"target\":[\"https:\\\/\\\/admin.acceleratingscience.com\\\/clinical-conversations\\\/how-does-a-covid-19-test-work\\\/\"]}]},{\"@type\":\"ImageObject\",\"inLanguage\":\"en-US\",\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/clinical-conversations\\\/how-does-a-covid-19-test-work\\\/#primaryimage\",\"url\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/clinical-conversations\\\/wp-content\\\/uploads\\\/sites\\\/18\\\/2020\\\/12\\\/COL011389-Coronavirus-310x274-1.jpg\",\"contentUrl\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/clinical-conversations\\\/wp-content\\\/uploads\\\/sites\\\/18\\\/2020\\\/12\\\/COL011389-Coronavirus-310x274-1.jpg\",\"width\":310,\"height\":274,\"caption\":\"Coronavirus\"},{\"@type\":\"BreadcrumbList\",\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/clinical-conversations\\\/how-does-a-covid-19-test-work\\\/#breadcrumb\",\"itemListElement\":[{\"@type\":\"ListItem\",\"position\":1,\"name\":\"Home\",\"item\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/clinical-conversations\\\/\"},{\"@type\":\"ListItem\",\"position\":2,\"name\":\"How does a SARS-CoV-2 (COVID-19) test work?\"}]},{\"@type\":\"WebSite\",\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/clinical-conversations\\\/#website\",\"url\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/clinical-conversations\\\/\",\"name\":\"Clinical Conversations\",\"description\":\"Molecular and antibody testing content for clinical lab professionals and clinicians\",\"potentialAction\":[{\"@type\":\"SearchAction\",\"target\":{\"@type\":\"EntryPoint\",\"urlTemplate\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/clinical-conversations\\\/?s={search_term_string}\"},\"query-input\":{\"@type\":\"PropertyValueSpecification\",\"valueRequired\":true,\"valueName\":\"search_term_string\"}}],\"inLanguage\":\"en-US\"},{\"@type\":\"Person\",\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/clinical-conversations\\\/#\\\/schema\\\/person\\\/acd0b707cacf9a1ce3873f1b01d8f540\",\"name\":\"onbeat\",\"url\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/author\\\/onbeat\\\/\"}]}<\/script>\n<!-- \/ Yoast SEO Premium plugin. -->","yoast_head_json":{"title":"How does a COVID-19 test work? | Ask a Scientist","description":"Understand how the tests used to diagnose COVID-19 are developed, how testing is being performed and how results are being generated.","robots":{"index":"index","follow":"follow","max-snippet":"max-snippet:-1","max-image-preview":"max-image-preview:large","max-video-preview":"max-video-preview:-1"},"canonical":"https:\/\/www.thermofisher.com\/blog\/clinical-conversations\/how-does-a-covid-19-test-work\/","og_locale":"en_US","og_type":"article","og_title":"How does a SARS-CoV-2 (COVID-19) test work?","og_description":"Understand how the tests used to diagnose COVID-19 are developed, how testing is being performed and how results are being generated.","og_url":"https:\/\/admin.acceleratingscience.com\/clinical-conversations\/how-does-a-covid-19-test-work\/","og_site_name":"Clinical Conversations","article_publisher":"https:\/\/www.facebook.com\/thermofisher","article_published_time":"2020-04-02T23:11:58+00:00","article_modified_time":"2022-10-06T23:21:05+00:00","og_image":[{"width":310,"height":274,"url":"https:\/\/admin.acceleratingscience.com\/clinical-conversations\/wp-content\/uploads\/sites\/18\/2020\/12\/COL011389-Coronavirus-310x274-1.jpg","type":"image\/jpeg"}],"author":"onbeat","twitter_card":"summary_large_image","twitter_creator":"@thermofisher","twitter_site":"@thermofisher","twitter_misc":{"Written by":"onbeat","Est. reading time":"7 