{"id":1461,"date":"2018-10-19T00:00:00","date_gmt":"2018-10-18T15:00:00","guid":{"rendered":"https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/cad_hplc\/"},"modified":"2023-12-28T10:55:43","modified_gmt":"2023-12-28T01:55:43","slug":"cad_hplc","status":"publish","type":"post","link":"https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/cad_hplc\/","title":{"rendered":"\u8ab0\u3067\u3082\u308f\u304b\u308b\uff01\u8377\u96fb\u5316\u7c92\u5b50\u691c\u51fa\u5668\u306e\u57fa\u790e"},"content":{"rendered":"<h2>\u306f\u3058\u3081\u306b<\/h2>\n<p>\u6db2\u4f53\u30af\u30ed\u30de\u30c8\u30b0\u30e9\u30d5\u30a3\u30fc\uff08HPLC\uff09\u3067\u4f7f\u308f\u308c\u3066\u3044\u308b\u691c\u51fa\u5668\u3092\u8abf\u3079\u3066\u3044\u3066\u300c\u8377\u96fb\u5316\u7c92\u5b50\u691c\u51fa\u5668\u300d\u3068\u3044\u3046\u3042\u307e\u308a\u805e\u304d\u306a\u308c\u306a\u3044\u691c\u51fa\u5668\u3092\u898b\u3064\u3051\u305f\u3053\u3068\u304c\u3042\u308b\u65b9\u3082\u591a\u3044\u306e\u3067\u306f\u306a\u3044\u3067\u3057\u3087\u3046\u304b\uff1f<\/p>\n<p><strong>\u300c\u8377\u96fb\u5316\u7c92\u5b50\u691c\u51fa\u5668\u306b\u3064\u3044\u3066\u77e5\u308a\u305f\u3044\u300d<\/strong><\/p>\n<p><strong>\u300c\u4f7f\u3044\u52dd\u624b\u306e\u826f\u3044\u6c4e\u7528\u30bf\u30a4\u30d7\u306e\u691c\u51fa\u5668\u3092\u63a2\u3057\u3066\u3044\u308b\u300d<\/strong><\/p>\n<p>\u3053\u306e\u8a18\u4e8b\u306f\u305d\u3093\u306a\u65b9\u306b\u5411\u3051\u305f\u8a18\u4e8b\u3067\u3059\u3002<\/p>\n<p><a class=\"btn btn-red\" href=\"https:\/\/www.thermofisher.com\/jp\/ja\/home\/global\/forms\/industrial\/042421-cad-lab-form-blog.html?ICID=cmd_hplc_r04_jp_cad_hplc_Social_LAB\" target=\"_blank\" rel=\"noopener noreferrer\">PDF\u30c0\u30a6\u30f3\u30ed\u30fc\u30c9<\/a><\/p>\n<p>\u306f\u3058\u3081\u307e\u3057\u3066\u3002<br \/>\n\u79c1\u306f\u30b5\u30fc\u30e2\u30d5\u30a3\u30c3\u30b7\u30e3\u30fc\u30b5\u30a4\u30a8\u30f3\u30c6\u30a3\u30d5\u30a3\u30c3\u30af\u682a\u5f0f\u4f1a\u793e \u30af\u30ed\u30de\u30c8\u30b0\u30e9\u30d5\u30a3\u30fc&amp;MS\u4e8b\u696d\u90e8\u306b\u6240\u5c5e\u3057\u3066\u3044\u308b\u7530\u53e3\u3068\u7533\u3057\u307e\u3059\u3002<\/p>\n<p>\u304a\u5ba2\u69d8\u306e\u30af\u30ed\u30de\u30c8\u30b0\u30e9\u30d5\u30a3\u30fc&amp;MS\u30e9\u30a4\u30d5\u304c\u3069\u3046\u3057\u305f\u3089\u3088\u308a\u697d\u3057\u3044\u3082\u306e\u306b\u306a\u308b\u306e\u304b\u3001\u611b\u3059\u308b\u30af\u30ed\u30de\u30c8\u30b0\u30e9\u30d5\u30a3\u30fc\u6280\u8853\u3068\u3068\u3082\u306b\u65e5\u3005\u3055\u307e\u3056\u307e\u306a\u3053\u3068\u306b\u596e\u95d8\u3057\u3066\u304a\u308a\u307e\u3059\u3002\u5b9f\u306f\u4ed5\u4e8b\u67c4\u3001\u30bb\u30df\u30ca\u30fc\u306a\u3069\u3067\u8377\u96fb\u5316\u7c92\u5b50\u691c\u51fa\u5668\u306b\u3064\u3044\u3066\u3054\u8aac\u660e\u3055\u305b\u3066\u9802\u304f\u3053\u3068\u3082\u591a\u3044\u306e\u3067\u3059\u304c\u3001\u3053\u306e\u691c\u51fa\u5668\u306e\u8208\u5473\u6df1\u3044\u3068\u3053\u308d\u306f<strong>\u4ed6\u306e\u691c\u51fa\u5668\u306b\u306f\u306a\u3044\u7279\u5fb4<\/strong>\u304c\u3042\u308b\u3068\u3044\u3046\u3068\u3053\u308d\u3067\u3059\u3002<\/p>\n<p>\u305d\u3093\u306a\u8377\u96fb\u5316\u7c92\u5b50\u691c\u51fa\u5668\u3001\u3055\u307e\u3056\u307e\u306a\u5206\u91ce\u3067\u30eb\u30fc\u30c1\u30f3\u5206\u6790\u306b\u4f7f\u7528\u3055\u308c\u3066\u3044\u308b\u3053\u3068\u3092\u3054\u5b58\u77e5\u3067\u3057\u3087\u3046\u304b\uff1f<\/p>\n<p>\u3053\u306e\u8a18\u4e8b\u3067\u306f\u8377\u96fb\u5316\u7c92\u5b50\u691c\u51fa\u5668\u306e\u57fa\u790e\u3068\u984c\u3057\u307e\u3057\u3066\u3001\u305d\u306e\u7279\u5fb4\u304c\u3088\u304f\u5206\u304b\u308b\u3088\u3046\u306b\u3001HPLC\u3067\u4e00\u822c\u7684\u306b\u4f7f\u7528\u3055\u308c\u308b\u691c\u51fa\u5668\u3068\u306e\u6bd4\u8f03\u306b\u52a0\u3048\u3001\u4ee3\u8868\u7684\u306a\u30a2\u30d7\u30ea\u30b1\u30fc\u30b7\u30e7\u30f3\u4f8b\u3082\u3054\u7d39\u4ecb\u3057\u307e\u3059\u3002\u30af\u30ed\u30de\u30c8\u30b0\u30e9\u30d5\u30a3\u30fc\u6280\u8853\u3092\u611b\u3059\u308b\u8005\u3068\u3057\u3066\u306e\u8996\u70b9\u3067\u3001\u4ed6\u306e\u30b5\u30a4\u30c8\u306b\u306f\u8f09\u3063\u3066\u3044\u306a\u3044\u3088\u3046\u306a\u8a73\u3057\u3044\u60c5\u5831\u3092\u5206\u304b\u308a\u3084\u3059\u304f\u304a\u8a71\u3057\u3088\u3046\u3068\u601d\u3044\u307e\u3059\u3002<\/p>\n<h2>\u6db2\u4f53\u30af\u30ed\u30de\u30c8\u30b0\u30e9\u30d5\u30a3\u30fc\u306e\u691c\u51fa\u5668<\/h2>\n<p>\u4e00\u822c\u7684\u306b\u6db2\u4f53\u30af\u30ed\u30de\u30c8\u30b0\u30e9\u30d5\u30a3\u30fc\u3067\u5e83\u304f\u4f7f\u308f\u308c\u3066\u3044\u308b\u691c\u51fa\u5668\u306b\u306f\u4ee5\u4e0b\u306e\u3088\u3046\u306a\u7a2e\u985e\u304c\u3042\u308a\u307e\u3059\u3002\u5316\u5408\u7269\u304c\u691c\u51fa\u3067\u304d\u308b\u304b\u3069\u3046\u304b\u3067\u9078\u629e\u3059\u308b\u691c\u51fa\u5668\u306f\u5909\u308f\u308b\u305f\u3081\u3001\u7c21\u5358\u306b<strong>\u5404\u691c\u51fa\u5668\u306e\u7279\u5fb4\u3068\u691c\u51fa\u3067\u304d\u308b\u7269\u8cea<\/strong>\u306b\u3064\u3044\u3066\u307e\u3068\u3081\u307e\u3057\u305f\u3002<\/p>\n<ul>\n<li>\u7d2b\u5916\u53ef\u8996\u5438\u5149\u5149\u5ea6\uff08UV\/Vis\uff09\u691c\u51fa\u5668 \uff1a190\uff5e900 nm\u306e\u5149\u3092\u5438\u53ce\u3059\u308b\u7269\u8cea\u304c\u6e2c\u5b9a\u5bfe\u8c61\u3002\u7279\u5b9a\u306e\u6ce2\u9577\u306e\u5149\u3092\u5438\u53ce\u3059\u308b\u7269\u8cea\u3060\u3051\u3092\u691c\u51fa\u3059\u308b\u306e\u3067\u9078\u629e\u6027\uff08\u7269\u8cea\u306b\u5bfe\u3059\u308b\u8b58\u5225\u80fd\u529b\uff09\u304c\u3042\u308b\u3002<\/li>\n<li>\u86cd\u5149\u691c\u51fa\u5668\uff1a\u5149\u3092\u5438\u53ce\uff08\u52b1\u8d77\uff09\u3001\u5149\u3092\u767a\u3059\u308b\uff08\u86cd\u5149\uff09\u7269\u8cea\u304c\u6e2c\u5b9a\u5bfe\u8c61\u3002\u7279\u5b9a\u6ce2\u9577\u3067\u52b1\u8d77\u30fb\u86cd\u5149\u3059\u308b\u7269\u8cea\u3060\u3051\u3092\u691c\u51fa\u3059\u308b\u306e\u3067\u9078\u629e\u6027\u304c\u9ad8\u3044\u3002<\/li>\n<li>\u8cea\u91cf\u5206\u6790\u8a08\uff1a\u30a4\u30aa\u30f3\u5316\u3059\u308b\u7269\u8cea\u304c\u6e2c\u5b9a\u5bfe\u8c61\u3002\u7269\u8cea\u306e\u8cea\u91cf\u3068\u96fb\u8377\u306b\u5fdc\u3058\u305f\u691c\u51fa\u304c\u3067\u304d\u308b\u305f\u3081\u3001\u9078\u629e\u6027\u304c\u3068\u3066\u3082\u9ad8\u3044\u3002\u3067\u3082\u3001\u5024\u6bb5\u3082\u9ad8\u3044\u3002<\/li>\n<li>\u793a\u5dee\u5c48\u6298\u7387\uff08RI\uff09\u691c\u51fa\u5668\uff1a\u79fb\u52d5\u76f8\u3068\u7570\u306a\u308b\u5149\u306e\u5c48\u6298\u7387\u3092\u6709\u3059\u308b\u7269\u8cea\u304c\u6e2c\u5b9a\u5bfe\u8c61\u3002\u307b\u3068\u3093\u3069\u306e\u7269\u8cea\u306f\u79fb\u52d5\u76f8\u3068\u7570\u306a\u308b\u5c48\u6298\u7387\u3092\u6301\u3064\u306e\u3067\u3001\u3042\u3089\u3086\u308b\u6210\u5206\u304c\u691c\u51fa\u53ef\u80fd\uff08\u6c4e\u7528\u6027\u304c\u9ad8\u3044\uff09\u3002<\/li>\n<\/ul>\n<h2>\u9078\u629e\u6027\u3068\u6c4e\u7528\u6027<\/h2>\n<p>\u3053\u306e\u3088\u3046\u306bHPLC\u306e\u691c\u51fa\u5668\u306b\u306f\u3055\u307e\u3056\u307e\u306a\u7279\u5fb4\u304c\u3042\u308a\u307e\u3059\u3002<\/p>\n<p>\u9078\u629e\u6027\u3092\u6301\u3064\u691c\u51fa\u5668\u306e\u5834\u5408\u3001\u611f\u5ea6\u3092\u5f97\u308b\u3068\u3044\u3046\u9762\u3067\u306f\u5927\u304d\u306a\u5229\u70b9\u304c\u3042\u308b\u306e\u3067\u3059\u304c\u3001\u3069\u3046\u3057\u3066\u3082\u82e6\u624b\u306a\u3082\u306e\uff08\u691c\u51fa\u3067\u304d\u306a\u3044\u6210\u5206\uff09\u304c\u51fa\u3066\u304d\u3066\u3057\u307e\u3046\u3068\u3044\u3046\u6b20\u70b9\u304c\u3042\u308a\u307e\u3059\u3002<span