사양
정확도±5% using SI-traceable mass foil set (for mass measurement)
인증/적합성U.S. EPA PM-2.5 equivalent monitor (EQPM-1609-183)
통신 프로토콜MODBUS, TCP/IP, Gesytec (Bayern-Hessen), streaming data and NTP (Network Time Protocol) protocols. Simultaneous connections from different locations over Ethernet
농도 측정 범위60 to 3,600 seconds and 24-hour
데이터 평균Every 1 second
깊이(영국식 단위)21.16 in.
깊이(미터법)53.8 cm
설명5014iQ Beta Attenuation Monitor
전기 요구사항Designed to meet CE: EN61326: 1997 + A1: 1998 + A2: 2001 + A3: 2003, EN61010-1 UL: 61010-1: 2004; CSA: C22.2 No. 61010-1: 2004; FCC: Part 15 Subpart B, Class B, UKCA
높이(영국식 단위)12.2 in.
높이(미터법)31 cm
측정 범위0 to 10,000 μg/m3
출력 - Standard ports 3 full-speed USB ports (one in front, two in rear) 1 gigabit ethernet port
- Analog I/O (optional) 4 isolated voltage inputs 0–10 V 6 isolated analog voltages outputs, with 4 selectable ranges 6 isolated analog current outputs, with 2 selectable ranges
- Digital I/O (optional) 16 digital inputs (TTL) 8 solenoid driver outputs,10 digital reed relay contact outputs
- Serial ports (optional) 1 RS-232/485 port, 1 RS-485 external accessory port
정밀도±2.0 μg/m3, <80 μg/m3, 4–5 μg/m3 > 80 μg/m3 (24-hour average)
분해능0.1 μg/m3
샘플링 흐름 및 제어 - Sample flow precision: ±2% of measured value
- Sample flow accuracy: <5% of measured value
와트 - 100–240 VAC, 50–60 Hz recommended, 805 watts (115 V)
- 880 watts maximum (220–240 V (instrument, heater & pump)
중량(영국식 단위)< 55.1 lbs.
중량(미터법)< 25 kg
폭(영국식 단위)16.73 in.
폭(미터법)42.5 cm
유량 속도1 m3/h (16.67 L/min) measured across an internal subsonic orifice, user selectable 14 to 20 lpm
가스 측정Carbon-14 (C-14), < 3.7 MBq (<100 μCi)
온도(영국식 단위) 작동The temperature of sampled air may vary -30 to 50 °C. 5014iQ Monitor must be weather protected within range 4 to 40 °C, an optional complete outdoor enclosure provides complete weather protection
Unit SizeEach