{"id":6865,"date":"2020-06-01T15:09:45","date_gmt":"2020-06-01T06:09:45","guid":{"rendered":"http:\/\/admin.acceleratingscience.com\/learning-at-the-bench\/?p=6865"},"modified":"2025-12-19T16:44:41","modified_gmt":"2025-12-19T07:44:41","slug":"airfilter_ce_gsd_ts_1","status":"publish","type":"post","link":"https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/airfilter_ce_gsd_ts_1\/","title":{"rendered":"\u30ad\u30e3\u30d4\u30e9\u30ea\u30fc\u30b7\u30fc\u30b1\u30f3\u30b5\u306e\u30a8\u30a2\u30d5\u30a3\u30eb\u30bf\u30fc\u3092\u78ba\u8a8d\u3055\u308c\u3066\u3044\u307e\u3059\u304b\uff1f"},"content":{"rendered":"<p>\u30ad\u30e3\u30d4\u30e9\u30ea\u30fc\u30b7\u30fc\u30b1\u30f3\u30b5\uff08\u30b8\u30a7\u30cd\u30c6\u30a3\u30c3\u30af\u30a2\u30ca\u30e9\u30a4\u30b6\uff09\u306f\u5b9a\u671f\u7684\u306a\u30e1\u30f3\u30c6\u30ca\u30f3\u30b9\u3092\u884c\u3046\u3053\u3068\u3067\u3001\u3088\u308a\u304d\u308c\u3044\u3067\u5b89\u5b9a\u3057\u305f\u30c7\u30fc\u30bf\u3092\u5f97\u308b\u3053\u3068\u304c\u3067\u304d\u307e\u3059\u3002<br \/>\n\u305d\u3053\u3067\u4eca\u56de\u306f\u3001\u30a8\u30a2\u30d5\u30a3\u30eb\u30bf\u30fc\u306e\u30e1\u30f3\u30c6\u30ca\u30f3\u30b9\u3064\u3044\u3066\u3001\u3054\u7d39\u4ecb\u3057\u307e\u3059\u3002<\/p>\n<p>\u25bc\u3053\u3093\u306a\u65b9\u306b\u304a\u3059\u3059\u3081\u3067\u3059\uff01<br \/>\n\u30fb\u30c7\u30fc\u30bf\u306b\u60f3\u5b9a\u5916\u306e\u5909\u7570\u3084\u30b9\u30d1\u30a4\u30af\u30ce\u30a4\u30ba\u591a\u3044\u65b9<br \/>\n\u30fb\u30dd\u30ea\u30de\u30fc\u52a3\u5316\u304c\u65e9\u3044\u3068\u611f\u3058\u3089\u308c\u308b\u65b9<br \/>\n\u30fb\u30ea\u30c6\u30fc\u30ca\u30fc\u304c\u9ed2\u305a\u3080\u65b9<\/p>\n<h2>\u30a8\u30a2\u30d5\u30a3\u30eb\u30bf\u30fc\u306e\u5f79\u5272<\/h2>\n<p>\u30a8\u30a2\u30d5\u30a3\u30eb\u30bf\u30fc\u306f\u3044\u305a\u308c\u306e\u30ad\u30e3\u30d4\u30e9\u30ea\u30fc\u30b7\u30fc\u30b1\u30f3\u30b5\uff08\u30b8\u30a7\u30cd\u30c6\u30a3\u30c3\u30af\u30a2\u30ca\u30e9\u30a4\u30b6\uff09\u306b\u3082\u88c5\u5099\u3055\u308c\u3066\u3044\u307e\u3059\u3002\u305d\u3057\u3066\u30a8\u30a2\u30d5\u30a3\u30eb\u30bf\u30fc\u306f\u88c5\u7f6e\u5185\u306b\u5faa\u74b0\u3059\u308b\u51b7\u5374\u7528\u7a7a\u6c17\u306e\u53d6\u308a\u8fbc\u307f\u306b\u969b\u3057\u3001\u307b\u3053\u308a\u3092\u9664\u304f\u5f79\u5272\u3092\u6301\u3063\u3066\u3044\u307e\u3059\u3002<\/p>\n<h2>\u30a8\u30a2\u30d5\u30a3\u30eb\u30bf\u30fc\u306b\u8d77\u56e0\u3059\u308b\u30c8\u30e9\u30d6\u30eb<\/h2>\n<p>\u76ee\u8a70\u307e\u308a\u3057\u305f\u30a8\u30a2\u30d5\u30a3\u30eb\u30bf\u30fc\u306f\u30c7\u30fc\u30bf\u30c8\u30e9\u30d6\u30eb\u306e\u539f\u56e0\u306b\u306a\u308a\u307e\u3059\u3002<br \/>\n\u30fb\u76ee\u8a70\u307e\u308a\u3057\u305f\u307b\u3053\u308a\u306f\u88c5\u7f6e\u5185\u3092\u5faa\u74b0\u3057\u3001<br \/>\n\u30fb\u691c\u51fa\u90e8\u4f4d\u3092\u6c5a\u3057\u3001\u611f\u5ea6\u4f4e\u4e0b\u306b\u3064\u306a\u304c\u308a\u307e\u3059\u3002<br \/>\n\u30fb\u691c\u51fa\u90e8\u4f4d\u3092\u901a\u904e\u3057\u30b9\u30d1\u30a4\u30af\u30ce\u30a4\u30ba\u307e\u305f\u306f\u60f3\u5b9a\u5916\u306e\u5909\u7570\u3068\u3057\u3066\u691c\u51fa\u3055\u308c\u307e\u3059\u3002<br \/>\n\u30fb\u96fb\u6975\u90e8\u4f4d\u306b\u89e6\u308c\u3001\u6cf3\u52d5\u3092\u4e71\u3057\u307e\u3059\u3002<br \/>\n\u30fb\u76ee\u8a70\u307e\u308a\u306e\u305f\u3081\u65b0\u3057\u3044\u7a7a\u6c17\u3092\u53d6\u308a\u8fbc\u3081\u305a\u3001\u6696\u307e\u3063\u305f\u7a7a\u6c17\u304c\u5eab\u5185\u3092\u5faa\u74b0\u3057\u3001<br \/>\n\u30fb\u30dd\u30ea\u30de\u30fc\u6e29\u5ea6\u3092\u4e0a\u6607\u3055\u305b\u3001\u52a3\u5316\u3092\u65e9\u3081\u307e\u3059\u3002<br \/>\n\u30fb\u30d0\u30c3\u30d5\u30a1\u30fc\u306e\u84b8\u767a\u3092\u65e9\u3081\u307e\u3059\u3002<\/p>\n<h2>\u30a8\u30a2\u30d5\u30a3\u30eb\u30bf\u30fc\u306e\u30e1\u30f3\u30c6\u30ca\u30f3\u30b9<\/h2>\n<p>\u30a8\u30a2\u30d5\u30a3\u30eb\u30bf\u30fc\u306e\u30e1\u30f3\u30c6\u30ca\u30f3\u30b9\u306f\u65b0\u3057\u3044\u3082\u306e\u306b\u4ea4\u63db\u3092\u63a8\u5968\u3057\u3066\u304a\u308a\u307e\u3059\u304c\u3001\u5fdc\u6025\u63aa\u7f6e\u3068\u3057\u3066\u6e05\u6383\u3044\u305f\u3060\u304f\u4e8b\u3082\u53ef\u80fd\u3067\u3059\u3002\u64cd\u4f5c\u306b\u3042\u305f\u308a\u88c5\u7f6e\u306e\u96fb\u6e90\u306f\u5207\u3063\u3066\u304f\u3060\u3055\u3044\u3002<br \/>\n\u30fbApplied Biosystems\u2122 3130\/3730 Genetic Analyzer\u30b7\u30ea\u30fc\u30ba<br \/>\n\u672c\u4f53\u5e95\u9762\uff08\u5de6\u4e0b\uff09\u306e\u4e8c\u3064\u306e\u306d\u3058\u3092\u7de9\u3081\u3001\u30a8\u30a2\u30d5\u30a3\u30eb\u30bf\u30fc\u3092\u5f15\u304d\u51fa\u3057\u307e\u3059\u3002<br \/>\n\u4e0b\u306e\u9762\u304b\u3089\u6383\u9664\u6a5f\u3067\u307b\u3053\u308a\u3092\u5438\u3044\u3001\u5fdc\u6025\u63aa\u7f6e\u3092\u65bd\u3057\u307e\u3059\u3002<br \/>\n\u30fbApplied Biosystems\u2122 3500 Genetic Analyzer\u30b7\u30ea\u30fc\u30ba<br \/>\n\u672c\u4f53\u80cc\u9762\u306e\u30a2\u30c0\u30d7\u30bf\u30fc\u3054\u3068\u5916\u3057\u307e\u3059\u3002<br \/>\n\u6d41\u6c34\uff08\u30b7\u30e3\u30ef\u30fc\uff09\u3067\u9759\u304b\u306b\u307b\u3053\u308a\u3092\u6d41\u3057\u3001\u5341\u5206\u4e7e\u71e5\u3055\u305b\u3066\u53d6\u308a\u4ed8\u3051\u307e\u3059\u3002<\/p>\n<h2>\u307e\u3068\u3081<\/h2>\n<p>\u30fb\u3088\u308a\u826f\u3044\u30c7\u30fc\u30bf\u3092\u5f97\u308b\u305f\u3081\u306b\u30a8\u30a2\u30d5\u30a3\u30eb\u30bf\u30fc\u3092\u5b9a\u671f\u7684\u306b\u70b9\u691c\u30fb\u6e05\u6383\u3057\u307e\u3057\u3087\u3046\u3002<br \/>\n\u30fb\u30a8\u30a2\u30d5\u30a3\u30eb\u30bf\u30fc\u306e\u30e1\u30f3\u30c6\u30ca\u30f3\u30b9\u306f\u3068\u3066\u3082\u7c21\u5358\u3067\u3059\u3002<\/p>\n<p>\u30e1\u30f3\u30c6\u30ca\u30f3\u30b9\u3067\u304a\u56f0\u308a\u306e\u65b9\u306f\u3001<a href=\"https:\/\/www.thermofisher.com\/jp\/ja\/home\/products-and-services\/services\/instrument-qualification-services.html?CID=gsd_cap_sbu_r04_jp_cp1349_pjt5618_gsd00000_0so_blg_op_lgn_kt_s00_dne6_Social_LAB\" target=\"_blank\" rel=\"noopener noreferrer\">\u4fdd\u5b88\u70b9\u691c\u30b5\u30fc\u30d3\u30b9<\/a>\u3092\u691c\u8a0e\u3057\u3066\u307f\u3066\u304f\u3060\u3055\u3044\u3002<\/p>\n<p><a href=\"https:\/\/www.thermofisher.com\/jp\/ja\/home\/global\/forms\/pcr-inquiry.html?CID=gsd_cap_sbu_r04_jp_cp1349_pjt5618_gsd00000_0so_blg_op_lgn_kt_s00_dne6_Social_LAB\" target=\"_blank\" rel=\"noopener noreferrer\">\u907a\u4f1d\u5b50\u89e3\u6790\u6a5f\u5668\u306b\u95a2\u3059\u308b\u304a\u554f\u3044\u5408\u308f\u305b\u30fb\u8a2a\u554f\u4f9d\u983c\u30fb\u8cc7\u6599\u8acb\u6c42\u306f\u3053\u3061\u3089<\/a><\/p>\n<h2>\uff3b\u7121\u6599\u516c\u958b\u4e2d\uff3d\u30b5\u30f3\u30ac\u30fc\u30b7\u30fc\u30b1\u30f3\u30b9\u306e\u57fa\u790e<\/h2>\n<p>\u30b7\u30fc\u30b1\u30f3\u30b9\u306e\u30ef\u30fc\u30af\u30d5\u30ed\u30fc\u3084\u30c7\u30fc\u30bf\u304b\u3089\u306e\u30c8\u30e9\u30d6\u30eb\u30b7\u30e5\u30fc\u30c8\u306b\u3064\u3044\u3066\u306f\u3001\u3053\u3061\u3089\u306e\u30ac\u30a4\u30c9\u306b\u3082\u8a73\u3057\u3044\u8a18\u8f09\u304c\u3054\u3056\u3044\u307e\u3059\u3002\u305c\u3072\u3054\u6d3b\u7528\u304f\u3060\u3055\u3044\u3002<\/p>\n<p><a class=\"btn btn-red\" href=\"https:\/\/www.thermofisher.com\/jp\/ja\/home\/global\/forms\/sanger-sequencing-guide-download.html?CID=gsd_cap_sbu_r04_jp_cp1349_pjt5618_gsd00000_0so_blg_op_lgn_kt_s00_dne6_Social_LAB\" target=\"_blank\" rel=\"noopener noreferrer\">\u30c0\u30a6\u30f3\u30ed\u30fc\u30c9\u3059\u308b<\/a><\/p>\n<p>&nbsp;<\/p>\n<p>\u7814\u7a76\u7528\u306b\u306e\u307f\u4f7f\u7528\u3067\u304d\u307e\u3059\u3002\u8a3a\u65ad\u7528\u306b\u306f\u4f7f\u7528\u3044\u305f\u3060\u3051\u307e\u305b\u3093\u3002<\/p>\n","protected":false},"excerpt":{"rendered":"<p>\u30ad\u30e3\u30d4\u30e9\u30ea\u30fc\u30b7\u30fc\u30b1\u30f3\u30b5\uff08\u30b8\u30a7\u30cd\u30c6\u30a3\u30c3\u30af\u30a2\u30ca\u30e9\u30a4\u30b6\uff09\u306f\u5b9a\u671f\u7684\u306a\u30e1\u30f3\u30c6\u30ca\u30f3\u30b9\u3092\u884c\u3046\u3053\u3068\u3067\u3001\u3088\u308a\u304d\u308c\u3044\u3067\u5b89\u5b9a\u3057\u305f\u30c7\u30fc\u30bf\u3092\u5f97\u308b\u3053\u3068\u304c\u3067\u304d\u307e\u3059\u3002 \u305d\u3053\u3067\u4eca\u56de\u306f\u3001\u30a8\u30a2\u30d5\u30a3\u30eb\u30bf\u30fc\u306e\u30e1\u30f3\u30c6\u30ca\u30f3\u30b9\u3064\u3044\u3066\u3001\u3054\u7d39\u4ecb\u3057\u307e\u3059\u3002 \u25bc\u3053\u3093\u306a\u65b9\u306b\u304a\u3059\u3059<\/p>\n","protected":false},"author":975,"featured_media":0,"comment_status":"open","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_kad_blocks_custom_css":"","_kad_blocks_head_custom_js":"","_kad_blocks_body_custom_js":"","_kad_blocks_footer_custom_js":"","_monsterinsights_skip_tracking":false,"_genesis_hide_title":false,"_genesis_hide_breadcrumbs":false,"_genesis_hide_singular_image":false,"_genesis_hide_footer_widgets":false,"_genesis_custom_body_class":"","_genesis_custom_post_class":"","_genesis_layout":"","_jetpack_newsletter_access":"","_jetpack_dont_email_post_to_subs":false,"_jetpack_newsletter_tier_id":0,"_jetpack_memberships_contains_paywalled_content":false,"_jetpack_feature_clip_id":0,"_jetpack_memberships_contains_paid_content":false,"footnotes":"","jetpack_post_was_ever_published":false},"categories":[156],"tags":[172],"division":[],"class_list":{"0":"post-6865","1":"post","2":"type-post","3":"status-publish","4":"format-standard","6":"category-mol-bio","7":"tag-ce","8":"entry"},"_selected_authors":[],"_selected_reviewers":[],"acf":[],"yoast_head":"<!