{"id":1258,"date":"2013-09-12T10:13:43","date_gmt":"2013-09-12T14:13:43","guid":{"rendered":"http:\/\/admin.acceleratingscience.com\/?p=1258"},"modified":"2026-01-06T15:11:37","modified_gmt":"2026-01-06T15:11:37","slug":"comparing-sample-preparation-methods","status":"publish","type":"post","link":"https:\/\/www.thermofisher.com\/blog\/mining\/comparing-sample-preparation-methods\/","title":{"rendered":"Comparing Sample Preparation Methods"},"content":{"rendered":"<p><div style=\"width: 290px\" class=\"wp-caption alignright\"><img loading=\"lazy\" decoding=\"async\" title=\"Prepping Mining Samples Improves XRF Results\" alt=\"Prepping Mining Samples Improves XRF Results\" src=\"http:\/\/admin.acceleratingscience.com\/wp-content\/uploads\/2013\/09\/p1020763.jpg\" width=\"280\" height=\"210\" \/><p class=\"wp-caption-text\">Prepping Mining Samples Improves XRF Results<\/p><\/div>  In my last <a href=\"http:\/\/admin.acceleratingscience.com\/mining\/importance-sample-preparation-mining\/\">article<\/a>, I showed how adequate sample preparation is recommended to significantly improve results when using portable XRF in the mining field \u2013 especially for the XRF analysis of elements lighter than calcium.  In this article, I\u2019ll be discussing two different types of equipment and methods for preparing your samples: a direct rock sampler and an electric hammer mill.  The direct rock sampler allows the collection of fine powders from a rock face or even drill core. The device can be attached to either an AC (110\/220V) or DC (18-24V) powered commercial angle grinder. The special diamond cutting blades easily generate a sample by cutting slots, or even making a line cut, on the rock face or drill cores. The powder is collected in a sample tube. This material can then be cupped or pelletized for XRF analysis. This type of sample collection and preparation can result in a spatially resolved, quantitative rock-face analysis in minutes, in the field.  Take a look at this <a href=\"http:\/\/www.youtube.com\/watch?v=hUKkpd2qEoc&amp;feature=share&amp;list=PLbTZpxgfR_tx0Vuu-BYaYG5IbeFjrt5jC\" target=\"_blank\">short direct rock sampler video<\/a> to see a demonstration of how to properly collect samples using this tool. To obtain best assays from drill cores, this method of sample preparation is strongly recommended.  The electric hammer mill is a powerful device, weighing less than 5 kg, but provides rapid milling of samples and delivers a homogenized, fine powder in just seconds. This mill operates at 150W at 12VDC, so you can use it for hours and run it off a conventional automobile battery. First you need to chip pieces off the rock. Rock chips must be less than 8 mm to feed the mill, so you should use a \u201ccrusher tool\u201d to reduce the size of larger rocks. When you make your favorite smoothie, you don\u2019t drop in giant hunks of ice; the large ice pieces could jam the blades and your drink would be called a chunky instead of a smoothie. It\u2019s the same way with milling your samples. Similar to the direct rock sampler method, the fine powder that results from the milling can also be cupped or pelletized.  Here\u2019s a video demonstration on how to <a href=\"http:\/\/www.youtube.com\/watch?v=UKnkKp3iAdo&amp;feature=share&amp;list=PLbTZpxgfR_tx0Vuu-BYaYG5IbeFjrt5jC\" target=\"_blank\">collect samples using an electric hammer mill<\/a>.  Both of these methods convert the sample into a fine powder so it can then become a pellet. Just like when children fill a pail with beach sand and then flip it over to become a cylindrical castle, you can use a press &#8212; just like a pail. Just a few things to keep in mind&#8230; The press should be on a firm surface. When compacting the powder, tap gently on the sides to ensure the powder slides to the bottom of the press, then gently tap on the top until the powder doesn&#8217;t compact anymore. Give the press one solid firm hit on the top to form the pellet. That\u2019s ONE solid hit; if you hit it several times you will destroy the pellet and have to start all over again. Remember that you are producing solid powder tablets without binder so you have to be firm but gentle. How many times did you see that beach sand castle slide into a heap because the sand wasn&#8217;t compacted enough in the pail?  