style=\"text-decoration: underline\">\u691c\u51fa\u3067\u304d\u308b\u304b\u3069\u3046\u304b\u306f\u5316\u5408\u7269\u306e\u7279\u6027\u306b\u3088\u3063\u3066\u6c7a\u307e\u3063\u3066\u3057\u307e\u3046\u306e\u3067\u3059<\/span>\u3002<\/p>\n<p>\u4f8b\u3048\u3070\u3001UV\u5438\u53ce\u304c\u4e0d\u5341\u5206\u306a\u6210\u5206\u3084\u3001\u86cd\u5149\u3092\u767a\u3057\u306a\u3044\u6210\u5206\u3001\u3055\u3089\u306b\u30a4\u30aa\u30f3\u5316\u3057\u306a\u3044\u6210\u5206\u3068\u3044\u3046\u306e\u306f\u305f\u304f\u3055\u3093\u5b58\u5728\u3057\u307e\u3059\u3002\u3082\u3057\u904b\u60aa\u304f\u305d\u3046\u3044\u3063\u305f\u7269\u8cea\u306b\u51fa\u4f1a\u3063\u305f\u3089\u3069\u3046\u3057\u305f\u3089\u3088\u3044\u306e\u3067\u3057\u3087\u3046\u304b\uff1f<\/p>\n<p>\u300c\u8a98\u5c0e\u4f53\u5316\u300d\u3068\u3044\u3046\u9078\u629e\u80a2\u304c\u9078\u3079\u308b\u5834\u5408\u3082\u3042\u308a\u307e\u3059\u304c\u3001\u8a98\u5c0e\u4f53\u5316\u8a66\u85ac\u306e\u9078\u629e\u3084\u524d\u51e6\u7406\u306e\u624b\u9593\u3092\u8003\u3048\u308b\u3068\u3001\u6b63\u76f4\u56f0\u308b\u3053\u3068\u304c\u591a\u3044\u3067\u3059\u3002<\/p>\n<p>\u305d\u306e\u3088\u3046\u306a\u5834\u5408\u3001\u6c4e\u7528\u6027\u304c\u9ad8\u3044RI\u691c\u51fa\u5668\u3067\u3042\u308c\u3070\u3001\u5316\u5408\u7269\u306e\u7279\u6027\u3092\u6c17\u306b\u3059\u308b\u3053\u3068\u306a\u304f\u7c21\u5358\u306b\u5206\u6790\u3067\u304d\u307e\u3059\u3002\u3057\u304b\u3057\u3001RI\u306f\u4ed6\u306e\u691c\u51fa\u5668\u306b\u6bd4\u3079\u308b\u3068\u611f\u5ea6\u304c\u4f4e\u304f\u3001\u307e\u305f\u30b0\u30e9\u30b8\u30a8\u30f3\u30c8\u5206\u6790\u304c\u51fa\u6765\u307e\u305b\u3093\u3002\u305d\u306e\u7d50\u679c\u3001\u5341\u5206\u306a\u611f\u5ea6\u3084\u5206\u96e2\u3092\u5f97\u308b\u3053\u3068\u304c\u96e3\u3057\u3044\u3068\u3044\u3046\u8ab2\u984c\u304c\u3042\u308a\u307e\u3059\u3002<\/p>\n<p>\u3053\u306e\u3088\u3046\u306a\u8ab2\u984c\u306b\u76f4\u9762\u3057\u305f\u5834\u5408\u3001RI\u691c\u51fa\u5668\u4ee5\u5916\u306e\u9078\u629e\u80a2\u3068\u3057\u3066\u601d\u3044\u51fa\u3057\u3066\u9802\u304d\u305f\u3044\u6c4e\u7528\u691c\u51fa\u5668\u304c\u8377\u96fb\u5316\u7c92\u5b50\u691c\u51fa\u5668\u306b\u306a\u308a\u307e\u3059\u3002<\/p>\n<h2>\u8377\u96fb\u5316\u7c92\u5b50\u691c\u51fa\u5668\u306e\u539f\u7406\u3068\u7279\u5fb4<\/h2>\n<p>\u8377\u96fb\u5316\u7c92\u5b50\u691c\u51fa\u5668\uff08CAD\uff1aCharged Aerosol Detector\uff09\u306f\u3001\u5316\u5408\u7269\u306e\u7279\u6027\u306b\u5927\u304d\u304f\u5de6\u53f3\u3055\u308c\u306a\u3044\u691c\u51fa\u539f\u7406\u3068\u611f\u5ea6\u3092\u6301\u3061\u3001\u30b0\u30e9\u30b8\u30a8\u30f3\u30c8\u306b\u3088\u3063\u3066\u30ab\u30e9\u30e0\u306e\u5206\u96e2\u6027\u80fd\u3092\u6700\u5927\u9650\u6d3b\u304b\u3059\u3053\u3068\u306e\u3067\u304d\u308b\u6c4e\u7528\u691c\u51fa\u5668\u3067\u3059\u3002<\/p>\n<p>\u539f\u7406\u306b\u3064\u3044\u3066\u691c\u51fa\u307e\u3067\u306e\u30b9\u30c6\u30c3\u30d7\u3068\u4f75\u305b\u3066\u7c21\u5358\u306b\u8aac\u660e\u3057\u307e\u3059\u3002<a href=\"https:\/\/www.thermofisher.com\/blog\/wp-content\/uploads\/sites\/13\/2019\/07\/9be9e9560e88dfed50ec85a2294d365f.jpg\"><img loading=\"lazy\" decoding=\"async\" class=\"size-large wp-image-4331 aligncenter\" src=\"https:\/\/www.thermofisher.com\/blog\/wp-content\/uploads\/sites\/13\/2019\/07\/9be9e9560e88dfed50ec85a2294d365f-1024x622.jpg\" alt=\"\u8377\u96fb\u5316\u7c92\u5b50\u691c\u51fa\u5668\u306e\u57fa\u790e\" width=\"1024\" height=\"622\" \/><\/a><\/p>\n<h5>Step1\uff1a\u5674\u9727\u5316\uff20\u30cd\u30d6\u30e9\u30a4\u30b6\u30fc<\/h5>\n<p>\u30ab\u30e9\u30e0\u304b\u3089\u306e\u6eb6\u51fa\u6db2\u3092\u7a92\u7d20\u30ac\u30b9\u3067\u5674\u9727\u3059\u308b\u4e8b\u306b\u3088\u308a\u3001\u7d30\u304b\u306a\u6db2\u6ef4\u3092\u3064\u304f\u308a\u307e\u3059\u3002<\/p>\n<h5>Step2\uff1a\u8131\u6eb6\u5a92\u30fb\u7c92\u5b50\u5316\uff20\u30c9\u30e9\u30a4\u30f3\u30b0\u30c1\u30e5\u30fc\u30d6<\/h5>\n<p>\u52a0\u6e29\u306b\u3088\u308a\u6db2\u6ef4\u306e\u6eb6\u5a92\u304c\u6c17\u5316\u3057\uff08\u8131\u6eb6\u5a92\uff09\u3001\u6b8b\u3063\u305f\u6210\u5206\u304c\u7c92\u5b50\u306b\u306a\u308a\u307e\u3059\u3002<\/p>\n<h5>Step3\uff1a\u8377\u96fb\u5316\uff20\u30df\u30ad\u30b7\u30f3\u30b0\u30c1\u30e3\u30f3\u30d0\u30fc<\/h5>\n<p>\u751f\u6210\u3057\u305f\u7c92\u5b50\u306b\u30c1\u30e3\u30fc\u30b8\u3057\u305f\u7a92\u7d20\u30ac\u30b9\u3092\u885d\u7a81\u3055\u305b\u3001\u7c92\u5b50\u8868\u9762\u306b\u9759\u96fb\u7684\u306a\u30d7\u30e9\u30b9\u306e\u96fb\u8377\u3092\u307e\u3068\u308f\u305b\u307e\u3059\u3002<\/p>\n<h5>Step\uff14\uff1a\u691c\u51fa\uff20\u30b3\u30ec\u30af\u30bf\u30fc<\/h5>\n<p>\u307e\u3068\u308f\u305b\u305f\u96fb\u8377\u3092\u653e\u96fb\u3057\u3001\u96fb\u6d41\u5024\u3068\u3057\u3066\u8a08\u6e2c\u3057\u307e\u3059\u3002<\/p>\n<p>CAD\u306f\u30cd\u30d6\u30e9\u30a4\u30b6\u30fc\u3067\u5674\u9727\u3057\u305f\u5f8c\u3001\u8131\u6eb6\u5a92\u3057\u3066\u304b\u3089\u691c\u51fa\u3092\u884c\u3044\u307e\u3059\u3002\u305d\u306e\u305f\u3081\u3001\u63ee\u767a\u6027\u7269\u8cea\u306f\u691c\u51fa\u3067\u304d\u307e\u305b\u3093\u304c\u3001<strong>\u4e0d\u63ee\u767a\u6027\u7269\u8cea\u3067\u3042\u308c\u3070\u3059\u3079\u3066\u691c\u51fa\u5bfe\u8c61<\/strong>\u306b\u306a\u308a\u307e\u3059\u3002\u307e\u305f\u3001RI\u3068\u306f\u7570\u306a\u308a\u3001\u30b0\u30e9\u30b8\u30a8\u30f3\u30c8\u5206\u6790\u304c\u53ef\u80fd\u3067\u3059\u3002\u3055\u3089\u306b\u3001CAD\u3067\u306f\u7c92\u5b50\u306b\u96fb\u8377\u3092\u4e0e\u3048\u305d\u306e\u96fb\u8377\u91cf\u3092\u6e2c\u5b9a\u3059\u308b\u3053\u3068\u304b\u3089\u3001<strong>\u7c92\u5b50\u306e\u5927\u304d\u3055\u3084\u91cf\uff08\u6210\u5206\u91cf\uff09\u306b\u5fdc\u3058\u305f\u9762\u7a4d\u5024\u3084\u9ad8\u3055\u