{"id":15913,"date":"2026-08-11T07:00:00","date_gmt":"2026-08-11T12:00:00","guid":{"rendered":"https:\/\/www.thermofisher.com\/blog\/materials\/?p=15913"},"modified":"2026-08-11T07:16:22","modified_gmt":"2026-08-11T12:16:22","slug":"simplify-quality-control-testing-with-uv-vis-custom-methods-for-pass-fail-evaluation","status":"publish","type":"post","link":"https:\/\/www.thermofisher.com\/blog\/materials\/simplify-quality-control-testing-with-uv-vis-custom-methods-for-pass-fail-evaluation\/","title":{"rendered":"Simplify Quality Control Testing with UV-Vis Custom Methods for Pass\/Fail Evaluation"},"content":{"rendered":"\n<div class=\"wp-block-tf-blocks-intro-summary\">\n<h2 class=\"wp-block-heading\" id=\"h-article-summary\">Article Summary<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">UV-Vis spectroscopy enables fast, non-destructive quality control by measuring sample absorbance and comparing results against predefined acceptance criteria. Automated custom methods simplify pass\/fail testing by detecting concentration errors and contamination and reducing manual interpretation, thereby helping laboratories improve consistency and efficiency across routine manufacturing workflows.<\/p>\n<\/div>\n\n\n\n<div class=\"wp-block-tf-blocks-definition-block\">\n<h2 class=\"wp-block-heading\" id=\"h-simplify-quality-control-testing-with-uv-vis-for-pass-fail-evaluation\">Simplify Quality Control Testing with UV-Vis for Pass\/Fail Evaluation<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">Quality control is a critical part of manufacturing, helping ensure every product meets established specifications before it reaches customers. Whether verifying the concentration of an active ingredient or detecting potential cross-contamination between products, fast and reliable analytical testing can improve process consistency while reducing costly errors.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">UV-Visible (UV-Vis) spectroscopy offers a simple, non-destructive way to monitor many solution-based products. With newer <a href=\"https:\/\/www.thermofisher.com\/order\/catalog\/product\/840-411100?icid=CAD_blog_materials_2026Aug\" id=\"https:\/\/www.thermofisher.com\/order\/catalog\/product\/840-411100?SID=srch-srp-840-411100\" target=\"_blank\" rel=\"noreferrer noopener\">advanced spectrophotometer models<\/a>, laboratories can go beyond basic absorbance measurement workflows by creating custom methods that automatically evaluate samples against predefined acceptance criteria, providing clear pass\/fail results in seconds.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\" id=\"h-using-uv-vis-spectroscopy-for-quality-control\">Using UV-Vis spectroscopy for quality control<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">Many compounds\u2014including dyes, flavor additives, caffeine, and other UV-Vis absorbing substances\u2014produce characteristic absorption spectra that can be used to identify and quantify them.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">According to Beer\u2019s Law, absorbance is directly proportional to concentration. This relationship allows UV-Vis spectroscopy to determine whether the concentration of a particular species of interest falls within an acceptable concentration range for the product. Because the technique is non-destructive, samples remain available for additional testing or downstream processing if needed.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">These capabilities make <a href=\"http:\/\/thermofisher.com\/uvvis?icid=CAD_blog_materials_2026Aug\">UV-Vis spectroscopy<\/a> a valuable tool for routine quality control across a variety of manufacturing environments.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\" id=\"h-automating-quality-decisions-with-custom-methods\">Automating quality decisions with custom methods<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">While absorbance measurements provide valuable analytical data, many manufacturing workflows require a simple answer: Does this sample pass or fail my criteria?<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">A spectrophotometer with a Custom Method application addresses this question, allowing users to create workflows that automatically process the data as needed:<\/p>\n\n\n\n<ul class=\"wp-block-list\">\n<li>Calculate sample concentration<\/li>\n\n\n\n<li>Compare results against predefined acceptance limits<\/li>\n\n\n\n<li>Detect potential contamination<\/li>\n\n\n\n<li>Display immediate pass\/fail results<\/li>\n\n\n\n<li>Reduce manual calculations and operator interpretation<\/li>\n<\/ul>\n\n\n\n<p class=\"wp-block-paragraph\">This approach helps standardize routine testing while improving efficiency and consistency across users.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\" id=\"h-demonstrating-product-quality-testing\">Demonstrating product quality testing<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">To illustrate the workflow, laboratory testing was performed in which two commercially available sports drinks with different color profiles were analyzed. One beverage served as the target product, while the second represented a potential contaminant that could be introduced during manufacturing. Both beverages could be analyzed independently, and then mixed samples of different proportions could be prepared to test how much of the contaminant was detectable alongside the target.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Using UV-Vis spectroscopy, characteristic absorbance peaks were identified for the target product, and calibration data were used to establish acceptable concentration limits. A second calculation monitored wavelengths associated with the potential contaminant, enabling the instrument to determine whether contamination had occurred.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">A custom method combined these calculations into an automated quality control workflow that evaluated every sample against predefined pass\/fail criteria. When programmed properly, the custom method could simply report to the user whether the sample concentration and purity fell within established limits or not, without requiring the user to perform a series of calculations for each sample tested.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\" id=\"h-distinguishing-between-concentration-errors-and-contamination\">Distinguishing between concentration errors and contamination<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">One advantage of custom methods is the ability to identify different types of quality issues within the same analysis.<br>In the demonstration referred to above, the custom workflow successfully distinguished between three categories of samples:<\/p>\n\n\n\n<ul class=\"wp-block-list\">\n<li>Samples that were improperly diluted or mixed\u2014Results from these showed concentrations outside the acceptable range.