To see a side-by-side comparison of steps, see the diagram page of<span>\u00a0the\u00a0<\/span><a href=\"http:\/\/www.niton.com\/docs\/document-library\/spec-sheet_samplepreptools_mining_final_2012feb14-(2).pdf?sfvrsn=2\">mining sample prep tools specification sheet<\/a><span>.\u00a0<\/span>  <span><span>Both of these methods make it easy to set up a full laboratory in the field and help ensure you get the most accurate results as possible. \u00a0<\/span><\/span>  &nbsp;<\/p>\n","protected":false},"excerpt":{"rendered":"<p>In my last article, I showed how adequate sample preparation is recommended to significantly improve results when using portable XRF in the mining field \u2013 especially for the XRF analysis of elements lighter than calcium. In this article, I\u2019ll be discussing two different types of equipment and methods for preparing your samples: a direct rock<\/p>\n","protected":false},"author":41,"featured_media":1257,"comment_status":"open","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_kad_blocks_custom_css":"","_kad_blocks_head_custom_js":"","_kad_blocks_body_custom_js":"","_kad_blocks_footer_custom_js":"","_monsterinsights_skip_tracking":false,"_genesis_hide_title":false,"_genesis_hide_breadcrumbs":false,"_genesis_hide_singular_image":false,"_genesis_hide_footer_widgets":false,"_genesis_custom_body_class":"","_genesis_custom_post_class":"","_genesis_layout":"","_jetpack_newsletter_access":"","_jetpack_dont_email_post_to_subs":false,"_jetpack_newsletter_tier_id":0,"_jetpack_memberships_contains_paywalled_content":false,"_jetpack_memberships_contains_paid_content":false,"footnotes":""},"categories":[6,11],"tags":[71,16,182,181,76,23],"division":[],"class_list":{"0":"post-1258","1":"post","2":"type-post","3":"status-publish","4":"format-standard","5":"has-post-thumbnail","7":"category-elemental-geochemical-analysis","8":"category-portable-xrf","9":"tag-light-elements","10":"tag-mining-2","11":"tag-rock-sampler","12":"tag-sample-preparation-mining","13":"tag-x-ray-fluorescence","14":"tag-xrf","15":"entry"},"_selected_authors":[1199],"_selected_reviewers":[],"acf":[],"yoast_head":"<!-- This site is optimized with the Yoast SEO Premium plugin v27.8 (Yoast SEO v27.8) - https:\/\/yoast.com\/product\/yoast-seo-premium-wordpress\/ -->\n<title>Comparing Sample Preparation Methods<\/title>\n<meta name=\"description\" content=\"Adequate sample preparation is recommended to improve results with portable XRF in the mining field \u2013 especially for elements lighter than calcium.\" \/>\n<meta name=\"robots\" content=\"index, follow, max-snippet:-1, max-image-preview:large, max-video-preview:-1\" \/>\n<link rel=\"canonical\" href=\"https:\/\/www.thermofisher.com\/blog\/mining\/comparing-sample-preparation-methods\/\" \/>\n<meta property=\"og:locale\" content=\"en_US\" \/>\n<meta property=\"og:type\" content=\"article\" \/>\n<meta property=\"og:title\" content=\"Comparing Sample Preparation Methods\" \/>\n<meta property=\"og:description\" content=\"Adequate sample preparation is recommended to improve results with portable XRF in the mining field \u2013 especially for elements lighter than calcium.\" \/>\n<meta property=\"og:url\" content=\"https:\/\/www.thermofisher.com\/blog\/mining\/comparing-sample-preparation-methods\/\" \/>\n<meta property=\"og:site_name\" content=\"Advancing Mining\" \/>\n<meta property=\"article:publisher\" content=\"https:\/\/www.facebook.com\/thermofisher\" \/>\n<meta property=\"article:published_time\" content=\"2013-09-12T14:13:43+00:00\" \/>\n<meta property=\"article:modified_time\" content=\"2026-01-06T15:11:37+00:00\" \/>\n<meta property=\"og:image\" content=\"http:\/\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2013\/09\/p1020763-1024x768.jpg\" \/>\n\t<meta property=\"og:image:width\" content=\"1024\" \/>\n\t<meta property=\"og:image:height\" content=\"768\" \/>\n\t<meta property=\"og:image:type\" content=\"image\/jpeg\" \/>\n<meta name=\"author\" content=\"Ali Somarin\" \/>\n<meta name=\"twitter:card\" content=\"summary_large_image\" \/>\n<meta name=\"twitter:creator\" content=\"@thermofisher\" \/>\n<meta