{"id":2952,"date":"2014-08-28T08:00:41","date_gmt":"2014-08-28T12:00:41","guid":{"rendered":"http:\/\/admin.acceleratingscience.com\/?p=2952"},"modified":"2026-01-19T21:55:35","modified_gmt":"2026-01-19T21:55:35","slug":"gold-and-silver-pollutants-identified-using-xrf","status":"publish","type":"post","link":"https:\/\/www.thermofisher.com\/blog\/mining\/gold-and-silver-pollutants-identified-using-xrf\/","title":{"rendered":"Gold and Silver \u201cPollutants\u201d Identified Using XRF"},"content":{"rendered":"<p><img loading=\"lazy\" decoding=\"async\" style=\"border: 0;float: right;margin: 10px\" title=\"Crystals galena sphalerite and quartz\" src=\"http:\/\/admin.acceleratingscience.com\/wp-content\/uploads\/2014\/08\/117479246.jpg\" alt=\"Crystals galena sphalerite and quartz\" width=\"326\" height=\"244\" \/>Lead and zinc are non-ferrous metals often found together in sulfide ores. The most common lead- and zinc-containing ores are galena and sphalerite. A lead-zinc ore may also contain gold and silver, lead sulfide, zinc sulfide, iron sulfide, iron carbonate, and quartz. Lead and zinc ores are mined primarily underground in countries including China, Australia, Canada, the Unites States, and Peru.  According to the U.S. Geological Survey, <a href=\"http:\/\/minerals.usgs.gov\/minerals\/pubs\/commodity\/zinc\/mcs-2014-zinc.pdf\">13.5 million tons of zinc<\/a> and an estimated <a href=\"http:\/\/minerals.usgs.gov\/minerals\/pubs\/commodity\/lead\/mcs-2014-lead.pdf\">5.40 million tons of lead <\/a>were mined worldwide in 2013. In the U.S. alone, the value of zinc mined in 2013, based on zinc contained in concentrate, was about $1.60 billion and the 2013 value of recoverable mined lead, based on the average North American producer price, was about $829 million. Zinc is used as a protective coating on steel and iron, while most lead is used to make lead-acid batteries.  Lead and zinc can be produced pyrometallurgically or hydrometallurgically, depending on the type of ore used as a charge. In the pyrometallurgical process, ore concentrate containing lead, zinc, or both is fed, in some cases after sintering, into a primary smelter. Lead concentrations can be 50-70%, and the sulfur content of sulfidic ores is in the range of 15-20%. Zinc concentration is in the range of 40-60%, with sulfur content in sulfidic ores in the range of 26-34%. During sintering, a blast of hot air or oxygen is used to oxidize the sulfur present in the feed to sulfur dioxide (SO<sub>2<\/sub>). Blast furnaces are used in conventional processes for reduction and refining of lead compounds to produce lead bullion. This refining process removes any remaining nonlead materials (e.g., gold, silver, bismuth, zinc, and metal oxides such as oxides of antimony, arsenic, tin, and copper).  While the gold and silver are considered to be pollutants in the refinement process, these elements can be more profitable than the primary ore elements. It is for this reason that the accurate trace analysis for silver and gold in the ore material is essential. The quickest and best method for this analysis is wavelength dispersive x-ray fluorescence (WDXRF).  WDXRF technology is well established for elemental analysis of samples at various stages during mineral processing. WDXRF offers high sensitivity down to low atomic number elements, high repeatability and element selectivity. WDXRF is also favored for its wide dynamic range and ability to achieve the performance levels needed for routine industrial applications.  So how do <a href=\"http:\/\/www.thermoscientific.com\/en\/product\/arl-perform-x-sequential-x-ray-fluorescence-spectrometer.html\">WDXRF instruments<\/a> measure up? \u00a0Read the\u00a0<a href=\"http:\/\/www.thermoscientific.com\/content\/dam\/tfs\/ATG\/CAD\/CAD%20Documents\/Application%20&amp;%20Technical%20Notes\/Elemental%20Analysis\/XRF\/XR-AN41673-ARL%20PERFORM'X-Precious%20element%20analysis-0913.pdf\">analysis of silver and gold in lead and zinc ore minerals<\/a>\u00a0study data and analysis.