Search Thermo Fisher Scientific
Search Thermo Fisher Scientific
Detectable Analytes | Hydrogen Sulfide, Sulfur Dioxide |
Inputs | 16 Digital Inputs (Standard), 8 0 to 10 VDC Analog Inputs (Optional) |
Linearity | ±1% full scale <100 ppm |
Response Time | 80 sec. (10 sec. average time); 110 sec. (60 sec. average time); 320 sec. (300 sec. average time) |
Span Drift | ±1% full scale |
Temperature (English) Operating | 68°F to 86°F |
Temperature (Metric) Operating | 20°C to 30°C |
Zero Noise | Manual Mode SO2 or Combine Sulfur: 1.0 ppb (10 sec. averaging time); 0.5 ppb (60 sec. averaging time); 0.25 ppb (300 sec. averaging time); Automatic Mode SO2 or H2S: 3.0 ppb (10 sec. averaging time); 1.5 ppb (60 sec. averaging time); 0.75 ppb (300 sec. averaging time) |
Certifications/Compliance | ETL, CSA, UL, CE, UKCA |
Depth (English) | 23 in. |
Depth (Metric) | 584 mm |
Description | Model 450i Hydrogen Sulfide - Sulfur Dioxide Analyzer |
Detection Limit | Manual Mode SO2 or Combine Sulfur: 2.0 ppb (10 second averaging time); 1.0 ppb (60 second averaging time); 0.5 ppb (300 second averaging time); Automatic Mode SO2 or H2S: 6.0 ppb (10 second averaging time); 2.0 ppb (60 second averaging time); 1.5 ppb (300 seconds) |
Height (English) | 8.62 in. |
Height (Metric) | 219 mm |
Measurement Range | Preset: 0 to 0.05, 0.1, 0.2, 0.5, 1, 2, 5, and 10 ppm; 0 to 0.2; Extended: 0 to 0.5, 1, 2, 5, 10, 20, 50 and 100 ppm; 0-2, 5, 10, 20, 50, 100, 200, and 250 mg/m3; Custom: 0 to 0.05 to 100 ppm; 0 to 0.2 to 250 mg/m3 |
Mounting Options | Bench, EIA, retrofit |
Outputs | Selectable Voltage, RS232/RS485, TCP/IP, 10 Status Relays, and Power Fail Indication (standard); 0 to 20 or 4 to 20mA Isolated Current Outout (optional) |
Precision | 1% of reading or 1 ppb (whichever is greater) |
Voltage | 100/115/220/240 V |
Weight (English) | 48 lb. |
Weight (Metric) | 21.8 kg |
Width (English) | 16.75 in. |
Width (Metric) | 425 mm |
Zero Drift | <1ppb |
Unit Size | Each |
Catalog Number | Specifications | Unit Size | Description | Price (USD) | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
450I | Each | Model 450i Hydrogen Sulfide - Sulfur Dioxide Analyzer | Request A Quote | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Key Features:
iSeries Advantage: