Thermo Scientific™ TLD-700 Thermoluminescent Dosimetry Material consists of Lithium Fluoride (Lithium isotope) Li:Mg, Ti. It is suitable for gamma and beta dosimetry applications.

 
Catalog Number
SNO74435
Unit Size
Each
Description
TLD-700 chip (6.35 x 6.35 x 0.89 mm)
Price (USD)
Full specifications
Radiation TypeBeta, Gamma
DescriptionTLD-700 chip (6.35 x 6.35 x 0.89 mm)
Emission Spectra3500 to 6000 Å (4000 maximum)
Energy Response1.25 keV/60Co
Height (Metric)0.89 mm
MaterialLithium Fluoride (7Li isotope) LiF:Mg, Ti
Sensitivity1.0 at 60Co relative to LiF
Width (Metric)6.35 mm
Product TypeThermoluminescent Dosimetry Material
Length (Metric)6.35 mm
Measurement Ranges10 pGy to 10 Gy
Unit SizeEach
Catalog Number
Unit Size
Each
Description
TLD-700 chip (3.2 x 3.2 x 0.15 mm)
Price (USD)
Full specifications
Radiation TypeBeta, Gamma
DescriptionTLD-700 chip (3.2 x 3.2 x 0.15 mm)
Emission Spectra3500 to 6000 Å (4000 maximum)
Energy Response1.25 keV/60Co
Height (Metric)0.15 mm
MaterialLithium Fluoride (7Li isotope) LiF:Mg, Ti
Sensitivity1.0 at 60Co relative to LiF
Width (Metric)3.2 mm
Product TypeThermoluminescent Dosimetry Material
Length (Metric)3.2 mm
Measurement Ranges10 pGy to 10 Gy
Unit SizeEach
Catalog Number
Unit Size
Each
Description
TLD-700 chip (3.2 x 3.2 x 0.38 mm)
Price (USD)
Full specifications
Radiation TypeBeta, Gamma
DescriptionTLD-700 chip (3.2 x 3.2 x 0.38 mm)
Emission Spectra3500 to 6000 Å (4000 maximum)
Energy Response1.25 keV/60Co
Height (Metric)0.38 mm
MaterialLithium Fluoride (7Li isotope) LiF:Mg, Ti
Sensitivity1.0 at 60Co relative to LiF
Width (Metric)3.2 mm
Product TypeThermoluminescent Dosimetry Material
Length (Metric)3.2 mm
Measurement Ranges10 pGy to 10 Gy
Unit SizeEach
Catalog Number
Unit Size
Each
Description
TLD-700 chip (3.2 x 3.2 x 0.89 mm)
Price (USD)
Full specifications
Radiation TypeBeta, Gamma
DescriptionTLD-700 chip (3.2 x 3.2 x 0.89 mm)
Emission Spectra3500 to 6000 Å (4000 maximum)
Energy Response1.25 keV/60Co
Height (Metric)0.89 mm
MaterialLithium Fluoride (7Li isotope) LiF:Mg, Ti
Sensitivity1.0 at 60Co relative to LiF
Width (Metric)3.2 mm
Product TypeThermoluminescent Dosimetry Material
Length (Metric)3.2 mm
Measurement Ranges10 pGy to 10 Gy
Unit SizeEach
Catalog Number
Unit Size
Each
Description
TLD-700 disk (Diameter: 3.6 x 0.38 mm)
Price (USD)
Full specifications
Radiation TypeBeta, Gamma
DescriptionTLD-700 disk (Diameter: 3.6 x 0.38 mm)
Emission Spectra3500 to 6000 Å (4000 maximum)
Energy Response1.25 keV/60Co
Height (Metric)0.38 mm
MaterialLithium Fluoride (7Li isotope) LiF:Mg, Ti
Sensitivity1.0 at 60Co relative to LiF
Product TypeThermoluminescent Dosimetry Material
Measurement Ranges10 pGy to 10 Gy
Diameter3.6 mm
Unit SizeEach
Catalog Number
Unit Size
Each
Description