304c\u5f97\u3089\u308c\u307e\u3059\u3002<\/strong><\/p>\n<p>\u540c\u3058\u5206\u6790\u6761\u4ef6\u3067\u691c\u51fa\u5668\u306b\u5165\u308b\u91cf\u304c\u540c\u3058\u3067\u3042\u308c\u3070\u3001CAD\u306f\u7269\u8cea\u306e\u7279\u6027\u306b\u95a2\u308f\u3089\u305a<strong>\u307b\u307c\u540c\u4e00\u306e\u9762\u7a4d\u5024<\/strong>\u3092\u793a\u3057\u307e\u3059\u3002<\/p>\n<p>\u9078\u629e\u6027\u3092\u6709\u3059\u308b\u691c\u51fa\u5668\u3067\u306f\u5316\u5b66\u69cb\u9020\u306b\u4f9d\u5b58\u3057\u3066\u5fdc\u7b54\u611f\u5ea6\u304c\u7570\u306a\u308b\u306e\u3067\u3001\u7121\u4f5c\u70ba\u306b\u7269\u8cea\u3092\u9078\u629e\u30fb\u6e2c\u5b9a\u3059\u308b\u3068\u6fc3\u5ea6\u304c\u4e00\u7dd2\u3067\u3082\u3001\u5316\u5408\u7269\u306b\u3088\u3063\u3066\u9762\u7a4d\u5024\u3084\u9ad8\u3055\u304c\u5909\u308f\u308a\u307e\u3059\u3002\u4e00\u4f8b\u3068\u3057\u3066\u3001UV\u691c\u51fa\u5668\u3067\u3042\u308c\u3070\u5316\u5408\u7269\u306b\u3088\u3063\u306660\uff05\u7a0b\u5ea6\u9762\u7a4d\u5024\u304c\u5909\u308f\u308a\u307e\u3059\u304c\u3001CAD\u3067\u306f10%\u7a0b\u5ea6\u3068\u306a\u308a\u307e\u3059\u3002<\/p>\n<p><a href=\"https:\/\/www.thermofisher.com\/blog\/wp-content\/uploads\/sites\/13\/2019\/07\/8d2cf6dde49984ce011a7a4d9d5fb3f6.jpg\"><img loading=\"lazy\" decoding=\"async\" class=\"aligncenter wp-image-4329 \" src=\"https:\/\/www.thermofisher.com\/blog\/wp-content\/uploads\/sites\/13\/2019\/07\/8d2cf6dde49984ce011a7a4d9d5fb3f6.jpg\" alt=\"\u5404\u6210\u5206\u306e\u9762\u7a4d\u5024\" width=\"705\" height=\"473\" \/><\/a><\/p>\n<p>\u3053\u306e\u7279\u5fb4\u3092\u5229\u7528\u3059\u308b\u3068\u3001\u5225\u306e\u7269\u8cea\u3067\u4f5c\u6210\u3057\u305f\u691c\u91cf\u7dda\u3092\u4f7f\u3063\u3066\u5b9a\u91cf\uff08\u534a\u5b9a\u91cf\uff09\u304c\u884c\u3048\u308b\u306e\u3067\u3001\u6a19\u6e96\u8a66\u6599\u306e\u306a\u3044\u7269\u8cea\u306e\u91cf\u3092\u76f8\u5bfe\u9762\u7a4d\u5024\u304b\u3089\u63a8\u5b9a\u3059\u308b\u3053\u3068\u3082\u53ef\u80fd\u3067\u3059\u3002<\/p>\n<p>\u3061\u306a\u307f\u306b\u3001\u691c\u51fa\u4e0b\u9650\u306f\u6570\u767eppb\u7a0b\u5ea6\u3067\u6bd4\u8f03\u7684\u826f\u597d\u306a\u691c\u51fa\u611f\u5ea6\u304c\u5f97\u3089\u308c\u307e\u3059\u3002<\/p>\n<p>\u3053\u306e\u3088\u3046\u306b\u3001CAD\u306f\u5e45\u5e83\u3044\u5316\u5408\u7269\u306b\u5bfe\u5fdc\u3067\u304d\u308b\u3068\u3068\u3082\u306b\u5316\u5408\u7269\u306b\u4f9d\u5b58\u3057\u306a\u3044\u4e00\u8cab\u3057\u305f\u5fdc\u7b54\u6027\u3092\u6709\u3057\u305f\u30e6\u30cb\u30fc\u30af\u306a\u691c\u51fa\u5668\u306b\u306a\u308a\u307e\u3059\u3002<\/p>\n<p>&nbsp;<\/p>\n<h2>\u8377\u96fb\u5316\u7c92\u5b50\u691c\u51fa\u5668\u306e\u30a2\u30d7\u30ea\u30b1\u30fc\u30b7\u30e7\u30f3<\/h2>\n<p>\u691c\u51fa\u5bfe\u8c61\u7269\u304c\u5e45\u5e83\u3044\u3053\u3068\u304b\u3089\u3001CAD\u306e\u30a2\u30d7\u30ea\u2015\u30b1\u30fc\u30b7\u30e7\u30f3\u306f\u6709\u6a5f\u5316\u5408\u7269\u304b\u3089\u7121\u6a5f\u30a4\u30aa\u30f3\u307e\u3067\u591a\u5c90\u306b\u6e21\u308a\u307e\u3059\u3002<\/p>\n<p><a href=\"https:\/\/www.thermofisher.com\/blog\/wp-content\/uploads\/sites\/13\/2019\/07\/c3e345582c8d3cc0cdd21dec849afe25.jpg\"><img loading=\"lazy\" decoding=\"async\" class=\"aligncenter wp-image-4844\" src=\"https:\/\/www.thermofisher.com\/blog\/wp-content\/uploads\/sites\/13\/2019\/07\/c3e345582c8d3cc0cdd21dec849afe25-1024x576.jpg\" alt=\"\u4ee3\u8868\u7684\u306a\u30a2\u30d7\u30ea\u30b1\u30fc\u30b7\u30e7\u30f3\" width=\"707\" height=\"378\" \/><\/a><\/p>\n<p>&nbsp;<\/p>\n<p>\u7cd6\u5206\u6790\u3084\u3001\u533b\u85ac\u54c1\u3068\u30ab\u30a6\u30f3\u30bf\u30fc\u30a4\u30aa\u30f3\u3001\u7121\u6a5f\u30a4\u30aa\u30f3\u306e\u4e00\u6589\u5206\u6790\u3001\u754c\u9762\u6d3b\u6027\u5264\u5206\u6790\u306a\u3069\u306f\u3055\u307e\u3056\u307e\u306a\u304a\u5ba2\u69d8\u3067\u3054\u5229\u7528\u3044\u305f\u3060\u3044\u3066\u3044\u308b\u30a2\u30d7\u30ea\u30b1\u30fc\u30b7\u30e7\u30f3\u306b\u306a\u308a\u307e\u3059\u3002<\/p>\n<p>\u4e00\u65b9\u3067\u3001\u6210\u5206\u306e\u542b\u6709\u91cf\u306b\u6bd4\u4f8b\u3057\u305f\u5fdc\u7b54\u6027\uff08\u91cd\u91cf\u5fdc\u7b54\u6027\uff09\u3092\u5229\u7528\u3057\u305f\u30a2\u30d7\u30ea\u30b1\u30fc\u30b7\u30e7\u30f3\u3068\u3057\u3066\u306f\u3001\u6a19\u6e96\u8a66\u6599\u304c\u5b58\u5728\u3057\u306a\u3044\u5316\u5408\u7269\u3001\u4f8b\u3048\u3070<strong>\u4e0d\u7d14\u7269\u3084\u5929\u7136\u7269\u306e\u542b\u91cf\u78ba\u8a8d<\/strong>\u304c\u6319\u3052\u3089\u308c\u307e\u3059\u3002\u305d\u308c\u4ee5\u5916\u306b\u3082\u3001\u88fd\u5264\u306e\u82db\u9177\u8a66\u9a13\u306a\u3069\u3001<strong>\u7d99\u6642\u7684\u306a\u7269\u8cea\u306e\u91cf\u7684\u5909\u5316\u306e\u30e2\u30cb\u30bf\u30fc<\/strong>\u306b\u5229\u7528\u3055\u308c\u3066\u3044\u307e\u3059\u3002<\/p>\n<p>\u3053\u306e\u3088\u3046\u306b\u30a2\u30a4\u30c7\u30a3\u30a2\u6b21\u7b2c\u3067\u3055\u307e\u3056\u307e\u306a\u7528\u9014\u306b\u5fdc\u7528\u3067\u304d\u308bCAD\u3067\u3059\u304c\u3001\u5f0a\u793eLC\u4ee5\u5916\u306b\u3082\u3001Shimadzu\uff08\u5cf6\u6d25\uff09\u88fd\u3084Waters\u88fd\u306a\u3069\u3001\u4ed6\u793eHPLC\u30b7\u30b9\u30c6\u30e0\u306b\u3082\u63a5\u7d9a\u53ef\u80fd\u3067\u3059\u3002<\/p>\n<h2>\u307e\u3068\u3081<\/h2>\n<p>CAD\u306e\u539f\u7406\u306b\u3064\u3044\u3066\u7c21\u5358\u306b\u3054\u7d39\u4ecb\u3057\u3066\u304d\u307e\u3057\u305f\u3002\u30e9\u30dc\u3067\u5e83\u304f\u4f7f\u308f\u308c\u3066\u3044\u308b\u691c\u51fa\u5668\u3067\u56f0\u3063\u305f\u3068\u304d\u306f\u3001\u662f\u975eCAD\u3092\u601d\u3044\u51fa\u3057\u3066\u4e0b\u3055\u3044\uff01<\/p>\n<p>\u3010CAD\u306e\u7279\u5fb4\u3011<\/p>\n<ul>\n<li>\u30ab\u30e9\u30e0\u691c\u51fa\u7269\u306e\u5674\u9727\u30fb\u6c17\u5316\u306b\u3088\u308a\u751f\u6210\u3057\u305f\u7c92\u5b50\u306b\u9759\u96fb\u7684\u306a\u30d7\u30e9\u30b9\u30c1\u30e3\u30fc\u30b8\u3092\u4e0e\u3048\u3001\u305d\u306e\u96fb\u8377\u91cf\u3092\u8a08\u6e2c\u3059\u308b\u539f\u7406\u3092\u6709\u3059\u308b<\/li>\n<li>\u3042\u3089\u3086\u308b\u4e0d\u63ee\u767a\u6027\u7269\u8cea\u304c\u691c\u51fa\u5bfe\u8c61<\/li>\n<li>\u5316\u5b66\u69cb\u9020\u306b\u4f9d\u5b58\u3057\u306a\u3044\u4e00\u8cab\u3057\u305f\u5fdc\u7b54\u6027<\/li>\n<\/ul>\n<p>CAD\u306b\u3064\u3044\u3066\u3082\u3046\u5c11\u3057\u8a73\u3057\u3044\u60c5\u5831\u3092\u77e5\u308a\u305f\u3044\u3068\u3044\u3046\u65b9\u306f\u3001\u95a2\u9023\u60c5\u5831\u3092\u3054\u89a7\u3044\u305f\u3060\u304f\u304b\u3001\u304a\u6c17\u8efd\u306b\u304a\u554f\u3044\u5408\u308f\u305b\u3044\u305f\u3060\u3051\u308c\u3070\u3068\u601d\u3044\u307e\u3059\u3002<\/p>\n<h2>\u95a2\u9023\u60c5\u5831<\/h2>\n<p><a