<\/p>\n","protected":false},"excerpt":{"rendered":"<p>Lead and zinc are non-ferrous metals often found together in sulfide ores. The most common lead- and zinc-containing ores are galena and sphalerite. A lead-zinc ore may also contain gold and silver, lead sulfide, zinc sulfide, iron sulfide, iron carbonate, and quartz. Lead and zinc ores are mined primarily underground in countries including China, Australia,<\/p>\n","protected":false},"author":54,"featured_media":3049,"comment_status":"open","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_kad_blocks_custom_css":"","_kad_blocks_head_custom_js":"","_kad_blocks_body_custom_js":"","_kad_blocks_footer_custom_js":"","_monsterinsights_skip_tracking":false,"_genesis_hide_title":false,"_genesis_hide_breadcrumbs":false,"_genesis_hide_singular_image":false,"_genesis_hide_footer_widgets":false,"_genesis_custom_body_class":"","_genesis_custom_post_class":"","_genesis_layout":"","_jetpack_newsletter_access":"","_jetpack_dont_email_post_to_subs":false,"_jetpack_newsletter_tier_id":0,"_jetpack_memberships_contains_paywalled_content":false,"_jetpack_hide_featured_image":false,"_jetpack_featured_image_duplicate_dismissed":0,"_jetpack_memberships_contains_paid_content":false,"footnotes":""},"categories":[6],"tags":[],"division":[],"class_list":["post-2952","post","type-post","status-publish","format-standard","has-post-thumbnail","category-elemental-geochemical-analysis","entry"],"_selected_authors":[1252],"_selected_reviewers":[],"acf":[],"yoast_head":"<!-- This site is optimized with the Yoast SEO Premium plugin v28.9 (Yoast SEO v28.9) - https:\/\/yoast.com\/product\/yoast-seo-premium-wordpress\/ -->\n<title>Gold and Silver \u201cPollutants\u201d Identified Using XRF<\/title>\n<meta name=\"description\" content=\"WDXRF instruments measure up in the analysis of silver and gold in lead and zinc ore mining minerals.\" \/>\n<meta name=\"robots\" content=\"index, follow, max-snippet:-1, max-image-preview:large, max-video-preview:-1\" \/>\n<link rel=\"canonical\" href=\"https:\/\/www.thermofisher.com\/blog\/mining\/gold-and-silver-pollutants-identified-using-xrf\/\" \/>\n<meta property=\"og:locale\" content=\"en_US\" \/>\n<meta property=\"og:type\" content=\"article\" \/>\n<meta property=\"og:title\" content=\"Gold and Silver \u201cPollutants\u201d Identified Using XRF\" \/>\n<meta property=\"og:description\" content=\"WDXRF instruments measure up in the analysis of silver and gold in lead and zinc ore mining minerals.\" \/>\n<meta property=\"og:url\" content=\"https:\/\/www.thermofisher.com\/blog\/mining\/gold-and-silver-pollutants-identified-using-xrf\/\" \/>\n<meta property=\"og:site_name\" content=\"Advancing Mining\" \/>\n<meta property=\"article:publisher\" content=\"https:\/\/www.facebook.com\/thermofisher\" \/>\n<meta property=\"article:published_time\" content=\"2014-08-28T12:00:41+00:00\" \/>\n<meta property=\"article:modified_time\" content=\"2026-01-19T21:55:35+00:00\" \/>\n<meta property=\"og:image\" content=\"http:\/\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2014\/08\/117479246-1024x768.jpg\" \/>\n\t<meta property=\"og:image:width\" content=\"1024\" \/>\n\t<meta property=\"og:image:height\" content=\"768\" \/>\n\t<meta property=\"og:image:type\" content=\"image\/jpeg\" \/>\n<meta name=\"author\" content=\"Christopher Shaffer\" \/>\n<meta name=\"twitter:card\" content=\"summary_large_image\" \/>\n<meta name=\"twitter:creator\" content=\"@thermofisher\" \/>\n<meta name=\"twitter:site\" content=\"@thermofisher\" \/>\n<meta name=\"twitter:label1\" content=\"Written by\" \/>\n\t<meta name=\"twitter:data1\" content=\"Christopher Shaffer\" \/>\n\t<meta name=\"twitter:label2\" content=\"Est. reading time\" \/>\n\t<meta name=\"twitter:data2\" content=\"3 minutes\" \/>\n<script type=\"application\/ld+json\" class=\"yoast-schema-graph\">{\"@context\":\"https:\\\/\\\/schema.org\",\"@graph\":[{\"@type\":\"Article\",\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/mining\\\/gold-and-silver-pollutants-identified-using-xrf\\\/#article\",\"isPartOf\":{\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/mining\\\/gold-and-silver-pollutants-identified-using-xrf\\\/\"},\"author\":{\"name\":\"Christopher