Thermo Scientific™

Dionex™ IonPac™ AS31 Analytical and Guard Columns

Catalog number: 303147
Thermo Scientific™

Dionex™ IonPac™ AS31 Analytical and Guard Columns

Catalog number: 303147

Achieve faster separation of haloacetic acids (HAAs), bromate, and dalapon in drinking water prior to detection with ion chromatography (IC) and mass spectrometry (MS) with the new Thermo Scientific™ Dionex™ IonPac™ AS31 column. Specifically designed for fast analysis, the Dionex IonPac AS31 provides approximately 39% shorter run times relative to the Dionex IonPac AS24, reducing U.S. EPA Method 557 run time from 57 minutes to approximately 35 minutes. This new hydroxide-selective anion-exchange column has relatively high capacity to allow large loop injections for trace analysis (µg/L) without sample pre-treatment.

 
Catalog Number
303147
Unit Size
Each
Type
Microbore Analytical Column
Length (Metric)
250 mm
Particle Size
6 µm
Price (USD)
1,387.91
Your Price:
Availability
***
Full specifications
Quantity
Packing MaterialEthylvinylbenzene/Divinylbenzene
Column Capacity116 µeq
Column ChemistryHydroxide
Flow Rate0.3 mL/min.
For Use With (Application)Anion Exchange
For Use With (Equipment)Dionex IonPac AG31 Microbore Guard Column
HydrophobicityLow
Max. Pressure5000 psi (345 bar)
Mobile Phase CompatibilitypH 0 to 14, 0 to 100% Organic Solvents (Acetonitrile, Methanol, 2-Propanol)
Particle Size6 µm
Stationary PhaseSuper Macroporous Resin
TypeMicrobore Analytical Column
Length (Metric)250 mm
Diameter (Metric)2 mm
Crosslinking55% DVB
Column ConstructionPEEK
Anion Exchange Capacity116 µeq
Unit SizeEach
Catalog Number
Unit Size
Each
Type
Microbore Guard Column
Length (Metric)
50 mm
Particle Size
11 µm
Price (USD)
460.37
Your Price:
Availability
***
Full specifications
Quantity
Packing MaterialEthylvinylbenzene/Divinylbenzene
Column Capacity1.5 µeq
Column ChemistryHydroxide
Flow Rate0.30 to 0.40 mL/min.
For Use With (Application)Anion Exchange
For Use With (Equipment)Dionex IonPac AS31 Microbore Analytical Column
HydrophobicityLow
Max. Pressure5000 psi (345 bar)
Mobile Phase CompatibilitypH 0 to 14, 0 to 100% Organic Solvents (Acetonitrile, Methanol, 2-Propanol)
Particle Size11 µm
Stationary PhaseMicroporous Resin
TypeMicrobore Guard Column
Length (Metric)50 mm
Diameter (Metric)2 mm
Crosslinking55% DVB
Column ConstructionPEEK
Anion Exchange Capacity1.5 µeq
Unit SizeEach
Showing 2 of 2
Catalog NumberSpecificationsUnit SizeTypeLength (Metric)Particle SizePrice (USD)AvailabilityQuantity
303147Full specifications
EachMicrobore Analytical Column250 mm6 µm
1,387.91
Your Price:
***
Packing MaterialEthylvinylbenzene/Divinylbenzene
Column Capacity116 µeq
Column ChemistryHydroxide
Flow Rate0.3 mL/min.
For Use With (Application)Anion Exchange
For Use With (Equipment)Dionex IonPac AG31 Microbore Guard Column
HydrophobicityLow
Max. Pressure5000 psi (345 bar)
Mobile Phase CompatibilitypH 0 to 14, 0 to 100% Organic Solvents (Acetonitrile, Methanol, 2-Propanol)
Particle Size6 µm
Stationary PhaseSuper Macroporous Resin
TypeMicrobore Analytical Column
Length (Metric)250 mm
Diameter (Metric)2 mm
Crosslinking55% DVB
Column ConstructionPEEK
Anion Exchange Capacity116 µeq
Unit SizeEach
303148Full specifications
EachMicrobore Guard Column50 mm11 µm
460.37
Your Price:
***
Packing MaterialEthylvinylbenzene/Divinylbenzene
Column Capacity1.5 µeq
Column ChemistryHydroxide
Flow Rate0.30 to 0.40 mL/min.
For Use With (Application)Anion Exchange
For Use With (Equipment)Dionex IonPac AS31 Microbore Analytical Column
HydrophobicityLow
Max. Pressure5000 psi (345 bar)
Mobile Phase CompatibilitypH 0 to 14, 0 to 100% Organic Solvents (Acetonitrile, Methanol, 2-Propanol)
Particle Size11 µm
Stationary PhaseMicroporous Resin
TypeMicrobore Guard Column
Length (Metric)50 mm
Diameter (Metric)2 mm
Crosslinking55% DVB
Column ConstructionPEEK
Anion Exchange Capacity1.5 µeq
Unit SizeEach
Showing 2 of 2
Superior Chromatographic Performance
  • Can meet or exceed the performance requirements of U.S. EPA Method 557.
  • High capacity (116 µeq/column) allows injection of complex matrices without overloading the column while maintaining excellent peak shapes for lower concentration analytes.
  • Requires no sample pre-treatment when determining haloacetic acids in high-ionic strength matrices.
  • Column selectivity has been optimized at an operating temperature of 15°C to ensure reproducible recovery for haloacetic acids.
Use with a High-Pressure IC system

A high-pressure IC system, such as the Thermo Scientific™ Dionex™ ICS-5000+ HPIC™ system or Dionex ICS-6000 HPIC system is required to operate this column under standard conditions.