name=\"twitter:site\" content=\"@thermofisher\" \/>\n<meta name=\"twitter:label1\" content=\"Written by\" \/>\n\t<meta name=\"twitter:data1\" content=\"Ali Somarin\" \/>\n\t<meta name=\"twitter:label2\" content=\"Est. reading time\" \/>\n\t<meta name=\"twitter:data2\" content=\"3 minutes\" \/>\n<script type=\"application\/ld+json\" class=\"yoast-schema-graph\">{\"@context\":\"https:\\\/\\\/schema.org\",\"@graph\":[{\"@type\":\"Article\",\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/mining\\\/comparing-sample-preparation-methods\\\/#article\",\"isPartOf\":{\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/mining\\\/comparing-sample-preparation-methods\\\/\"},\"author\":{\"name\":\"Ali Somarin\",\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/mining\\\/#\\\/schema\\\/person\\\/221f9fc8c2093ee2fe5c4df3861fe565\"},\"headline\":\"Comparing Sample Preparation Methods\",\"datePublished\":\"2013-09-12T14:13:43+00:00\",\"dateModified\":\"2026-01-06T15:11:37+00:00\",\"mainEntityOfPage\":{\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/mining\\\/comparing-sample-preparation-methods\\\/\"},\"wordCount\":602,\"commentCount\":0,\"image\":{\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/mining\\\/comparing-sample-preparation-methods\\\/#primaryimage\"},\"thumbnailUrl\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/mining\\\/wp-content\\\/uploads\\\/sites\\\/3\\\/2013\\\/09\\\/p1020763.jpg\",\"keywords\":[\"light elements\",\"mining\",\"rock sampler\",\"sample preparation mining\",\"x-ray fluorescence\",\"xrf\"],\"articleSection\":[\"Elemental\\\/Geochemical Analysis\",\"Portable XRF\"],\"inLanguage\":\"en-US\",\"potentialAction\":[{\"@type\":\"CommentAction\",\"name\":\"Comment\",\"target\":[\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/mining\\\/comparing-sample-preparation-methods\\\/#respond\"]}]},{\"@type\":\"WebPage\",\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/mining\\\/comparing-sample-preparation-methods\\\/\",\"url\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/mining\\\/comparing-sample-preparation-methods\\\/\",\"name\":\"Comparing Sample Preparation Methods\",\"isPartOf\":{\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/mining\\\/#website\"},\"primaryImageOfPage\":{\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/mining\\\/comparing-sample-preparation-methods\\\/#primaryimage\"},\"image\":{\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/mining\\\/comparing-sample-preparation-methods\\\/#primaryimage\"},\"thumbnailUrl\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/mining\\\/wp-content\\\/uploads\\\/sites\\\/3\\\/2013\\\/09\\\/p1020763.jpg\",\"datePublished\":\"2013-09-12T14:13:43+00:00\",\"dateModified\":\"2026-01-06T15:11:37+00:00\",\"author\":{\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/mining\\\/#\\\/schema\\\/person\\\/221f9fc8c2093ee2fe5c4df3861fe565\"},\"description\":\"Adequate sample preparation is recommended to improve results with portable XRF in the mining field \u2013 especially for elements lighter than calcium.\",\"breadcrumb\":{\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/mining\\\/comparing-sample-preparation-methods\\\/#breadcrumb\"},\"inLanguage\":\"en-US\",\"potentialAction\":[{\"@type\":\"ReadAction\",\"target\":[\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/mining\\\/comparing-sample-preparation-methods\\\/\"]}]},{\"@type\":\"ImageObject\",\"inLanguage\":\"en-US\",\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/mining\\\/comparing-sample-preparation-methods\\\/#primaryimage\",\"url\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/mining\\\/wp-content\\\/uploads\\\/sites\\\/3\\\/2013\\\/09\\\/p1020763.jpg\",\"contentUrl\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/mining\\\/wp-content\\\/uploads\\\/sites\\\/3\\\/2013\\\/09\\\/p1020763.jpg\",\"width\":4000,\"height\":3000},{\"@type\":\"BreadcrumbList\",\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/mining\\\/comparing-sample-preparation-methods\\\/#breadcrumb\",\"itemListElement\":[{\"@type\":\"ListItem\",\"position\":1,\"name\":\"Home\",\"item\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/mining\\\/\"},{\"@type\":\"ListItem\",\"position\":2,\"name\":\"Comparing