| MICROSVAOED | Full specifications | Each | Micro sievert with audio, extended detector | Request A Quote |
Description | Micro sievert with audio, extended detector | Shock | 100 g per lightweight machine of MIL-STD 202 C, method 202 B | Energy Range | 40 keV to 1.3 MeV | Display Type | Ruggedized, recessed, high-torque 1mA meter protected by impact resistant Lexan polycarbonate window with 85.1mm (3.35 in.) scale marked: MICRO REM: 0 to 200 μSv/h, with “Bat. ok«, “HV ok” check bands. MICRO SIEVERT: 0 to 2.0 μSv/h, with “Bat. ok«, “HV ok” check bands. | Battery Type | 29 V, MN1604, or equal | Product Type | Tissue-Equivalent Survey Meter | Operating Temperature (Metric) | −20°C to +50°C | Operating Temperature (English) | -4°F to +122°F | Response Time | Optimized for each range, 0-90% of final reading as follows: Range: : Time: X0.1 : <15 seconds X1 : <15 seconds X10 : <5 seconds X100 : <2 seconds X1000 : <2 seconds | Detector Type | Internal, tissue equivalent, organic scintillator | Geotropism | Within ± 2% of full scale | Humidity | <5% change in reading from 10% to 95% | For Use With (Application) | Routine low/medium level gammadose rate surveysConfirming radiation boundariesMonitoring items for unrestricted releaseX-ray surveys, using the expanded low-energy range option | Vibration | 5 g in each of three mutually orthogonal axes at one or more frequencies from 10 to 33 Hz. | Accuracy | Within 10% of reading for 13cCs between 20% and 100% of full scale on any range | Radiation Type | Gamma, X-Ray | Battery Life | >100 hr. | Dose Range | Five linear ranges:
| Unit Size | Each |
|
| MICROSIEVERTAO | Full specifications | Each | Micro sievert with audio | Request A Quote |
Description | Micro sievert with audio | Shock | 100 g per lightweight machine of MIL-STD 202 C, method 202 B | Energy Range | 40 keV to 1.3 MeV | Display Type | Ruggedized, recessed, high-torque 1mA meter protected by impact resistant Lexan polycarbonate window with 85.1mm (3.35 in.) scale marked: MICRO REM: 0 to 200 μSv/h, with “Bat. ok”, “HV ok” check bands. MICRO SIEVERT: 0 to 2.0 μSv/h, with “Bat. ok”, “HV ok” check bands. | Battery Type | 29 V, MN1604, or equal | Product Type | Tissue-Equivalent Survey Meter | Operating Temperature (Metric) | −20°C to +50°C | Operating Temperature (English) | -4°F to +122°F | Response Time | Optimized for each range, 0-90% of final reading as follows: Range: : Time: X0.1 : <15 seconds X1 : <15 seconds X10 : <5 seconds X100 : <2 seconds X1000 : <2 seconds | Detector Type | NaI | Geotropism | Within ± 2% of full scale | Humidity | <5% change in reading from 10% to 95% | For Use With (Application) | Routine low/medium level gammadose rate surveysConfirming radiation boundariesMonitoring items for unrestricted releaseX-ray surveys, using the expanded low-energy range option | Vibration | 5 g in each of three mutually orthogonal axes at one or more frequencies from 10 to 33 Hz. | Accuracy | Within 10% of reading for 13cCs between 20% and 100% of full scale on any range | Radiation Type | Gamma, X-Ray | Battery Life | >100 hr. | Dose Range | Five linear ranges:
| Unit Size | Each |
|
| MICROSVAOLEED also known as MICROSV/AOLEED | Full specifications | Each | Micro sievert LE with audio, extended detector | Request A Quote |
Description | Micro sievert LE with audio, extended detector | Shock | 100 g per lightweight machine of MIL-STD 202 C, method 202 B | Energy Range | 40 keV to 1.3 MeV | Display Type | Ruggedized, recessed, high-torque 1mA meter protected by impact resistant Lexan polycarbonate window with 85.1mm (3.35 in.) scale marked: MICRO REM: 0 to 200 μSv/h, with “Bat. ok«, “HV ok” check bands. MICRO SIEVERT: 0 to 2.0 μSv/h, with “Bat. ok«, “HV ok” check bands. | Battery Type | 29 V, MN1604, or equal | Product Type | Tissue-Equivalent Survey Meter | Operating Temperature (Metric) | −20°C to +50°C | Operating Temperature (English) | -4°F to +122°F | Response Time | Optimized for each range, 0-90% of final reading as follows: Range: : Time: X0.1 : <15 seconds X1 : <15 seconds X10 : <5 seconds X100 : <2 seconds X1000 : <2 seconds | Detector Type | NaI | Geotropism | Within ± 2% of full scale | Humidity | <5% change in reading from 10% to 95% | For Use With (Application) | Routine low/medium level gammadose rate surveysConfirming radiation boundariesMonitoring items for unrestricted releaseX-ray surveys, using the expanded low-energy range option | Vibration | 5 g in each of three mutually orthogonal axes at one or more frequencies from 10 to 33 Hz. | Accuracy | Within 10% of reading for 13cCs between 20% and 100% of full scale on any range | Radiation Type | Gamma, X-Ray | Battery Life | >100 hr. | Dose Range | Five linear ranges:
