Megaplex™ Primer Pools, Human Pools Set v3.0
Product Image
Applied Biosystems™

Megaplex™ Primer Pools, Human Pools Set v3.0

Green features
Megaplex™ Primer Pools, Human Pools Set v3.0는 Megaplex™ RT Primers Human Pool A v2.1, Megaplex™ RT Primers Human Pool B자세히 알아보기
Have Questions?
카탈로그 번호수량
444475050 reactions
카탈로그 번호 4444750
제품 가격(KRW)
2,506,000
Each
카트에 추가하기
수량:
50 reactions
제품 가격(KRW)
2,506,000
Each
카트에 추가하기
Megaplex™ Primer Pools, Human Pools Set v3.0는 Megaplex™ RT Primers Human Pool A v2.1, Megaplex™ RT Primers Human Pool B v3.0, Megaplex™ PreAmp Primers, Human Pool A v2.1 Megaplex™ PreAmp Primers, Human Pool B v3.0를 포함한 하나의 제품 번호로 이용할 수 있습니다. Megaplex™ Primer Pools, Human Pools Set v3.0는 754개 고유한 microRNAs의 RT primer에 4개의 control을 더 포함합니다. 분석작업을 하실때 sensitivity가 가장 중요한 경우, Megaplex™ PreAmp Primer가 저발현 microRNAs 검출력을 유의하게 증강시켜 1 ng 정도의 적은 총 RNA 주입으로 종합적인 발현 프로필을 생성할 수 있습니다. Megaplex™ Primer Pools, Human Pools Set v3.0은 TaqMan® Array Human MicroRNA Card Set v3.0에 사용하도록 만들어졌으나 개별 TaqMan® MicroRNA Assays를 사용한 실시간 분석용 cDNA를 준비하는데도 사용할 수 있습니다.
For Research Use Only. Not for use in diagnostic procedures.
사양
그린 기능Sustainable packaging
반응 수50 Reactions
프라이머RT Primers, PreAmp Primers
제품라인Megaplex
제품 유형Primer Pool
수량50 reactions
Human
충분50 Reactions
용도(애플리케이션)miRNA analysis
GC-Rich PCR PerformanceLow
PCR 방법qPCR
Unit SizeEach
구성 및 보관
Store at -20°C

자주 묻는 질문(FAQ)

Can Ct's greater than a cut-off be considered valid results?

Yes they can. However, it is important to recognize the true linearity and detection limits of your assay: Ct values above the cut-off can indicate non-specific amplification, unless your NTC is a true- no-target control, and you have run a statistically significant number of replicates. Any results with Ct above the recommended cut-off need to be validated with individual assays on plates.

What is a good Ct cut-off for the TaqMan MicroRNA Array Cards and TaqMan Advanced miRNA Array Cards? In other words, beyond what Ct should I not trust the data?

The typical Ct cut-off on TaqMan Array Cards is 32, which is equivalent to Ct 35 on a plate (10 µl reaction). Previous studies show that if you use pre-amplification, a Ct cut-off of 29 or 30 can be used to reduce numbers of false positives (see Technical Note Optimized protocols for human or rodent microRNA profiling with precious samples). To ensure that you have selected a correct cut-off, you should run replicates of the same sample and use Ct cut-off before you see an increase in the Standard Deviation.

Why are the Ct values for my endogenous controls are highly variable across the sample set?

Highly variable values for endogenous controls is most likely due to biological variation. Use an alternative endogenous control such as a non-variable miRNA.

Why is the Ct value for the No Template Control (NTC) <35 for some assays?

Please review the following possible causes and solutions:
- There are non-specific interactions between primers. For NTC information on a specific assay, refer to the Megaplex Assay Performance File.
- The cDNA template is contaminated. Please follow established PCR good laboratory practices.
- The preamplification product was not diluted properly before real-time PCR. Follow the recommendations in the protocol. Please see "Dilute the preamplification reaction" on page 18.

Why did I not detect any miRNA?

It is possible that the threshold is set too high to detect miRNA in samples with low expression. Identify an appropriate NOAMP flag threshold if the relative threshold algorithm (Crt) is used. Alternatively, identify an appropriate threshold if Ct is used.