-- This site is optimized with the Yoast SEO Premium plugin v27.8 (Yoast SEO v27.8) - https:\/\/yoast.com\/product\/yoast-seo-premium-wordpress\/ -->\n<title>\u30ad\u30e3\u30d4\u30e9\u30ea\u30fc\u30b7\u30fc\u30b1\u30f3\u30b5\u306e\u30a8\u30a2\u30d5\u30a3\u30eb\u30bf\u30fc\u3092\u78ba\u8a8d\u3055\u308c\u3066\u3044\u307e\u3059\u304b\uff1f - Learning at the Bench<\/title>\n<meta name=\"robots\" content=\"index, follow, max-snippet:-1, max-image-preview:large, max-video-preview:-1\" \/>\n<link rel=\"canonical\" href=\"https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/airfilter_ce_gsd_ts_1\/\" \/>\n<meta property=\"og:locale\" content=\"ja_JP\" \/>\n<meta property=\"og:type\" content=\"article\" \/>\n<meta property=\"og:title\" content=\"\u30ad\u30e3\u30d4\u30e9\u30ea\u30fc\u30b7\u30fc\u30b1\u30f3\u30b5\u306e\u30a8\u30a2\u30d5\u30a3\u30eb\u30bf\u30fc\u3092\u78ba\u8a8d\u3055\u308c\u3066\u3044\u307e\u3059\u304b\uff1f\" \/>\n<meta property=\"og:description\" content=\"\u30ad\u30e3\u30d4\u30e9\u30ea\u30fc\u30b7\u30fc\u30b1\u30f3\u30b5\uff08\u30b8\u30a7\u30cd\u30c6\u30a3\u30c3\u30af\u30a2\u30ca\u30e9\u30a4\u30b6\uff09\u306f\u5b9a\u671f\u7684\u306a\u30e1\u30f3\u30c6\u30ca\u30f3\u30b9\u3092\u884c\u3046\u3053\u3068\u3067\u3001\u3088\u308a\u304d\u308c\u3044\u3067\u5b89\u5b9a\u3057\u305f\u30c7\u30fc\u30bf\u3092\u5f97\u308b\u3053\u3068\u304c\u3067\u304d\u307e\u3059\u3002 \u305d\u3053\u3067\u4eca\u56de\u306f\u3001\u30a8\u30a2\u30d5\u30a3\u30eb\u30bf\u30fc\u306e\u30e1\u30f3\u30c6\u30ca\u30f3\u30b9\u3064\u3044\u3066\u3001\u3054\u7d39\u4ecb\u3057\u307e\u3059\u3002 \u25bc\u3053\u3093\u306a\u65b9\u306b\u304a\u3059\u3059\" \/>\n<meta property=\"og:url\" content=\"https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/airfilter_ce_gsd_ts_1\/\" \/>\n<meta property=\"og:site_name\" content=\"Learning at the Bench\" \/>\n<meta property=\"article:publisher\" content=\"https:\/\/www.facebook.com\/ThermoFisherJapan\" \/>\n<meta property=\"article:published_time\" content=\"2020-06-01T06:09:45+00:00\" \/>\n<meta property=\"article:modified_time\" content=\"2025-12-19T07:44:41+00:00\" \/>\n<meta name=\"author\" content=\"LatB Staff\" \/>\n<meta name=\"twitter:card\" content=\"summary_large_image\" \/>\n<meta name=\"twitter:creator\" content=\"@ThermoFisherJP\" \/>\n<meta name=\"twitter:site\" content=\"@ThermoFisherJP\" \/>\n<meta name=\"twitter:label1\" content=\"\u57f7\u7b46\u8005\" \/>\n\t<meta name=\"twitter:data1\" content=\"LatB Staff\" \/>\n\t<meta name=\"twitter:label2\" content=\"\u63a8\u5b9a\u8aad\u307f\u53d6\u308a\u6642\u9593\" \/>\n\t<meta name=\"twitter:data2\" content=\"3\u5206\" \/>\n<script type=\"application\/ld+json\" class=\"yoast-schema-graph\">{\"@context\":\"https:\\\/\\\/schema.org\",\"@graph\":[{\"@type\":\"Article\",\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/learning-at-the-bench\\\/airfilter_ce_gsd_ts_1\\\/#article\",\"isPartOf\":{\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/learning-at-the-bench\\\/airfilter_ce_gsd_ts_1\\\/\"},\"author\":{\"name\":\"LatB Staff\",\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/learning-at-the-bench\\\/#\\\/schema\\\/person\\\/bdbe319a2001f279ddce6ab8eaad1253\"},\"headline\":\"\u30ad\u30e3\u30d4\u30e9\u30ea\u30fc\u30b7\u30fc\u30b1\u30f3\u30b5\u306e\u30a8\u30a2\u30d5\u30a3\u30eb\u30bf\u30fc\u3092\u78ba\u8a8d\u3055\u308c\u3066\u3044\u307e\u3059\u304b\uff1f\",\"datePublished\":\"2020-06-01T06:09:45+00:00\",\"dateModified\":\"2025-12-19T07:44:41+00:00\",\"mainEntityOfPage\":{\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/learning-at-the-bench\\\/airfilter_ce_gsd_ts_1\\\/\"},\"wordCount\":9,\"commentCount\":0,\"keywords\":[\"\u30ad\u30e3\u30d4\u30e9\u30ea\u30fc\u96fb\u6c17\u6cf3\u52d5\"],\"articleSection\":[\"\u5206\u5b50\u751f\u7269\u5b66\u5b9f\u9a13\u95a2\u9023\"],\"inLanguage\":\"ja\",\"potentialAction\":[{\"@type\":\"CommentAction\",\"name\":\"Comment\",\"target\":[\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/learning-at-the-bench\\\/airfilter_ce_gsd_ts_1\\\/#respond\"]}]},{\"@type\":\"WebPage\",\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/learning-at-the-bench\\\/airfilter_ce_gsd_ts_1\\\/\",\"url\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/learning-at-the-bench\\\/airfilter_ce_gsd_ts_1\\\/\",\"name\":\"\u30ad\u30e3\u30d4\u30e9\u30ea\u30fc\u30b7\u30fc\u30b1\u30f3\u30b5\u306e\u30a8\u30a2\u30d5\u30a3\u30eb\u30bf\u30fc\u3092\u78ba\u8a8d\u3055\u308c\u3066\u3044\u307e\u3059\u304b\uff1f - Learning at the Bench\",\"isPartOf\":{\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/learning-at-the-bench\\\/#website\"},\"datePublished\":\"2020-06-01T06:09:45+00:00\",\"dateModified\":\"2025-12-19T07:44:41+00:00\",\"author\":{\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/learning-at-the-bench\\\/#\\\/schema\\\/person\\\/bdbe319a2001f279ddce6ab8eaad1253\"},\"breadcrumb\":{\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/learning-at-the-bench\\\/airfilter_ce_gsd_ts_1\\\/#breadcrumb\"},\"inLanguage\":\"ja\",\"potentialAction\":[{\"@type\":\"ReadAction\",\"target\":[\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/learning-at-the-bench\\\/airfilter_ce_gsd_ts_1\\\/\"]}]},{\"@type\":\"BreadcrumbList\",\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/learning-at-the-bench\\\/airfilter_ce_gsd_ts_1\\\/#breadcrumb\",\"itemListElement\":[{\"@type\":\"ListItem\",\"position\":1,\"name\":\"Home\",\"item\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/learning-at-the-bench\\\/\"},{\"@type\":\"ListItem\",\"position\":2,\"name\":\"\u30ad\u30e3\u30d4\u30e9\u30ea\u30fc\u30b7\u30fc\u30b1\u30f3\u30b5\u306e\u30a8\u30a2\u30d5\u30a3\u30eb\u30bf\u30fc\u3092\u78ba\u8a8d\u3055\u308c\u3066\u3044\u307e\u3059\u304b\uff1f\"}]},{\"@type\":\"WebSite\",\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/learning-at-the-bench\\\/#website\",\"url\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/learning-at-the-bench\\\/\",\"name\":\"Learning at the Bench\",\"description\":\"\",\"potentialAction\":[{\"@type\":\"SearchAction\",\"target\":{\"@type\":\"EntryPoint\",\"urlTemplate\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/learning-at-the-bench\\\/?s={search_term_string}\"},\"query-input\":{\"@type\":\"PropertyValueSpecification\",\"valueRequired\":true,\"valueName\":\"search_term_string\"}}],\"inLanguage\":\"ja\"},{\"@type\":\"Person\",\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/learning-at-the-bench\\\/#\\\/schema\\\/person\\\/bdbe319a2001f279ddce6ab8eaad1253\",\"name\":\"LatB Staff\",\"image\":{\"@type\":\"ImageObject\",\"inLanguage\":\"ja\",\"@id\":\"https:\\\/\\\/secure.gravatar.com\\\/avatar\\\/2b3fdaccc53ad9dae08baccbcf9e8774350bd5682b89e353a9b7a999834e9306?s=96&d=mm&r=g\",\"url\":\"https:\\\/\\\/secure.gravatar.com\\\/avatar\\\/2b3fdaccc53ad9dae08baccbcf9e8774350bd5682b89e353a9b7a999834e9306?s=96&d=mm&r=g\",\"contentUrl\":\"https:\\\/\\\/secure.gravatar.com\\\/avatar\\\/2b3fdaccc53ad9dae08baccbcf9e8774350bd5682b89e353a9b7a999834e9306?s=96&d=mm&r=g\",\"caption\":\"LatB Staff\"},\"url\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/author\\\/lgbadmin\\\/\"}]}<\/script>\n<!-- \/ Yoast SEO Premium plugin. -->","yoast_head_json":{"title":"\u30ad\u30e3\u30d4\u30e9\u30ea\u30fc\u30b7\u30fc\u30b1\u30f3\u30b5\u306e\u30a8\u30a2\u30d5\u30a3\u30eb\u30bf\u30fc\u3092\u78ba\u8a8d\u3055\u308c\u3066\u3044\u307e\u3059\u304b\uff1f - Learning at the Bench","robots":{"index":"index","follow":"follow","max-snippet":"max-snippet:-1","max-image-preview":"max-image-preview:large","max-video-preview":"max-video-preview:-1"},"canonical":"https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/airfilter_ce_gsd_ts_1\/","og_locale":"ja_JP","og_type":"article","og_title":"\u30ad\u30e3\u30d4\u30e9\u30ea\u30fc\u30b7\u30fc\u30b1\u30f3\u30b5\u306e\u30a8\u30a2\u30d5\u30a3\u30eb\u30bf\u30fc\u3092\u78ba\u8a8d\u3055\u308c\u3066\u3044\u307e\u3059\u304b\uff1f","og_description":"\u30ad\u30e3\u30d4\u30e9\u30ea\u30fc\u30b7\u30fc\u30b1\u30f3\u30b5\uff08\u30b8\u30a7\u30cd\u30c6\u30a3\u30c3\u30af\u30a2\u30ca\u30e9\u30a4\u30b6\uff09\u306f\u5b9a\u671f\u7684\u306a\u30e1\u30f3\u30c6\u30ca\u30f3\u30b9\u3092\u884c\u3046\u3053\u3068\u3067\u3001\u3088\u308a\u304d\u308c\u3044\u3067\u5b89\u5b9a\u3057\u305f\u30c7\u30fc\u30bf\u3092\u5f97\u308b\u3053\u3068\u304c\u3067\u304d\u307e\u3059\u3002 \u305d\u3053\u3067\u4eca\u56de\u306f\u3001\u30a8\u30a2\u30d5\u30a3\u30eb\u30bf\u30fc\u306e\u30e1\u30f3\u30c6\u30ca\u30f3\u30b9\u3064\u3044\u3066\u3001\u3054\u7d39\u4ecb\u3057\u307e\u3059\u3002 \u25bc\u3053\u3093\u306a\u65b9\u306b\u304a\u3059\u3059","og_url":"https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/airfilter_ce_gsd_ts_1\/","og_site_name":"Learning at the Bench","article_publisher":"https:\/\/www.facebook.com\/ThermoFisherJapan","article_published_time":"2020-06-01T06:09:45+00:00","article_modified_time":"2025-12-19T07:44:41+00:00","author":"LatB Staff","twitter_card":"summary_large_image","twitter_creator":"@ThermoFisherJP","twitter_site":"@ThermoFisherJP","twitter_misc":{"\u57f7\u7b46\u8005":"LatB Staff","\u63a8\u5b9a\u8aad\u307f\u53d6\u308a\u6642\u9593":"3\u5206"},"schema":{"@context":"https:\/\/schema.org","@graph":[{"@type":"Article","@id":"https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/airfilter_ce_gsd_ts_1\/#article","isPartOf":{"@id":"https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/airfilter_ce_gsd_ts_1\/"},"author":{"name":"LatB Staff","@id":"https:\/\/admin.acceleratingscience.com\/learning-at-the-bench\/#\/schema\/person\/bdbe319a2001f279ddce6ab8eaad1253"},"headline":"\u30ad\u30e3\u30d4\u30e9\u30ea\u30fc\u30b7\u30fc\u30b1\u30f3\u30b5\u306e\u30a8\u30a2\u30d5\u30a3\u30eb\u30bf\u30fc\u3092\u78ba\u8a8d\u3055\u308c\u3066\u3044\u307e\u3059\u304b\uff1f","datePublished":"2020-06-01T06:09:45+00:00","dateModified":"2025-12-19T07:44:41+00:00","mainEntityOfPage":{"@id":"https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/airfilter_ce_gsd_ts_1\/"},"wordCount":9,"commentCount":0,"keywords":["\u30ad\u30e3\u30d4\u30e9\u30ea\u30fc\u96fb\u6c17\u6cf3\u52d5"],"articleSection":["\u5206\u5b50\u751f\u7269\u5b66\u5b9f\u9a13\u95a2\u9023"],"inLanguage":"ja","potentialAction":[{"@type":"CommentAction","name":"Comment","target":["https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/airfilter_ce_gsd_ts_1\/#respond"]}]},{"@type":"WebPage","@id":"https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/airfilter_ce_gsd_ts_1\/","url":"https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/airfilter_ce_gsd_ts_1\/","name":"\u30ad\u30e3\u30d4\u30e9\u30ea\u30fc\u30b7\u30fc\u30b1\u30f3\u30b5\u306e\u30a8\u30a2\u30d5\u30a3\u30eb\u30bf\u30fc\u3092\u78ba\u8a8d\u3055\u308c\u3066\u3044\u307e\u3059\u304b\uff1f - Learning at the Bench","isPartOf":{"@id":"https:\/\/admin.acceleratingscience.com\/learning-at-the-bench\/#website"},"datePublished":"2020-06-01T06:09:45+00:00","dateModified":"2025-12-19T07:44:41+00:00","author":{"@id":"https:\/\/admin.acceleratingscience.com\/learning-at-the-bench\/#\/schema\/person\/bdbe319a2001f279ddce6ab8eaad1253"},"breadcrumb":{"@id":"https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/airfilter_ce_gsd_ts_1\/#breadcrumb"},"inLanguage":"ja","potentialAction":[{"@type":"ReadAction","target":["https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/airfilter_ce_gsd_ts_1\/"]}]},{"@type":"BreadcrumbList","@id":"https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/airfilter_ce_gsd_ts_1\/#breadcrumb","itemListElement":[{"@type":"ListItem","position":1,"name":"Home","item":"https:\/\/admin.acceleratingscience.com\/learning-at-the-bench\/"},{"@type":"ListItem","position":2,"name":"\u30ad\u30e3\u30d4\u30e9\u30ea\u30fc\u30b7\u30fc\u30b1\u30f3\u30b5\u306e\u30a8\u30a2\u30d5\u30a3\u30eb\u30bf\u30fc\u3092\u78ba\u8a8d\u3055\u308c\u3066\u3044\u307e\u3059\u304b\uff1f"}]},{"@type":"WebSite","@id":"https:\/\/admin.acceleratingscience.com\/learning-at-the-bench\/#website","url":"https:\/\/admin.acceleratingscience.com\/learning-at-the-bench\/","name":"Learning at the Bench","description":"","potentialAction":[{"@type":"SearchAction","target":{"@type":"EntryPoint","urlTemplate":"https:\/\/admin.acceleratingscience.com\/learning-at-the-bench\/?s={search_term_string}"},"query-input":{"@type":"PropertyValueSpecification","valueRequired":true,"valueName":"search_term_string"}}],"inLanguage":"ja"},{"@type":"Person","@id":"https:\/\/admin.acceleratingscience.com\/learning-at-the-bench\/#\/schema\/person\/bdbe319a2001f279ddce6ab8eaad1253","name":"LatB