href=\"https:\/\/www.thermofisher.com\/order\/catalog\/product\/VF-D20-A?ICID=cmd_hplc_r04_jp_cad_hplc_Social_LAB\" target=\"_blank\" rel=\"noopener noreferrer\">\u88fd\u54c1\u7d39\u4ecb<\/a><\/p>\n<p><a href=\"https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/lc-cad_voc\/?ICID=cmd_hplc_r04_jp_cad_hplc_Social_LAB\" target=\"_blank\" rel=\"noopener noreferrer\">\u304a\u5ba2\u69d8\u4e8b\u4f8b\uff1a\u534a\u5c0e\u4f53\u88fd\u9020<\/a><\/p>\n<p>&nbsp;<\/p>\n<p>\u7814\u7a76\u7528\u306b\u306e\u307f\u4f7f\u7528\u3067\u304d\u307e\u3059\u3002\u8a3a\u65ad\u7528\u306b\u306f\u4f7f\u7528\u3044\u305f\u3060\u3051\u307e\u305b\u3093\u3002<\/p>\n","protected":false},"excerpt":{"rendered":"<p>\u306f\u3058\u3081\u306b \u6db2\u4f53\u30af\u30ed\u30de\u30c8\u30b0\u30e9\u30d5\u30a3\u30fc\uff08HPLC\uff09\u3067\u4f7f\u308f\u308c\u3066\u3044\u308b\u691c\u51fa\u5668\u3092\u8abf\u3079\u3066\u3044\u3066\u300c\u8377\u96fb\u5316\u7c92\u5b50\u691c\u51fa\u5668\u300d\u3068\u3044\u3046\u3042\u307e\u308a\u805e\u304d\u306a\u308c\u306a\u3044\u691c\u51fa\u5668\u3092\u898b\u3064\u3051\u305f\u3053\u3068\u304c\u3042\u308b\u65b9\u3082\u591a\u3044\u306e\u3067\u306f\u306a\u3044\u3067\u3057\u3087\u3046\u304b\uff1f \u300c\u8377\u96fb\u5316\u7c92\u5b50\u691c\u51fa\u5668\u306b\u3064\u3044\u3066\u77e5\u308a\u305f\u3044\u300d \u300c\u4f7f<\/p>\n","protected":false},"author":988,"featured_media":4940,"comment_status":"open","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_kad_blocks_custom_css":"","_kad_blocks_head_custom_js":"","_kad_blocks_body_custom_js":"","_kad_blocks_footer_custom_js":"","_monsterinsights_skip_tracking":false,"_genesis_hide_title":false,"_genesis_hide_breadcrumbs":false,"_genesis_hide_singular_image":false,"_genesis_hide_footer_widgets":false,"_genesis_custom_body_class":"","_genesis_custom_post_class":"","_genesis_layout":"","_jetpack_newsletter_access":"","_jetpack_dont_email_post_to_subs":false,"_jetpack_newsletter_tier_id":0,"_jetpack_memberships_contains_paywalled_content":false,"_jetpack_feature_clip_id":0,"_jetpack_memberships_contains_paid_content":false,"footnotes":"","jetpack_post_was_ever_published":false},"categories":[157],"tags":[131],"division":[],"class_list":{"0":"post-1461","1":"post","2":"type-post","3":"status-publish","4":"format-standard","5":"has-post-thumbnail","7":"category-analysis","8":"tag-cad","9":"entry"},"_selected_authors":"","_selected_reviewers":"","acf":[],"yoast_head":"<!-- This site is optimized with the Yoast SEO Premium plugin v27.8 (Yoast SEO v27.8) - https:\/\/yoast.com\/product\/yoast-seo-premium-wordpress\/ -->\n<title>\u8ab0\u3067\u3082\u308f\u304b\u308b\uff01\u8377\u96fb\u5316\u7c92\u5b50\u691c\u51fa\u5668\uff08\u5e2f\u96fb\u30a8\u30a2\u30ed\u30be\u30eb\u691c\u51fa\u5668\uff09\u306e\u57fa\u790e - Learning at the Bench<\/title>\n<meta name=\"description\" content=\"\u300c\u8377\u96fb\u5316\u7c92\u5b50\u691c\u51fa\u5668\uff08\u5e2f\u96fb\u30a8\u30a2\u30ed\u30be\u30eb\u691c\u51fa\u5668\uff09\u306b\u3064\u3044\u3066\u77e5\u308a\u305f\u3044\u300d\u300c\u4f7f\u3044\u52dd\u624b\u306e\u826f\u3044\u6c4e\u7528\u30bf\u30a4\u30d7\u306e\u691c\u51fa\u5668\u3092\u63a2\u3057\u3066\u3044\u308b\u300d\u305d\u3093\u306a\u65b9\u306b\u5411\u3051\u305f\u8a18\u4e8b\u3067\u3059\u3002\u86cd\u5149\u691c\u51fa\u5668\u3084\u793a\u5dee\u5c48\u6298\u7387\uff08RI\uff09\u691c\u51fa\u5668\u306a\u3069\u3001HPLC\u306e\u691c\u51fa\u5668\u5229\u7528\u306b\u304a\u3051\u308b\u8ab2\u984c\u89e3\u6c7a\u306e\u30b5\u30dd\u30fc\u30c8\u3092\u3057\u307e\u3059\u3002\" \/>\n<meta name=\"robots\" content=\"index, follow, max-snippet:-1, max-image-preview:large, max-video-preview:-1\" \/>\n<link rel=\"canonical\" href=\"https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/cad_hplc\/\" \/>\n<meta property=\"og:locale\" content=\"ja_JP\" \/>\n<meta property=\"og:type\" content=\"article\" \/>\n<meta property=\"og:title\" content=\"\u8ab0\u3067\u3082\u308f\u304b\u308b\uff01\u8377\u96fb\u5316\u7c92\u5b50\u691c\u51fa\u5668\u306e\u57fa\u790e\" \/>\n<meta property=\"og:description\" content=\"\u300c\u8377\u96fb\u5316\u7c92\u5b50\u691c\u51fa\u5668\uff08\u5e2f\u96fb\u30a8\u30a2\u30ed\u30be\u30eb\u691c\u51fa\u5668\uff09\u306b\u3064\u3044\u3066\u77e5\u308a\u305f\u3044\u300d\u300c\u4f7f\u3044\u52dd\u624b\u306e\u826f\u3044\u6c4e\u7528\u30bf\u30a4\u30d7\u306e\u691c\u51fa\u5668\u3092\u63a2\u3057\u3066\u3044\u308b\u300d\u305d\u3093\u306a\u65b9\u306b\u5411\u3051\u305f\u8a18\u4e8b\u3067\u3059\u3002\u86cd\u5149\u691c\u51fa\u5668\u3084\u793a\u5dee\u5c48\u6298\u7387\uff08RI\uff09\u691c\u51fa\u5668\u306a\u3069\u3001HPLC\u306e\u691c\u51fa\u5668\u5229\u7528\u306b\u304a\u3051\u308b\u8ab2\u984c\u89e3\u6c7a\u306e\u30b5\u30dd\u30fc\u30c8\u3092\u3057\u307e\u3059\u3002\" \/>\n<meta property=\"og:url\" content=\"https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/cad_hplc\/\" \/>\n<meta property=\"og:site_name\" content=\"Learning at the Bench\" \/>\n<meta property=\"article:publisher\" content=\"https:\/\/www.facebook.com\/ThermoFisherJapan\" \/>\n<meta property=\"article:published_time\" content=\"2018-10-18T15:00:00+00:00\" \/>\n<meta property=\"article:modified_time\" content=\"2023-12-28T01:55:43+00:00\" \/>\n<meta property=\"og:image\" content=\"http:\/\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2019\/07\/136641032-RESIZED-2.png\" \/>\n\t<meta property=\"og:image:width\" content=\"1200\" \/>\n\t<meta property=\"og:image:height\" content=\"630\" \/>\n\t<meta property=\"og:image:type\" content=\"image\/png\" \/>\n<meta name=\"author\" content=\"LATB Staff\" \/>\n<meta name=\"twitter:card\" content=\"summary_large_image\" \/>\n<meta name=\"twitter:creator\" content=\"@ThermoFisherJP\" \/>\n<meta name=\"twitter:site\" content=\"@ThermoFisherJP\" \/>\n<meta name=\"twitter:label1\" content=\"\u57f7\u7b46\u8005\" \/>\n\t<meta name=\"twitter:data1\" content=\"LATB Staff\" \/>\n\t<meta name=\"twitter:label2\" content=\"\u63a8\u5b9a\u8aad\u307f\u53d6\u308a\u6642\u9593\" \/>\n\t<meta name=\"twitter:data2\" content=\"9\u5206\" \/>\n<script type=\"application\/ld+json\" class=\"yoast-schema-graph\">{\"@context\":\"https:\\\/\\\/schema.org\",\"@graph\":[{\"@type\":\"Article\",\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/learning-at-the-bench\\\/cad_hplc\\\/#article\",\"isPartOf\":{\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/learning-at-the-bench\\\/cad_hplc\\\/\"},\"author\":{\"name\":\"LATB