FHT 762 Wendi-2 Wide-Energy Neutron Detector
FHT 762 Wendi-2 Wide-Energy Neutron Detector
Thermo Scientific™

FHT 762 Wendi-2 Wide-Energy Neutron Detector

Features high sensitivity and an excellent energy and angular response provided by the WENDI-2 design originally developed at Los Alamos.
Have Questions?
Número de catálogo FHT762WENDI2
Precio (CLP)
-

Thermo Scientific™ FHT 762 Wendi-2 Wide-Energy Neutron Detector features high sensitivity and an excellent energy and angular response provided by the WENDI-2 design originally developed at Los Alamos.

Fits to FHT 6020 area monitor and FH 40 G survey meter.

Combines excellent energy response in the “normal” energy range up to 15 MeV with a close match to H™(10) behavior up to 5 GeV.
  • Thermal to 5 GeV energy range according to H™(10), ICRP 74
  • High sensitivity due to large He-3 tube
  • Excellent gamma rejection
  • Design was originally developed at Los Alamos
  • No significant spill-over needs to be considered for gamma dose rates up to 1 Sv/h

Compatible with:

FHT 6020 area monitor, FH 40 G survey meter

Especificaciones
Certificaciones/conformidadICRP 74 / FHT 6020 area monitor, FH 40 G survey meter
DescripciónFHT 762 Wendi-2
Diámetro (imperial)9 in.
Diámetro (métrico)320 mm
Altura (imperial)12.6 in.
HumedadUp to 90%, Non-Condensing
Linealidad±20% (Angular dependence: ±20% in all directions)
Temperatura de funcionamiento (métrico)−30°C to 50°C
Presión500 to 1500 hPa
SensibilidadSensitivity: 0.84 cps/(µSv/h) Cf-252Gamma sensitivity: 1 to 5µSv/h at 100 mSv/h, 662keV
Peso (imperial)29.8 lb.
Intervalo de energía25 meV to 5 GeV
Altura (métrico)230 mm
Peso (métrico)13.5 kg
Unit SizeEach