Thermo Scientific™

Prima BT Bench Top Process Mass Spectrometer

Catalog number: PRIMABT
Thermo Scientific™

Prima BT Bench Top Process Mass Spectrometer

Catalog number: PRIMABT

Achieve high scanning speeds and reproducible measurements with the Thermo Scientific™ Prima BT Bench Top Process Mass Spectrometer. Specifically designed for process development laboratories, this real-time process gas analyzer is a highly precise, multi-port magnetic sector analyzer.

 
Catalog Number
PRIMABT
Unit Size
Each
Description
Prima BT Bench Top Process Gas Analysis Mass Spectrometer
Price (USD)
Full specifications
Ambient Temperature12 to 30 °C
Analysis TimeTypically 0.3 to 1.0 seconds per gas component (configurable by software)
DescriptionPrima BT Bench Top Process Gas Analysis Mass Spectrometer
Communication ProtocolsModbus serial RS232/422/485, Siemens 3964R, VGCP and PVGCP
Height (Metric)0.53 m (1.7 ft)
Mass RangeAdjustable, default is 1-150 amu at 1000 eV ion acceleration voltage, (at 750 eV ion acceleration voltage, mass range is 1-200 amu)
Max. No. of ComponentsNot limited by software
Max. No. of Peaks per ComponentNot limited by software
Power115/230 VAC, consumption 1.5 kVA
Precision≤ 0.1% relative (typical, application dependent)
Stability≤ 1% relative over one week (typical, application dependent)
Weight (Metric)100 kg (220 lb); additional floor mounting rotary pump 30 kg (66 lb)
Width (Metric)0.86 m (2.8 ft)
Area Classification OptionsGeneral Purpose
Analyzer TypeScanning Magnetic Sector Mass Spectrometer with Enclosed Electron Impact Dual Filament Ion Source
Dynamic RangeFaraday: 10 ppm – 100% (theoretical, application dependent)
Optional single MCP SEM: 1 ppm – 1000 ppm (theoretical, application dependent)
Optional twin MCP SEM: 10 ppb – 1000 ppm (theoretical, application dependent)
Depth0.56 m (1.8 ft)
No. of Sample Stream InletsUp to 15
No. of Calibration Gas InletsUp to 6
Unit SizeEach
Showing 1 of 1
Catalog NumberSpecificationsUnit SizeDescriptionPrice (USD)
PRIMABTFull specifications
EachPrima BT Bench Top Process Gas Analysis Mass SpectrometerRequest A Quote
Ambient Temperature12 to 30 °C
Analysis TimeTypically 0.3 to 1.0 seconds per gas component (configurable by software)
DescriptionPrima BT Bench Top Process Gas Analysis Mass Spectrometer
Communication ProtocolsModbus serial RS232/422/485, Siemens 3964R, VGCP and PVGCP
Height (Metric)0.53 m (1.7 ft)
Mass RangeAdjustable, default is 1-150 amu at 1000 eV ion acceleration voltage, (at 750 eV ion acceleration voltage, mass range is 1-200 amu)
Max. No. of ComponentsNot limited by software
Max. No. of Peaks per ComponentNot limited by software
Power115/230 VAC, consumption 1.5 kVA
Precision≤ 0.1% relative (typical, application dependent)
Stability≤ 1% relative over one week (typical, application dependent)
Weight (Metric)100 kg (220 lb); additional floor mounting rotary pump 30 kg (66 lb)
Width (Metric)0.86 m (2.8 ft)
Area Classification OptionsGeneral Purpose
Analyzer TypeScanning Magnetic Sector Mass Spectrometer with Enclosed Electron Impact Dual Filament Ion Source
Dynamic RangeFaraday: 10 ppm – 100% (theoretical, application dependent)
Optional single MCP SEM: 1 ppm – 1000 ppm (theoretical, application dependent)
Optional twin MCP SEM: 10 ppb – 1000 ppm (theoretical, application dependent)
Depth0.56 m (1.8 ft)
No. of Sample Stream InletsUp to 15
No. of Calibration Gas InletsUp to 6
Unit SizeEach
Showing 1 of 1
The Prima BT Process Mass Spectrometer offers robust performance in the presence of off-gases and vapors that have the potential to contaminate internal vacuum components. As the ions are extracted from the ion source at high energy, excellent stability is achieved for low molecular weight compounds such as hydrogen and helium.
 
Download the specification sheet located in the Resources tab above for more detailed information on this process gas analyzer.
 
Key Features
  • Scanning magnetic-sector technology
  • 16-port Rapid Multistream Sampler (RMS)
  • Six-port automatic calibration manifold
  • High analytical precision, accuracy and stability
  • Thermo Scientific™ GasWorks Software
Warranty and Service Offering
To maintain optimal product performance, you need immediate access to experts worldwide, as well as priority status when your analyzer needs repair or replacement. We offer comprehensive, flexible support solutions for all phases of the product life cycle. Through predictable, fixed-cost pricing, our services help protect the return on investment and total cost of ownership of your Thermo Scientific products.

Figures

Frequently asked questions (FAQs)

Citations & References