Shaffer\",\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/mining\\\/#\\\/schema\\\/person\\\/6bf7f4fdbbfbfce814fa4d081be51a2e\"},\"headline\":\"Gold and Silver \u201cPollutants\u201d Identified Using XRF\",\"datePublished\":\"2014-08-28T12:00:41+00:00\",\"dateModified\":\"2026-01-19T21:55:35+00:00\",\"mainEntityOfPage\":{\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/mining\\\/gold-and-silver-pollutants-identified-using-xrf\\\/\"},\"wordCount\":428,\"commentCount\":0,\"image\":{\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/mining\\\/gold-and-silver-pollutants-identified-using-xrf\\\/#primaryimage\"},\"thumbnailUrl\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/mining\\\/wp-content\\\/uploads\\\/sites\\\/3\\\/2014\\\/08\\\/117479246.jpg\",\"articleSection\":[\"Elemental\\\/Geochemical Analysis\"],\"inLanguage\":\"en-US\",\"potentialAction\":[{\"@type\":\"CommentAction\",\"name\":\"Comment\",\"target\":[\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/mining\\\/gold-and-silver-pollutants-identified-using-xrf\\\/#respond\"]}]},{\"@type\":\"WebPage\",\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/mining\\\/gold-and-silver-pollutants-identified-using-xrf\\\/\",\"url\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/mining\\\/gold-and-silver-pollutants-identified-using-xrf\\\/\",\"name\":\"Gold and Silver \u201cPollutants\u201d Identified Using XRF\",\"isPartOf\":{\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/mining\\\/#website\"},\"primaryImageOfPage\":{\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/mining\\\/gold-and-silver-pollutants-identified-using-xrf\\\/#primaryimage\"},\"image\":{\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/mining\\\/gold-and-silver-pollutants-identified-using-xrf\\\/#primaryimage\"},\"thumbnailUrl\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/mining\\\/wp-content\\\/uploads\\\/sites\\\/3\\\/2014\\\/08\\\/117479246.jpg\",\"datePublished\":\"2014-08-28T12:00:41+00:00\",\"dateModified\":\"2026-01-19T21:55:35+00:00\",\"author\":{\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/mining\\\/#\\\/schema\\\/person\\\/6bf7f4fdbbfbfce814fa4d081be51a2e\"},\"description\":\"WDXRF instruments measure up in the analysis of silver and gold in lead and zinc ore mining minerals.\",\"breadcrumb\":{\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/mining\\\/gold-and-silver-pollutants-identified-using-xrf\\\/#breadcrumb\"},\"inLanguage\":\"en-US\",\"potentialAction\":[{\"@type\":\"ReadAction\",\"target\":[\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/mining\\\/gold-and-silver-pollutants-identified-using-xrf\\\/\"]}]},{\"@type\":\"ImageObject\",\"inLanguage\":\"en-US\",\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/mining\\\/gold-and-silver-pollutants-identified-using-xrf\\\/#primaryimage\",\"url\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/mining\\\/wp-content\\\/uploads\\\/sites\\\/3\\\/2014\\\/08\\\/117479246.jpg\",\"contentUrl\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/mining\\\/wp-content\\\/uploads\\\/sites\\\/3\\\/2014\\\/08\\\/117479246.jpg\",\"width\":3264,\"height\":2448,\"caption\":\"Crystals galena sphalerite and quartz\"},{\"@type\":\"BreadcrumbList\",\"@id\":\"https:\\\/\\\/www.thermofisher.com\\\/blog\\\/mining\\\/gold-and-silver-pollutants-identified-using-xrf\\\/#breadcrumb\",\"itemListElement\":[{\"@type\":\"ListItem\",\"position\":1,\"name\":\"Home\",\"item\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/mining\\\/\"},{\"@type\":\"ListItem\",\"position\":2,\"name\":\"Gold and Silver \u201cPollutants\u201d Identified Using XRF\"}]},{\"@type\":\"WebSite\",\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/mining\\\/#website\",\"url\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/mining\\\/\",\"name\":\"Advancing