TLD-700 disk (Diameter: 4.5 x 0.6 mm)
Price (USD)
Full specifications
Radiation TypeBeta, Gamma
DescriptionTLD-700 disk (Diameter: 4.5 x 0.6 mm)
Emission Spectra3500 to 6000 Å (4000 maximum)
Energy Response1.25 keV/60Co
Height (Metric)0.6 mm
MaterialLithium Fluoride (7Li isotope) LiF:Mg, Ti
Sensitivity1.0 at 60Co relative to LiF
Product TypeThermoluminescent Dosimetry Material
Measurement Ranges10 pGy to 10 Gy
Diameter4.5 mm
Unit SizeEach
Catalog Number
Unit Size
Each
Description
TLD-700 Pelletized chip (3.2 x 3.2 x 0.89 mm)
Price (USD)
Full specifications
Radiation TypeBeta, Gamma
DescriptionTLD-700 Pelletized chip (3.2 x 3.2 x 0.89 mm)
Emission Spectra3500 to 6000 Å (4000 maximum)
Energy Response1.25 keV/60Co
Height (Metric)0.89 mm
MaterialLithium Fluoride (7Li isotope) LiF:Mg, Ti
Sensitivity1.0 at 60Co relative to LiF
Width (Metric)3.2 mm
Product TypeThermoluminescent Dosimetry Material
Length (Metric)3.2 mm
Measurement Ranges10 pGy to 10 Gy
Unit SizeEach
Catalog Number
Unit Size
Each
Description
TLD-700 rod (Diameter: 1× 2 mm)
Price (USD)
Full specifications
Radiation TypeBeta, Gamma
DescriptionTLD-700 rod (Diameter: 1× 2 mm)
Emission Spectra3500 to 6000 Å (4000 maximum)
Energy Response1.25 keV/60Co
Height (Metric)2 mm
MaterialLithium Fluoride (7Li isotope) LiF:Mg, Ti
Sensitivity1.0 at 60Co relative to LiF
Product TypeThermoluminescent Dosimetry Material
Measurement Ranges10 pGy to 10 Gy
Diameter1 mm
Unit SizeEach
Catalog Number
Unit Size
Each
Description
TLD-700 rod (Diameter: 1× 6 mm)
Price (USD)
Full specifications
Radiation TypeBeta, Gamma
DescriptionTLD-700 rod (Diameter: 1× 6 mm)
Emission Spectra3500 to 6000 Å (4000 maximum)
Energy Response1.25 keV/60Co
Height (Metric)6 mm
MaterialLithium Fluoride (7Li isotope) LiF:Mg, Ti
Sensitivity1.0 at 60Co relative to LiF
Product TypeThermoluminescent Dosimetry Material
Measurement Ranges10 pGy to 10 Gy
Diameter1 mm
Unit SizeEach
Catalog Number
Unit Size
Each
Description
TLD-700 rod (Diameter: 1× 3 mm)
Price (USD)
Full specifications
Radiation TypeBeta, Gamma
DescriptionTLD-700 rod (Diameter: 1× 3 mm)
Emission Spectra3500 to 6000 Å (4000 maximum)
Energy Response1.25 keV/60Co
Height (Metric)3 mm
MaterialLithium Fluoride (7Li isotope) LiF:Mg, Ti
Sensitivity1.0 at 60Co relative to LiF
Product TypeThermoluminescent Dosimetry Material
Measurement Ranges10 pGy to 10 Gy
Diameter1 mm
Unit SizeEach
Showing 10 of 17Show more
Catalog NumberSpecificationsUnit SizeDescriptionPrice (USD)
SNO74435Full specifications
EachTLD-700 chip (6.35 x 6.35 x 0.89 mm)Request A Quote
Radiation TypeBeta, Gamma
DescriptionTLD-700 chip (6.35 x 6.35 x 0.89 mm)
Emission Spectra3500 to 6000 Å (4000 maximum)
Energy Response1.25 keV/60Co
Height (Metric)0.89 mm
MaterialLithium Fluoride (7Li isotope) LiF:Mg, Ti