<\/li>\n\n\n\n<li>Samples that contained contamination\u2014Even when contamination was not visually apparent, UV-Vis analysis was able to detect the presence of species not identified as the desired product.<\/li>\n\n\n\n<li>Samples that met both concentration and purity requirements\u2014In such cases, the custom workflow validated<\/li>\n<\/ul>\n\n\n\n<p class=\"wp-block-paragraph\">This capability provides operators with more meaningful information than a simple absorbance measurement, helping determine whether a manufacturing issue is related to formulation or cross-contamination.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\" id=\"h-enabling-faster-decisions-and-greater-consistency\">Enabling faster decisions and greater consistency<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">By embedding acceptance criteria directly into the analytical method, the use of custom methods simplifies routine quality testing. This brings several benefits:<\/p>\n\n\n\n<ul class=\"wp-block-list\">\n<li>Automated pass\/fail assessments<\/li>\n\n\n\n<li>Reduced manual data interpretation<\/li>\n\n\n\n<li>Consistent decision-making across operators<\/li>\n\n\n\n<li>Faster routine quality checks<\/li>\n\n\n\n<li>Flexible custom methods for application-specific workflows<\/li>\n<\/ul>\n\n\n\n<p class=\"wp-block-paragraph\">Because custom methods can be tailored to specific products and analytical requirements, laboratories can build standardized quality control procedures that align with their manufacturing processes.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\" id=\"h-supporting-routine-manufacturing-quality-control\">Supporting routine manufacturing quality control<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">Although this demonstration used colored sports drinks as an example, the same approach can be applied to monitoring many solution-based products that contain UV-Vis absorbing compounds, including these potential applications:<\/p>\n\n\n\n<ul class=\"wp-block-list\">\n<li>Food and beverage formulations<\/li>\n\n\n\n<li>Pharmaceutical solutions<\/li>\n\n\n\n<li>Chemical manufacturing processes<\/li>\n\n\n\n<li>Flavor and dye concentrations<\/li>\n\n\n\n<li>Other solution-phase products requiring routine quality verification<\/li>\n<\/ul>\n\n\n\n<p class=\"wp-block-paragraph\">By combining UV-Vis spectroscopy with customizable analytical workflows, laboratories can automate routine testing while maintaining confidence in product quality.<strong>\u00a0<\/strong><\/p>\n<\/div>\n\n\n\n<div class=\"wp-block-tf-blocks-takeaway-summary\">\n<h2 class=\"wp-block-heading\" id=\"h-conclusion\">Conclusion<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">Custom methods enable laboratories to automate concentration calculations, contamination checks, and pass\/fail decisions. This automation helps streamline manufacturing workflows while supporting consistent product quality.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Reliable quality control depends on fast, consistent analytical testing that delivers actionable results. UV-Vis spectroscopy provides a non-destructive approach for verifying product concentration and detecting potential contamination, while automated custom methods streamline routine quality assessments with predefined pass\/fail criteria. By embedding acceptance limits directly into analytical workflows, laboratories can reduce manual interpretation, improve consistency across operators, and accelerate manufacturing quality control. This combination of UV-Vis analysis and workflow automation supports standardized testing across food and beverage, pharmaceutical, chemical, and other solution-based manufacturing applications, helping organizations maintain product quality with greater confidence.<\/p>\n<\/div>\n\n\n\n<div class=\"wp-block-tf-blocks-internal-links\">\n<p class=\"wp-block-paragraph\"><em>Further Reading<\/em><\/p>\n\n\n\n<ul class=\"wp-block-list\">\n<li>Application note: Setting pass\/fail criteria with GENESYS custom method application: Sports drink quality testing<br><a href=\"https:\/\/documents.thermofisher.com\/TFS-Assets\/CAD\/Flyers\/genesys-smart-qc-fl54698-en.pdf\" target=\"_blank\" rel=\"noreferrer noopener\">https:\/\/documents.thermofisher.com\/TFS-Assets\/CAD\/Flyers\/genesys-smart-qc-fl54698-en.pdf?icid=CAD_blog_materials_2026Aug<\/a><\/li>\n\n\n\n<li>Application note: Unbiased Color Analysis using UV-Visible Spectrophotometers<br><a href=\"https:\/\/documents.thermofisher.com\/TFS-Assets\/CAD\/Application-Notes\/unbiased-color-analysis-using-uv-visible-spectrophotometers-an1406-en.pdf\" target=\"_blank\" rel=\"noreferrer noopener\">https:\/\/documents.thermofisher.com\/TFS-Assets\/CAD\/Application-Notes\/unbiased-color-analysis-using-uv-visible-spectrophotometers-an1406-en.pdf?icid=CAD_blog_materials_2026Aug<\/a><\/li>\n\n\n\n<li>Application note: Utilizing UV-Visible Spectroscopy for Color Analysis of Fabrics<br><a href=\"https:\/\/documents.thermofisher.com\/TFS-Assets%2FCAD%2FApplication-Notes%2Fcolor-analysis-of-fabric-app-note-1529047.pdf\" target=\"_blank\" rel=\"noreferrer noopener\">https:\/\/documents.thermofisher.com\/TFS-Assets%2FCAD%2FApplication-Notes%2Fcolor-analysis-of-fabric-app-note-1529047.pdf?icid=CAD_blog_materials_2026Aug<\/a><\/li>\n\n\n\n<li>Academic lesson plan: Food Dyes and Beer\u2019s Law<br><a href=\"https:\/\/documents.thermofisher.com\/TFS-Assets\/MSD\/Scientific-Resources\/FL53216-food-dyes-beers-law-lesson-plan-uv-vis.pdf\" target=\"_blank\" rel=\"noreferrer noopener\">https:\/\/documents.thermofisher.com\/TFS-Assets\/MSD\/Scientific-Resources\/FL53216-food-dyes-beers-law-lesson-plan-uv-vis.pdf?icid=CAD_blog_materials_2026Aug<\/a><\/li>\n\n\n\n<li>Online training course: Quantitative Analysis: Intro to Simple Beer&#8217;s Law <a href=\"https:\/\/unitylabservices.gosignmeup.com\/public\/Course\/browse?courseid=4888\">https:\/\/unitylabservices.gosignmeup.com\/public\/Course\/browse?courseid=4888<\/a><\/li>\n<\/ul>\n<\/div>\n\n\n\n<h2 class=\"wp-block-heading\" id=\"h-frequently-asked-questions\">Frequently Asked Questions<\/h2>\n\n\n<style>.kt-accordion-id15913_6c3e4f-03 .kt-accordion-inner-wrap{column-gap:var(--global-kb-gap-md, 2rem);row-gap:10px;}.kt-accordion-id15913_6c3e4f-03 .kt-accordion-panel-inner{border-top:0px solid transparent;border-right:1px solid transparent;border-bottom:1px solid transparent;border-left:1px solid transparent;background:#ffffff;padding-top:var(--global-kb-spacing-sm, 1.5rem);padding-right:var(--global-kb-spacing-sm, 1.5rem);padding-bottom:var(--global-kb-spacing-sm, 1.5rem);padding-left:var(--global-kb-spacing-sm, 1.5rem);}.kt-accordion-id15913_6c3e4f-03 > .kt-accordion-inner-wrap > .wp-block-kadence-pane > .kt-accordion-header-wrap > .kt-blocks-accordion-header{border-top:1px solid #eeeeee;border-right:1px solid #eeeeee;border-bottom:1px solid #eeeeee;border-left:2px solid #eeeeee;border-top-left-radius:0px;border-top-right-radius:0px;border-bottom-right-radius:0px;border-bottom-left-radius:0px;background:#ffffff;color:#444444;padding-top:14px;padding-right:16px;padding-bottom:14px;padding-left:16px;}.kt-accordion-id15913_6c3e4f-03:not( .kt-accodion-icon-style-basiccircle ):not( .kt-accodion-icon-style-xclosecircle ):not( .kt-accodion-icon-style-arrowcircle )  > .kt-accordion-inner-wrap > .wp-block-kadence-pane > .kt-accordion-header-wrap .kt-blocks-accordion-icon-trigger:after, .kt-accordion-id15913_6c3e4f-03:not( .kt-accodion-icon-style-basiccircle ):not( .kt-accodion-icon-style-xclosecircle ):not( .kt-accodion-icon-style-arrowcircle )  > .kt-accordion-inner-wrap > .wp-block-kadence-pane > .kt-accordion-header-wrap .kt-blocks-accordion-icon-trigger:before{background:#444444;}.kt-accordion-id15913_6c3e4f-03:not( .kt-accodion-icon-style-basic ):not( .kt-accodion-icon-style-xclose ):not( .kt-accodion-icon-style-arrow ) .kt-blocks-accordion-icon-trigger{background:#444444;}.kt-accordion-id15913_6c3e4f-03:not( .kt-accodion-icon-style-basic ):not( .kt-accodion-icon-style-xclose ):not( .kt-accodion-icon-style-arrow ) .kt-blocks-accordion-icon-trigger:after, .kt-accordion-id15913_6c3e4f-03:not( .kt-accodion-icon-style-basic ):not( .kt-accodion-icon-style-xclose ):not( .kt-accodion-icon-style-arrow ) .kt-blocks-accordion-icon-trigger:before{background:#ffffff;}.kt-accordion-id15913_6c3e4f-03 > .kt-accordion-inner-wrap > .wp-block-kadence-pane > .kt-accordion-header-wrap > .kt-blocks-accordion-header:hover, \n\t\t\t\tbody:not(.hide-focus-outline) .kt-accordion-id15913_6c3e4f-03 .kt-blocks-accordion-header:focus-visible{color:#444444;background:#ffffff;border-top-color:#878787;border-right-color:#878787;border-bottom-color:#878787;border-left-color:#878787;}.kt-accordion-id15913_6c3e4f-03:not( .kt-accodion-icon-style-basiccircle ):not( .kt-accodion-icon-style-xclosecircle ):not( .kt-accodion-icon-style-arrowcircle ) .kt-accordion-header-wrap .kt-blocks-accordion-header:hover .kt-blocks-accordion-icon-trigger:after, .kt-accordion-id15913_6c3e4f-03:not( .kt-accodion-icon-style-basiccircle ):not( .kt-accodion-icon-style-xclosecircle ):not( .kt-accodion-icon-style-arrowcircle ) .kt-accordion-header-wrap .kt-blocks-accordion-header:hover .kt-blocks-accordion-icon-trigger:before, body:not(.hide-focus-outline) .kt-accordion-id15913_6c3e4f-03:not( .kt-accodion-icon-style-basiccircle ):not( .kt-accodion-icon-style-xclosecircle ):not( .kt-accodion-icon-style-arrowcircle ) .kt-blocks-accordion--visible .kt-blocks-accordion-icon-trigger:after, body:not(.hide-focus-outline) .kt-accordion-id15913_6c3e4f-03:not( .kt-accodion-icon-style-basiccircle ):not( .kt-accodion-icon-style-xclosecircle ):not( .kt-accodion-icon-style-arrowcircle ) .kt-blocks-accordion-header:focus-visible .kt-blocks-accordion-icon-trigger:before{background:#444444;}.kt-accordion-id15913_6c3e4f-03:not( .kt-accodion-icon-style-basic ):not( .kt-accodion-icon-style-xclose ):not( .kt-accodion-icon-style-arrow ) .kt-accordion-header-wrap .kt-blocks-accordion-header:hover .kt-blocks-accordion-icon-trigger, body:not(.hide-focus-outline) .kt-accordion-id15913_6c3e4f-03:not( .kt-accodion-icon-style-basic ):not( .kt-accodion-icon-style-xclose ):not( .kt-accodion-icon-style-arrow ) .kt-accordion-header-wrap .kt-blocks-accordion-header:focus-visible .kt-blocks-accordion-icon-trigger{background:#444444;}.kt-accordion-id15913_6c3e4f-03:not( .kt-accodion-icon-style-basic ):not( .kt-accodion-icon-style-xclose ):not( .kt-accodion-icon-style-arrow ) .kt-accordion-header-wrap .kt-blocks-accordion-header:hover .kt-blocks-accordion-icon-trigger:after, .kt-accordion-id15913_6c3e4f-03:not( .kt-accodion-icon-style-basic ):not( .kt-accodion-icon-style-xclose ):not( .kt-accodion-icon-style-arrow ) .kt-accordion-header-wrap .kt-blocks-accordion-header:hover .kt-blocks-accordion-icon-trigger:before, body:not(.hide-focus-outline) .kt-accordion-id15913_6c3e4f-03:not( .kt-accodion-icon-style-basic ):not( .kt-accodion-icon-style-xclose ):not( .kt-accodion-icon-style-arrow ) .kt-accordion-header-wrap .kt-blocks-accordion-header:focus-visible .kt-blocks-accordion-icon-trigger:after, body:not(.hide-focus-outline) .kt-accordion-id15913_6c3e4f-03:not( .kt-accodion-icon-style-basic ):not( .kt-accodion-icon-style-xclose ):not( .kt-accodion-icon-style-arrow ) .kt-accordion-header-wrap .kt-blocks-accordion-header:focus-visible .kt-blocks-accordion-icon-trigger:before{background:#ffffff;}.kt-accordion-id15913_6c3e4f-03 .kt-accordion-header-wrap .kt-blocks-accordion-header:focus-visible,\n\t\t\t\t.kt-accordion-id15913_6c3e4f-03 > .kt-accordion-inner-wrap > .wp-block-kadence-pane > .kt-accordion-header-wrap > .kt-blocks-accordion-header.kt-accordion-panel-active{color:#444444;background:#ffffff;border-top-color:#eeeeee;border-right-color:#eeeeee;border-bottom-color:#eeeeee;border-left-color:#0e9cd1;}.kt-accordion-id15913_6c3e4f-03:not( .kt-accodion-icon-style-basiccircle ):not( .kt-accodion-icon-style-xclosecircle ):not( .kt-accodion-icon-style-arrowcircle )  > .kt-accordion-inner-wrap > .wp-block-kadence-pane > .kt-accordion-header-wrap > .kt-blocks-accordion-header.kt-accordion-panel-active .kt-blocks-accordion-icon-trigger:after, .kt-accordion-id15913_6c3e4f-03:not( .kt-accodion-icon-style-basiccircle ):not( .kt-accodion-icon-style-xclosecircle ):not( .kt-accodion-icon-style-arrowcircle )  > .kt-accordion-inner-wrap > .wp-block-kadence-pane > .kt-accordion-header-wrap > .kt-blocks-accordion-header.kt-accordion-panel-active .kt-blocks-accordion-icon-trigger:before{background:#444444;}.kt-accordion-id15913_6c3e4f-03:not( .kt-accodion-icon-style-basic ):not( .kt-accodion-icon-style-xclose ):not( .kt-accodion-icon-style-arrow ) .kt-blocks-accordion-header.kt-accordion-panel-active .kt-blocks-accordion-icon-trigger{background:#444444;}.kt-accordion-id15913_6c3e4f-03:not( .kt-accodion-icon-style-basic ):not( .kt-accodion-icon-style-xclose ):not( .kt-accodion-icon-style-arrow ) .kt-blocks-accordion-header.kt-accordion-panel-active .kt-blocks-accordion-icon-trigger:after, .kt-accordion-id15913_6c3e4f-03:not( .kt-accodion-icon-style-basic ):not( .kt-accodion-icon-style-xclose ):not( .kt-accodion-icon-style-arrow ) .kt-blocks-accordion-header.kt-accordion-panel-active .kt-blocks-accordion-icon-trigger:before{background:#ffffff;}@media all and (max-width: 1024px){.kt-accordion-id15913_6c3e4f-03 .kt-accordion-panel-inner{border-top:0px solid transparent;border-right:1px solid transparent;border-bottom:1px solid transparent;border-left:1px solid transparent;}}@media all and (max-width: 1024px){.kt-accordion-id15913_6c3e4f-03 > .kt-accordion-inner-wrap > .wp-block-kadence-pane > .kt-accordion-header-wrap > .kt-blocks-accordion-header{border-top:1px solid #eeeeee;border-right:1px solid #eeeeee;border-bottom:1px solid #eeeeee;border-left:2px solid #eeeeee;}}@media all and (max-width: 1024px){.kt-accordion-id15913_6c3e4f-03 > .kt-accordion-inner-wrap > .wp-block-kadence-pane > .kt-accordion-header-wrap > .kt-blocks-accordion-header:hover, \n\t\t\t\tbody:not(.hide-focus-outline) .kt-accordion-id15913_6c3e4f-03 .kt-blocks-accordion-header:focus-visible{border-top-color:#878787;border-right-color:#878787;border-bottom-color:#878787;border-left-color:#878787;}}@media all and (max-width: 1024px){.kt-accordion-id15913_6c3e4f-03 .kt-accordion-header-wrap .kt-blocks-accordion-header:focus-visible,\n\t\t\t\t.kt-accordion-id15913_6c3e4f-03 > .kt-accordion-inner-wrap > .wp-block-kadence-pane > .kt-accordion-header-wrap > .kt-blocks-accordion-header.kt-accordion-panel-active{border-top-color:#eeeeee;border-right-color:#eeeeee;border-bottom-color:#eeeeee;border-left-color:#0e9cd1;}}@media all and (max-width: 767px){.kt-accordion-id15913_6c3e4f-03 .kt-accordion-inner-wrap{display:block;}.kt-accordion-id15913_6c3e4f-03 .kt-accordion-inner-wrap .kt-accordion-pane:not(:first-child){margin-top:10px;}.kt-accordion-id15913_6c3e4f-03 .kt-accordion-panel-inner{border-top:0px solid transparent;border-right:1px solid transparent;border-bottom:1px solid transparent;border-left:1px solid transparent;}.kt-accordion-id15913_6c3e4f-03 > .kt-accordion-inner-wrap > .wp-block-kadence-pane > .kt-accordion-header-wrap > .kt-blocks-accordion-header{border-top:1px solid #eeeeee;border-right:1px solid #eeeeee;border-bottom:1px solid #eeeeee;border-left:2px solid #eeeeee;}.kt-accordion-id15913_6c3e4f-03 > .kt-accordion-inner-wrap > .wp-block-kadence-pane > .kt-accordion-header-wrap > .kt-blocks-accordion-header:hover, \n\t\t\t\tbody:not(.hide-focus-outline) .kt-accordion-id15913_6c3e4f-03 .kt-blocks-accordion-header:focus-visible{border-top-color:#878787;border-right-color:#878787;border-bottom-color:#878787;border-left-color:#878787;}.kt-accordion-id15913_6c3e4f-03 .kt-accordion-header-wrap .kt-blocks-accordion-header:focus-visible,\n\t\t\t\t.kt-accordion-id15913_6c3e4f-03 > .kt-accordion-inner-wrap > .wp-block-kadence-pane > .kt-accordion-header-wrap > .kt-blocks-accordion-header.kt-accordion-panel-active{border-top-color:#eeeeee;border-right-color:#eeeeee;border-bottom-color:#eeeeee;border-left-color:#0e9cd1;}}<\/style>\n<div class=\"wp-block-kadence-accordion alignnone\"><div class=\"kt-accordion-wrap kt-accordion-id15913_6c3e4f-03 kt-accordion-has-10-panes kt-active-pane-0 kt-accordion-block kt-pane-header-alignment-left kt-accodion-icon-style-arrow kt-accodion-icon-side-right\" style=\"max-width:none\"><div class=\"kt-accordion-inner-wrap\" data-allow-multiple-open=\"false\" data-start-open=\"0\">\n<div class=\"wp-block-kadence-pane kt-accordion-pane kt-accordion-pane-1 kt-pane15913_509d09-9a\"><h2 class=\"kt-accordion-header-wrap\"><button class=\"kt-blocks-accordion-header kt-acccordion-button-label-show\" type=\"button\"><span class=\"kt-blocks-accordion-title-wrap\"><span class=\"kt-blocks-accordion-title\"><strong>How can UV-Vis spectroscopy be used for quality control testing?<\/strong><\/span><\/span><span class=\"kt-blocks-accordion-icon-trigger\"><\/span><\/button><\/h2><div class=\"kt-accordion-panel kt-accordion-panel-hidden\"><div class=\"kt-accordion-panel-inner\">\n<ul class=\"wp-block-list\">\n<li>UV-Vis spectroscopy measures the absorbance of compounds that absorb ultraviolet or visible light. By comparing absorbance values to predefined acceptance criteria, laboratories can verify product concentration and identify samples that may require additional evaluation.<\/li>\n<\/ul>\n<\/div><\/div><\/div>\n\n\n\n<div class=\"wp-block-kadence-pane kt-accordion-pane kt-accordion-pane-2 kt-pane15913_c91323-8a\"><h2 class=\"kt-accordion-header-wrap\"><button class=\"kt-blocks-accordion-header kt-acccordion-button-label-show\" type=\"button\"><span class=\"kt-blocks-accordion-title-wrap\"><span class=\"kt-blocks-accordion-title\"><strong>What is a custom pass\/fail method?<\/strong><\/span><\/span><span class=\"kt-blocks-accordion-icon-trigger\"><\/span><\/button><\/h2><div class=\"kt-accordion-panel kt-accordion-panel-hidden\"><div class=\"kt-accordion-panel-inner\">\n<ul class=\"wp-block-list\">\n<li>A custom pass\/fail method is a user-defined workflow that automatically evaluates analytical results against predefined acceptance limits. The method can calculate sample concentration, assess quality criteria, and display a pass or fail result without requiring manual calculations.<\/li>\n<\/ul>\n<\/div><\/div><\/div>\n\n\n\n<div class=\"wp-block-kadence-pane kt-accordion-pane kt-accordion-pane-3 kt-pane15913_c61ca1-2f\"><h2 class=\"kt-accordion-header-wrap\"><button class=\"kt-blocks-accordion-header kt-acccordion-button-label-show\" type=\"button\"><span class=\"kt-blocks-accordion-title-wrap\"><span class=\"kt-blocks-accordion-title\"><strong><strong>How can the custom methods on a UV-Vis spectrophotometer detect product contamination?<\/strong><\/strong><\/span><\/span><span class=\"kt-blocks-accordion-icon-trigger\"><\/span><\/button><\/h2><div class=\"kt-accordion-panel kt-accordion-panel-hidden\"><div class=\"kt-accordion-panel-inner\">\n<ul class=\"wp-block-list\">\n<li>Custom methods can be designed to monitor absorbance at wavelengths associated with target compounds or potential contaminants. This approach can help identify samples that may contain unexpected materials when appropriate analytical methods have been developed.<\/li>\n<\/ul>\n<\/div><\/div><\/div>\n\n\n\n<div class=\"wp-block-kadence-pane kt-accordion-pane kt-accordion-pane-4 kt-pane15913_05dd0a-8a\"><h2 class=\"kt-accordion-header-wrap\"><button class=\"kt-blocks-accordion-header kt-acccordion-button-label-show\" type=\"button\"><span class=\"kt-blocks-accordion-title-wrap\"><span class=\"kt-blocks-accordion-title\"><strong>How does a spectrophotometer distinguish between concentration errors and contamination?<\/strong><\/span><\/span><span class=\"kt-blocks-accordion-icon-trigger\"><\/span><\/button><\/h2><div class=\"kt-accordion-panel kt-accordion-panel-hidden\"><div class=\"kt-accordion-panel-inner\">\n<ul class=\"wp-block-list\">\n<li>Concentration measurements deal with varying levels of the same analyte. Contamination is caused by the presence of an unwanted compound, which will lead to a different absorbance spectrum that can be distinguished from the preferred product. A custom method can combine multiple calculations within a single analysis and compare these two things. For example, one calculation can evaluate whether a sample falls within an acceptable concentration range, while another monitors absorbance associated with a potential contaminant.<\/li>\n<\/ul>\n<\/div><\/div><\/div>\n\n\n\n<div class=\"wp-block-kadence-pane kt-accordion-pane kt-accordion-pane-5 kt-pane15913_513acb-8c\"><h2 class=\"kt-accordion-header-wrap\"><button class=\"kt-blocks-accordion-header kt-acccordion-button-label-show\" type=\"button\"><span class=\"kt-blocks-accordion-title-wrap\"><span class=\"kt-blocks-accordion-title\"><strong><strong>Why use automated pass\/fail methods instead of manual data interpretation?