Mining\",\"description\":\"\",\"potentialAction\":[{\"@type\":\"SearchAction\",\"target\":{\"@type\":\"EntryPoint\",\"urlTemplate\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/mining\\\/?s={search_term_string}\"},\"query-input\":{\"@type\":\"PropertyValueSpecification\",\"valueRequired\":true,\"valueName\":\"search_term_string\"}}],\"inLanguage\":\"en-US\"},{\"@type\":\"Person\",\"@id\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/mining\\\/#\\\/schema\\\/person\\\/6bf7f4fdbbfbfce814fa4d081be51a2e\",\"name\":\"Christopher Shaffer\",\"image\":{\"@type\":\"ImageObject\",\"inLanguage\":\"en-US\",\"@id\":\"https:\\\/\\\/secure.gravatar.com\\\/avatar\\\/1f1923cff250c2d1b62ec0eb8fc5dc5e0eab78190d631bedff2410403189d09e?s=96&d=mm&r=g\",\"url\":\"https:\\\/\\\/secure.gravatar.com\\\/avatar\\\/1f1923cff250c2d1b62ec0eb8fc5dc5e0eab78190d631bedff2410403189d09e?s=96&d=mm&r=g\",\"contentUrl\":\"https:\\\/\\\/secure.gravatar.com\\\/avatar\\\/1f1923cff250c2d1b62ec0eb8fc5dc5e0eab78190d631bedff2410403189d09e?s=96&d=mm&r=g\",\"caption\":\"Christopher Shaffer\"},\"url\":\"https:\\\/\\\/admin.acceleratingscience.com\\\/author\\\/chris-shafferthermofisher-com\\\/\"}]}<\/script>\n<!-- \/ Yoast SEO Premium plugin. -->","yoast_head_json":{"title":"Gold and Silver \u201cPollutants\u201d Identified Using XRF","description":"WDXRF instruments measure up in the analysis of silver and gold in lead and zinc ore mining minerals.","robots":{"index":"index","follow":"follow","max-snippet":"max-snippet:-1","max-image-preview":"max-image-preview:large","max-video-preview":"max-video-preview:-1"},"canonical":"https:\/\/www.thermofisher.com\/blog\/mining\/gold-and-silver-pollutants-identified-using-xrf\/","og_locale":"en_US","og_type":"article","og_title":"Gold and Silver \u201cPollutants\u201d Identified Using XRF","og_description":"WDXRF instruments measure up in the analysis of silver and gold in lead and zinc ore mining minerals.","og_url":"https:\/\/www.thermofisher.com\/blog\/mining\/gold-and-silver-pollutants-identified-using-xrf\/","og_site_name":"Advancing Mining","article_publisher":"https:\/\/www.facebook.com\/thermofisher","article_published_time":"2014-08-28T12:00:41+00:00","article_modified_time":"2026-01-19T21:55:35+00:00","og_image":[{"width":1024,"height":768,"url":"http:\/\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2014\/08\/117479246-1024x768.jpg","type":"image\/jpeg"}],"author":"Christopher Shaffer","twitter_card":"summary_large_image","twitter_creator":"@thermofisher","twitter_site":"@thermofisher","twitter_misc":{"Written by":"Christopher Shaffer","Est. reading time":"3 minutes"},"schema":{"@context":"https:\/\/schema.org","@graph":[{"@type":"Article","@id":"https:\/\/www.thermofisher.com\/blog\/mining\/gold-and-silver-pollutants-identified-using-xrf\/#article","isPartOf":{"@id":"https:\/\/www.thermofisher.com\/blog\/mining\/gold-and-silver-pollutants-identified-using-xrf\/"},"author":{"name":"Christopher Shaffer","@id":"https:\/\/admin.acceleratingscience.com\/mining\/#\/schema\/person\/6bf7f4fdbbfbfce814fa4d081be51a2e"},"headline":"Gold and Silver \u201cPollutants\u201d Identified Using XRF","datePublished":"2014-08-28T12:00:41+00:00","dateModified":"2026-01-19T21:55:35+00:00","mainEntityOfPage":{"@id":"https:\/\/www.thermofisher.com\/blog\/mining\/gold-and-silver-pollutants-identified-using-xrf\/"},"wordCount":428,"commentCount":0,"image":{"@id":"https:\/\/www.thermofisher.com\/blog\/mining\/gold-and-silver-pollutants-identified-using-xrf\/#primaryimage"},"thumbnailUrl":"https:\/\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2014\/08\/117479246.jpg","articleSection":["Elemental\/Geochemical Analysis"],"inLanguage":"en-US","potentialAction":[{"@type":"CommentAction","name":"Comment","target":["https:\/\/www.thermofisher.com\/blog\/mining\/gold-and-silver-pollutants-identified-using-xrf\/#respond"]}]},{"@type":"WebPage","@id":"https:\/\/www.thermofisher.com\/blog\/mining\/gold-and-silver-pollutants-identified-using-xrf\/","url":"https:\/\/www.thermofisher.com\/blog\/mining\/gold-and-silver-pollutants-identified-using-xrf\/","name":"Gold