Especificações
AccuracyWithin 10% of reading for 13cCs between 20% and 100% of full scale on any range
Battery Life>100 hr.
Control Module Details8-Position rotary switch: Off; Bat. (checks the condition of the batteries); HV (Audits the detector high voltage); X1000, X100, X10, X1, Xo.1 (Counting ranges = multiply the meter reading by these figures)
DescriptionTissue Equivalent Survey Meter
Detector TypeNaI
Display TypeRuggedized, recessed, high-torque 1mA meter protected by impact resistant Lexan polycarbonate window with 85.1mm (3.35 in.) scale marked:MICRO REM: 0 to 200 µSv/h, with 'Bat. ok', 'HV ok' check bands.
Dose RangeX0.1: 0 to 20 µrem/h; X1: 0 to 200 µrem/h; X10: 0 to 2000 µrem/h; X100: 0 to 20,000 µrem/h; X1000: 0 to 200,000 µrem/h
For Use With (Application)Routine low/medium level gammadose rate surveysConfirming radiation boundariesMonitoring items for unrestricted releaseX-ray surveys, using the expanded low-energy range option
GeotropismWithin ± 2% of full scale
Humidity<5% change in reading from 10% to 95%
IncludesAudio Capability
Operating Temperature (English)-4°F to +122°F
Operating Temperature (Metric)−20°C to +50°C
Response TimeOptimized for each range, 0-90% of final reading as follows: Range: : Time: X0.1 : <15 sec.; X1 : <15 sec.; X10 : <5 sec.; X100 : <2 X1000 : <2 sec.
Shock100 g per lightweight machine of MIL-STD 202 C, method 202 B
Vibration5 g in each of three mutually orthogonal axes at one or more frequencies from 10 to 33 Hz.
Battery Type29 V, MN1604, or equal
Energy Range40 keV to 1.3 MeV
Product TypeTissue-Equivalent Survey Meter
Radiation TypeGamma, X-Ray
Unit SizeEach