Sensitivity1.0 at 60Co relative to LiF
Width (Metric)6.35 mm
Product TypeThermoluminescent Dosimetry Material
Length (Metric)6.35 mm
Measurement Ranges10 pGy to 10 Gy
Unit SizeEach
SNO78806Full specifications
EachTLD-700 chip (3.2 x 3.2 x 0.15 mm)Request A Quote
Radiation TypeBeta, Gamma
DescriptionTLD-700 chip (3.2 x 3.2 x 0.15 mm)
Emission Spectra3500 to 6000 Å (4000 maximum)
Energy Response1.25 keV/60Co
Height (Metric)0.15 mm
MaterialLithium Fluoride (7Li isotope) LiF:Mg, Ti
Sensitivity1.0 at 60Co relative to LiF
Width (Metric)3.2 mm
Product TypeThermoluminescent Dosimetry Material
Length (Metric)3.2 mm
Measurement Ranges10 pGy to 10 Gy
Unit SizeEach
SNO78815Full specifications
EachTLD-700 chip (3.2 x 3.2 x 0.38 mm)Request A Quote
Radiation TypeBeta, Gamma
DescriptionTLD-700 chip (3.2 x 3.2 x 0.38 mm)
Emission Spectra3500 to 6000 Å (4000 maximum)
Energy Response1.25 keV/60Co
Height (Metric)0.38 mm
MaterialLithium Fluoride (7Li isotope) LiF:Mg, Ti
Sensitivity1.0 at 60Co relative to LiF
Width (Metric)3.2 mm
Product TypeThermoluminescent Dosimetry Material
Length (Metric)3.2 mm
Measurement Ranges10 pGy to 10 Gy
Unit SizeEach
SNO78835Full specifications
EachTLD-700 chip (3.2 x 3.2 x 0.89 mm)Request A Quote
Radiation TypeBeta, Gamma
DescriptionTLD-700 chip (3.2 x 3.2 x 0.89 mm)
Emission Spectra3500 to 6000 Å (4000 maximum)
Energy Response1.25 keV/60Co
Height (Metric)0.89 mm
MaterialLithium Fluoride (7Li isotope) LiF:Mg, Ti
Sensitivity1.0 at 60Co relative to LiF
Width (Metric)3.2 mm
Product TypeThermoluminescent Dosimetry Material
Length (Metric)3.2 mm
Measurement Ranges10 pGy to 10 Gy
Unit SizeEach
SNP73615Full specifications
EachTLD-700 disk (Diameter: 3.6 x 0.38 mm)Request A Quote
Radiation TypeBeta, Gamma
DescriptionTLD-700 disk (Diameter: 3.6 x 0.38 mm)
Emission Spectra3500 to 6000 Å (4000 maximum)
Energy Response1.25 keV/60Co
Height (Metric)0.38 mm
MaterialLithium Fluoride (7Li isotope) LiF:Mg, Ti
Sensitivity1.0 at 60Co relative to LiF
Product TypeThermoluminescent Dosimetry Material
Measurement Ranges10 pGy to 10 Gy
Diameter3.6 mm
Unit SizeEach
SNP74524Full specifications
EachTLD-700 disk (Diameter: 4.5 x 0.6 mm)Request A Quote
Radiation TypeBeta, Gamma
DescriptionTLD-700 disk (Diameter: 4.5 x 0.6 mm)
Emission Spectra3500 to 6000 Å (4000 maximum)
Energy Response1.25 keV/60Co
Height (Metric)0.6 mm
MaterialLithium Fluoride (7Li isotope) LiF:Mg, Ti
Sensitivity1.0 at 60Co relative to LiF
Product TypeThermoluminescent Dosimetry Material
Measurement Ranges10 pGy to 10 Gy
Diameter4.5 mm
Unit SizeEach
SNP78835Full specifications
EachTLD-700 Pelletized chip (3.2 x 3.2 x 0.89 mm)Request A Quote
Radiation TypeBeta, Gamma
DescriptionTLD-700 Pelletized chip (3.2 x 3.2 x 0.89 mm)
Emission Spectra3500 to 6000 Å (4000 maximum)
Energy Response1.25 keV/60Co