and Silver \u201cPollutants\u201d Identified Using XRF","isPartOf":{"@id":"https:\/\/admin.acceleratingscience.com\/mining\/#website"},"primaryImageOfPage":{"@id":"https:\/\/www.thermofisher.com\/blog\/mining\/gold-and-silver-pollutants-identified-using-xrf\/#primaryimage"},"image":{"@id":"https:\/\/www.thermofisher.com\/blog\/mining\/gold-and-silver-pollutants-identified-using-xrf\/#primaryimage"},"thumbnailUrl":"https:\/\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2014\/08\/117479246.jpg","datePublished":"2014-08-28T12:00:41+00:00","dateModified":"2026-01-19T21:55:35+00:00","author":{"@id":"https:\/\/admin.acceleratingscience.com\/mining\/#\/schema\/person\/6bf7f4fdbbfbfce814fa4d081be51a2e"},"description":"WDXRF instruments measure up in the analysis of silver and gold in lead and zinc ore mining minerals.","breadcrumb":{"@id":"https:\/\/www.thermofisher.com\/blog\/mining\/gold-and-silver-pollutants-identified-using-xrf\/#breadcrumb"},"inLanguage":"en-US","potentialAction":[{"@type":"ReadAction","target":["https:\/\/www.thermofisher.com\/blog\/mining\/gold-and-silver-pollutants-identified-using-xrf\/"]}]},{"@type":"ImageObject","inLanguage":"en-US","@id":"https:\/\/www.thermofisher.com\/blog\/mining\/gold-and-silver-pollutants-identified-using-xrf\/#primaryimage","url":"https:\/\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2014\/08\/117479246.jpg","contentUrl":"https:\/\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2014\/08\/117479246.jpg","width":3264,"height":2448,"caption":"Crystals galena sphalerite and quartz"},{"@type":"BreadcrumbList","@id":"https:\/\/www.thermofisher.com\/blog\/mining\/gold-and-silver-pollutants-identified-using-xrf\/#breadcrumb","itemListElement":[{"@type":"ListItem","position":1,"name":"Home","item":"https:\/\/admin.acceleratingscience.com\/mining\/"},{"@type":"ListItem","position":2,"name":"Gold and Silver \u201cPollutants\u201d Identified Using XRF"}]},{"@type":"WebSite","@id":"https:\/\/admin.acceleratingscience.com\/mining\/#website","url":"https:\/\/admin.acceleratingscience.com\/mining\/","name":"Advancing Mining","description":"","potentialAction":[{"@type":"SearchAction","target":{"@type":"EntryPoint","urlTemplate":"https:\/\/admin.acceleratingscience.com\/mining\/?s={search_term_string}"},"query-input":{"@type":"PropertyValueSpecification","valueRequired":true,"valueName":"search_term_string"}}],"inLanguage":"en-US"},{"@type":"Person","@id":"https:\/\/admin.acceleratingscience.com\/mining\/#\/schema\/person\/6bf7f4fdbbfbfce814fa4d081be51a2e","name":"Christopher Shaffer","image":{"@type":"ImageObject","inLanguage":"en-US","@id":"https:\/\/secure.gravatar.com\/avatar\/1f1923cff250c2d1b62ec0eb8fc5dc5e0eab78190d631bedff2410403189d09e?s=96&d=mm&r=g","url":"https:\/\/secure.gravatar.com\/avatar\/1f1923cff250c2d1b62ec0eb8fc5dc5e0eab78190d631bedff2410403189d09e?s=96&d=mm&r=g","contentUrl":"https:\/\/secure.gravatar.com\/avatar\/1f1923cff250c2d1b62ec0eb8fc5dc5e0eab78190d631bedff2410403189d09e?s=96&d=mm&r=g","caption":"Christopher Shaffer"},"url":"https:\/\/admin.acceleratingscience.com\/author\/chris-shafferthermofisher-com\/"}]}},"taxonomy_info":{"category":[{"value":6,"label":"Elemental\/Geochemical Analysis"}]},"featured_image_src_large":["https:\/\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2014\/08\/117479246-1024x768.jpg",760,570,true],"author_info":{"display_name":"Christopher Shaffer","author_link":"https:\/\/admin.acceleratingscience.com\/author\/chris-shafferthermofisher-com\/"},"comment_info":0,"category_info":[{"term_id":6,"name":"Elemental\/Geochemical Analysis","slug":"elemental-geochemical-analysis","term_group":0,"term_taxonomy_id":6,"taxonomy":"category","description":"","parent":0,"count":144,"filter":"raw","meta":[],"cat_ID":6,"category_count":144,"category_description":"","cat_name":"Elemental\/Geochemical