이 유리 시린지는 POP-4™ Performance Optimized Polymer, POP-6™ Performance Optimized Polymer, GeneScan™ polymer로 310 Genetic Analyzer capillary를 충진하는데 적합합니다.호환 제품310자세히 알아보기
Have Questions?
카탈로그 번호수량
43044711 syringe
카탈로그 번호 4304471
제품 가격(KRW)
254,000
Each
카트에 추가하기
수량:
1 syringe
제품 가격(KRW)
254,000
Each
카트에 추가하기
이 유리 시린지는 POP-4™ Performance Optimized Polymer, POP-6™ Performance Optimized Polymer, GeneScan™ polymer로 310 Genetic Analyzer capillary를 충진하는데 적합합니다.

호환 제품
310 Genetic Analyzer Capillary, 47 cm (36 cm wtr), 310 Running Buffer, 10X, BigDye™ Terminator v1.1 Cycle Sequencing Kit, BigDye™ Terminator v3.1 Cycle Sequencing Kit, Genetic Analyzer Sample Tubes, 0.5 ml, Hi-Di™ Formamide, 3100⁄3100-Avant™ Genetic Analyzers용 POP-4™ Polymer, 310 Genetic Analyzer용 POP-4™ Polymer, 310 Genetic Analyzer용 POP-6™ Polymer, 0.5 ml Sample Tubes용 Septa.

연구용으로만 사용할 수 있습니다. 진단 절차에 사용할 수 없습니다.
For Research Use Only. Not for use in diagnostic procedures.
사양
제품 유형Syringe
수량1 syringe
용도 (장비)310 GeneticAnalyzer
물질Glass
부피(미터법)1 mL
Unit SizeEach

자주 묻는 질문(FAQ)

How often should I change my polymer and buffer? 

Polymer in the syringe should be changed every three days. Do not return unused polymer to the bottle.

Buffer and water should be changed every 48 hours or, if the run lasts for more than 48 hours, after the run is complete.

Find additional tips, troubleshooting help, and resources within our Capillary Electrophoresis Instruments Support Center.

How often do I really need to clean the pump block?

The pump block needs to be cleaned when:

Installing a syringe.

Switching applications (i.e., from sequencing to fragment analysis).

Shutting down the instrument for a long period of time.

If the instrument has run for more than 4 days without any of the above occurring.

Appearance of unusual spikes in the electropherogram.

If the syringe containing the polymer has been on the instrument for more than a week, it is important to clean the syringe and the block, and to change the polymer. At room temperature, enough urea decomposition can occur in one week to cause transient current increases during electrophoreses. If polymer is allowed to dry in the pump block channel, the pump block may be ruined.

Find additional tips, troubleshooting help, and resources within our Capillary Electrophoresis Instruments Support Center.

What happens to the 310 Genetic Analyzer if it runs out of polymer during the run?

A message will appear at the start of the run telling you if you have enough polymer to complete the run. If you should continue without adding more polymer, then the run will be cancelled at the injection where it ran out and an entry will be made into the run log that there is no polymer. The following injections will appear to run for 1–2 minutes each and will be recorded into the run log. No data will be collected for these injections.

Find additional tips, troubleshooting help, and resources within our Capillary Electrophoresis Instruments Support Center.