KingFisher™ Flex 24 Deep Well head
KingFisher™ Flex 24 Deep Well head
Thermo Scientific™

KingFisher™ Flex 24 Deep Well head

This is a magnetic head for the KingFisher™ Flex Magnetic Particle Purification System (Cat. Nos. 5400630 and 5400640). This magnetic자세히 알아보기
Have Questions?
카탈로그 번호수량
24074440Each
카탈로그 번호 24074440
제품 가격(KRW)
-
수량:
Each
This is a magnetic head for the KingFisher™ Flex Magnetic Particle Purification System (Cat. Nos. 5400630 and 5400640). This magnetic head can serve as a replacement head for the 24 deep-well format system or a supplemental head for the 96 deep-well format system.
For Research Use Only. Not for use in diagnostic procedures.
사양
용도(장비)KingFisher™ Flex
제품라인KingFisher
수량Each
배송 조건Room Temperature
용량24 Samples
형식24 Deep-well Head
Unit SizeEach

자주 묻는 질문(FAQ)

I see that the KingFisher 24 Deep Well Head has a processing volume of 5 mL. Does that mean that I can process up to 5 mL of my sample?

No, the processing volume refers to the maximum volume that can be held in the 24 Deep Well plates. This incorporates the volume of binding buffers and beads, so when scaling up reaction volumes, it is important to make sure that the total volume does not exceed 5 mL.

My KingFisher Flex run was stopped due to an error message. Can I start the protocol from where it stopped?

No, once a protocol was stopped, it will restart from the beginning. The extraction procedure can either be finished manually by using a magnetic stand, or a modified version of the original KingFisher Flex script would have to be created that starts off at the point where the run was stopped.

I have a 12-plate protocol but there are only 8 slots on the KingFisher Flex instrument turntable. What should I do with the remaining 4 plates?

Set up the protocol on the BindIt software. After the first 8 plates are done, the instrument will pause and instruct you to switch out the plates.

Is it always necessary to use the transport locks on the KingFisher Flex instrument?

The transport locks are only necessary when relocating the instrument.

Are the volumes of reagents in each well critical for a successful KingFisher Flex run?

We strongly recommend that you maintain the specified volumes within the defined limits to avoid spillover and to help ensure the best performance at the most efficient level.