Analysis","category_nicename":"elemental-geochemical-analysis","category_parent":0}],"tag_info":false,"jetpack-related-posts":[{"id":13488,"url":"https:\/\/www.thermofisher.com\/blog\/mining\/the-role-of-edxrf-spectrometers-in-the-extraction-of-copper-from-ore\/","url_meta":{"origin":2952,"position":0},"title":"The Role of EDXRF Spectrometers in the Extraction of Copper from Ore","author":"Marlene Gasdia-Cochrane","date":"November 4, 2025","format":false,"excerpt":"Copper extraction is a complex process requiring precise control to ensure high purity. Energy Dispersive X-ray Fluorescence (EDXRF) spectrometers play a vital role by providing fast, non-destructive elemental analysis of ores, concentrates, and intermediate products. This enables mining operations to optimize extraction efficiency, maintain quality assurance, and meet stringent industry\u2026","rel":"","context":"In &quot;Elemental\/Geochemical Analysis&quot;","block_context":{"text":"Elemental\/Geochemical Analysis","link":"https:\/\/admin.acceleratingscience.com\/mining\/elemental-geochemical-analysis\/"},"img":{"alt_text":"a rock showing signs of copper and gold.","src":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2025\/10\/iStock-1309478061.jpg?resize=350%2C200&ssl=1","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2025\/10\/iStock-1309478061.jpg?resize=350%2C200&ssl=1 1x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2025\/10\/iStock-1309478061.jpg?resize=525%2C300&ssl=1 1.5x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2025\/10\/iStock-1309478061.jpg?resize=700%2C400&ssl=1 2x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2025\/10\/iStock-1309478061.jpg?resize=1050%2C600&ssl=1 3x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2025\/10\/iStock-1309478061.jpg?resize=1400%2C800&ssl=1 4x"},"classes":[]},{"id":13544,"url":"https:\/\/www.thermofisher.com\/blog\/mining\/did-you-read-these-mining-and-minerals-articles-yet\/","url_meta":{"origin":2952,"position":1},"title":"Did You Read These Mining and Minerals Articles Yet?","author":"Marlene Gasdia-Cochrane","date":"December 16, 2025","format":false,"excerpt":"According to Google, here are the Top 10 read articles this year from Advancing Mining. Some have been around for quite awhile, but are still as relevant as ever. If you missed any of the latest articles, just click\u00a0here\u00a0to see the most recently published articles. You can even filter by\u2026","rel":"","context":"In &quot;General\/Industry&quot;","block_context":{"text":"General\/Industry","link":"https:\/\/admin.acceleratingscience.com\/mining\/general-industry-mining\/"},"img":{"alt_text":"Top 10, words, isolated on a dark blue background. Alphabet on stone blocks.Rating. Rating concept. Background.","src":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2025\/12\/iStock-1411361298.jpg?resize=350%2C200&ssl=1","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2025\/12\/iStock-1411361298.jpg?resize=350%2C200&ssl=1 1x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2025\/12\/iStock-1411361298.jpg?resize=525%2C300&ssl=1 1.5x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2025\/12\/iStock-1411361298.jpg?resize=700%2C400&ssl=1 2x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2025\/12\/iStock-1411361298.jpg?resize=1050%2C600&ssl=1 3x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2025\/12\/iStock-1411361298.jpg?resize=1400%2C800&ssl=1 4x"},"classes":[]},{"id":14189,"url":"https:\/\/www.thermofisher.com\/blog\/mining\/understanding-xrf-testing-for-phosphate-ore-quality-and-elemental-composition\/","url_meta":{"origin":2952,"position":2},"title":"Understanding XRF Testing for Phosphate Ore Quality and Elemental Composition","author":"Marlene Gasdia-Cochrane","date":"November 3, 2026","format":false,"excerpt":"Article Summary Phosphate ore is the primary source of phosphorus used in fertilizers, making reliable phosphate analysis important to global agriculture and food production. As higher-grade deposits become less available, mining operations increasingly need to process lower-grade ores with more variable chemical compositions. Handheld X-ray