Harshaw TLD Model 5500 Readers
Harshaw TLD Model 5500 Readers
Thermo Scientific™

Harshaw TLD Model 5500 Readers

Measure and evaluate doses with an automatic, compact table-top system that provides unattended operation for up to 50 thermoluminescence dosimetry (TLD) detectors.
Have Questions?
보기 방식 변경buttonViewtableView
카탈로그 번호설명
HARSHAWTLD5500Model 5500 TLD Reader Instrument
5500TLD-DS33 Year Extended Depot Service Warranty
5500READERMAModel 5500 TLD Reader Instrument, Macro Pick Version
카탈로그 번호 HARSHAWTLD5500
제품 가격(KRW)
-
견적 요청하기
설명:
Model 5500 TLD Reader Instrument

Get near-tissue equivalent thermoluminescence dosimetry (TLD) performance with a high-capacity reader for all types of TLD elements and health physics applications. The Thermo Scientific™ HARSHAW TLD Model 5500 Automatic Reader provides unattended operation for up to 50 dosimeters and accommodates TL chips, disks, rods and cubes in a broad variety of sizes. This reader is suitable for all kinds of applications that require TLD detectors. Research applications are: radiation measurement verification, total body irradiation dose measurement, and many others where measurement quality assurance is critical.

The HARSHAW TLD Model 5500 Manual Reader includes a carrier disk for up to 50 TL detectors, a linear, programmable contactless nitrogen heating system, and a cooled photomultiplier tube with associated electronics to measure the TL light output. The operational software, Thermo Scientific™ WinREMS™, which runs on a separate computer, provides the user interface, the reader control and the applications software.

  • Multiple, programmable, linear heating cycles including pre-read anneal, acquire and post-read anneal cycles.
  • Typical cycle time per detector: typically 30 seconds.
  • Heating temperature capability up to 400°C.
  • Thermoelectric PMT cooler for maximum gain stability.
  • Seven decade glow curve acquisition range.
  • Options include glow curve deconvolution software (WinGCF Software), various types of neutral density filter to extend the high measurement range, and TL element specific carrier disks for best possible reproducibility.
  • Warm-up time: 30 minutes.
  • Easy to operate, service and maintain.
  • Size: 38 × 46 × 48cm.
  • 40kg.

Dosimetry Performance Using LiF:Mg,Ti (TLD-100/600/700):

  • Nearly tissue equivalent.
  • Radiation types: photons (E>5keV), electrons (E>70keV), neutrons (from thermal energies up to ∽100MeV, TLD-600 only).
  • Measurement range: 10 μGy to 1 Gy (linear), 1 Gy to 20 Gy (supralinear).
  • Detection threshold: <10 μGy based on 2.26 standard deviations for 10 sequential readings of an unexposed dosimeter.
  • Repeatability: Less than 2% variation based on one standard deviation for 10 sequential measurements of 1 mGy 137-Cs.
  • Residual TL signal: <0.2% of reading.
  • Fading: <20% in 3 months without thermal treatment; <5% in 3 months using pre-read anneal or glow curve deconvolution software (WinGCF Software).
  • Batch uniformity: ±30% from the factory standard.
  • Reuse: typically %500.

Instrument Performance:

  • Dynamic range: 7 decades.
  • HV stability: ±0.005% per hour, ±0.02% per 8 hours.
  • Linearity: better than 1%.
  • Test light: temperature-stabilized LED.
  • Electrical supply: selectable, 110/120VAC ±10%, 60Hz or 220/240VAC ±10%, 50Hz
  • Nitrogen supply:
    Purified (99.995%), dry.
    Pressure: 2.8 to 6.3kg/cm2.
    Flow Rate: internally regulated to 400 liters/h if specified pressure is provided.
  • Operating temperature: 15° to 40°C.
  • Storage temperature: -10° to +60°C.
  • Humidity: Functions within specification after 24 hour exposure to 90% humidity and subsequent 6-hour recovery with use of the nitrogen supply.
  • Ambient light exposure: Functions within specifications when exposed to ambient light up to 1,000 Lux with cover on the reader.
  • Shock: Withstands a 1-cm drop onto a concrete surface.

Ordering Alerts:

Computer requirements: Any commercially available PC system with Windows™ 7, XP, 2000, or NT 4. 32 bit OS only. RS232 strongly recommended. (Please contact your sales representative for approved USB to RS232 converters).

Compatible with:

Thermo Scientific TLD Materials, Thermo Scientific TLD Accessories

Recommended for:

Dose verification and research studies, quality assurance, whole body dosimetry, extremity dosimetry, environmental dosimetry, high dose applications (radiation hardness testing), and many more.


The intended use for Harshaw TLD Readers and their associated dosimeters are as follows: The Thermo Scientific Model 5500 Harshaw TLD Reader is a PC-driven, manually operated, tabletop instrument for thermoluminescent dosimetry (TLD) measurement. It economically provides both high performance and high reliability, and it complies with the latest International Standards Organization (ISO) requirements. The 5500 reads one dosimeter per loading and accommodates a variety of TL configurations, including chips, disks, rods, and powder.

These products are not intended for in vitro diagnostic purposes in accordance with our product documentation, manuals, and labels. They are designated for General Laboratory Use Only. Our Products have not been tested or validated for such applications and their use for in vitro diagnostic purposes may result in inaccurate results and, more seriously, health and safety risks. For General Lab Use Only - Not For Diagnostic Procedures.

사양
주변 광선 노출Functions within specifications when exposed to ambient light up to 1000 Lux with cover on the reader
배치 균일성±30% from the factory standard
설명Model 5500 TLD Reader Instrument
검출기 유형Thermoluminescence (TL) Detector
선량 범위Measurement range: 10 μGy to 1 Gy (linear), 1 Gy to 20 Gy (supralinear)
Detection threshold: <10 μGy based on 2.26 standard deviations for 10 sequential readings of an unexposed dosimeter
선량계 재사용성>500
동적 범위7 decades
유속Flow Rate: internally regulated to 400 liters/h if specified pressure is provided
높이(영국식 단위)1.89 in.
높이(미터법)48 cm
습도Functions within specifications after 24-hour exposure to 95% and subsequent 6-hour recovery
작동 온도(영국식 단위)59°F to 104°F
작동 온도(미터법)15°C to 40°C
PMTsThermoelectric PMT cooler for Maximum gain stability
전원 공급 장치100/120/220/240 VAC
제품 유형Radiation Dosimeter Reader
민감도Dependent on config. Used
보관 온도(영국식 단위)14°F to 140°F
보관 온도(미터법)-10°C to +60°C
시간 및 온도 프로필Multiple, programmable, linear heating cycles include pre-read anneal, acquire and post-read anneal cycles; Typical cycle time per detector: typically 30 seconds
중량(영국식 단위)88.18 lb.
중량(미터법)40 kg
용도 (장비)Thermo Scientific TLD Card Dosimeters, Thermo Scientific TLD Materials, Thermo Scientific TLD Accessories
선형성Better than 1%
방사선 유형Photons, Electrons, Neutrons
참조 라이트Temperature-stabilized LED
반복성for 1 mGy (100 mrad) 137 Cs doses, >2% standard deviation of 10 sequential measurements
안정성HV stability: ±0.005% per hour, ±0.02% per 8 hours
Unit SizeEach