| VH-D20-A | Full specifications | Each | Vanquish Charged Aerosol Detector H with Concentric Flow Nebulizer | Charged Aerosol Detector | 100-240 VAC, 50/60 Hz, max. 150 W/255 VA | 1 USB port (USB 2.0, “B” type) 1 USB hub with 3 ports (USB 2.0, “A” type) to connect further Vanquish modules | Request A Quote |
Analog Outputs | 0-1 V DC with full scale output range: 1 pA to 500 pA in 1-2-5 sequence (installable option) | Communications | 1 USB port (USB 2.0, “B” type) 1 USB hub with 3 ports (USB 2.0, “A” type) to connect further Vanquish modules | Description | Vanquish Charged Aerosol Detector H with Concentric Flow Nebulizer | Data Collection Rate | 200 Hz max | Detection Mode | Charged Aerosol Detection | Type | Charged Aerosol Detector | Filter | Filter Time Constant: 4 th order low-pass Bessel digital with time constraint based user selections: 0.1, 0.2, 0.5, 1.0, 2.0, 3.6, 5.0, 10.0 sec | For Use With (Equipment) | Vanquish UHPLC System | Gas Requirements | Nitrogen or Compressed Air (Nitrogen recommended) Inlet Pressure: 482-551 kPa (4.8-5.5 bar, 70-80 psi) | GLP Features | Predictive Performance functions: Gas filter change interval, service monitoring period, system parameters logged in the Chromeleon Chromatography Data System Audit Trail. | Inlet Gas Pressure | Electronically controlled pressure regulation system | PC Connection | 1 USB port (USB 2.0, “B” type) 1 USB hub with 3 ports (USB 2.0, “A” type) to connect further Vanquish modules | Range | Dynamic Range: Up to 4 orders of magnitude | Safety Features | Power-up diagnostics, inlet gas over-pressure, high evaporation temp, pump flow shutoff, fluid leak detection and safe leak handling | Temperature Control | Evaporation temperature: Settable Range: Ambient +5°C to +100°C | Dimensions (L x W x H) | 192 mm x 420 mm x 620 mm (7.6 in. x 16.5 in. x 24.4 in.) Weight: 17.5 kg (38.6 lb.) | Electrical Requirements | 100-240 VAC, 50/60 Hz, max. 150 W/255 VA | Unit Size | Each |
|
| VF-D20-A | Full specifications | Each | Vanquish Charged Aerosol Detector F with Concentric Flow Nebulizer | Charged Aerosol Detector | 100-240 VAC, 50/60 Hz, max. 150 W/255 VA | 1 USB port (USB 2.0, “B” type) 1 USB hub with 3 ports (USB 2.0, “A” type) to connect further Vanquish modules | Request A Quote |
Analog Outputs | 0-1 V DC with full scale output range: 1 pA to 500 pA in 1-2-5 sequence (installable option) | Communications | 1 USB port (USB 2.0, “B” type) 1 USB hub with 3 ports (USB 2.0, “A” type) to connect further Vanquish modules | Description | Vanquish Charged Aerosol Detector F with Concentric Flow Nebulizer | Data Collection Rate | 100 Hz max | Detection Mode | Charged Aerosol Detection | Type | Charged Aerosol Detector | Filter | Filter Time Constant: 4 th order low-pass Bessel digital with time constraint based user selections: 0.1, 0.2, 0.5, 1.0, 2.0, 3.6, 5.0, 10.0 sec | For Use With (Equipment) | Vanquish UHPLC System | Gas Requirements | Nitrogen or Compressed Air (Nitrogen recommended) Inlet Pressure: 482-551 kPa (4.8-5.5 bar, 70-80 psi) | GLP Features | Predictive Performance functions: Gas filter change interval, service monitoring period, system parameters logged in the Chromeleon Chromatography Data System Audit Trail. | Inlet Gas Pressure | Electronically controlled pressure regulation system | PC Connection | 1 USB port (USB 2.0, “B” type) 1 USB hub with 3 ports (USB 2.0, “A” type) to connect further Vanquish modules | Range | Dynamic Range: Up to 4 orders of magnitude | Safety Features | Power-up diagnostics, inlet gas over-pressure, high evaporation temp, pump flow shutoff, fluid leak detection and safe leak handling | Temperature Control | Evaporation temperature: Select: +35°C, +50°C, +70°C | Dimensions (L x W x H) | 192 mm x 420 mm x 620 mm (7.6 in. x 16.5 in. x 24.4 in.) Weight: 17.5 kg (38.6 lb.) | Electrical Requirements | 100-240 VAC, 50/60 Hz, max. 150 W/255 VA | Unit Size | Each |
|