Staff","image":{"@type":"ImageObject","inLanguage":"ja","@id":"https:\/\/secure.gravatar.com\/avatar\/2b3fdaccc53ad9dae08baccbcf9e8774350bd5682b89e353a9b7a999834e9306?s=96&d=mm&r=g","url":"https:\/\/secure.gravatar.com\/avatar\/2b3fdaccc53ad9dae08baccbcf9e8774350bd5682b89e353a9b7a999834e9306?s=96&d=mm&r=g","contentUrl":"https:\/\/secure.gravatar.com\/avatar\/2b3fdaccc53ad9dae08baccbcf9e8774350bd5682b89e353a9b7a999834e9306?s=96&d=mm&r=g","caption":"LatB Staff"},"url":"https:\/\/admin.acceleratingscience.com\/author\/lgbadmin\/"}]}},"taxonomy_info":{"category":[{"value":156,"label":"\u5206\u5b50\u751f\u7269\u5b66\u5b9f\u9a13\u95a2\u9023"}],"post_tag":[{"value":172,"label":"\u30ad\u30e3\u30d4\u30e9\u30ea\u30fc\u96fb\u6c17\u6cf3\u52d5"}]},"featured_image_src_large":false,"author_info":{"display_name":"LatB Staff","author_link":"https:\/\/admin.acceleratingscience.com\/author\/lgbadmin\/"},"comment_info":0,"category_info":[{"term_id":156,"name":"\u5206\u5b50\u751f\u7269\u5b66\u5b9f\u9a13\u95a2\u9023","slug":"mol-bio","term_group":0,"term_taxonomy_id":156,"taxonomy":"category","description":"","parent":0,"count":538,"filter":"raw","meta":[],"cat_ID":156,"category_count":538,"category_description":"","cat_name":"\u5206\u5b50\u751f\u7269\u5b66\u5b9f\u9a13\u95a2\u9023","category_nicename":"mol-bio","category_parent":0}],"tag_info":[{"term_id":172,"name":"\u30ad\u30e3\u30d4\u30e9\u30ea\u30fc\u96fb\u6c17\u6cf3\u52d5","slug":"ce","term_group":0,"term_taxonomy_id":172,"taxonomy":"post_tag","description":"","parent":0,"count":34,"filter":"raw","meta":[]}],"jetpack-related-posts":[{"id":24647,"url":"https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/sanger-seq-mutational-analysis-gsd-ts-ce-26007\/","url_meta":{"origin":6865,"position":0},"title":"\u30de\u30a4\u30ca\u30fc\u5909\u7570\u3092\u691c\u51fa\uff01Minor Variant Finder\u30bd\u30d5\u30c8\u30a6\u30a8\u30a2","author":"LatB Staff","date":"2026\u5e742\u67089\u65e5","format":false,"excerpt":"\u30ad\u30e3\u30d4\u30e9\u30ea\u30fc\u30b7\u30fc\u30b1\u30f3\u30b5\u3067\u306f\u300125%\u7a0b\u5ea6\u306e\u5272\u5408\u306e\u5909\u7570\u3092\u691c\u51fa\u3059\u308b\u3053\u3068\u304c\u3067\u304d\u307e\u3059\u304c\u3001\u3053\u308c\u3088\u308a\u3082\u4f4e\u3044\u5272\u5408\u306e\u5909\u2026","rel":"","context":"\u5206\u5b50\u751f\u7269\u5b66\u5b9f\u9a13\u95a2\u9023","block_context":{"text":"\u5206\u5b50\u751f\u7269\u5b66\u5b9f\u9a13\u95a2\u9023","link":"https:\/\/admin.acceleratingscience.com\/learning-at-the-bench\/mol-bio\/"},"img":{"alt_text":"","src":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2026\/01\/sanger-seq-mutational-analysis-gsd-ts-ce-26007_eyecatch.jpg?resize=350%2C200&ssl=1","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2026\/01\/sanger-seq-mutational-analysis-gsd-ts-ce-26007_eyecatch.jpg?resize=350%2C200&ssl=1 1x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2026\/01\/sanger-seq-mutational-analysis-gsd-ts-ce-26007_eyecatch.jpg?resize=525%2C300&ssl=1 1.5x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2026\/01\/sanger-seq-mutational-analysis-gsd-ts-ce-26007_eyecatch.jpg?resize=700%2C400&ssl=1 2x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2026\/01\/sanger-seq-mutational-analysis-gsd-ts-ce-26007_eyecatch.jpg?resize=1050%2C600&ssl=1 3x"},"classes":[]},{"id":25090,"url":"https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/battery-development-elemental-analysis-csm-26016\/","url_meta":{"origin":6865,"position":1},"title":"\u30ea\u30c1\u30a6\u30e0\u30a4\u30aa\u30f3\u96fb\u6c60\u88fd\u9020\uff1a\u4f7f\u7528\u3055\u308c\u308b\u5316\u5b66\u5206\u6790\u6280\u8853","author":"LatB Staff","date":"2026\u5e744\u670820\u65e5","format":false,"excerpt":"\u9ad8\u6027\u80fd\u96fb\u6c60\u306b\u5bfe\u3059\u308b\u9700\u8981\u306e\u6025\u5897\u306f\u3001\u751f\u7523\u306e\u3042\u3089\u3086\u308b\u6bb5\u968e\u306b\u304a\u3051\u308b\u5927\u898f\u6a21\u306a\u6295\u8cc7\u3092\u4fc3\u3057\u3066\u3044\u307e\u3059\u3002\u96fb\u6c60\u7814\u7a76\u304a\u3088\u3073\u2026","rel":"","context":"\u5206\u6790\u30fb\u6e2c\u5b9a\u95a2\u9023","block_context":{"text":"\u5206\u6790\u30fb\u6e2c\u5b9a\u95a2\u9023","link":"https:\/\/admin.acceleratingscience.com\/learning-at-the-bench\/analysis\/"},"img":{"alt_text":"","src":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2026\/04\/battery-development-elemental-analysis-csm-26016_eyecatch.jpg?resize=350%2C200&ssl=1","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2026\/04\/battery-development-elemental-analysis-csm-26016_eyecatch.jpg?resize=350%2C200&ssl=1 