Columna preempaquetada MabCaptureC™ MiniChrom™
Thermo Scientific™

Columna preempaquetada MabCaptureC™ MiniChrom™

Las columnas Thermo Scientific™ MabCaptureC™ MiniChrom™ están diseñadas para el cribado de resina a escala de laboratorio, el desarrollo de procesos y la preparación de muestras.
Have Questions?
Cambiar vistabuttonViewtableView
Número de catálogoVolumen del lecho
59436620055 mL
59436620011 mL
Número de catálogo 5943662005
Precio (CLP)
1.863.254
Each
Volumen del lecho:
5 mL
Precio (CLP)
1.863.254
Each

Las columnas MabCaptureC™ MiniChrom™ están diseñadas para el cribado de resina a escala de laboratorio, el desarrollo de procesos y la preparación de muestras.

These pre-packed columns are ready to connect and have the following features:

  • Packed with MabCaptureC™ Affinity Matrix, offering high dynamic binding capacity and high resolution over a wide range of flow rates
  • Direct connection to standard chromatography systems

The 1 mL columns can be used for quick screenings of application feasibility and lab scale purification on a convenient and easy-to-use pre-packed column. The 10-cm bed height of the 5 mL column allows initial process development at a bench scale.

The pre-packed MabCaptureC™ Affinity Matrix contains engineered rProtein A that binds at the CH2-CH3 interface of the Fc-region of IgGs and shows cross binding to the VH3 domain of IgGs.

Features of MabCaptureC™ Affinity Matrix include:

  • High dynamic mAb binding capacity of > 50 g/L at 4.8 min residence time
  • Purification of mAbs and a subpopulation of scFv's and Fab fragments containing VH3 binding domains
  • Clean and sanitize with standard sodium hydroxide agents for chromatography products
  • Excellent scalability
  • Non-animal derived

Main resin characteristics:

  • Matrix: agarose-based, epoxide activated
  • Average particle size: 75 ± 10 μm
  • Ligand: recombinant C domain of Protein A
  • Ligand coupling method: epoxide coupling
  • Binding capacity: >55 g IgG/L resin at 5 min residence time and 10% breakthrough; >75 g IgG/L resin at 10 min. residence time and 10% breakthrough
  • Elution conditions: 0.1 M glycine pH 3.0, or 20–100 mM sodium citrate or acetate, pH 3.0
  • Cleaning conditions: 0.2 M sodium hydroxide for 15 minutes every cycle
  • Column dimensions: 0.8 x 2 cm (1 mL), 0.8 x 10 cm (5 mL)
  • Operating pressure: ≤ 3 bar
  • Formulation buffer: 20% (v/v) ethanol
Para uso exclusivo en investigación. No apto para uso en procedimientos diagnósticos.
Especificaciones
Volumen del lecho5 mL
Capacidad de unión>55 g IgG/L resin (5 min. residence time and 10% breakthrough), >75 g IgG/L resin (10 min. residence time and 10% breakthrough)
Diámetro (métrico)0.8 cm
Longitud (métrico)10 cm
Tipo de ligandorProteína A modificada
Matrizbasado en agarosa, activado con epóxido
Presión máx. de funcionamiento≤3 bar
Cantidad5 mL
Fase estacionariaGránulos de agarosa
ObjetivoAcM IgG y una subpoblación de scFv y Fab que contienen el dominio VH3
Tamaño de partículas75 ±10 μm
Unit SizeEach