사양
알람/고장 릴레이Signal horn: OUTO
Malfunction: OUT1
Alarm 1 (pre-Alarm): OUT2
Alarm 2: (main Alarm): OUT3
Measurement in progress: OUT4
주변 온도Operating: Min. +15°C; Max. +45°C
백그라운드 수준Typ. 0.2 cps in compensation window, amb. Dose rate level approx. 100nSv/h
0.07 cps in iodine window
연결USB port, ethernet port, USB stick
깊이(영국식 단위)31.5 in. without packing
깊이(미터법)80 cm without packing
설명Fixed Iodine Monitor
검출기 효율0.093 Cps/Bq
디지털 입력Push button “Ack alarm”: IN0
Signal “Air flow ok”: IN1
Signal “Heating ok”: IN2
Reserved: IN3
디스플레이 유형Windows-based touch screen panel PC
효율Typ. 0.093 cps/Bq125I
전기 요구사항Operating voltage: 900 V
환경 조건Permitted temp. gradient: Max: 15 K/h
Sampled air intake temperature: Min. -10°C, Max. +45°C with optional heating on
Barometric pressure: Min. 500, Max. 1050 hPa
Rel. humidity of sampled air at intake: Min. 20, Max. 80 % R.H. non-condensing
필터 유형Glass fiber GF10, grade H12 for prevention of contamination and pollution of the iodine cartridge by aerosols
Air intake flange: DN 16; KF
Air intake filter diameter: 50 mm; S grade
Filter cartridge diameter: 58 mm; charcoal
Cartridge service life: several weeks depending on the iodine conc. of the ambient air
Cartridge activity load: 1 MBq
높이(영국식 단위)63 in. without packing
입력6 counters in total
4 counters are used for the three energy windows: Background energy window left; iodine energy window; background energy window right
측정 범위Min. 100; Max. 105 Bq/m3 related to 60 minutes sampling time
Measuring Principle: The sample air is passed through a 2 ¼ “ standard filter cartridge filled with iodine adsorbent. Elementary as well as organically bound iodine are precipitated. The filter cartridge is manually exchanged. A NaI scintillation detector located in the axis of the filter cartridge continuously measures the low-energy gamma quanta emitted by the accumulated I-125 activity. The pulse height is discriminated by an appropriately set iodine window.
전력 소비300 VA
펌프Air flow rate: typ. 4, Max. 5 m3/h
Supply voltage: 230V AC, single phase
Supply frequency: 50 Hz
Power consumption: 250 VA
Protection class: IP54
Noise level: 68 dB at 1 m distance
Maintenance: 6 months
분해능10% for 137 Cs
차폐 유형25 mm, lead in all directions
Omni-directional lead shielding made of ring elements above the filter drawer with top lid. A built-in lutetium adapter is used for continuous peak-stabilization is integrated in the lead shield.
중량(영국식 단위)375 lb. net, without packling
폭(영국식 단위)23.6 in. without packing
폭(미터법)60 cm (Without Packing)
검출 한계Minimum Detectable Activity Concentration (MDAC) provided according to ISO 11929
<0.6 Bq/m3 related to 125I with t1 = 5 min, t2 = 30 min. at a constant dose rate level of 100 nSv/h
검출기 유형NaI
Diameter (Metric) Detector Window50.8 mm
용도(애플리케이션)Iodine Monitoring
높이(미터법)160 cm (Without Packing)
제품 유형Iodine Monitor
방사선 유형131I
중량(미터법)170 kg (net, without packing)
Unit SizeEach