Sample Preparation Methods\"}]},{\"@type\":\"WebSite\",\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/mining\\\/#website\",\"url\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/mining\\\/\",\"name\":\"Advancing Mining\",\"description\":\"\",\"potentialAction\":[{\"@type\":\"SearchAction\",\"target\":{\"@type\":\"EntryPoint\",\"urlTemplate\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/mining\\\/?s={search_term_string}\"},\"query-input\":{\"@type\":\"PropertyValueSpecification\",\"valueRequired\":true,\"valueName\":\"search_term_string\"}}],\"inLanguage\":\"en-US\"},{\"@type\":\"Person\",\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/mining\\\/#\\\/schema\\\/person\\\/221f9fc8c2093ee2fe5c4df3861fe565\",\"name\":\"Ali Somarin\",\"image\":{\"@type\":\"ImageObject\",\"inLanguage\":\"en-US\",\"@id\":\"https:\\\/\\\/secure.gravatar.com\\\/avatar\\\/17fb0b34b2d8eeab22a9d52c1513fb313468707e8095c3538834fa1405bd7323?s=96&d=mm&r=g\",\"url\":\"https:\\\/\\\/secure.gravatar.com\\\/avatar\\\/17fb0b34b2d8eeab22a9d52c1513fb313468707e8095c3538834fa1405bd7323?s=96&d=mm&r=g\",\"contentUrl\":\"https:\\\/\\\/secure.gravatar.com\\\/avatar\\\/17fb0b34b2d8eeab22a9d52c1513fb313468707e8095c3538834fa1405bd7323?s=96&d=mm&r=g\",\"caption\":\"Ali Somarin\"},\"url\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/author\\\/alireza-somarinthermofisher-com\\\/\"}]}<\/script>\n<!-- \/ Yoast SEO Premium plugin. -->","yoast_head_json":{"title":"Comparing Sample Preparation Methods","description":"Adequate sample preparation is recommended to improve results with portable XRF in the mining field \u2013 especially for elements lighter than calcium.","robots":{"index":"index","follow":"follow","max-snippet":"max-snippet:-1","max-image-preview":"max-image-preview:large","max-video-preview":"max-video-preview:-1"},"canonical":"https:\/\/www.thermofisher.com\/blog\/mining\/comparing-sample-preparation-methods\/","og_locale":"en_US","og_type":"article","og_title":"Comparing Sample Preparation Methods","og_description":"Adequate sample preparation is recommended to improve results with portable XRF in the mining field \u2013 especially for elements lighter than calcium.","og_url":"https:\/\/www.thermofisher.com\/blog\/mining\/comparing-sample-preparation-methods\/","og_site_name":"Advancing Mining","article_publisher":"https:\/\/www.facebook.com\/thermofisher","article_published_time":"2013-09-12T14:13:43+00:00","article_modified_time":"2026-01-06T15:11:37+00:00","og_image":[{"width":1024,"height":768,"url":"http:\/\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2013\/09\/p1020763-1024x768.jpg","type":"image\/jpeg"}],"author":"Ali Somarin","twitter_card":"summary_large_image","twitter_creator":"@thermofisher","twitter_site":"@thermofisher","twitter_misc":{"Written by":"Ali Somarin","Est. reading time":"3 minutes"},"schema":{"@context":"https:\/\/schema.org","@graph":[{"@type":"Article","@id":"https:\/\/www.thermofisher.com\/blog\/mining\/comparing-sample-preparation-methods\/#article","isPartOf":{"@id":"https:\/\/www.thermofisher.com\/blog\/mining\/comparing-sample-preparation-methods\/"},"author":{"name":"Ali Somarin","@id":"https:\/\/admin.acceleratingscience.com\/mining\/#\/schema\/person\/221f9fc8c2093ee2fe5c4df3861fe565"},"headline":"Comparing Sample Preparation Methods","datePublished":"2013-09-12T14:13:43+00:00","dateModified":"2026-01-06T15:11:37+00:00","mainEntityOfPage":{"@id":"https:\/\/www.thermofisher.com\/blog\/mining\/comparing-sample-preparation-methods\/"},"wordCount":602,"commentCount":0,"image":{"@id":"https:\/\/www.thermofisher.com\/blog\/mining\/comparing-sample-preparation-methods\/#primaryimage"},"thumbnailUrl":"https:\/\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2013\/09\/p1020763.jpg","keywords":["light elements","mining","rock sampler","sample preparation mining","x-ray fluorescence","xrf"],"articleSection":["Elemental\/Geochemical Analysis","Portable