UltiMate™ 3000 Fluorescence Detectors
Thermo Scientific™

UltiMate™ 3000 Fluorescence Detectors

Develop highly selective methods with sensitivity three to six orders of magnitude greater than UV detection with this fluorescence detector.
Have Questions?
變更視圖buttonViewtableView
產品號碼Detection ModeModel
5078.0015Single ChannelFLD-3100 w/ Dual PMT
5078.00204 ChannelsFLD-3400 RS
5078.0010Single ChannelFLD-3100
產品號碼 5078.0015
價格 (HKD)
-
Detection Mode:
Single Channel
Model:
FLD-3100 w/ Dual PMT

Develop highly selective methods for trace analytes or novel molecules with sensitivity three to six orders of magnitude greater than UV detection. The Thermo Scientific™ Dionex™ UltiMate™ 3000 Fluorescence Detectors offer four channel acquisition at up to 200Hz so you can optimize quantification on the most sensitive channel and single-channel detection at up to 100Hz. 2D or 3D synchro, excitation, emission scans, zero order mode, or multichannel acquisition offer flexibility for method development or routine sample characterization. A variable emission filter allows real-time, compound-related sensitivity optimization.

Recommended Use

The Thermo Scientific™ UltiMate 3000 Rapid Separation Fluorescence Detector FLD-3400 is recommended for method development labs that need flexibility in detecting novel compounds that don’t absorb in the UV-vis region of the spectrum. It is ideal for the analysis of carbohydrates, glycans, and polyaromatic hydrocarbons (PAHs) in combination with any Thermo Scientific™ Dionex™ UltiMate 3000 system. The ability to cleave and derivitize glycans from intact proteins or monoclonal antibodies eliminates interferences from coeluting matrix peptides.

Sensitive Fluorescence Performance

  • Data rates at up to 200Hz detect the narrowest peaks in UHPLC separations.
  • Maximum stray light suppression provides excellent sensitivity.
  • Low limits of detection are possible with a Raman S/N: >550 ASTM (>2100 using dark signal as noise reference).
  • Active flow cell temperature control ensures stable fluorophore activity independent of ambient temperature.
  • Long-life xenon flash lamp delivers the highest sensitivity and long-term operation.
  • Optional second photomultiplier (PMT) for Dual-PMT operation extends the wavelength range up to 900nm without sacrificing sensitivity in the standard wavelength range.

Easy-to-Use Fluorescence Detector

  • Large front panel display allows easy monitoring of detector status.
  • Flow cell syringe injection kit enables cost-effective off-line measurements.
  • Flow cell optimized for trace analysis with narrow-bore UHPLC and conventional HPLC columns.
  • Automate qualification monitoring and tracking of flow cells for regulatory compliance.
  • Thermo Scientific™ Dionex™ Chromeleon™ Chromatography Data System software offers performance optimization and instrument wellness monitoring for ease of use and maximum uptime.
規格
Descriptionwith Dual-PMT
Detection ModeSingle Channel
DisplayLCD Monitor
FilterFixed
Flow CellWithout Flow Cell
Flow Cell Max. Pressure290 psi
Flow Cell Thermostatting15°C Above Ambient -50°C
Flow Cell Volume8 μL
GLP FeaturesAutomatic Instrument Qualification (AutoQ), System Wellness and Qualification Monitoring with Chromeleon Software, Lamp and Cell ID Chips.
LampXenon Flash Lamp
Lamp Pulse Frequency300 Hz
ModelFLD-3100 w/ Dual PMT
No. of Channels1
PC ConnectionUSB
Sensitivity>550 ASTM
Spectral Bandwidth20 nm
Temperature ControlYes
Wavelength Accuracy±2 nm
Wavelength Range200 to 880 nm
Weight (Metric)18 kg
Data Collection Rate100 Hz Max.
Detector TypeHigh-Sensitivity Fluorescence
TypeFluorescence Detector
Unit SizeEach