1x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2026\/04\/battery-development-elemental-analysis-csm-26016_eyecatch.jpg?resize=525%2C300&ssl=1 1.5x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2026\/04\/battery-development-elemental-analysis-csm-26016_eyecatch.jpg?resize=700%2C400&ssl=1 2x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2026\/04\/battery-development-elemental-analysis-csm-26016_eyecatch.jpg?resize=1050%2C600&ssl=1 3x"},"classes":[]},{"id":24587,"url":"https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/twin-screw-extrusion-extruder-cad-26003\/","url_meta":{"origin":6865,"position":2},"title":"\u4e8c\u8ef8\u62bc\u51fa\u6a5f\u306e\u57fa\u790e\u2468 \u5f8c\u5de5\u7a0b\u3068\u30b5\u30f3\u30d7\u30eb\u306e\u56de\u53ce \u30da\u30ec\u30bf\u30a4\u30ba","author":"LatB Staff","date":"2026\u5e741\u670822\u65e5","format":false,"excerpt":"\u66f4\u65b0\u65e5 : \u3053\u306e\u8a18\u4e8b\u306e\u8981\u70b9 \u30da\u30ec\u30bf\u30a4\u30ba\u306f\u4e8c\u8ef8\u62bc\u51fa\u6a5f\u306b\u3088\u308b\u30b3\u30f3\u30d1\u30a6\u30f3\u30c7\u30a3\u30f3\u30b0\u5f8c\u306e\u4ee3\u8868\u7684\u306a\u30b5\u30f3\u30d7\u30eb\u56de\u53ce\u6cd5\u2026","rel":"","context":"\u5206\u6790\u30fb\u6e2c\u5b9a\u95a2\u9023","block_context":{"text":"\u5206\u6790\u30fb\u6e2c\u5b9a\u95a2\u9023","link":"https:\/\/admin.acceleratingscience.com\/learning-at-the-bench\/analysis\/"},"img":{"alt_text":"","src":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2026\/01\/twin-screw-extrusion-extruder-cad-26003_eyeycatch.jpg?resize=350%2C200&ssl=1","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2026\/01\/twin-screw-extrusion-extruder-cad-26003_eyeycatch.jpg?resize=350%2C200&ssl=1 1x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2026\/01\/twin-screw-extrusion-extruder-cad-26003_eyeycatch.jpg?resize=525%2C300&ssl=1 1.5x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2026\/01\/twin-screw-extrusion-extruder-cad-26003_eyeycatch.jpg?resize=700%2C400&ssl=1 2x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2026\/01\/twin-screw-extrusion-extruder-cad-26003_eyeycatch.jpg?resize=1050%2C600&ssl=1 3x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2026\/01\/twin-screw-extrusion-extruder-cad-26003_eyeycatch.jpg?resize=1400%2C800&ssl=1 4x"},"classes":[]},{"id":24462,"url":"https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/battery-development-microscopy-csm-25076\/","url_meta":{"origin":6865,"position":3},"title":"\u5168\u56fa\u4f53\u96fb\u6c60\u306e\u9769\u65b0\u3092\u652f\u3048\u308b\u5148\u7aef\u7684\u7279\u6027\u8a55\u4fa1\u6280\u8853\u306e\u63a2\u7a76","author":"LatB Staff","date":"2025\u5e7412\u670817\u65e5","format":false,"excerpt":"\u5168\u56fa\u4f53\u96fb\u6c60\u306e\u958b\u767a\u3067\u306f\u3001\u4f7f\u7528\u3055\u308c\u308b\u6750\u6599\u306e\u7406\u89e3\u3092\u6df1\u3081\u308b\u305f\u3081\u306b\u3001\u5148\u7aef\u7684\u306a\u6027\u8a55\u4fa1\u6280\u8853\u304c\u4e0d\u53ef\u6b20\u3067\u3059\u3002\u3053\u308c\u3089\u306e\u624b\u2026","rel":"","context":"\u5206\u6790\u30fb\u6e2c\u5b9a\u95a2\u9023","block_context":{"text":"\u5206\u6790\u30fb\u6e2c\u5b9a\u95a2\u9023","link":"https:\/\/admin.acceleratingscience.com\/learning-at-the-bench\/analysis\/"},"img":{"alt_text":"","src":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2025\/12\/battery-development-microscopy-csm-25076_eyecatch.jpg?resize=350%2C200&ssl=1","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2025\/12\/battery-development-microscopy-csm-25076_eyecatch.jpg?resize=350%2C200&ssl=1 1x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2025\/12\/battery-development-microscopy-csm-25076_eyecatch.jpg?resize=525%2C300&ssl=1 1.5x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2025\/12\/battery-development-microscopy-csm-25076_eyecatch.jpg?resize=700%2C400&ssl=1 2x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2025\/12\/battery-development-microscopy-csm-25076_eyecatch.jpg?resize=1050%2C600&ssl=1 3x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2025\/12\/battery-development-microscopy-csm-25076_eyecatch.jpg?resize=1400%2C800&ssl=1 4x"},"classes":[]},{"id":24427,"url":"https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/twin-screw-extrusion-extruder-cad-25076\/","url_meta":{"origin":6865,"position":4},"title":"\u4e8c\u8ef8\u62bc\u51fa\u6a5f\u306e\u57fa\u790e\u2467\uff1a\u5f8c\u5de5\u7a0b\u3068\u30b5\u30f3\u30d7\u30eb\u306e\u56de\u53ce 