<\/strong><\/strong><\/span><\/span><span class=\"kt-blocks-accordion-icon-trigger\"><\/span><\/button><\/h2><div class=\"kt-accordion-panel kt-accordion-panel-hidden\"><div class=\"kt-accordion-panel-inner\">\n<ul class=\"wp-block-list\">\n<li>Automated pass\/fail methods can standardize routine quality assessments, reduce manual calculations, and support consistent decision-making across multiple users by applying the same acceptance criteria to every sample.<\/li>\n<\/ul>\n<\/div><\/div><\/div>\n\n\n\n<div class=\"wp-block-kadence-pane kt-accordion-pane kt-accordion-pane-6 kt-pane15913_3e9934-9d\"><h2 class=\"kt-accordion-header-wrap\"><button class=\"kt-blocks-accordion-header kt-acccordion-button-label-show\" type=\"button\"><span class=\"kt-blocks-accordion-title-wrap\"><span class=\"kt-blocks-accordion-title\"><strong>Can custom methods be tailored for different products or manufacturing workflows?<\/strong><\/span><\/span><span class=\"kt-blocks-accordion-icon-trigger\"><\/span><\/button><\/h2><div class=\"kt-accordion-panel kt-accordion-panel-hidden\"><div class=\"kt-accordion-panel-inner\">\n<ul class=\"wp-block-list\">\n<li>Yes. Custom methods can be configured with application-specific calculations and acceptance limits to support routine quality testing for a variety of solution-based products and manufacturing processes.<\/li>\n<\/ul>\n<\/div><\/div><\/div>\n\n\n\n<div class=\"wp-block-kadence-pane kt-accordion-pane kt-accordion-pane-7 kt-pane15913_b9ecdd-32\"><h2 class=\"kt-accordion-header-wrap\"><button class=\"kt-blocks-accordion-header kt-acccordion-button-label-show\" type=\"button\"><span class=\"kt-blocks-accordion-title-wrap\"><span class=\"kt-blocks-accordion-title\"><strong><strong>What types of products can be evaluated using custom UV-Vis methods?<\/strong><\/strong><\/span><\/span><span class=\"kt-blocks-accordion-icon-trigger\"><\/span><\/button><\/h2><div class=\"kt-accordion-panel kt-accordion-panel-hidden\"><div class=\"kt-accordion-panel-inner\">\n<ul class=\"wp-block-list\">\n<li>Custom UV-Vis methods can be developed for many solution-based products that contain UV-Vis absorbing compounds, including food and beverage formulations, pharmaceutical solutions, chemical products, dyes, and flavor additives.<\/li>\n<\/ul>\n<\/div><\/div><\/div>\n\n\n\n<div class=\"wp-block-kadence-pane kt-accordion-pane kt-accordion-pane-8 kt-pane15913_6eb05c-27\"><h2 class=\"kt-accordion-header-wrap\"><button class=\"kt-blocks-accordion-header kt-acccordion-button-label-show\" type=\"button\"><span class=\"kt-blocks-accordion-title-wrap\"><span class=\"kt-blocks-accordion-title\"><strong>How does UV-Vis spectroscopy support routine manufacturing quality control?<\/strong><\/span><\/span><span class=\"kt-blocks-accordion-icon-trigger\"><\/span><\/button><\/h2><div class=\"kt-accordion-panel kt-accordion-panel-hidden\"><div class=\"kt-accordion-panel-inner\">\n<ul class=\"wp-block-list\">\n<li>UV-Vis spectroscopy provides a non-destructive way to measure absorbance and determine whether products meet predefined quality specifications. When combined with automated methods, it can simplify routine testing and improve consistency across operators.<\/li>\n<\/ul>\n<\/div><\/div><\/div>\n\n\n\n<div class=\"wp-block-kadence-pane kt-accordion-pane kt-accordion-pane-9 kt-pane15913_d5f7f3-f2\"><h2 class=\"kt-accordion-header-wrap\"><button class=\"kt-blocks-accordion-header kt-acccordion-button-label-show\" type=\"button\"><span class=\"kt-blocks-accordion-title-wrap\"><span class=\"kt-blocks-accordion-title\"><strong><strong>What should I consider when selecting a UV-Vis spectrophotometer for quality control testing?<\/strong><\/strong><\/span><\/span><span class=\"kt-blocks-accordion-icon-trigger\"><\/span><\/button><\/h2><div class=\"kt-accordion-panel kt-accordion-panel-hidden\"><div class=\"kt-accordion-panel-inner\">\n<ul class=\"wp-block-list\">\n<li>Consider factors such as measurement performance, ease of method development, automation capabilities, data analysis features, and the ability to standardize routine quality control procedures across users and applications.<\/li>\n<\/ul>\n<\/div><\/div><\/div>\n<\/div><\/div><\/div>\n\n\n\n<p class=\"wp-block-paragraph\"><\/p>\n","protected":false},"excerpt":{"rendered":"<p>Frequently Asked Questions<\/p>\n","protected":false},"author":38,"featured_media":15918,"comment_status":"open","ping_status":"closed","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_kad_blocks_custom_css":"","_kad_blocks_head_custom_js":"","_kad_blocks_body_custom_js":"","_kad_blocks_footer_custom_js":"","_monsterinsights_skip_tracking":false,"_genesis_hide_title":false,"_genesis_hide_breadcrumbs":false,"_genesis_hide_singular_image":false,"_genesis_hide_footer_widgets":false,"_genesis_custom_body_class":"","_genesis_custom_post_class":"","_genesis_layout":"","footnotes":""},"categories":[28,980],"tags":[1426,1427,1421,1424,1422,1429,1428,1419,1430,1423],"division":[],"class_list":["post-15913","post","type-post","status-publish","format-standard","has-post-thumbnail","category-general-news-tips-trends","category-uv-vis","tag-bca-assay","tag-bicinchoninic-acid-assay","tag-manufacturing-quality-control","tag-pass-fail-analysis","tag-product-quality-control","tag-protein-concentration","tag-protein-quantification","tag-quality-control-testing","tag-total-protein-determination","tag-uv-vis-analysis","entry"],"_selected_authors":[1690],"_selected_reviewers":[],"acf":[],"yoast_head":"<!-- This site is optimized with the Yoast SEO Premium plugin v27.8 (Yoast SEO v27.8) - https:\/\/yoast.com\/product\/yoast-seo-premium-wordpress\/ -->\n<title>Simplify Quality Control Testing with UV-Vis Custom Methods for Pass\/Fail Evaluation - Advancing Materials<\/title>\n<meta name=\"description\" content=\"Learn how UV-Vis spectroscopy and automated pass\/fail methods simplify quality control, detect contamination, and improve consistency in manufacturing testing.\" \/>\n<meta name=\"robots\" content=\"index, follow, max-snippet:-1, max-image-preview:large, max-video-preview:-1\" \/>\n<link rel=\"canonical\" href=\"https:\/\/www.thermofisher.com\/blog\/materials\/simplify-quality-control-testing-with-uv-vis-custom-methods-for-pass-fail-evaluation\/\" \/>\n<meta property=\"og:locale\" content=\"en_US\" \/>\n<meta property=\"og:type\" content=\"article\" \/>\n<meta property=\"og:title\" content=\"Simplify Quality Control Testing with UV-Vis Custom Methods for Pass\/Fail Evaluation\" \/>\n<meta property=\"og:description\" content=\"Learn how UV-Vis spectroscopy and automated pass\/fail methods simplify quality control, detect contamination, and improve consistency in manufacturing testing.