Thermo Scientific™

Immuno Standard Module White and Black

Thermo Scientific™ Immuno Standard Module White and Black을 사용해 형광/발광 면역 분석법 및 결합 분석법을 수행할 수 있습니다.
Have Questions?
보기 방식 변경buttonViewtableView
카탈로그 번호색상부피(미터법) 웰표면 처리
437915Clear350 μLMaxiSorp
437591White350 μLMaxiSorp
437702White350 μLPolySorp
475515Black400 μLMaxiSorp
475523Black400 μLPolySorp
7566White380 μLHydrophobic
카탈로그 번호 437915
제품 가격(KRW)
924,000
Case of 60
카트에 추가하기
색상:
Clear
부피(미터법) 웰:
350 μL
표면 처리:
MaxiSorp
제품 가격(KRW)
924,000
Case of 60
카트에 추가하기
Thermo Scientific™ Immuno Standard Module White and Black을 사용해 형광/발광 면역 분석법 및 결합 분석법을 수행할 수 있습니다.
특징:
  • 8웰, 12웰 또는 16웰 스트립으로 제공
  • 최적의 분석을 위해 프레임이 있거나 없는 형태 선택
  • 편평 바닥 또는 C형 웰
  • 최적의 결합을 위한 세 가지 표면 옵션
  • 낮은 배경 형광으로 가능한 높은 신호 반사를 제공하는 흰색 형식(형광 관찰에 권장)
  • 형광 관찰 시 발생할 수 있는 후방 산란광을 줄이는 검정색 형식
  • 시간 분해 형광 분야에서만 사용되는 투명 형식
  • 표면 옵션:

  • Thermo Scientific™ PolySorp™ 및 Microlite™ 1+는 소수성 분자를 흡착하는 데 사용됩니다.
  • Microlite 2+는 약간의 친수성을 가지며 다양한 생체 분자와 결합합니다.
  • Thermo Scientific™ MaxiSorp™는 친수성이며 항체 샌드위치 분석에 이상적입니다.

  • 책임 보증 기간: 판매 약관에 제시된 모든 보증 요건 및 배제 조항에 따라 달라질 수 있으나 이 제품의 보증 기간은 제품을 선적한 날짜부터 이후 구십(90)일입니다.

    내부 연구 목적 사용 제한에 대한 면책 사항: 내부 연구 목적만을 위해 제품을 이용하는 구매자에 대한 어떠한 제한 사항도 이 문구와 관련된 웹 페이지의 제품에 적용되지 않습니다.

    사양
    결합 특성Hydrophilic
    분리No
    인증Yes
    검출 방법Colorimetric
    용도(애플리케이션)ELISA
    케이스당 수60
    팩당 수5
    포장60 Cases
    플레이트 블로킹None
    제품 유형Microplate Strip
    수량60/Cs.
    타겟 분자Antibodies (Glycoproteins)
    부피(미터법) 작동250 μL
    어레이8 x 12
    색상Clear
    형식Standard
    물질Polystyrene
    웰 수96
    멸균Non-sterile
    표면 처리MaxiSorp
    부피(미터법) 웰350 μL
    웰 형상Round, C Bottom
    Unit SizeCase of 60

    자주 묻는 질문(FAQ)

    What length of peptide is ideal for binding to the Thermo Scientific MaxiSorp surface? What are the detection limitations?

    We have tested and found that a 3 amino acid peptide (Pro, Leu, Gly) cannot be detected when passively adsorbed on the MaxiSorp surface. However, this peptide can be detected when covalently immobilized using CovaLink NH Modules and CovaLink NH2 Modules and Plates. Using covalent immobilization of small peptide residues, you can typically obtain a better orientation of the molecule and reduced problems with antibody recognition of the peptide due to masking of the epitope. We have discovered that a 7 amino acid peptide from the MHC Class II antigen can be detected when adsorbed on the MaxiSorp surface. We state that the detection limitation using the MaxiSorp surface is between 3 and 7 amino acid residues.
    One additional note is that detection is contingent upon the orientation of the peptide when immobilized. If the active site is inactivated or hidden at the site facing the solid phase, no detection signal is observed.

    Can I bind either single- or double-stranded DNA to the Thermo Scientific MaxiSorp surface?

    Single-stranded DNA can be adsorbed to our MaxiSorp surface using approximately 10 µg ssDNA per mL PBS, pH 8.2, although the stability is uncertain. Based on our experience, ssDNA immobilized on the MaxiSorp surface is so loosely bound that it can be removed by stringent washing.
    Double-stranded DNA will not bind to the MaxiSorp surface. DNA, however, can be covalently bound to Nunc NucleoLink Strips.

    For Nunc Immuno plates and modules, what are the advantages of one well geometry type over another? Which types should I use for various applications?

    The following list describes the geometries of wells available for Nunc Immuno-plates and modules:
    - Flat bottom (F): Allows maximum transmission of light. These plates can be read on a monochromatic reader.
    - Round bottom (U): This geometry optimizes washing and coating. These plates can be read using a dual wavelength reader.
    - "C" bottom (C): This design of the well is a combination of both flat and round bottoms. Basically, it is a flat bottomed well with curved edges at the bottom. These plates also can be read using a monochromatic reader combining optimal reading and washing.
    - StarWell: These wells have a modified "C" shape geometry with eight fins strategically placed at the bottom. This feature increases surface area, allowing more molecules to become immobilized which reduces incubation times.

    What is the difference between certified and non-certified MaxiSorp plates and modules?

    Both of these surfaces are identical. The only difference between them is that for the certified plates, a representative sample from each manufacturing lot undergoes a Binding Capacity test. This test is an ELISA-like assay used in our quality control laboratories to ensure binding capabilities.

    How can I reduce high background readings and non-specific binding, when performing ELISA using Nunc-Immuno MaxiSorp Plates?

    Assay sensitivity depends strongly on efficient removal of non-specific reacting molecules. High background readings and coating instability can be minimized by addition of a blocking step after the first coating. The excess surface is then occupied by indifferent molecules.
    We recommend washing three times after each coating step by using a solution of 0.15 M phosphate buffer (pH 7.2) with 0.2 M NaCl and 0.05% Tween 20.
    For blocking, we recommend using 0.5% BSA, 1% casein or 1% gelatin in 0.15 M phosphate buffer (pH 8.2) or carbonate buffer (pH 9.6).