Staff\",\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/learning-at-the-bench\\\/#\\\/schema\\\/person\\\/00a095a01d54afcdf9457465feb04f7d\"},\"headline\":\"\u8ab0\u3067\u3082\u308f\u304b\u308b\uff01\u8377\u96fb\u5316\u7c92\u5b50\u691c\u51fa\u5668\u306e\u57fa\u790e\",\"datePublished\":\"2018-10-18T15:00:00+00:00\",\"dateModified\":\"2023-12-28T01:55:43+00:00\",\"mainEntityOfPage\":{\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/learning-at-the-bench\\\/cad_hplc\\\/\"},\"wordCount\":45,\"commentCount\":0,\"image\":{\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/learning-at-the-bench\\\/cad_hplc\\\/#primaryimage\"},\"thumbnailUrl\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/learning-at-the-bench\\\/wp-content\\\/uploads\\\/sites\\\/13\\\/2019\\\/07\\\/136641032-RESIZED-2.png\",\"keywords\":[\"\u8377\u96fb\u5316\u7c92\u5b50\u691c\u51fa\u5668\"],\"articleSection\":[\"\u5206\u6790\u30fb\u6e2c\u5b9a\u95a2\u9023\"],\"inLanguage\":\"ja\",\"potentialAction\":[{\"@type\":\"CommentAction\",\"name\":\"Comment\",\"target\":[\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/learning-at-the-bench\\\/cad_hplc\\\/#respond\"]}]},{\"@type\":\"WebPage\",\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/learning-at-the-bench\\\/cad_hplc\\\/\",\"url\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/learning-at-the-bench\\\/cad_hplc\\\/\",\"name\":\"\u8ab0\u3067\u3082\u308f\u304b\u308b\uff01\u8377\u96fb\u5316\u7c92\u5b50\u691c\u51fa\u5668\uff08\u5e2f\u96fb\u30a8\u30a2\u30ed\u30be\u30eb\u691c\u51fa\u5668\uff09\u306e\u57fa\u790e - Learning at the Bench\",\"isPartOf\":{\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/learning-at-the-bench\\\/#website\"},\"primaryImageOfPage\":{\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/learning-at-the-bench\\\/cad_hplc\\\/#primaryimage\"},\"image\":{\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/learning-at-the-bench\\\/cad_hplc\\\/#primaryimage\"},\"thumbnailUrl\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/learning-at-the-bench\\\/wp-content\\\/uploads\\\/sites\\\/13\\\/2019\\\/07\\\/136641032-RESIZED-2.png\",\"datePublished\":\"2018-10-18T15:00:00+00:00\",\"dateModified\":\"2023-12-28T01:55:43+00:00\",\"author\":{\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/learning-at-the-bench\\\/#\\\/schema\\\/person\\\/00a095a01d54afcdf9457465feb04f7d\"},\"description\":\"\u300c\u8377\u96fb\u5316\u7c92\u5b50\u691c\u51fa\u5668\uff08\u5e2f\u96fb\u30a8\u30a2\u30ed\u30be\u30eb\u691c\u51fa\u5668\uff09\u306b\u3064\u3044\u3066\u77e5\u308a\u305f\u3044\u300d\u300c\u4f7f\u3044\u52dd\u624b\u306e\u826f\u3044\u6c4e\u7528\u30bf\u30a4\u30d7\u306e\u691c\u51fa\u5668\u3092\u63a2\u3057\u3066\u3044\u308b\u300d\u305d\u3093\u306a\u65b9\u306b\u5411\u3051\u305f\u8a18\u4e8b\u3067\u3059\u3002\u86cd\u5149\u691c\u51fa\u5668\u3084\u793a\u5dee\u5c48\u6298\u7387\uff08RI\uff09\u691c\u51fa\u5668\u306a\u3069\u3001HPLC\u306e\u691c\u51fa\u5668\u5229\u7528\u306b\u304a\u3051\u308b\u8ab2\u984c\u89e3\u6c7a\u306e\u30b5\u30dd\u30fc\u30c8\u3092\u3057\u307e\u3059\u3002\",\"breadcrumb\":{\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/learning-at-the-bench\\\/cad_hplc\\\/#breadcrumb\"},\"inLanguage\":\"ja\",\"potentialAction\":[{\"@type\":\"ReadAction\",\"target\":[\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/learning-at-the-bench\\\/cad_hplc\\\/\"]}]},{\"@type\":\"ImageObject\",\"inLanguage\":\"ja\",\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/learning-at-the-bench\\\/cad_hplc\\\/#primaryimage\",\"url\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/learning-at-the-bench\\\/wp-content\\\/uploads\\\/sites\\\/13\\\/2019\\\/07\\\/136641032-RESIZED-2.png\",\"contentUrl\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/learning-at-the-bench\\\/wp-content\\\/uploads\\\/sites\\\/13\\\/2019\\\/07\\\/136641032-RESIZED-2.png\",\"width\":1200,\"height\":630,\"caption\":\"\u8377\u96fb\u7c92\u5b50\u691c\u51fa\u5668\"},{\"@type\":\"BreadcrumbList\",\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/learning-at-the-bench\\\/cad_hplc\\\/#breadcrumb\",\"itemListElement\":[{\"@type\":\"ListItem\",\"position\":1,\"name\":\"Home\",\"item\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/learning-at-the-bench\\\/\"},{\"@type\":\"ListItem\",\"position\":2,\"name\":\"\u8ab0\u3067\u3082\u308f\u304b\u308b\uff01\u8377\u96fb\u5316\u7c92\u5b50\u691c\u51fa\u5668\u306e\u57fa\u790e\"}]},{\"@type\":\"WebSite\",\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/learning-at-the-bench\\\/#website\",\"url\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/learning-at-the-bench\\\/\",\"name\":\"Learning at the Bench\",\"description\":\"\",\"potentialAction\":[{\"@type\":\"SearchAction\",\"target\":{\"@type\":\"EntryPoint\",\"urlTemplate\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/learning-at-the-bench\\\/?s={search_term_string}\"},\"query-input\":{\"@type\":\"PropertyValueSpecification\",\"valueRequired\":true,\"valueName\":\"search_term_string\"}}],\"inLanguage\":\"ja\"},{\"@type\":\"Person\",\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/learning-at-the-bench\\\/#\\\/schema\\\/person\\\/00a095a01d54afcdf9457465feb04f7d\",\"name\":\"LATB Staff\",\"image\":{\"@type\":\"ImageObject\",\"inLanguage\":\"ja\",\"@id\":\"https:\\\/\\\/secure.gravatar.com\\\/avatar\\\/8803277856c8b753feb4217dff0f9953327452df1a615c3c88b2f9864efb2ddd?s=96&d=mm&r=g\",\"url\":\"https:\\\/\\\/secure.gravatar.com\\\/avatar\\\/8803277856c8b753feb4217dff0f9953327452df1a615c3c88b2f9864efb2ddd?s=96&d=mm&r=g\",\"contentUrl\":\"https:\\\/\\\/secure.gravatar.com\\\/avatar\\\/8803277856c8b753feb4217dff0f9953327452df1a615c3c88b2f9864efb2ddd?s=96&d=mm&r=g\",\"caption\":\"LATB Staff\"},\"url\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/author\\\/latbstaff\\\/\"}]}<\/script>\n<!