\" \/>\n<meta property=\"og:url\" content=\"https:\/\/www.thermofisher.com\/blog\/materials\/simplify-quality-control-testing-with-uv-vis-custom-methods-for-pass-fail-evaluation\/\" \/>\n<meta property=\"og:site_name\" content=\"Advancing Materials\" \/>\n<meta property=\"article:publisher\" content=\"https:\/\/www.facebook.com\/thermofisher\" \/>\n<meta property=\"article:published_time\" content=\"2026-08-11T12:00:00+00:00\" \/>\n<meta property=\"article:modified_time\" content=\"2026-08-11T12:16:22+00:00\" \/>\n<meta property=\"og:image\" content=\"https:\/\/admin.acceleratingscience.com\/materials\/wp-content\/uploads\/sites\/7\/2026\/08\/AdobeStock_1070627476-scaled-e1786131077251.jpeg\" \/>\n\t<meta property=\"og:image:width\" content=\"2556\" \/>\n\t<meta property=\"og:image:height\" content=\"1075\" \/>\n\t<meta property=\"og:image:type\" content=\"image\/jpeg\" \/>\n<meta name=\"author\" content=\"Marlene Gasdia-Cochrane\" \/>\n<meta name=\"twitter:card\" content=\"summary_large_image\" \/>\n<meta name=\"twitter:creator\" content=\"@thermofisher\" \/>\n<meta name=\"twitter:site\" content=\"@thermofisher\" \/>\n<meta name=\"twitter:label1\" content=\"Written by\" \/>\n\t<meta name=\"twitter:data1\" content=\"Marlene Gasdia-Cochrane\" \/>\n\t<meta name=\"twitter:label2\" content=\"Est. reading time\" \/>\n\t<meta name=\"twitter:data2\" content=\"7 minutes\" \/>\n<script type=\"application\/ld+json\" class=\"yoast-schema-graph\">{\"@context\":\"https:\\\/\\\/schema.org\",\"@graph\":[{\"@type\":\"Article\",\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/materials\\\/simplify-quality-control-testing-with-uv-vis-custom-methods-for-pass-fail-evaluation\\\/#article\",\"isPartOf\":{\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/materials\\\/simplify-quality-control-testing-with-uv-vis-custom-methods-for-pass-fail-evaluation\\\/\"},\"author\":{\"name\":\"Marlene Gasdia-Cochrane\",\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/materials\\\/#\\\/schema\\\/person\\\/fb6befa44b6a83c9a215be75d33bb8e8\"},\"headline\":\"Simplify Quality Control Testing with UV-Vis Custom Methods for Pass\\\/Fail Evaluation\",\"datePublished\":\"2026-08-11T12:00:00+00:00\",\"dateModified\":\"2026-08-11T12:16:22+00:00\",\"mainEntityOfPage\":{\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/materials\\\/simplify-quality-control-testing-with-uv-vis-custom-methods-for-pass-fail-evaluation\\\/\"},\"wordCount\":1456,\"commentCount\":0,\"image\":{\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/materials\\\/simplify-quality-control-testing-with-uv-vis-custom-methods-for-pass-fail-evaluation\\\/#primaryimage\"},\"thumbnailUrl\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/materials\\\/wp-content\\\/uploads\\\/sites\\\/7\\\/2026\\\/08\\\/AdobeStock_1070627476-scaled-e1786131077251.jpeg\",\"keywords\":[\"BCA assay\",\"bicinchoninic acid assay\",\"manufacturing quality control\",\"pass fail analysis\",\"product quality control\",\"protein concentration\",\"protein quantification\",\"quality control testing\",\"total protein determination\",\"uv vis analysis\"],\"articleSection\":[\"General News, Tips, and Trends\",\"UV Vis\"],\"inLanguage\":\"en-US\",\"potentialAction\":[{\"@type\":\"CommentAction\",\"name\":\"Comment\",\"target\":[\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/materials\\\/simplify-quality-control-testing-with-uv-vis-custom-methods-for-pass-fail-evaluation\\\/#respond\"]}]},{\"@type\":\"WebPage\",\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/materials\\\/simplify-quality-control-testing-with-uv-vis-custom-methods-for-pass-fail-evaluation\\\/\",\"url\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/materials\\\/simplify-quality-control-testing-with-uv-vis-custom-methods-for-pass-fail-evaluation\\\/\",\"name\":\"Simplify Quality Control Testing with UV-Vis Custom Methods for Pass\\\/Fail Evaluation - Advancing Materials\",\"isPartOf\":{\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/materials\\\/#website\"},\"primaryImageOfPage\":{\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/materials\\\/simplify-quality-control-testing-with-uv-vis-custom-methods-for-pass-fail-evaluation\\\/#primaryimage\"},\"image\":{\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/materials\\\/simplify-quality-control-testing-with-uv-vis-custom-methods-for-pass-fail-evaluation\\\/#primaryimage\"},\"thumbnailUrl\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/materials\\\/wp-content\\\/uploads\\\/sites\\\/7\\\/2026\\\/08\\\/AdobeStock_1070627476-scaled-e1786131077251.jpeg\",\"datePublished\":\"2026-08-11T12:00:00+00:00\",\"dateModified\":\"2026-08-11T12:16:22+00:00\",\"author\":{\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/materials\\\/#\\\/schema\\\/person\\\/fb6befa44b6a83c9a215be75d33bb8e8\"},\"description\":\"Learn how UV-Vis spectroscopy and automated pass\\\/fail methods simplify quality control, detect contamination, and improve consistency in manufacturing testing.\",\"breadcrumb\":{\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/materials\\\/simplify-quality-control-testing-with-uv-vis-custom-methods-for-pass-fail-evaluation\\\/#breadcrumb\"},\"inLanguage\":\"en-US\",\"potentialAction\":[{\"@type\":\"ReadAction\",\"target\":[\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/materials\\\/simplify-quality-control-testing-with-uv-vis-custom-methods-for-pass-fail-evaluation\\\/\"]}]},{\"@type\":\"ImageObject\",\"inLanguage\":\"en-US\",\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/materials\\\/simplify-quality-control-testing-with-uv-vis-custom-methods-for-pass-fail-evaluation\\\/#primaryimage\",\"url\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/materials\\\/wp-content\\\/uploads\\\/sites\\\/7\\\/2026\\\/08\\\/AdobeStock_1070627476-scaled-e1786131077251.jpeg\",\"contentUrl\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/materials\\\/wp-content\\\/uploads\\\/sites\\\/7\\\/2026\\\/08\\\/AdobeStock_1070627476-scaled-e1786131077251.jpeg\",\"width\":2556,\"height\":1075,\"caption\":\"A close-up of colorful, condensation-covered beverage bottles, showcasing refreshing sports drinks in vibrant hues.