XRF"],"inLanguage":"en-US","potentialAction":[{"@type":"CommentAction","name":"Comment","target":["https:\/\/www.thermofisher.com\/blog\/mining\/comparing-sample-preparation-methods\/#respond"]}]},{"@type":"WebPage","@id":"https:\/\/www.thermofisher.com\/blog\/mining\/comparing-sample-preparation-methods\/","url":"https:\/\/www.thermofisher.com\/blog\/mining\/comparing-sample-preparation-methods\/","name":"Comparing Sample Preparation Methods","isPartOf":{"@id":"https:\/\/admin.acceleratingscience.com\/mining\/#website"},"primaryImageOfPage":{"@id":"https:\/\/www.thermofisher.com\/blog\/mining\/comparing-sample-preparation-methods\/#primaryimage"},"image":{"@id":"https:\/\/www.thermofisher.com\/blog\/mining\/comparing-sample-preparation-methods\/#primaryimage"},"thumbnailUrl":"https:\/\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2013\/09\/p1020763.jpg","datePublished":"2013-09-12T14:13:43+00:00","dateModified":"2026-01-06T15:11:37+00:00","author":{"@id":"https:\/\/admin.acceleratingscience.com\/mining\/#\/schema\/person\/221f9fc8c2093ee2fe5c4df3861fe565"},"description":"Adequate sample preparation is recommended to improve results with portable XRF in the mining field \u2013 especially for elements lighter than calcium.","breadcrumb":{"@id":"https:\/\/www.thermofisher.com\/blog\/mining\/comparing-sample-preparation-methods\/#breadcrumb"},"inLanguage":"en-US","potentialAction":[{"@type":"ReadAction","target":["https:\/\/www.thermofisher.com\/blog\/mining\/comparing-sample-preparation-methods\/"]}]},{"@type":"ImageObject","inLanguage":"en-US","@id":"https:\/\/www.thermofisher.com\/blog\/mining\/comparing-sample-preparation-methods\/#primaryimage","url":"https:\/\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2013\/09\/p1020763.jpg","contentUrl":"https:\/\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2013\/09\/p1020763.jpg","width":4000,"height":3000},{"@type":"BreadcrumbList","@id":"https:\/\/www.thermofisher.com\/blog\/mining\/comparing-sample-preparation-methods\/#breadcrumb","itemListElement":[{"@type":"ListItem","position":1,"name":"Home","item":"https:\/\/admin.acceleratingscience.com\/mining\/"},{"@type":"ListItem","position":2,"name":"Comparing Sample Preparation Methods"}]},{"@type":"WebSite","@id":"https:\/\/admin.acceleratingscience.com\/mining\/#website","url":"https:\/\/admin.acceleratingscience.com\/mining\/","name":"Advancing Mining","description":"","potentialAction":[{"@type":"SearchAction","target":{"@type":"EntryPoint","urlTemplate":"https:\/\/admin.acceleratingscience.com\/mining\/?s={search_term_string}"},"query-input":{"@type":"PropertyValueSpecification","valueRequired":true,"valueName":"search_term_string"}}],"inLanguage":"en-US"},{"@type":"Person","@id":"https:\/\/admin.acceleratingscience.com\/mining\/#\/schema\/person\/221f9fc8c2093ee2fe5c4df3861fe565","name":"Ali Somarin","image":{"@type":"ImageObject","inLanguage":"en-US","@id":"https:\/\/secure.gravatar.com\/avatar\/17fb0b34b2d8eeab22a9d52c1513fb313468707e8095c3538834fa1405bd7323?s=96&d=mm&r=g","url":"https:\/\/secure.gravatar.com\/avatar\/17fb0b34b2d8eeab22a9d52c1513fb313468707e8095c3538834fa1405bd7323?s=96&d=mm&r=g","contentUrl":"https:\/\/secure.gravatar.com\/avatar\/17fb0b34b2d8eeab22a9d52c1513fb313468707e8095c3538834fa1405bd7323?s=96&d=mm&r=g","caption":"Ali Somarin"},"url":"https:\/\/admin.acceleratingscience.com\/author\/alireza-somarinthermofisher-com\/"}]}},"taxonomy_info":{"category":[{"value":6,"label":"Elemental\/Geochemical Analysis"},{"value":11,"label":"Portable XRF"}],"post_tag":[{"value":71,"label":"light elements"},{"value":16,"label":"mining"},{"value":182,"label":"rock sampler"},{"value":181,"label":"sample preparation mining"},{"value":76,"label":"x-ray fluorescence"},{"value":23,"label":"xrf"}]},"featured_image_src_large":["https:\/\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2013\/09\/p1020763-1024x768.jpg",760,570,true],"author_info":{"display_name":"Ali Somarin","author_link":"https:\/\/admin.acceleratingscience.com\/author\/alireza-somarinthermofisher-com\/"},"comment_info":0,"category_info":[{"term_id":6,"name":"Elemental\/Geochemical Analysis","slug":"elemental-geochemical-analysis","term_group":0,"term_taxonomy_id":6,"taxonomy":"category","description":"","parent":0,"count":144,"filter":"raw","meta":[],"cat_ID":6,"category_count":144,"category_description":"","cat_name":"Elemental\/Geochemical