Measure density or density-related variables with Thermo Scientific™ Sarasota Liquid Density Meters--FD910 and FD950*. These meters are for use within the oil and gas, petrochemical, chemical, food and beverage, and general process industries.

*Please note:  The FD960 has been discontinued although we continue to provide services and support for this product.
 

 
Catalog Number
14180
Unit Size
Each
Description
Liquid Density Meters
Price (USD)
Full specifications
Ambient Temperature-20° to +60°C (-4° to +140°F) ambient
Calibration CertificationCE marked; ISO 9001:2008
Certifications/ComplianceCE marked; ISO 9001:2008
ConnectionsScrew terminals; Cable entry: 2 x 0.75 in. NPT
DescriptionLiquid Density Meters
Electronics Housing MaterialCopper free aluminium grey epoxy finish; Plate glass window for local display option
Environmental RatingIP65 (NEMA 4X)
Flow RangeVertical installation: 0L/min to 300L/min (0 USG/min to 79 USG/min); Horizontal installation: 5L/min to 300L/min (1.3 USG/min to 79 USG/min)
Installation PositionVertical installation (standard), horizontal installation (optional); No instrument or pipework supports required
Local Display4-1/2-digit 0.3 in. (7.6mm) 7-segment LCD display. Resolution 0.1% or 0.01% depending on display variable.
Max. Operating PressureSarasota FD910 and FD950: as flange rating
Operating Density Range0  to 131.1 lb./ft.³
Other Wetted Parts MaterialSarasota FD910 (Class 150, 300) / Sarasota FD910 (Class 600) / FD950: Hastelloy C276
OutputsF option (frequency output): Frequency related to density on 2-wire current modulated loop 6 mA to 18 mA; 4-wire PT100; H option (headmounted electronics): Analog 4-20 mA related to density or density derived variable; HART protocol
Power SupplyF option (frequency output): 13-28VDC 10mA average (peak 18mA); H option (headmounted electronics) 2 x 13-28VDC 25mA; 4-20mA current pressure input available
Repeatability0.0012 lb./ft.³
Temperature Effect0.005kg/m³/°C (0.0002 lb./ ft.³/°F) note: correction coefficients applied
Temperature MeasurementHigh accuracy 1/3 DIN integral 4-wire PT100
Transducer Calibration AccuracyAvailable to ±0.0062 lb./ft.³
Weight (English) Net24 lb.
Weight (English) Shipping33 lb.
Weight (Metric) Net11kg
Weight (Metric) Shipping15kg
Item DescriptionLiquid density meters--FD910 and FD950
Unit SizeEach
Showing 1 of 1
Catalog NumberSpecificationsUnit SizeDescriptionPrice (USD)
14180Full specifications
EachLiquid Density MetersRequest A Quote
Ambient Temperature-20° to +60°C (-4° to +140°F) ambient
Calibration CertificationCE marked; ISO 9001:2008
Certifications/ComplianceCE marked; ISO 9001:2008
ConnectionsScrew terminals; Cable entry: 2 x 0.75 in. NPT
DescriptionLiquid Density Meters
Electronics Housing MaterialCopper free aluminium grey epoxy finish; Plate glass window for local display option
Environmental RatingIP65 (NEMA 4X)
Flow RangeVertical installation: 0L/min to 300L/min (0 USG/min to 79 USG/min); Horizontal installation: 5L/min to 300L/min (1.3 USG/min to 79 USG/min)
Installation PositionVertical installation (standard), horizontal installation (optional); No instrument or pipework supports required
Local Display4-1/2-digit 0.3 in. (7.6mm) 7-segment LCD display. Resolution 0.1% or 0.01% depending on display variable.
Max. Operating PressureSarasota FD910 and FD950: as flange rating
Operating Density Range0  to 131.1 lb./ft.³
Other Wetted Parts MaterialSarasota FD910 (Class 150, 300) / Sarasota FD910 (Class 600) / FD950: Hastelloy C276
OutputsF option (frequency output): Frequency related to density on 2-wire current modulated loop 6 mA to 18 mA; 4-wire PT100; H option (headmounted electronics): Analog 4-20 mA related to density or density derived variable; HART protocol
Power SupplyF option (frequency output): 13-28VDC 10mA average (peak 18mA); H option (headmounted electronics) 2 x 13-28VDC 25mA; 4-20mA current pressure input available
Repeatability0.0012 lb./ft.³
Temperature Effect0.005kg/m³/°C (0.0002 lb./ ft.³/°F) note: correction coefficients applied
Temperature MeasurementHigh accuracy 1/3 DIN integral 4-wire PT100
Transducer Calibration AccuracyAvailable to ±0.0062 lb./ft.³
Weight (English) Net24 lb.
Weight (English) Shipping33 lb.
Weight (Metric) Net11kg
Weight (Metric) Shipping15kg
Item DescriptionLiquid density meters--FD910 and FD950
Unit SizeEach
Showing 1 of 1
  • Continuous, online density monitoring
  • Measurement at process conditions
  • Straight through flow path
  • Materials to suit applications
  • Compact and lightweight, easy to install
  • Hazardous area approvals
  • Converter electronics to suit application

Sarasota FD910 Industrial Density Meter:

  • Provides high accuracy measurement of general process liquids and slurries
  • Delivers continuous online density or density related measurement with straight-through flow
  • Built to be insensitive to plant/pipelinevibrations
  • Wide operating temperature span and options to suit an extensive range of applications

Sarasota FD950 Chemical Density Meter:

  • Provides high accuracy measurement of aggressive liquids and slurries
  • Provides continuous online density measurement
  • Delivers straight-through flow in a compact and lightweight design
  • Accurate to 0.1kg/m

Recommended for:

  • Blending
  • Product identification
  • Interface detection
  • Dilution measurement
  • Process/quality control
  • SG measurement
  • Process efficiency
  • Waste minimization
  • Product consistency
  • Volumetric to mass flow metering systems