Septa Cathode Buffer Container, for 3500 and SeqStudio™ Flex
Applied Biosystems™
Septa Cathode Buffer Container, for 3500 and SeqStudio™ Flex
이는 3500 series Genetic Analyzers을 사용하기 위해 Cathode Buffer Container의 양쪽 구획에 사용하는 septa입니다.이 제품은 연구용으로만 사용가능합니다. 진단 절차에 사용할자세히 알아보기
Have Questions?
카탈로그 번호
수량
4410715
10 Containers
카탈로그 번호 4410715
제품 가격(KRW)
822,000
Each
카트에 추가하기
수량:
10 Containers
제품 가격(KRW)
822,000
Each
카트에 추가하기
이는 3500 series Genetic Analyzers을 사용하기 위해 Cathode Buffer Container의 양쪽 구획에 사용하는 septa입니다.
이 제품은 연구용으로만 사용가능합니다. 진단 절차에 사용할 수 없습니다.
For Research Use Only. Not for use in diagnostic procedures.
사양
용도(장비)3500 Series Genetic Analyzers, SeqStudio Flex Series Genetic Analyzers
제품 유형Septa Cathode Buffer Container
수량10 Containers
배송 조건Room Temperature
Unit SizeEach
구성 및 보관
This product contains 10 septa.
자주 묻는 질문(FAQ)
When I started my run on the Applied Biosystems 3500/3500xL Genetic Analyzer, an error occurred: "Unstable Electrophoresis current detected, check for air bubbles" but I don't see any bubbles. What else can cause this?
Some of the other causes of an Unstable Electrophoresis current detected error message are:
1.Leak on the system
2. Polymer that:
-Has expired
-Has been left on the instrument for more than the recommended time
-Is a mixture of expired polymer and non-expired polymer
3. Running Buffer that was:
-On the instrument longer than 2 weeks
-Not filled to the fill line or evaporated below the fill line
4. An arcing event that was not cleaned afterwards using the water wash wizard
5. Not performing regular maintenance on the instrument
6. Hardware issues
Inspect the system for leaks. If you do not see any leaks on the system, perform the wash the pump chamber and channels wizard using the conditioning pouch and place fresh, non-expired polymer and fresh Anode and Cathode buffer containers. If the problem persists, a service call may be required.
Do I need to replace the buffer every 14 days if the Applied Biosystems 3500/3500xL Genetic Analyzer is not in use?
If the instrument is not in use, it is not necessary to replace the buffer every 14 days. However, the buffer level needs to remain at the fill line to prevent the capillary or array from drying out. Top off the buffer volume with water if necessary. Replace the anode and cathode buffer containers prior to starting a new run.