Height (Metric)0.89 mm
MaterialLithium Fluoride (7Li isotope) LiF:Mg, Ti
Sensitivity1.0 at 60Co relative to LiF
Width (Metric)3.2 mm
Product TypeThermoluminescent Dosimetry Material
Length (Metric)3.2 mm
Measurement Ranges10 pGy to 10 Gy
Unit SizeEach
SNO70102Full specifications
EachTLD-700 rod (Diameter: 1× 2 mm)Request A Quote
Radiation TypeBeta, Gamma
DescriptionTLD-700 rod (Diameter: 1× 2 mm)
Emission Spectra3500 to 6000 Å (4000 maximum)
Energy Response1.25 keV/60Co
Height (Metric)2 mm
MaterialLithium Fluoride (7Li isotope) LiF:Mg, Ti
Sensitivity1.0 at 60Co relative to LiF
Product TypeThermoluminescent Dosimetry Material
Measurement Ranges10 pGy to 10 Gy
Diameter1 mm
Unit SizeEach
SNO70106Full specifications
EachTLD-700 rod (Diameter: 1× 6 mm)Request A Quote
Radiation TypeBeta, Gamma
DescriptionTLD-700 rod (Diameter: 1× 6 mm)
Emission Spectra3500 to 6000 Å (4000 maximum)
Energy Response1.25 keV/60Co
Height (Metric)6 mm
MaterialLithium Fluoride (7Li isotope) LiF:Mg, Ti
Sensitivity1.0 at 60Co relative to LiF
Product TypeThermoluminescent Dosimetry Material
Measurement Ranges10 pGy to 10 Gy
Diameter1 mm
Unit SizeEach
SNO70103Full specifications
EachTLD-700 rod (Diameter: 1× 3 mm)Request A Quote
Radiation TypeBeta, Gamma
DescriptionTLD-700 rod (Diameter: 1× 3 mm)
Emission Spectra3500 to 6000 Å (4000 maximum)
Energy Response1.25 keV/60Co
Height (Metric)3 mm
MaterialLithium Fluoride (7Li isotope) LiF:Mg, Ti
Sensitivity1.0 at 60Co relative to LiF
Product TypeThermoluminescent Dosimetry Material
Measurement Ranges10 pGy to 10 Gy
Diameter1 mm
Unit SizeEach
Showing 10 of 17Show more
Single element dosimeters and assemblies are widely used in many installations and processed using Harshaw TLD Systems. These systems include a range of readers and irradiators with compatible software to implement calibration, radiation evaluation and management, dose algorithms, health physics record keeping, finder database, glow curve analysis and chain of custody monitoring.
  • Nearly tissue-equivalent
  • ± 15% sample-to-sample uniformity
  • Repeatability to within 2% or better

Note:

Handling - Vacuum tweezers should always be used. (Avoid mechanical tweezers or fingers). Small scratches, loss of mass or foreign deposits affect light emission.

Annealing - For annealing temperatures up to 400°C, the containers should be made from high-temperature stainless steel or oxidized aluminum, preferably thin to assist rapid cooling following annealing. (Do not use non-oxidized aluminum). The use of a dedicated annealing oven reduces the risk of contamination by foreign material. Place the annealing containers on open oven racks with air space all round to avoid inconsistent heat gradients. (Do not stack containers or allow them to touch the oven walls).