-- \/ Yoast SEO Premium plugin. -->","yoast_head_json":{"title":"\u8ab0\u3067\u3082\u308f\u304b\u308b\uff01\u8377\u96fb\u5316\u7c92\u5b50\u691c\u51fa\u5668\uff08\u5e2f\u96fb\u30a8\u30a2\u30ed\u30be\u30eb\u691c\u51fa\u5668\uff09\u306e\u57fa\u790e - Learning at the Bench","description":"\u300c\u8377\u96fb\u5316\u7c92\u5b50\u691c\u51fa\u5668\uff08\u5e2f\u96fb\u30a8\u30a2\u30ed\u30be\u30eb\u691c\u51fa\u5668\uff09\u306b\u3064\u3044\u3066\u77e5\u308a\u305f\u3044\u300d\u300c\u4f7f\u3044\u52dd\u624b\u306e\u826f\u3044\u6c4e\u7528\u30bf\u30a4\u30d7\u306e\u691c\u51fa\u5668\u3092\u63a2\u3057\u3066\u3044\u308b\u300d\u305d\u3093\u306a\u65b9\u306b\u5411\u3051\u305f\u8a18\u4e8b\u3067\u3059\u3002\u86cd\u5149\u691c\u51fa\u5668\u3084\u793a\u5dee\u5c48\u6298\u7387\uff08RI\uff09\u691c\u51fa\u5668\u306a\u3069\u3001HPLC\u306e\u691c\u51fa\u5668\u5229\u7528\u306b\u304a\u3051\u308b\u8ab2\u984c\u89e3\u6c7a\u306e\u30b5\u30dd\u30fc\u30c8\u3092\u3057\u307e\u3059\u3002","robots":{"index":"index","follow":"follow","max-snippet":"max-snippet:-1","max-image-preview":"max-image-preview:large","max-video-preview":"max-video-preview:-1"},"canonical":"https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/cad_hplc\/","og_locale":"ja_JP","og_type":"article","og_title":"\u8ab0\u3067\u3082\u308f\u304b\u308b\uff01\u8377\u96fb\u5316\u7c92\u5b50\u691c\u51fa\u5668\u306e\u57fa\u790e","og_description":"\u300c\u8377\u96fb\u5316\u7c92\u5b50\u691c\u51fa\u5668\uff08\u5e2f\u96fb\u30a8\u30a2\u30ed\u30be\u30eb\u691c\u51fa\u5668\uff09\u306b\u3064\u3044\u3066\u77e5\u308a\u305f\u3044\u300d\u300c\u4f7f\u3044\u52dd\u624b\u306e\u826f\u3044\u6c4e\u7528\u30bf\u30a4\u30d7\u306e\u691c\u51fa\u5668\u3092\u63a2\u3057\u3066\u3044\u308b\u300d\u305d\u3093\u306a\u65b9\u306b\u5411\u3051\u305f\u8a18\u4e8b\u3067\u3059\u3002\u86cd\u5149\u691c\u51fa\u5668\u3084\u793a\u5dee\u5c48\u6298\u7387\uff08RI\uff09\u691c\u51fa\u5668\u306a\u3069\u3001HPLC\u306e\u691c\u51fa\u5668\u5229\u7528\u306b\u304a\u3051\u308b\u8ab2\u984c\u89e3\u6c7a\u306e\u30b5\u30dd\u30fc\u30c8\u3092\u3057\u307e\u3059\u3002","og_url":"https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/cad_hplc\/","og_site_name":"Learning at the Bench","article_publisher":"https:\/\/www.facebook.com\/ThermoFisherJapan","article_published_time":"2018-10-18T15:00:00+00:00","article_modified_time":"2023-12-28T01:55:43+00:00","og_image":[{"width":1200,"height":630,"url":"http:\/\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2019\/07\/136641032-RESIZED-2.png","type":"image\/png"}],"author":"LATB Staff","twitter_card":"summary_large_image","twitter_creator":"@ThermoFisherJP","twitter_site":"@ThermoFisherJP","twitter_misc":{"\u57f7\u7b46\u8005":"LATB Staff","\u63a8\u5b9a\u8aad\u307f\u53d6\u308a\u6642\u9593":"9\u5206"},"schema":{"@context":"https:\/\/schema.org","@graph":[{"@type":"Article","@id":"https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/cad_hplc\/#article","isPartOf":{"@id":"https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/cad_hplc\/"},"author":{"name":"LATB Staff","@id":"https:\/\/admin.acceleratingscience.com\/learning-at-the-bench\/#\/schema\/person\/00a095a01d54afcdf9457465feb04f7d"},"headline":"\u8ab0\u3067\u3082\u308f\u304b\u308b\uff01\u8377\u96fb\u5316\u7c92\u5b50\u691c\u51fa\u5668\u306e\u57fa\u790e","datePublished":"2018-10-18T15:00:00+00:00","dateModified":"2023-12-28T01:55:43+00:00","mainEntityOfPage":{"@id":"https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/cad_hplc\/"},"wordCount":45,"commentCount":0,"image":{"@id":"https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/cad_hplc\/#primaryimage"},"thumbnailUrl":"https:\/\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2019\/07\/136641032-RESIZED-2.png","keywords":["\u8377\u96fb\u5316\u7c92\u5b50\u691c\u51fa\u5668"],"articleSection":["\u5206\u6790\u30fb\u6e2c\u5b9a\u95a2\u9023"],"inLanguage":"ja","potentialAction":[{"@type":"CommentAction","name":"Comment","target":["https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/cad_hplc\/#respond"]}]},{"@type":"WebPage","@id":"https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/cad_hplc\/","url":"https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/cad_hplc\/","name":"\u8ab0\u3067\u3082\u308f\u304b\u308b\uff01\u8377\u96fb\u5316\u7c92\u5b50\u691c\u51fa\u5668\uff08\u5e2f\u96fb\u30a8\u30a2\u30ed\u30be\u30eb\u691c\u51fa\u5668\uff09\u306e\u57fa\u790e - Learning at the Bench","isPartOf":{"@id":"https:\/\/admin.acceleratingscience.com\/learning-at-the-bench\/#website"},"primaryImageOfPage":{"@id":"https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/cad_hplc\/#primaryimage"},"image":{"@id":"https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/cad_hplc\/#primaryimage"},"thumbnailUrl":"https:\/\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2019\/07\/136641032-RESIZED-2.png","datePublished":"2018-10-18T15:00:00+00:00","dateModified":"2023-12-28T01:55:43+00:00","author":{"@id":"https:\/\/admin.acceleratingscience.com\/learning-at-the-bench\/#\/schema\/person\/00a095a01d54afcdf9457465feb04f7d"},"description":"\u300c\u8377\u96fb\u5316\u7c92\u5b50\u691c\u51fa\u5668\uff08\u5e2f\u96fb\u30a8\u30a2\u30ed\u30be\u30eb\u691c\u51fa\u5668\uff09\u306b\u3064\u3044\u3066\u77e5\u308a\u305f\u3044\u300d\u300c\u4f7f\u3044\u52dd\u624b\u306e\u826f\u3044\u6c4e\u7528\u30bf\u30a4\u30d7\u306e\u691c\u51fa\u5668\u3092\u63a2\u3057\u3066\u3044\u308b\u300d\u305d\u3093\u306a\u65b9\u306b\u5411\u3051\u305f\u8a18\u4e8b\u3067\u3059\u3002\u86cd\u5149\u691c\u51fa\u5668\u3084\u793a\u5dee\u5c48\u6298\u7387\uff08RI\uff09\u691c\u51fa\u5668\u306a\u3069\u3001HPLC\u306e\u691c\u51fa\u5668\u5229\u7528\u306b\u304a\u3051\u308b\u8ab2\u984c\u89e3\u6c7a\u306e\u30b5\u30dd\u30fc\u30c8\u3092\u3057\u307e\u3059\u3002","breadcrumb":{"@id":"https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/cad_hplc\/#breadcrumb"},"inLanguage":"ja","potentialAction":[{"@type":"ReadAction","target":["https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/cad_hplc\/"]}]},{"@type":"ImageObject","inLanguage":"ja","@id":"https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/cad_hplc\/#primaryimage","url":"https:\/\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2019\/07\/136641032-RESIZED-2.png","contentUrl":"https:\/\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2019\/07\/136641032-RESIZED-2.png","width":1200,"height":630,"caption":"\u8377\u96fb\u7c92\u5b50\u691c\u51fa\u5668"},{"@type":"BreadcrumbList","@id":"https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/cad_hplc\/#breadcrumb","itemListElement":[{"@type":"ListItem","position":1,"name":"Home","item":"https:\/\/admin.acceleratingscience.com\/learning-at-the-bench\/"},{"@type":"ListItem","position":2,"name":"\u8ab0\u3067\u3082\u308f\u304b\u308b\uff01\u8377\u96fb\u5316\u7c92\u5b50\u691c\u51fa\u5668\u306e\u57fa\u790e"}]},{"@type":"WebSite","@id":"https:\/\/admin.acceleratingscience.com\/learning-at-the-bench\/#website","url":"https:\/\/admin.acceleratingscience.com\/learning-at-the-bench\/","name":"Learning at the Bench","description":"","potentialAction":[{"@type":"SearchAction","target":{"@type":"EntryPoint","urlTemplate":"https:\/\/admin.acceleratingscience.com\/learning-at-the-bench\/?s={search_term_string}"},"query-input":{"@type":"PropertyValueSpecification","valueRequired":true,"valueName":"search_term_string"}}],"inLanguage":"ja"},{"@type":"Person","@id":"https:\/\/admin.acceleratingscience.com\/learning-at-the-bench\/#\/schema\/person\/00a095a01d54afcdf9457465feb04f7d","name":"LATB