minutes"},"schema":{"@context":"https:\/\/schema.org","@graph":[{"@type":"Article","@id":"https:\/\/admin.acceleratingscience.com\/clinical-conversations\/how-does-a-covid-19-test-work\/#article","isPartOf":{"@id":"https:\/\/admin.acceleratingscience.com\/clinical-conversations\/how-does-a-covid-19-test-work\/"},"author":{"name":"onbeat","@id":"https:\/\/admin.acceleratingscience.com\/clinical-conversations\/#\/schema\/person\/acd0b707cacf9a1ce3873f1b01d8f540"},"headline":"How does a SARS-CoV-2 (COVID-19) test work?","datePublished":"2020-04-02T23:11:58+00:00","dateModified":"2022-10-06T23:21:05+00:00","mainEntityOfPage":{"@id":"https:\/\/admin.acceleratingscience.com\/clinical-conversations\/how-does-a-covid-19-test-work\/"},"wordCount":1501,"commentCount":1,"image":{"@id":"https:\/\/admin.acceleratingscience.com\/clinical-conversations\/how-does-a-covid-19-test-work\/#primaryimage"},"thumbnailUrl":"https:\/\/admin.acceleratingscience.com\/clinical-conversations\/wp-content\/uploads\/sites\/18\/2020\/12\/COL011389-Coronavirus-310x274-1.jpg","keywords":["content-Article","COVID-19","qPCR","sars-cov-2"],"articleSection":["Science and education"],"inLanguage":"en-US","potentialAction":[{"@type":"CommentAction","name":"Comment","target":["https:\/\/admin.acceleratingscience.com\/clinical-conversations\/how-does-a-covid-19-test-work\/#respond"]}]},{"@type":"WebPage","@id":"https:\/\/admin.acceleratingscience.com\/clinical-conversations\/how-does-a-covid-19-test-work\/","url":"https:\/\/admin.acceleratingscience.com\/clinical-conversations\/how-does-a-covid-19-test-work\/","name":"How does a COVID-19 test work? | Ask a Scientist","isPartOf":{"@id":"https:\/\/admin.acceleratingscience.com\/clinical-conversations\/#website"},"primaryImageOfPage":{"@id":"https:\/\/admin.acceleratingscience.com\/clinical-conversations\/how-does-a-covid-19-test-work\/#primaryimage"},"image":{"@id":"https:\/\/admin.acceleratingscience.com\/clinical-conversations\/how-does-a-covid-19-test-work\/#primaryimage"},"thumbnailUrl":"https:\/\/admin.acceleratingscience.com\/clinical-conversations\/wp-content\/uploads\/sites\/18\/2020\/12\/COL011389-Coronavirus-310x274-1.jpg","datePublished":"2020-04-02T23:11:58+00:00","dateModified":"2022-10-06T23:21:05+00:00","author":{"@id":"https:\/\/admin.acceleratingscience.com\/clinical-conversations\/#\/schema\/person\/acd0b707cacf9a1ce3873f1b01d8f540"},"description":"Understand how the tests used to diagnose COVID-19 are developed, how testing is being performed and how results are being generated.","breadcrumb":{"@id":"https:\/\/admin.acceleratingscience.com\/clinical-conversations\/how-does-a-covid-19-test-work\/#breadcrumb"},"inLanguage":"en-US","potentialAction":[{"@type":"ReadAction","target":["https:\/\/admin.acceleratingscience.com\/clinical-conversations\/how-does-a-covid-19-test-work\/"]}]},{"@type":"ImageObject","inLanguage":"en-US","@id":"https:\/\/admin.acceleratingscience.com\/clinical-conversations\/how-does-a-covid-19-test-work\/#primaryimage","url":"https:\/\/admin.acceleratingscience.com\/clinical-conversations\/wp-content\/uploads\/sites\/18\/2020\/12\/COL011389-Coronavirus-310x274-1.jpg","contentUrl":"https:\/\/admin.acceleratingscience.com\/clinical-conversations\/wp-content\/uploads\/sites\/18\/2020\/12\/COL011389-Coronavirus-310x274-1.jpg","width":310,"height":274,"caption":"Coronavirus"},{"@type":"BreadcrumbList","@id":"https:\/\/admin.acceleratingscience.com\/clinical-conversations\/how-does-a-covid-19-test-work\/#breadcrumb","itemListElement":[{"@type":"ListItem","position":1,"name":"Home","item":"https:\/\/admin.acceleratingscience.com\/clinical-conversations\/"},{"@type":"ListItem","position":2,"name":"How does a SARS-CoV-2 (COVID-19) test work?"