\"},{\"@type\":\"BreadcrumbList\",\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/materials\\\/simplify-quality-control-testing-with-uv-vis-custom-methods-for-pass-fail-evaluation\\\/#breadcrumb\",\"itemListElement\":[{\"@type\":\"ListItem\",\"position\":1,\"name\":\"Home\",\"item\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/materials\\\/\"},{\"@type\":\"ListItem\",\"position\":2,\"name\":\"Simplify Quality Control Testing with UV-Vis Custom Methods for Pass\\\/Fail Evaluation\"}]},{\"@type\":\"WebSite\",\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/materials\\\/#website\",\"url\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/materials\\\/\",\"name\":\"Advancing Materials\",\"description\":\"Spectroscopy, spectrometry, and materials science stories and solutions about advancing research and improving product development\",\"potentialAction\":[{\"@type\":\"SearchAction\",\"target\":{\"@type\":\"EntryPoint\",\"urlTemplate\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/materials\\\/?s={search_term_string}\"},\"query-input\":{\"@type\":\"PropertyValueSpecification\",\"valueRequired\":true,\"valueName\":\"search_term_string\"}}],\"inLanguage\":\"en-US\"},{\"@type\":\"Person\",\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/materials\\\/#\\\/schema\\\/person\\\/fb6befa44b6a83c9a215be75d33bb8e8\",\"name\":\"Marlene Gasdia-Cochrane\",\"image\":{\"@type\":\"ImageObject\",\"inLanguage\":\"en-US\",\"@id\":\"https:\\\/\\\/secure.gravatar.com\\\/avatar\\\/d4ecccff16bb946462f4389b39255e116430923ec42d08f298b5633eba2d07dd?s=96&d=mm&r=g\",\"url\":\"https:\\\/\\\/secure.gravatar.com\\\/avatar\\\/d4ecccff16bb946462f4389b39255e116430923ec42d08f298b5633eba2d07dd?s=96&d=mm&r=g\",\"contentUrl\":\"https:\\\/\\\/secure.gravatar.com\\\/avatar\\\/d4ecccff16bb946462f4389b39255e116430923ec42d08f298b5633eba2d07dd?s=96&d=mm&r=g\",\"caption\":\"Marlene Gasdia-Cochrane\"},\"url\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/author\\\/mgasdiacochrane\\\/\"}]}<\/script>\n<!-- \/ Yoast SEO Premium plugin. -->","yoast_head_json":{"title":"Simplify Quality Control Testing with UV-Vis Custom Methods for Pass\/Fail Evaluation - Advancing Materials","description":"Learn how UV-Vis spectroscopy and automated pass\/fail methods simplify quality control, detect contamination, and improve consistency in manufacturing testing.","robots":{"index":"index","follow":"follow","max-snippet":"max-snippet:-1","max-image-preview":"max-image-preview:large","max-video-preview":"max-video-preview:-1"},"canonical":"https:\/\/www.thermofisher.com\/blog\/materials\/simplify-quality-control-testing-with-uv-vis-custom-methods-for-pass-fail-evaluation\/","og_locale":"en_US","og_type":"article","og_title":"Simplify Quality Control Testing with UV-Vis Custom Methods for Pass\/Fail Evaluation","og_description":"Learn how UV-Vis spectroscopy and automated pass\/fail methods simplify quality control, detect contamination, and improve consistency in manufacturing testing.","og_url":"https:\/\/www.thermofisher.com\/blog\/materials\/simplify-quality-control-testing-with-uv-vis-custom-methods-for-pass-fail-evaluation\/","og_site_name":"Advancing Materials","article_publisher":"https:\/\/www.facebook.com\/thermofisher","article_published_time":"2026-08-11T12:00:00+00:00","article_modified_time":"2026-08-11T12:16:22+00:00","og_image":[{"width":2556,"height":1075,"url":"https:\/\/admin.acceleratingscience.com\/materials\/wp-content\/uploads\/sites\/7\/2026\/08\/AdobeStock_1070627476-scaled-e1786131077251.jpeg","type":"image\/jpeg"}],"author":"Marlene Gasdia-Cochrane","twitter_card":"summary_large_image","twitter_creator":"@thermofisher","twitter_site":"@thermofisher","twitter_misc":{"Written by":"Marlene Gasdia-Cochrane","Est. reading time":"7 minutes"},"schema":{"@context":"https:\/\/schema.org","@graph":[{"@type":"Article","@id":"https:\/\/www.thermofisher.com\/blog\/materials\/simplify-quality-control-testing-with-uv-vis-custom-methods-for-pass-fail-evaluation\/#article","isPartOf":{"@id":"https:\/\/www.thermofisher.com\/blog\/materials\/simplify-quality-control-testing-with-uv-vis-custom-methods-for-pass-fail-evaluation\/"},"author":{"name":"Marlene Gasdia-Cochrane","@id":"https:\/\/admin.acceleratingscience.com\/materials\/#\/schema\/person\/fb6befa44b6a83c9a215be75d33bb8e8"},"headline":"Simplify Quality Control Testing with UV-Vis Custom Methods for Pass\/Fail Evaluation","datePublished":"2026-08-11T12:00:00+00:00","dateModified":"2026-08-11T12:16:22+00:00","mainEntityOfPage":{"@id":"https:\/\/www.thermofisher.com\/blog\/materials\/simplify-quality-control-testing-with-uv-vis-custom-methods-for-pass-fail-evaluation\/"},"wordCount":1456,"commentCount":0,"image":{"@id":"https:\/\/www.thermofisher.com\/blog\/materials\/simplify-quality-control-testing-with-uv-vis-custom-methods-for-pass-fail-evaluation\/#primaryimage"},"thumbnailUrl":"https:\/\/admin.acceleratingscience.com\/materials\/wp-content\/uploads\/sites\/7\/2026\/08\/AdobeStock_1070627476-scaled-e1786131077251.jpeg","keywords":["BCA assay","bicinchoninic acid assay","manufacturing quality control","pass fail analysis","product quality control","protein concentration","protein quantification","quality control testing","total protein determination","uv vis analysis"],"articleSection":["General News, Tips, and Trends","UV Vis"],"inLanguage":"en-US","potentialAction":[{"@type":"CommentAction","name":"Comment","target":["https:\/\/www.thermofisher.com\/blog\/materials\/simplify-quality-control-testing-with-uv-vis-custom-methods-for-pass-fail-evaluation\/#respond"]}]},{"@type":"WebPage","@id":"https:\/\/www.thermofisher.com\/blog\/materials\/simplify-quality-control-testing-with-uv-vis-custom-methods-for-pass-fail-evaluation\/","url":"https:\/\/www.thermofisher.com\/blog\/materials\/simplify-quality-control-testing-with-uv-vis-custom-methods-for-pass-fail-evaluation\/","name":"Simplify Quality Control Testing with UV-Vis Custom Methods for Pass\/Fail Evaluation - Advancing Materials","isPartOf":{"@id":"https:\/\/admin.acceleratingscience.com\/materials\/#website"},"primaryImageOfPage":{"@id":"https:\/\/www.thermofisher.com\/blog\/materials\/simplify-quality-control-testing-with-uv-vis-custom-methods-for-pass-fail-evaluation\/#primaryimage"},"image":{"@id":"https:\/\/www.thermofisher.com\/blog\/materials\/simplify-quality-control-testing-with-uv-vis-custom-methods-for-pass-fail-evaluation\/#primaryimage"},"thumbnailUrl":"https:\/\/admin.acceleratingscience.com\/materials\/wp-content\/uploads\/sites\/7\/2026\/08\/AdobeStock_1070627476-scaled-e1786131077251.jpeg","datePublished":"2026-08-11T12:00:00+00:00","dateModified":"2026-08-11T12:16:22+00:00","author":{"@id":"https:\/\/admin.acceleratingscience.com\/materials\/#\/schema\/person\/fb6befa44b6a83c9a215be75d33bb8e8"},"description":"Learn how UV-Vis spectroscopy and automated pass\/fail methods simplify quality control, detect contamination, and improve consistency in manufacturing testing.","breadcrumb":{"@id":"https:\/\/www.thermofisher.com\/blog\/materials\/simplify-quality-control-testing-with-uv-vis-custom-methods-for-pass-fail-evaluation\/#breadcrumb"},"inLanguage":"en-US","potentialAction":[{"@type":"ReadAction","target":["https:\/\/www.thermofisher.com\/blog\/materials\/simplify-quality-control-testing-with-uv-vis-custom-methods-for-pass-fail-evaluation\/"]}]},{"@type":"ImageObject","inLanguage":"en-US","@id":"https:\/\/www.thermofisher.com\/blog\/materials\/simplify-quality-control-testing-with-uv-vis-custom-methods-for-pass-fail-evaluation\/#primaryimage","url":"https:\/\/admin.acceleratingscience.com\/materials\/wp-content\/uploads\/sites\/7\/2026\/08\/AdobeStock_1070627476-scaled-e1786131077251.jpeg","contentUrl":"https:\/\/admin.acceleratingscience.com\/materials\/wp-content\/uploads\/sites\/7\/2026\/08\/AdobeStock_1070627476-scaled-e1786131077251.jpeg","width":2556,"height":1075,"caption":"A close-up of colorful, condensation-covered beverage bottles, showcasing refreshing sports drinks in vibrant hues."