Staff","image":{"@type":"ImageObject","inLanguage":"ja","@id":"https:\/\/secure.gravatar.com\/avatar\/8803277856c8b753feb4217dff0f9953327452df1a615c3c88b2f9864efb2ddd?s=96&d=mm&r=g","url":"https:\/\/secure.gravatar.com\/avatar\/8803277856c8b753feb4217dff0f9953327452df1a615c3c88b2f9864efb2ddd?s=96&d=mm&r=g","contentUrl":"https:\/\/secure.gravatar.com\/avatar\/8803277856c8b753feb4217dff0f9953327452df1a615c3c88b2f9864efb2ddd?s=96&d=mm&r=g","caption":"LATB Staff"},"url":"https:\/\/admin.acceleratingscience.com\/author\/latbstaff\/"}]}},"taxonomy_info":{"category":[{"value":157,"label":"\u5206\u6790\u30fb\u6e2c\u5b9a\u95a2\u9023"}],"post_tag":[{"value":131,"label":"\u8377\u96fb\u5316\u7c92\u5b50\u691c\u51fa\u5668"}]},"featured_image_src_large":["https:\/\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2019\/07\/136641032-RESIZED-2-1024x538.png",760,399,true],"author_info":{"display_name":"LATB Staff","author_link":"https:\/\/admin.acceleratingscience.com\/author\/latbstaff\/"},"comment_info":0,"category_info":[{"term_id":157,"name":"\u5206\u6790\u30fb\u6e2c\u5b9a\u95a2\u9023","slug":"analysis","term_group":0,"term_taxonomy_id":157,"taxonomy":"category","description":"","parent":0,"count":106,"filter":"raw","meta":[],"cat_ID":157,"category_count":106,"category_description":"","cat_name":"\u5206\u6790\u30fb\u6e2c\u5b9a\u95a2\u9023","category_nicename":"analysis","category_parent":0}],"tag_info":[{"term_id":131,"name":"\u8377\u96fb\u5316\u7c92\u5b50\u691c\u51fa\u5668","slug":"cad","term_group":0,"term_taxonomy_id":131,"taxonomy":"post_tag","description":"","parent":0,"count":2,"filter":"raw","meta":[]}],"jetpack-related-posts":[{"id":25090,"url":"https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/battery-development-elemental-analysis-csm-26016\/","url_meta":{"origin":1461,"position":0},"title":"\u30ea\u30c1\u30a6\u30e0\u30a4\u30aa\u30f3\u96fb\u6c60\u88fd\u9020\uff1a\u4f7f\u7528\u3055\u308c\u308b\u5316\u5b66\u5206\u6790\u6280\u8853","author":"LatB Staff","date":"2026\u5e744\u670820\u65e5","format":false,"excerpt":"\u9ad8\u6027\u80fd\u96fb\u6c60\u306b\u5bfe\u3059\u308b\u9700\u8981\u306e\u6025\u5897\u306f\u3001\u751f\u7523\u306e\u3042\u3089\u3086\u308b\u6bb5\u968e\u306b\u304a\u3051\u308b\u5927\u898f\u6a21\u306a\u6295\u8cc7\u3092\u4fc3\u3057\u3066\u3044\u307e\u3059\u3002\u96fb\u6c60\u7814\u7a76\u304a\u3088\u3073\u2026","rel":"","context":"\u5206\u6790\u30fb\u6e2c\u5b9a\u95a2\u9023","block_context":{"text":"\u5206\u6790\u30fb\u6e2c\u5b9a\u95a2\u9023","link":"https:\/\/admin.acceleratingscience.com\/learning-at-the-bench\/analysis\/"},"img":{"alt_text":"","src":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2026\/04\/battery-development-elemental-analysis-csm-26016_eyecatch.jpg?resize=350%2C200&ssl=1","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2026\/04\/battery-development-elemental-analysis-csm-26016_eyecatch.jpg?resize=350%2C200&ssl=1 1x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2026\/04\/battery-development-elemental-analysis-csm-26016_eyecatch.jpg?resize=525%2C300&ssl=1 1.5x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2026\/04\/battery-development-elemental-analysis-csm-26016_eyecatch.jpg?resize=700%2C400&ssl=1 2x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2026\/04\/battery-development-elemental-analysis-csm-26016_eyecatch.jpg?resize=1050%2C600&ssl=1 3x"},"classes":[]},{"id":25592,"url":"https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/ion-chromatography-contaminant-analysis-cmd-26024\/","url_meta":{"origin":1461,"position":1},"title":"\u5316\u7ca7\u54c1\u4e2d\u306ePFAS\u5206\u6790\u3068\u306f\uff1f\u7dcf\u6709\u6a5f\u30d5\u30c3\u7d20\uff08TOF\uff09\u6e2c\u5b9a\u306b\u3088\u308b\u30b9\u30af\u30ea\u30fc\u30cb\u30f3\u30b0\u3068\u898f\u5236\u5bfe\u5fdc","author":"LatB Staff","date":"2026\u5e749\u670810\u65e5","format":false,"excerpt":"PFAS\uff08\u6709\u6a5f\u30d5\u30c3\u7d20\u5316\u5408\u7269\uff09\u306f\u3001\u300c\u6c38\u9060\u306e\u5316\u5b66\u7269\u8cea\uff08Forever Chemicals\uff09\u300d\u3068\u3082\u547c\u3070\u308c\u3001\u2026","rel":"","context":"\u5206\u6790\u30fb\u6e2c\u5b9a\u95a2\u9023","block_context":{"text":"\u5206\u6790\u30fb\u6e2c\u5b9a\u95a2\u9023","link":"https:\/\/admin.acceleratingscience.com\/learning-at-the-bench\/analysis\/"},"img":{"alt_text":"","src":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2026\/08\/ion-chromatography-contaminant-analysis-cmd-26024_eyecatch.png?resize=350%2C200&ssl=1","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2026\/08\/ion-chromatography-contaminant-analysis-cmd-26024_eyecatch.png?resize=350%2C200&ssl=1 1x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2026\/08\/ion-chromatography-contaminant-analysis-cmd-26024_eyecatch.png?resize=525%2C300&ssl=1 1.5x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2026\/08\/ion-chromatography-contaminant-analysis-cmd-26024_eyecatch.png?resize=700%2C400&ssl=1 2x"},"classes":[]},{"id":25362,"url":"https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/battery-development-ion-chromatography-csm-26020\/","url_meta":{"origin":1461,"position":2},"title":"IC-MS\/MS\u3092\u7528\u3044\u305f\u30ea\u30c1\u30a6\u30e0\u30a4\u30aa\u30f3\u96fb\u6c60\u8ca0\u6975\u306e\u52a3\u5316\u89e3\u6790","author":"LatB Staff","date":"2026\u5e746\u670826\u65e5","format":false,"excerpt":"\u539f\u6587\uff08\u82f1\u8a9e\uff09\uff1aExploring Lithium-Ion Battery Anode Degrada\u2026","rel":"","context":"\u5206\u5b50\u751f\u7269\u5b66\u5b9f\u9a13\u95a2\u9023","block_context":{"text":"\u5206\u5b50\u751f\u7269\u5b66\u5b9f\u9a13\u95a2\u9023","link":"https:\/\/admin.acceleratingscience.com\/learning-at-the-bench\/mol-bio\/"},"img":{"alt_text":"","src":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2026\/06\/battery-development-ion-chromatography-csm-26020_eyecatch.jpg?resize=350%2C200&ssl=1","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2026\/06\/battery-development-ion-chromatography-csm-26020_eyecatch.jpg?resize=350%2C200&ssl=1 1x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2026\/06\/battery-development-ion-chromatography-csm-26020_eyecatch.jpg?resize=525%2C300&ssl=1 1.5x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2026\/06\/battery-development-ion-chromatography-csm-26020_eyecatch.jpg?resize=700%2C400&ssl=1 