fluorescence (XRF) provides a fast,\u2026","rel":"","context":"In &quot;Exploration&quot;","block_context":{"text":"Exploration","link":"https:\/\/admin.acceleratingscience.com\/mining\/exploration\/"},"img":{"alt_text":"An aerial view of an open pit phosphate mine","src":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2026\/10\/iStock-177760247_phosphatemine.jpg?resize=350%2C200&ssl=1","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2026\/10\/iStock-177760247_phosphatemine.jpg?resize=350%2C200&ssl=1 1x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2026\/10\/iStock-177760247_phosphatemine.jpg?resize=525%2C300&ssl=1 1.5x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2026\/10\/iStock-177760247_phosphatemine.jpg?resize=700%2C400&ssl=1 2x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2026\/10\/iStock-177760247_phosphatemine.jpg?resize=1050%2C600&ssl=1 3x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2026\/10\/iStock-177760247_phosphatemine.jpg?resize=1400%2C800&ssl=1 4x"},"classes":[]},{"id":2707,"url":"https:\/\/www.thermofisher.com\/blog\/mining\/potash-a-look-at-the-worlds-most-popular-fertilizer\/","url_meta":{"origin":2952,"position":3},"title":"Potash: A Look at the World\u2019s Most Popular Fertilizer","author":"Ali Somarin","date":"April 8, 2026","format":false,"excerpt":"Potash Explained Potash is one of the most essential fertilizers in modern agriculture, supplying potassium\u2014an irreplaceable nutrient that supports plant growth, crop yield, and overall food production. According to the U.S. Geological Survey (USGS), about 85% of potash is used in fertilizers, underscoring its critical role in global farming systems.\u2026","rel":"","context":"In &quot;Data Management\/Lab Automation&quot;","block_context":{"text":"Data Management\/Lab Automation","link":"https:\/\/admin.acceleratingscience.com\/mining\/data-management-lab-automation\/"},"img":{"alt_text":"potash evaporation ponds at mining site","src":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2014\/06\/iStock-1362567340.jpg?resize=350%2C200&ssl=1","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2014\/06\/iStock-1362567340.jpg?resize=350%2C200&ssl=1 1x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2014\/06\/iStock-1362567340.jpg?resize=525%2C300&ssl=1 1.5x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2014\/06\/iStock-1362567340.jpg?resize=700%2C400&ssl=1 2x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2014\/06\/iStock-1362567340.jpg?resize=1050%2C600&ssl=1 3x, https:\/\/i0.wp.com\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2014\/06\/iStock-1362567340.jpg?resize=1400%2C800&ssl=1 4x"},"classes":[]}],"jetpack_sharing_enabled":true,"jetpack_featured_media_url":"https:\/\/admin.acceleratingscience.com\/mining\/wp-content\/uploads\/sites\/3\/2014\/08\/117479246.jpg","_links":{"self":[{"href":"https:\/\/www.thermofisher.com\/blog\/mining\/wp-json\/wp\/v2\/posts\/2952","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/www.thermofisher.com\/blog\/mining\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/www.thermofisher.com\/blog\/mining\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/mining\/wp-json\/wp\/v2\/users\/54"}],"replies":[{"embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/mining\/wp-json\/wp\/v2\/comments?post=2952"}],"version-history":[{"count":0,"href":"https:\/\/www.thermofisher.com\/blog\/mining\/wp-json\/wp\/v2\/posts\/2952\/revisions"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/mining\/wp-json\/wp\/v2\/media\/3049"}],"wp:attachment":[{"href":"https:\/\/www.thermofisher.com\/blog\/mining\/wp-json\/wp\/v2\/media?parent=2952"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/mining\/wp-json\/wp\/v2\/categories?post=2952"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/mining\/wp-json\/wp\/v2\/tags?post=2952"},{"taxonomy":"division","embeddable":true,"href":"https:\/\/www.thermofisher.com\/blog\/mining\/wp-json\/wp\/v2\/division?post=2952"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}