Ion OneTouch™ DL Upgrade Kit
Product Image
Ion Torrent™

Ion OneTouch™ DL Upgrade Kit

Ion OneTouch™ DL Upgrade Kit는 Ion OneTouch™ System에 한 번 'DL' 업그레이드하여 Ion OneTouch™ 200 Template Kit v2 DL를 사용할자세히 알아보기
Have Questions?
카탈로그 번호수량
44745231 Kit
카탈로그 번호 4474523
제품 가격(KRW)
-
수량:
1 Kit
Ion OneTouch™ DL Upgrade Kit는 Ion OneTouch™ System에 한 번 'DL' 업그레이드하여 Ion OneTouch™ 200 Template Kit v2 DL를 사용할 수 있게 합니다. Ion OneTouch™ DL Upgrade Kit는 사용자가 설치할 수 있으며 수 분 내에 업그레이드가 완료됩니다. Kit 구성: Injector hub, tubing catch, wrench, firmware upgrade USB 플래시 드라이브. Ion OneTouch™ System 1대당 업그레이드 kit 1개가 필요합니다.

Ion OneTouch™ DL Upgrade Kit의 혜택:

• 기기 효율성을 개선합니다.
• 기기 성능을 개선합니다.
• Ion OneTouch™ 200 Template Kit v2 DL Kit와 호환됩니다.

참고:
이 업그레이드는 Ion OneTouch™ 200 Template Kit v2 DL를 사용할 OneTouch™ System에 실시해야 합니다.

For Research Use Only. Not for use in diagnostic procedures.
사양
용도(장비)Ion OneTouch™ System
제품 유형DL Upgrade Kit
수량1 Kit
배송 조건Room Temperature
Unit SizeEach
구성 및 보관
Includes injector hub, tubing catch, wrench, and firmware upgrade loaded USB flash drive.

Store at room temperature.

자주 묻는 질문(FAQ)

What should I do if the centrifuge is still running even though the time remaining is 0:00 at the final spin of the Ion OneTouch 2?

The "time remaining" is not entirely accurate. The centrifuge may keep rotating for about 15 minutes after the time has become 0:00. If it rotates for more than 15 minutes, please contact Technical Support.

Find additional tips, troubleshooting help, and resources within our Next-Generation Sequencing Support Center.

What should I do if the Recovery Solution does not come out from Ion OneTouch DL?

The Ion OneTouch DL cannot be repaired. However, you may order an Ion OneTouch DL Upgrade Kit and exchange some parts which may fix the problem.

Find additional tips, troubleshooting help, and resources within our Next-Generation Sequencing Support Center.

What should I do if gas leaks from the Ion OneTouch DL after I open the gas main?

Please check the following:

1) That the connection tube between Ion OneTouch DL is not loose
2) That the outlet of the gas tank regulator is at 30 psi

If the tube is not loose and the regulator value is correct, please pull the valve (the gold ring) behind the Ion OneTouch DL straight down a couple of times.

If the pressure is too high, you will need to make adjustments. For details, please contact Technical Support.

Find additional tips, troubleshooting help, and resources within our Next-Generation Sequencing Support Center.

What should I do if an "Over Pressure" error appears after I start an Ion OneTouch 2 run?

The pressure inside the instrument may be above normal. If the Ion OneTouch 2 firmware is older than v540, please update it to the latest version.

Find additional tips, troubleshooting help, and resources within our Next-Generation Sequencing Support Center.

What are these black precipitates in the recovery tube and waste of the Ion OneTouch System?

It may be rubber from the packing. The amount varies among instruments and it is not anything to be concerned about, as it will be removed by the Ion OneTouch ES.

Find additional tips, troubleshooting help, and resources within our Next-Generation Sequencing Support Center.