3D\u30d7\u30ea\u30f3\u30c6\u30a3\u30f3\u30b0\u7528\u30d5\u30a3\u30e9\u30e1\u30f3\u30c8\u306e\u4f5c\u88fd","author":"LatB Staff","date":"2025\u5e7412\u670816\u65e5","format":false,"excerpt":"\u4e8c\u8ef8\u62bc\u51fa\u6a5f\u306e\u57fa\u790e\u30b7\u30ea\u30fc\u30ba\u3067\u306f\u88c5\u7f6e\u306e\u4ed5\u7d44\u307f\u3084\u7279\u9577\u3001\u30d1\u30e9\u30e1\u30fc\u30bf\u30fc\u8a2d\u5b9a\u3084\u305d\u306e\u610f\u5473\u3001\u5f8c\u5de5\u7a0b\u30aa\u30d7\u30b7\u30e7\u30f3\u3092\u7d44\u307f\u5408\u2026","rel":"","context":"\u305d\u306e\u4ed6","block_context":{"text":"\u305d\u306e\u4ed6","link":"https:\/\/admin.acceleratingscience.com\/learning-at-the-bench\/else\/"},"img":{"alt_text":"","src":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2025\/12\/0a52b388a96d915f746c8911521e3771.png?resize=350%2C200&ssl=1","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2025\/12\/0a52b388a96d915f746c8911521e3771.png?resize=350%2C200&ssl=1 1x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2025\/12\/0a52b388a96d915f746c8911521e3771.png?resize=525%2C300&ssl=1 1.5x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2025\/12\/0a52b388a96d915f746c8911521e3771.png?resize=700%2C400&ssl=1 2x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2025\/12\/0a52b388a96d915f746c8911521e3771.png?resize=1050%2C600&ssl=1 3x"},"classes":[]},{"id":24108,"url":"https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/cell-culture-cell-count-bid-ts-mbr-25065-2\/","url_meta":{"origin":6865,"position":5},"title":"\u3010\u7b2c6\u56de\u3011 \u30a6\u30a7\u30b9\u30bf\u30f3\u30d6\u30ed\u30c3\u30c6\u30a3\u30f3\u30b0\u6cd5\u306b\u3088\u308b \u300e\u30a8\u30af\u30bd\u30bd\u30fc\u30e0\u4e2d\u306e\u30bf\u30f3\u30d1\u30af\u8cea\uff08Protein cargo\uff09\u300f\u306e\u89e3\u6790","author":"LatB Staff","date":"2026\u5e747\u670815\u65e5","format":false,"excerpt":"\u30a8\u30af\u30bd\u30bd\u30fc\u30e0\uff08\u30a8\u30ad\u30bd\u30bd\u30fc\u30e0\uff09\u306f\u3001\u7d30\u80de\u304b\u3089\u5206\u6ccc\u3055\u308c\u308b\u5c0f\u3055\u306a\u819c\u5c0f\u80de\u3067\u3001\u6838\u9178\uff08RNA\uff0fDNA\uff09\u3060\u3051\u3067\u306a\u304f\u3001\u2026","rel":"","context":"\u5206\u5b50\u751f\u7269\u5b66\u5b9f\u9a13\u95a2\u9023","block_context":{"text":"\u5206\u5b50\u751f\u7269\u5b66\u5b9f\u9a13\u95a2\u9023","link":"https:\/\/admin.acceleratingscience.com\/learning-at-the-bench\/mol-bio\/"},"img":{"alt_text":"","src":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2025\/11\/cell-culture-cell-count-bid-ts-mbr-25065_eyecatch.jpg?resize=350%2C200&ssl=1","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2025\/11\/cell-culture-cell-count-bid-ts-mbr-25065_eyecatch.jpg?resize=350%2C200&ssl=1 1x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2025\/11\/cell-culture-cell-count-bid-ts-mbr-25065_eyecatch.jpg?resize=525%2C300&ssl=1 1.5x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2025\/11\/cell-culture-cell-count-bid-ts-mbr-25065_eyecatch.jpg?resize=700%2C400&ssl=1 2x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2025\/11\/cell-culture-cell-count-bid-ts-mbr-25065_eyecatch.jpg?resize=1050%2C600&ssl=1 3x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/learning-at-the-bench\/wp-content\/uploads\/sites\/13\/2025\/11\/cell-culture-cell-count-bid-ts-mbr-25065_eyecatch.jpg?resize=1400%2C800&ssl=1 4x"},"classes":[]}],"jetpack_sharing_enabled":true,"amp_enabled":true,"jetpack_featured_media_url":"","_links":{"self":[{"href":"https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/wp-json\/wp\/v2\/posts\/6865","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/wp-json\/wp\/v2\/users\/975"}],"replies":[{"embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/wp-json\/wp\/v2\/comments?post=6865"}],"version-history":[{"count":0,"href":"https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/wp-json\/wp\/v2\/posts\/6865\/revisions"}],"wp:attachment":[{"href":"https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/wp-json\/wp\/v2\/media?parent=6865"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/wp-json\/wp\/v2\/categories?post=6865"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/wp-json\/wp\/v2\/tags?post=6865"},{"taxonomy":"division","embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/learning-at-the-bench\/wp-json\/wp\/v2\/division?post=6865"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}