2x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2026\/06\/battery-development-ion-chromatography-csm-26020_eyecatch.jpg?resize=1050%2C600&ssl=1 3x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2026\/06\/battery-development-ion-chromatography-csm-26020_eyecatch.jpg?resize=1400%2C800&ssl=1 4x"},"classes":[]},{"id":24462,"url":"https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/battery-development-microscopy-csm-25076\/","url_meta":{"origin":1461,"position":3},"title":"\u5168\u56fa\u4f53\u96fb\u6c60\u306e\u9769\u65b0\u3092\u652f\u3048\u308b\u5148\u7aef\u7684\u7279\u6027\u8a55\u4fa1\u6280\u8853\u306e\u63a2\u7a76","author":"LatB Staff","date":"2025\u5e7412\u670817\u65e5","format":false,"excerpt":"\u5168\u56fa\u4f53\u96fb\u6c60\u306e\u958b\u767a\u3067\u306f\u3001\u4f7f\u7528\u3055\u308c\u308b\u6750\u6599\u306e\u7406\u89e3\u3092\u6df1\u3081\u308b\u305f\u3081\u306b\u3001\u5148\u7aef\u7684\u306a\u6027\u8a55\u4fa1\u6280\u8853\u304c\u4e0d\u53ef\u6b20\u3067\u3059\u3002\u3053\u308c\u3089\u306e\u624b\u2026","rel":"","context":"\u5206\u6790\u30fb\u6e2c\u5b9a\u95a2\u9023","block_context":{"text":"\u5206\u6790\u30fb\u6e2c\u5b9a\u95a2\u9023","link":"https:\/\/admin.acceleratingscience.com\/learning-at-the-bench\/analysis\/"},"img":{"alt_text":"","src":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2025\/12\/battery-development-microscopy-csm-25076_eyecatch.jpg?resize=350%2C200&ssl=1","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2025\/12\/battery-development-microscopy-csm-25076_eyecatch.jpg?resize=350%2C200&ssl=1 1x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2025\/12\/battery-development-microscopy-csm-25076_eyecatch.jpg?resize=525%2C300&ssl=1 1.5x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2025\/12\/battery-development-microscopy-csm-25076_eyecatch.jpg?resize=700%2C400&ssl=1 2x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2025\/12\/battery-development-microscopy-csm-25076_eyecatch.jpg?resize=1050%2C600&ssl=1 3x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2025\/12\/battery-development-microscopy-csm-25076_eyecatch.jpg?resize=1400%2C800&ssl=1 4x"},"classes":[]},{"id":25673,"url":"https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/ion-chromatography-semiconductors-cmd-26028\/","url_meta":{"origin":1461,"position":4},"title":"\u534a\u5c0e\u4f53\u88fd\u9020\u306b\u304a\u3051\u308bIC\u306e\u5f79\u5272","author":"LatB Staff","date":"2026\u5e7410\u67082\u65e5","format":false,"excerpt":"\u30b9\u30de\u30fc\u30c8\u30d5\u30a9\u30f3\u3084AI\u30b5\u30fc\u30d0\u30fc\u3001\u81ea\u52d5\u8eca\u306a\u3069\u3001\u79c1\u305f\u3061\u306e\u8eab\u306e\u56de\u308a\u306b\u3042\u308b\u96fb\u5b50\u6a5f\u5668\u306e\u591a\u304f\u306f\u534a\u5c0e\u4f53\u306b\u3088\u3063\u3066\u652f\u3048\u3089\u2026","rel":"","context":"\u5206\u6790\u30fb\u6e2c\u5b9a\u95a2\u9023","block_context":{"text":"\u5206\u6790\u30fb\u6e2c\u5b9a\u95a2\u9023","link":"https:\/\/admin.acceleratingscience.com\/learning-at-the-bench\/analysis\/"},"img":{"alt_text":"","src":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2026\/09\/ion-chromatography-semiconductors-cmd-26028_eyecatch.jpg?resize=350%2C200&ssl=1","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2026\/09\/ion-chromatography-semiconductors-cmd-26028_eyecatch.jpg?resize=350%2C200&ssl=1 1x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2026\/09\/ion-chromatography-semiconductors-cmd-26028_eyecatch.jpg?resize=525%2C300&ssl=1 1.5x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2026\/09\/ion-chromatography-semiconductors-cmd-26028_eyecatch.jpg?resize=700%2C400&ssl=1 2x"},"classes":[]},{"id":24647,"url":"https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/sanger-seq-mutational-analysis-gsd-ts-ce-26007\/","url_meta":{"origin":1461,"position":5},"title":"\u30de\u30a4\u30ca\u30fc\u5909\u7570\u3092\u691c\u51fa\uff01Minor Variant Finder\u30bd\u30d5\u30c8\u30a6\u30a8\u30a2","author":"LatB Staff","date":"2026\u5e742\u67089\u65e5","format":false,"excerpt":"\u30ad\u30e3\u30d4\u30e9\u30ea\u30fc\u30b7\u30fc\u30b1\u30f3\u30b5\u3067\u306f\u300125%\u7a0b\u5ea6\u306e\u5272\u5408\u306e\u5909\u7570\u3092\u691c\u51fa\u3059\u308b\u3053\u3068\u304c\u3067\u304d\u307e\u3059\u304c\u3001\u3053\u308c\u3088\u308a\u3082\u4f4e\u3044\u5272\u5408\u306e\u5909\u2026","rel":"","context":"\u5206\u5b50\u751f\u7269\u5b66\u5b9f\u9a13\u95a2\u9023","block_context":{"text":"\u5206\u5b50\u751f\u7269\u5b66\u5b9f\u9a13\u95a2\u9023","link":"https:\/\/admin.acceleratingscience.com\/learning-at-the-bench\/mol-bio\/"},"img":{"alt_text":"","src":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2026\/01\/sanger-seq-mutational-analysis-gsd-ts-ce-26007_eyecatch.jpg?resize=350%2C200&ssl=1","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2026\/01\/sanger-seq-mutational-analysis-gsd-ts-ce-26007_eyecatch.jpg?resize=350%2C200&ssl=1 1x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2026\/01\/sanger-seq-mutational-analysis-gsd-ts-ce-26007_eyecatch.jpg?resize=525%2C300&ssl=1 1.5x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2026\/01\/sanger-seq-mutational-analysis-gsd-ts-ce-26007_eyecatch.jpg?resize=700%2C400&ssl=1 2x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2026\/01\/sanger-seq-mutational-analysis-gsd-ts-ce-26007_eyecatch.jpg?resize=1050%2C600&ssl=1 3x"},"classes":[]}],"jetpack_sharing_enabled":true,"amp_enabled":true,"jetpack_featured_media_url":"https:\/\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2019\/07\/136641032-RESIZED-2.png","_links":{"self":[{"href":"https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/wp-json\/wp\/v2\/posts\/1461","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/wp-json\/wp\/v2\/users\/988"}],"replies":[{"embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/wp-json\/wp\/v2\/comments?post=1461"}],"version-history":[{"count":0,"href":"https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/wp-json\/wp\/v2\/posts\/1461\/revisions"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/wp-json\/wp\/v2\/media\/4940"}],"wp:attachment":[{"href":"https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/wp-json\/wp\/v2\/media?parent=1461"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/wp-json\/wp\/v2\/categories?post=1461"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/wp-json\/wp\/v2\/tags?post=1461"},{"taxonomy":"division","embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/wp-json\/wp\/v2\/division?post=1461"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}