}]},{"@type":"WebSite","@id":"https:\/\/admin.acceleratingscience.com\/clinical-conversations\/#website","url":"https:\/\/admin.acceleratingscience.com\/clinical-conversations\/","name":"Clinical Conversations","description":"Molecular and antibody testing content for clinical lab professionals and clinicians","potentialAction":[{"@type":"SearchAction","target":{"@type":"EntryPoint","urlTemplate":"https:\/\/admin.acceleratingscience.com\/clinical-conversations\/?s={search_term_string}"},"query-input":{"@type":"PropertyValueSpecification","valueRequired":true,"valueName":"search_term_string"}}],"inLanguage":"en-US"},{"@type":"Person","@id":"https:\/\/admin.acceleratingscience.com\/clinical-conversations\/#\/schema\/person\/acd0b707cacf9a1ce3873f1b01d8f540","name":"onbeat","url":"https:\/\/admin.acceleratingscience.com\/author\/onbeat\/"}]}},"taxonomy_info":{"category":[{"value":10,"label":"Science and education"}],"post_tag":[{"value":121,"label":"content-Article"},{"value":22,"label":"COVID-19"},{"value":90,"label":"qPCR"},{"value":26,"label":"sars-cov-2"}]},"featured_image_src_large":["https:\/\/admin.acceleratingscience.com\/clinical-conversations\/wp-content\/uploads\/sites\/18\/2020\/12\/COL011389-Coronavirus-310x274-1.jpg",310,274,false],"author_info":{"display_name":"onbeat","author_link":"https:\/\/admin.acceleratingscience.com\/author\/onbeat\/"},"comment_info":5,"category_info":[{"term_id":10,"name":"Science and education","slug":"science-and-education","term_group":0,"term_taxonomy_id":10,"taxonomy":"category","description":"","parent":0,"count":118,"filter":"raw","meta":[],"cat_ID":10,"category_count":118,"category_description":"","cat_name":"Science and education","category_nicename":"science-and-education","category_parent":0}],"tag_info":[{"term_id":121,"name":"content-Article","slug":"content-article","term_group":0,"term_taxonomy_id":121,"taxonomy":"post_tag","description":"","parent":0,"count":93,"filter":"raw","meta":[]},{"term_id":22,"name":"COVID-19","slug":"covid-19","term_group":0,"term_taxonomy_id":22,"taxonomy":"post_tag","description":"","parent":0,"count":102,"filter":"raw","meta":[]},{"term_id":90,"name":"qPCR","slug":"qpcr","term_group":0,"term_taxonomy_id":90,"taxonomy":"post_tag","description":"","parent":0,"count":32,"filter":"raw","meta":[]},{"term_id":26,"name":"sars-cov-2","slug":"sars-cov-2","term_group":0,"term_taxonomy_id":26,"taxonomy":"post_tag","description":"","parent":0,"count":73,"filter":"raw","meta":[]}],"_links":{"self":[{"href":"https:\/\/www.thermofisher.com\/blog\/clinical-conversations\/wp-json\/wp\/v2\/posts\/249","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/www.thermofisher.com\/blog\/clinical-conversations\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/www.thermofisher.com\/blog\/clinical-conversations\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/clinical-conversations\/wp-json\/wp\/v2\/users\/1556"}],"replies":[{"embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/clinical-conversations\/wp-json\/wp\/v2\/comments?post=249"}],"version-history":[{"count":0,"href":"https:\/\/www.thermofisher.com\/blog\/clinical-conversations\/wp-json\/wp\/v2\/posts\/249\/revisions"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/clinical-conversations\/wp-json\/wp\/v2\/media\/15840"}],"wp:attachment":[{"href":"https:\/\/www.thermofisher.com\/blog\/clinical-conversations\/wp-json\/wp\/v2\/media?parent=249"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/clinical-conversations\/wp-json\/wp\/v2\/categories?post=249"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/clinical-conversations\/wp-json\/wp\/v2\/tags?post=249"},{"taxonomy":"division","embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/clinical-conversations\/wp-json\/wp\/v2\/division?post=249"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}