Analysis","category_nicename":"elemental-geochemical-analysis","category_parent":0},{"term_id":11,"name":"Portable XRF","slug":"portable-xrf","term_group":0,"term_taxonomy_id":11,"taxonomy":"category","description":"","parent":0,"count":76,"filter":"raw","meta":[],"cat_ID":11,"category_count":76,"category_description":"","cat_name":"Portable XRF","category_nicename":"portable-xrf","category_parent":0}],"tag_info":[{"term_id":71,"name":"light elements","slug":"light-elements","term_group":0,"term_taxonomy_id":71,"taxonomy":"post_tag","description":"","parent":0,"count":2,"filter":"raw","meta":[]},{"term_id":16,"name":"mining","slug":"mining-2","term_group":0,"term_taxonomy_id":16,"taxonomy":"post_tag","description":"","parent":0,"count":111,"filter":"raw","meta":[]},{"term_id":182,"name":"rock sampler","slug":"rock-sampler","term_group":0,"term_taxonomy_id":182,"taxonomy":"post_tag","description":"","parent":0,"count":1,"filter":"raw","meta":[]},{"term_id":181,"name":"sample preparation mining","slug":"sample-preparation-mining","term_group":0,"term_taxonomy_id":181,"taxonomy":"post_tag","description":"","parent":0,"count":2,"filter":"raw","meta":[]},{"term_id":76,"name":"x-ray fluorescence","slug":"x-ray-fluorescence","term_group":0,"term_taxonomy_id":76,"taxonomy":"post_tag","description":"","parent":0,"count":8,"filter":"raw","meta":[]},{"term_id":23,"name":"xrf","slug":"xrf","term_group":0,"term_taxonomy_id":23,"taxonomy":"post_tag","description":"","parent":0,"count":46,"filter":"raw","meta":[]}],"jetpack-related-posts":[{"id":13986,"url":"https:\/\/www.thermofisher.com\/blog\/mining\/best-practices-for-accurate-in-situ-geochemical-analysis\/","url_meta":{"origin":1258,"position":0},"title":"Best Practices for Accurate In Situ Geochemical Analysis","author":"Marlene Gasdia-Cochrane","date":"March 24, 2026","format":false,"excerpt":"In situ geochemical analysis refers to elemental testing performed directly on geological materials in their natural location, avoiding the delays and costs associated with laboratory testing.\u00a0 While in situ geochemical analysis delivers immediate elemental data, the reliability of those results depends heavily on proper field methodology. Portable analytical technologies used\u2026","rel":"","context":"In &quot;Exploration&quot;","block_context":{"text":"Exploration","link":"https:\/\/admin.acceleratingscience.com\/mining\/exploration\/"},"img":{"alt_text":"Mining truck on a dirt road in the Andes, Chile (near Embalse El Yeso, Caj\u00f3n del Maipo, San Jos\u00e9 de Maipo)","src":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2026\/03\/iStock-2222354034.jpg?resize=350%2C200&ssl=1","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2026\/03\/iStock-2222354034.jpg?resize=350%2C200&ssl=1 1x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2026\/03\/iStock-2222354034.jpg?resize=525%2C300&ssl=1 1.5x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2026\/03\/iStock-2222354034.jpg?resize=700%2C400&ssl=1 2x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2026\/03\/iStock-2222354034.jpg?resize=1050%2C600&ssl=1 3x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2026\/03\/iStock-2222354034.jpg?resize=1400%2C800&ssl=1 4x"},"classes":[]},{"id":1624,"url":"https:\/\/www.thermofisher.com\/blog\/mining\/a-halloween-treat-free-periodic-table\/","url_meta":{"origin":1258,"position":1},"title":"A Halloween Treat &#8211; Free Periodic Table","author":"Marlene Gasdia-Cochrane","date":"October 30, 2025","format":false,"excerpt":"No tricks here. Download this handy periodic table and X-ray energy reference guide to help in analyzing your mining samples. \u00a0It's our treat. Then visit our Mining and Minerals, Cement Production, and Coal Power pages for more goodies. Download Periodic Table.