| Unit Size | Each |
|
| MICROSVAOLE | Full specifications | Each | Micro sievert LE with audio | Request A Quote |
Description | Micro sievert LE with audio | Shock | 100 g per lightweight machine of MIL-STD 202 C, method 202 B | Energy Range | 40 keV to 1.3 MeV | Display Type | Ruggedized, recessed, high-torque 1mA meter protected by impact resistant Lexan polycarbonate window with 85.1mm (3.35 in.) scale marked: MICRO REM: 0 to 200 μSv/h, with “Bat. ok«, “HV ok” check bands. MICRO SIEVERT: 0 to 2.0 μSv/h, with “Bat. ok«, “HV ok” check bands. | Battery Type | 29 V, MN1604, or equal | Product Type | Tissue-Equivalent Survey Meter | Operating Temperature (Metric) | −20°C to +50°C | Operating Temperature (English) | -4°F to +122°F | Response Time | Optimized for each range, 0-90% of final reading as follows: Range: : Time: X0.1 : <15 seconds X1 : <15 seconds X10 : <5 seconds X100 : <2 seconds X1000 : <2 seconds | Detector Type | NaI | Geotropism | Within ± 2% of full scale | Humidity | <5% change in reading from 10% to 95% | For Use With (Application) | Routine low/medium level gammadose rate surveysConfirming radiation boundariesMonitoring items for unrestricted releaseX-ray surveys, using the expanded low-energy range option | Vibration | 5 g in each of three mutually orthogonal axes at one or more frequencies from 10 to 33 Hz. | Accuracy | Within 10% of reading for 13cCs between 20% and 100% of full scale on any range | Radiation Type | Gamma, X-Ray | Battery Life | >100 hr. | Dose Range | Five linear ranges:
| Unit Size | Each |
|
| MICROREMAOLEED | Full specifications | Each | LE version survey meter with audio, extended detector | Request A Quote |
Description | LE version survey meter with audio, extended detector | Shock | 100 g per lightweight machine of MIL-STD 202 C, method 202 B | Energy Range | 40 keV to 1.3 MeV | Display Type | Ruggedized, recessed, high-torque 1mA meter protected by impact resistant Lexan polycarbonate window with 85.1mm (3.35 in.) scale marked:MICRO REM: 0 to 200 µSv/h, with 'Bat. ok', 'HV ok' check bands. | Battery Type | 29 V, MN1604, or equal | Product Type | Tissue-Equivalent Survey Meter | Operating Temperature (Metric) | −20°C to +50°C | Operating Temperature (English) | -4°F to +122°F | Response Time | Optimized for each range, 0-90% of final reading as follows: Range: : Time: X0.1 : <15 seconds X1 : <15 seconds X10 : <5 seconds X100 : <2 seconds X1000 : <2 seconds | Detector Type | NaI | Geotropism | Within ± 2% of full scale | Humidity | <5% change in reading from 10% to 95% | For Use With (Application) | Routine low/medium level gammadose rate surveysConfirming radiation boundariesMonitoring items for unrestricted releaseX-ray surveys, using the expanded low-energy range option | Vibration | 5 g in each of three mutually orthogonal axes at one or more frequencies from 10 to 33 Hz. | Accuracy | Within 10% of reading for 13cCs between 20% and 100% of full scale on any range | Radiation Type | Gamma, X-Ray | Battery Life | >100 hr. | Dose Range | Five linear ranges:MICRO REM: X0.1: 0 to 20 µrem/h X1: 0 to 200 µrem/h X10: 0 to 2000 µrem/h X100: 0 to 20,000 µrem/h X1000: 0 to 200,000µrem/h | Unit Size | Each |
|
| MICROREMAO | Full specifications | Each | Tissue equivalent survey meter with audio capability | Request A Quote |
Description | Tissue equivalent survey meter with audio capability | Shock | 100 g per lightweight machine of MIL-STD 202 C, method 202 B | Energy Range | 40 keV to 1.3 MeV | Control Module Details | 8-Position rotary switch: Off; Bat. (checks the condition of the batteries); HV (Audits the detector high voltage); X1000, X100, X10, X1, Xo.1 (Counting ranges = multiply the meter reading by these figures) | Display Type | Ruggedized, recessed, high-torque 1mA meter protected by impact resistant Lexan polycarbonate window with 85.1mm (3.35 in.) scale marked:MICRO REM: 0 to 200 µSv/h, with 'Bat. ok', 'HV ok' check bands. | Battery Type | 29 V, MN1604, or equal | Product Type | Tissue-Equivalent Survey Meter | Operating Temperature (Metric) | −20°C to +50°C | Operating Temperature (English) | -4°F to +122°F | Response Time | Optimized for each range, 0-90% of final reading as follows: Range: : Time: X0.1 : <15 sec.; X1 : <15 sec.; X10 : <5 sec.; X100 : <2 X1000 : <2 sec. | Detector Type | NaI | Geotropism | Within ± 2% of full scale | Humidity | <5% change in reading from 10% to 95% | For Use With (Application) | Routine low/medium level gammadose rate surveysConfirming radiation boundariesMonitoring items for unrestricted releaseX-ray surveys, using the expanded low-energy range option | Vibration | 5 g in each of three mutually orthogonal axes at one or more frequencies from 10 to 33 Hz. | Accuracy | Within 10% of reading for 13cCs between 20% and 100% of full scale on any range | Radiation Type | Gamma, X-Ray | Battery Life | >100 hr. | Dose Range | X0.1: 0 to 20 µrem/h; X1: 0 to 200 µrem/h; X10: 0 to 2000 µrem/h; X100: 0 to 20,000 µrem/h; X1000: 0 to 200,000 µrem/h | Unit Size | Each |
|
| MICROREMAOLE | Full specifications | Each | Low energy version of Micro Rem survey meter with audio | Request A Quote |
Description | Low energy version of Micro Rem survey meter with audio | Shock | 100 g per lightweight machine of MIL-STD 202 C, method 202 B | Energy Range | 40 keV to 1.3 MeV | Display Type | Ruggedized, recessed, high-torque 1mA meter protected by impact resistant Lexan polycarbonate window with 85.1mm (3.35 in.) scale marked:MICRO REM: 0 to 200 µSv/h, with 'Bat. ok', 'HV ok' check bands. | Battery Type | 29 V, MN1604, or equal | Product Type | Tissue-Equivalent Survey Meter | Operating Temperature (Metric) | −20°C to +50°C | Operating Temperature (English) | -4°F to +122°F | Response Time | Optimized for each range, 0-90% of final reading as follows: Range: : Time: X0.1 : <15 seconds X1 : <15 seconds X10 : <5 seconds X100 : <2 seconds X1000 : <2 seconds | Detector Type | NaI | Geotropism | Within ± 2% of full scale | Humidity | <5% change in reading from 10% to 95% | For Use With (Application) | Routine low/medium level gammadose rate surveysConfirming radiation boundariesMonitoring items for unrestricted releaseX-ray surveys, using the expanded low-energy range option | Vibration | 5 g in each of three mutually orthogonal axes at one or more frequencies from 10 to 33 Hz. | Accuracy | Within 10% of reading for 13cCs between 20% and 100% of full scale on any range | Radiation Type | Gamma, X-Ray | Battery Life | >100 hr. | Dose Range | Five linear ranges:MICRO REM: X0.1: 0 to 20 µrem/h X1: 0 to 200 µrem/h X10: 0 to 2000 µrem/h X100: 0 to 20,000 µrem/h X1000: 0 to 200,000µrem/h | Unit Size | Each |
|
| MICROREMDS3 | Full specifications | Each | 3 year extended depot service for Micro Rem | Request A Quote |
Description | 3 year extended depot service for Micro Rem | Product Type | Tissue-Equivalent Survey Meter | Unit Size | Each |
|
| MICROREMAOED | Full specifications | Each | Survey meter with audio capability, extended detector | Request A Quote |
Description | Survey meter with audio capability, extended detector | Shock | 100 g per lightweight machine of MIL-STD 202 C, method 202 B | Energy Range | 40 keV to 1.3 MeV | Display Type | Ruggedized, recessed, high-torque 1mA meter protected by impact resistant Lexan polycarbonate window with 85.1mm (3.35 in.) scale marked:MICRO REM: 0 to 200 µSv/h, with 'Bat. ok', 'HV ok' check bands. | Battery Type | 29 V, MN1604, or equal | Product Type | Tissue-Equivalent Survey Meter | Operating Temperature (Metric) | −20°C to +50°C | Operating Temperature (English) | -4°F to +122°F | Response Time | Optimized for each range, 0-90% of final reading as follows: Range: : Time: X0.1 : <15 seconds X1 : <15 seconds X10 : <5 seconds X100 : <2 seconds X1000 : <2 seconds | Detector Type | NaI | Geotropism | Within ± 2% of full scale | Humidity | <5% change in reading from 10% to 95% | For Use With (Application) | Routine low/medium level gammadose rate surveysConfirming radiation boundariesMonitoring items for unrestricted releaseX-ray surveys, using the expanded low-energy range option | Vibration | 5 g in each of three mutually orthogonal axes at one or more frequencies from 10 to 33 Hz. | Accuracy | Within 10% of reading for 13cCs between 20% and 100% of full scale on any range | Radiation Type | Gamma, X-Ray | Battery Life | >100 hr. | Dose Range | Five linear ranges:MICRO REM: X0.1: 0 to 20 µrem/h X1: 0 to 200 µrem/h X10: 0 to 2000 µrem/h X100: 0 to 20,000 µrem/h X1000: 0 to 200,000µrem/h | Unit Size | Each |
|