},{"@type":"BreadcrumbList","@id":"https:\/\/www.thermofisher.com\/blog\/materials\/simplify-quality-control-testing-with-uv-vis-custom-methods-for-pass-fail-evaluation\/#breadcrumb","itemListElement":[{"@type":"ListItem","position":1,"name":"Home","item":"https:\/\/admin.acceleratingscience.com\/materials\/"},{"@type":"ListItem","position":2,"name":"Simplify Quality Control Testing with UV-Vis Custom Methods for Pass\/Fail Evaluation"}]},{"@type":"WebSite","@id":"https:\/\/admin.acceleratingscience.com\/materials\/#website","url":"https:\/\/admin.acceleratingscience.com\/materials\/","name":"Advancing Materials","description":"Spectroscopy, spectrometry, and materials science stories and solutions about advancing research and improving product development","potentialAction":[{"@type":"SearchAction","target":{"@type":"EntryPoint","urlTemplate":"https:\/\/admin.acceleratingscience.com\/materials\/?s={search_term_string}"},"query-input":{"@type":"PropertyValueSpecification","valueRequired":true,"valueName":"search_term_string"}}],"inLanguage":"en-US"},{"@type":"Person","@id":"https:\/\/admin.acceleratingscience.com\/materials\/#\/schema\/person\/fb6befa44b6a83c9a215be75d33bb8e8","name":"Marlene Gasdia-Cochrane","image":{"@type":"ImageObject","inLanguage":"en-US","@id":"https:\/\/secure.gravatar.com\/avatar\/d4ecccff16bb946462f4389b39255e116430923ec42d08f298b5633eba2d07dd?s=96&d=mm&r=g","url":"https:\/\/secure.gravatar.com\/avatar\/d4ecccff16bb946462f4389b39255e116430923ec42d08f298b5633eba2d07dd?s=96&d=mm&r=g","contentUrl":"https:\/\/secure.gravatar.com\/avatar\/d4ecccff16bb946462f4389b39255e116430923ec42d08f298b5633eba2d07dd?s=96&d=mm&r=g","caption":"Marlene Gasdia-Cochrane"},"url":"https:\/\/admin.acceleratingscience.com\/author\/mgasdiacochrane\/"}]}},"taxonomy_info":{"category":[{"value":28,"label":"General News, Tips, and Trends"},{"value":980,"label":"UV Vis"}],"post_tag":[{"value":1426,"label":"BCA assay"},{"value":1427,"label":"bicinchoninic acid assay"},{"value":1421,"label":"manufacturing quality control"},{"value":1424,"label":"pass fail analysis"},{"value":1422,"label":"product quality control"},{"value":1429,"label":"protein concentration"},{"value":1428,"label":"protein quantification"},{"value":1419,"label":"quality control testing"},{"value":1430,"label":"total protein determination"},{"value":1423,"label":"uv vis analysis"}]},"featured_image_src_large":["https:\/\/admin.acceleratingscience.com\/materials\/wp-content\/uploads\/sites\/7\/2026\/08\/AdobeStock_1070627476-scaled-e1786131077251-1024x431.jpeg",760,320,true],"author_info":{"display_name":"Marlene Gasdia-Cochrane","author_link":"https:\/\/admin.acceleratingscience.com\/author\/mgasdiacochrane\/"},"comment_info":3,"category_info":[{"term_id":28,"name":"General News, Tips, and Trends","slug":"general-news-tips-trends","term_group":0,"term_taxonomy_id":28,"taxonomy":"category","description":"","parent":0,"count":72,"filter":"raw","meta":[],"cat_ID":28,"category_count":72,"category_description":"","cat_name":"General News, Tips, and Trends","category_nicename":"general-news-tips-trends","category_parent":0},{"term_id":980,"name":"UV Vis","slug":"uv-vis","term_group":0,"term_taxonomy_id":980,"taxonomy":"category","description":"","parent":968,"count":9,"filter":"raw","meta":[],"cat_ID":980,"category_count":9,"category_description":"","cat_name":"UV Vis","category_nicename":"uv-vis","category_parent":968}],"tag_info":[{"term_id":1426,"name":"BCA assay","slug":"bca-assay","term_group":0,"term_taxonomy_id":1426,"taxonomy":"post_tag","description":"","parent":0,"count":1,"filter":"raw","meta":[]},{"term_id":1427,"name":"bicinchoninic acid assay","slug":"bicinchoninic-acid-assay","term_group":0,"term_taxonomy_id":1427,"taxonomy":"post_tag","description":"","parent":0,"count":1,"filter":"raw","meta":[]},{"term_id":1421,"name":"manufacturing quality control","slug":"manufacturing-quality-control","term_group":0,"term_taxonomy_id":1421,"taxonomy":"post_tag","description":"","parent":0,"count":1,"filter":"raw","meta":[]},{"term_id":1424,"name":"pass fail analysis","slug":"pass-fail-analysis","term_group":0,"term_taxonomy_id":1424,"taxonomy":"post_tag","description":"","parent":0,"count":1,"filter":"raw","meta":[]},{"term_id":1422,"name":"product quality control","slug":"product-quality-control","term_group":0,"term_taxonomy_id":1422,"taxonomy":"post_tag","description":"","parent":0,"count":1,"filter":"raw","meta":[]},{"term_id":1429,"name":"protein concentration","slug":"protein-concentration","term_group":0,"term_taxonomy_id":1429,"taxonomy":"post_tag","description":"","parent":0,"count":1,"filter":"raw","meta":[]},{"term_id":1428,"name":"protein quantification","slug":"protein-quantification","term_group":0,"term_taxonomy_id":1428,"taxonomy":"post_tag","description":"","parent":0,"count":1,"filter":"raw","meta":[]},{"term_id":1419,"name":"quality control testing","slug":"quality-control-testing","term_group":0,"term_taxonomy_id":1419,"taxonomy":"post_tag","description":"","parent":0,"count":1,"filter":"raw","meta":[]},{"term_id":1430,"name":"total protein determination","slug":"total-protein-determination","term_group":0,"term_taxonomy_id":1430,"taxonomy":"post_tag","description":"","parent":0,"count":1,"filter":"raw","meta":[]},{"term_id":1423,"name":"uv vis analysis","slug":"uv-vis-analysis","term_group":0,"term_taxonomy_id":1423,"taxonomy":"post_tag","description":"","parent":0,"count":1,"filter":"raw","meta":[]}],"_links":{"self":[{"href":"https:\/\/www.thermofisher.com\/blog\/materials\/wp-json\/wp\/v2\/posts\/15913","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/www.thermofisher.com\/blog\/materials\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/www.thermofisher.com\/blog\/materials\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/materials\/wp-json\/wp\/v2\/users\/38"}],"replies":[{"embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/materials\/wp-json\/wp\/v2\/comments?post=15913"}],"version-history":[{"count":0,"href":"https:\/\/www.thermofisher.com\/blog\/materials\/wp-json\/wp\/v2\/posts\/15913\/revisions"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/materials\/wp-json\/wp\/v2\/media\/15918"}],"wp:attachment":[{"href":"https:\/\/www.thermofisher.com\/blog\/materials\/wp-json\/wp\/v2\/media?parent=15913"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/materials\/wp-json\/wp\/v2\/categories?post=15913"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/materials\/wp-json\/wp\/v2\/tags?post=15913"},{"taxonomy":"division","embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/materials\/wp-json\/wp\/v2\/division?post=15913"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}