\u00a0 \u00a0","rel":"","context":"In &quot;Elemental\/Geochemical Analysis&quot;","block_context":{"text":"Elemental\/Geochemical Analysis","link":"https:\/\/admin.acceleratingscience.com\/mining\/elemental-geochemical-analysis\/"},"img":{"alt_text":"halloween scene with plastic pumpkin and candy","src":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2025\/10\/No-Tricks-.jpg?resize=350%2C200&ssl=1","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2025\/10\/No-Tricks-.jpg?resize=350%2C200&ssl=1 1x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2025\/10\/No-Tricks-.jpg?resize=525%2C300&ssl=1 1.5x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2025\/10\/No-Tricks-.jpg?resize=700%2C400&ssl=1 2x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2025\/10\/No-Tricks-.jpg?resize=1050%2C600&ssl=1 3x"},"classes":[]},{"id":13977,"url":"https:\/\/www.thermofisher.com\/blog\/mining\/what-is-in-situ-geochemical-analysis-and-why-it-is-critical-for-modern-field-operations\/","url_meta":{"origin":1258,"position":2},"title":"What Is In Situ Geochemical Analysis \u2014 And Why It Is Critical for Modern Field Operations","author":"Marlene Gasdia-Cochrane","date":"March 13, 2026","format":false,"excerpt":"In situ geochemical analysis has become a cornerstone of modern mineral exploration, mining, and environmental fieldwork. As project timelines tighten and operational costs increase, organizations can no longer afford delays caused by traditional laboratory testing workflows. In situ analysis \u2014 the ability to measure elemental composition directly at the sampling\u2026","rel":"","context":"In &quot;Exploration&quot;","block_context":{"text":"Exploration","link":"https:\/\/admin.acceleratingscience.com\/mining\/exploration\/"},"img":{"alt_text":"Close up of industrial worker holding mineral ore in gloved hands on mining site","src":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2026\/03\/iStock-896157842.jpg?resize=350%2C200&ssl=1","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2026\/03\/iStock-896157842.jpg?resize=350%2C200&ssl=1 1x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2026\/03\/iStock-896157842.jpg?resize=525%2C300&ssl=1 1.5x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2026\/03\/iStock-896157842.jpg?resize=700%2C400&ssl=1 2x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2026\/03\/iStock-896157842.jpg?resize=1050%2C600&ssl=1 3x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2026\/03\/iStock-896157842.jpg?resize=1400%2C800&ssl=1 4x"},"classes":[]},{"id":13544,"url":"https:\/\/www.thermofisher.com\/blog\/mining\/did-you-read-these-mining-and-minerals-articles-yet\/","url_meta":{"origin":1258,"position":3},"title":"Did You Read These Mining and Minerals Articles Yet?","author":"Marlene Gasdia-Cochrane","date":"December 16, 2025","format":false,"excerpt":"According to Google, here are the Top 10 read articles this year from Advancing Mining. Some have been around for quite awhile, but are still as relevant as ever. If you missed any of the latest articles, just click\u00a0here\u00a0to see the most recently published articles. You can even filter by\u2026","rel":"","context":"In &quot;General\/Industry&quot;","block_context":{"text":"General\/Industry","link":"https:\/\/admin.acceleratingscience.com\/mining\/general-industry-mining\/"},"img":{"alt_text":"Top 10, words, isolated on a dark blue background. Alphabet on stone blocks.Rating. Rating concept. Background.","src":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2025\/12\/iStock-1411361298.jpg?resize=350%2C200&ssl=1","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2025\/12\/iStock-1411361298.jpg?resize=350%2C200&ssl=1 1x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2025\/12\/iStock-1411361298.jpg?resize=525%2C300&ssl=1 1.5x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2025\/12\/iStock-1411361298.jpg?resize=700%2C400&ssl=1 2x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2025\/12\/iStock-1411361298.jpg?resize=1050%2C600&ssl=1 3x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2025\/12\/iStock-1411361298.jpg?resize=1400%2C800&ssl=1 4x"},"classes":[]},{"id":2707,"url":"https:\/\/www.thermofisher.com\/blog\/mining\/potash-a-look-at-the-worlds-most-popular-fertilizer\/","url_meta":{"origin":1258,"position":4},"title":"Potash: A Look at the World\u2019s Most Popular Fertilizer","author":"Ali Somarin","date":"April 8, 2026","format":false,"excerpt":"Potash Explained Potash is one of the most essential fertilizers in modern agriculture, supplying potassium\u2014an irreplaceable nutrient that supports plant growth, crop yield, and overall food production. According to the U.S. Geological Survey (USGS), about 85% of potash is used in fertilizers, underscoring its critical role in global farming systems.\u2026","rel":"","context":"In &quot;Data Management\/Lab Automation&quot;","block_context":{"text":"Data Management\/Lab Automation","link":"https:\/\/admin.acceleratingscience.com\/mining\/data-management-lab-automation\/"},"img":{"alt_text":"potash evaporation ponds at mining site","src":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2014\/06\/iStock-1362567340.jpg?resize=350%2C200&ssl=1","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2014\/06\/iStock-1362567340.jpg?resize=350%2C200&ssl=1 1x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2014\/06\/iStock-1362567340.jpg?resize=525%2C300&ssl=1 1.5x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2014\/06\/iStock-1362567340.jpg?resize=700%2C400&ssl=1 2x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2014\/06\/iStock-1362567340.jpg?resize=1050%2C600&ssl=1 3x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2014\/06\/iStock-1362567340.jpg?resize=1400%2C800&ssl=1 4x"},"classes":[]},{"id":13488,"url":"https:\/\/www.thermofisher.com\/blog\/mining\/the-role-of-edxrf-spectrometers-in-the-extraction-of-copper-from-ore\/","url_meta":{"origin":1258,"position":5},"title":"The Role of EDXRF Spectrometers in the Extraction of Copper from Ore","author":"Marlene Gasdia-Cochrane","date":"November 4, 2025","format":false,"excerpt":"Copper extraction is a complex process requiring precise control to ensure high purity. Energy Dispersive X-ray Fluorescence (EDXRF) spectrometers play a vital role by providing fast, non-destructive elemental analysis of ores, concentrates, and intermediate products. This enables mining operations to optimize extraction efficiency, maintain quality assurance, and meet stringent industry\u2026","rel":"","context":"In &quot;Elemental\/Geochemical Analysis&quot;","block_context":{"text":"Elemental\/Geochemical Analysis","link":"https:\/\/admin.acceleratingscience.com\/mining\/elemental-geochemical-analysis\/"},"img":{"alt_text":"a rock showing signs of copper and gold.","src":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2025\/10\/iStock-1309478061.jpg?resize=350%2C200&ssl=1","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2025\/10\/iStock-1309478061.jpg?resize=350%2C200&ssl=1 1x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2025\/10\/iStock-1309478061.jpg?resize=525%2C300&ssl=1 1.5x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2025\/10\/iStock-1309478061.jpg?resize=700%2C400&ssl=1 2x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2025\/10\/iStock-1309478061.jpg?resize=1050%2C600&ssl=1 3x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2025\/10\/iStock-1309478061.jpg?resize=1400%2C800&ssl=1 4x"},"classes":[]}],"jetpack_sharing_enabled":true,"jetpack_featured_media_url":"https:\/\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2013\/09\/p1020763.jpg","_links":{"self":[{"href":"https:\/\/www.thermofisher.com\/blog\/mining\/wp-json\/wp\/v2\/posts\/1258","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/www.thermofisher.com\/blog\/mining\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/www.thermofisher.com\/blog\/mining\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/mining\/wp-json\/wp\/v2\/users\/41"}],"replies":[{"embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/mining\/wp-json\/wp\/v2\/comments?post=1258"}],"version-history":[{"count":0,"href":"https:\/\/www.thermofisher.com\/blog\/mining\/wp-json\/wp\/v2\/posts\/1258\/revisions"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/mining\/wp-json\/wp\/v2\/media\/1257"}],"wp:attachment":[{"href":"https:\/\/www.thermofisher.com\/blog\/mining\/wp-json\/wp\/v2\/media?parent=1258"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/mining\/wp-json\/wp\/v2\/categories?post=1258"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/mining\/wp-json\/wp\/v2\/tags?post=1258"},{"taxonomy":"division","embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/mining\/wp-json\/wp\/v2\/division?post=1258"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}