호환성 | Robot |
검출기 유형 | Photomultiplier Tube (PMT) |
인큐베이터 온도 | 45°C |
인터페이스 | PC Software (Thermo Scientific SkanIt Software) |
광원 | Xenon Flash Lamp, LED |
측정 속도 | 15 s for 96-well Plate, 45 s for 384-well Plate, 135 s for 1536-Well Plate (Minimum Times) |
측정 기술 | Absorbance, Fluorescence Intensity, Luminescence, Time-resolved Fluorescence, AlphaScreen |
플레이트 유형 | 6 to 1536-well Plates (Absorbance 6 to 384-well Plates) |
제품라인 | Varioskan |
Reading Options | Top and Bottom |
판독 범위 | 0 to 6 Abs |
쉐이킹 | Orbital |
Software Type | SkanIt Software |
스펙트럼 스캔 속도 | <2.2 s/Well 400 to 500 nm, 1 Flash, 2 nm Step |
파장 선택 | Absorbance & Spectral Scanning: Double Monochromator, Fluorescence Intensity: Double Excitation and Emission Monochromator, AlphaScreen & Time-Resolved Fluorescence: Filter |
정확도 | 0.003 Abs or ±2%, at 200 to 399 nm (0 to 2 Abs) 0.003 Abs or ±1%, at 400 to 1000 nm (0 to 3 Abs) |
깊이(영국식 단위) | 23 in. |
깊이(미터법) | 58 cm |
동적 범위 | Fluorescence intensity: top reading > 6 decades, bottom reading > 5.5 decades Luminescence: > 7 decades |
높이(영국식 단위) | 20 in. |
높이(미터법) | 51 cm |
선형성 | 96-well plate: ±2% (0 to 4 Abs at 450 nm) 384-well plate: ±2% (0 to 3 Abs at 450 nm) |
측정 유형 | Endpoint, Kinetic, Spectra, Multipoint, Kinetic Spectra |
No. of Dispensers | None |
정밀도 | SD < 0.001 Abs or CV < 0.5%, at 450 nm (0 to 3 Abs) |
민감도 | Fluorescence intensity (top reading): < 0.4 fmol fluorescein/well (black 384-well plate) Fluorescence intensity (bottom reading): < 4 fmol fluorescein/well, (clear bottom black 384 square well plate) Time-resolved fluorescence: < 1 amol/well (white low volume 384 well plate) Luminescence: < 7 amol ATP/well (white 384-well plate) AlphaScreen: < 100 amol phosphotyrosine/well (384-well plate) |
소프트웨어 언어 버전 | English, German, French, Spanish, Portuguese, Italian, Japansese, Chinese and Russian |
파장 범위 | Absorbance: 200 to 1000 nm Fluorescence intensity: Excitation: 200 to 1000 nm, Emission: 270 to 840 nm Timetoresolved fluorescence: Excitation 334 nm, Emission 400 to 700 nm Luminescence: 360 to 670 nm AlphaScreen: Excitation 680 nm, Emission 400 to 660 nm |
중량(영국식 단위) | 125 lb. |
중량(미터법) | 57 kg |
폭(영국식 단위) | 21 in. |
폭(미터법) | 53 cm |
Unit Size | Each |
호환성 | Robot |
검출기 유형 | Photomultiplier Tube (PMT) |
인큐베이터 온도 | 45°C |
인터페이스 | PC Software (Thermo Scientific SkanIt Software) |
광원 | Xenon Flash Lamp |
측정 속도 | 15 s for 96-well Plate, 45 s for 384-well Plate, 135 s for 1536-Well Plate (Minimum Times) |
측정 기술 | Absorbance, Fluorescence Intensity |
플레이트 유형 | 6 to 1536-well Plates (Absorbance 6 to 384-well Plates) |
제품라인 | Varioskan |
Reading Options | Top |
판독 범위 | 0 to 6 Abs |
쉐이킹 | Orbital |
Software Type | SkanIt Software |
스펙트럼 스캔 속도 | <2.2 s/Well 400 to 500 nm, 1 Flash, 2 nm Step |
파장 선택 | Absorbance & Spectral Scanning: Double Monochromator, Fluorescence Intensity: Double Excitation and Emission Monochromator |
정확도 | 0.003 Abs or ±2%, at 200 to 399 nm (0 to 2 Abs) 0.003 Abs or ±1%, at 400 to 1000 nm (0 to 3 Abs) |
깊이(영국식 단위) | 23 in. |
깊이(미터법) | 58 cm |
동적 범위 | Fluorescence intensity: top reading > 6 decades |
높이(영국식 단위) | 20 in. |
높이(미터법) | 51 cm |
헤르츠 | 50/60 Hz |
선형성 | 96-well plate: ±2% (0 to 4 Abs at 450 nm) 384-well plate: ±2% (0 to 3 Abs at 450 nm) |
측정 유형 | Endpoint, Kinetic, Spectra, Multipoint, Kinetic Spectra |
No. of Dispensers | None |
정밀도 | SD < 0.001 Abs or CV < 0.5%, at 450 nm (0 to 3 Abs) |
민감도 | Fluorescence intensity (top reading): < 0.4 fmol fluorescein/well (black 384-well plate) |
소프트웨어 언어 버전 | English, German, French, Spanish, Portuguese, Italian, Japansese, Chinese and Russian |
전압 | 100/240 V |
파장 범위 | Absorbance: 200 to 1000 nm Fluorescence intensity: Excitation: 200 to 1000 nm, Emission: 270 to 840 nm |
중량(영국식 단위) | 119 lb. |
중량(미터법) | 54 kg |
폭(영국식 단위) | 21 in. |
폭(미터법) | 53 cm |
Unit Size | Each |
호환성 | Robot |
검출기 유형 | Photomultiplier Tube (PMT) |
인큐베이터 온도 | 45°C |
인터페이스 | PC Software (Thermo Scientific SkanIt Software) |
광원 | Xenon Flash Lamp |
측정 속도 | 15 s for 96-well Plate, 45 s for 384-well Plate, 135 s for 1536-Well Plate (Minimum Times) |
측정 기술 | Absorbance, Fluorescence Intensity, Luminescence |
플레이트 유형 | 6 to 1536-well Plates (Absorbance 6 to 384-well Plates) |
제품라인 | Varioskan |
Reading Options | Top |
판독 범위 | 0 to 6 Abs |
쉐이킹 | Orbital |
Software Type | SkanIt Software |
스펙트럼 스캔 속도 | <2.2 s/Well 400 to 500 nm, 1 Flash, 2 nm Step |
파장 선택 | Absorbance & Spectral Scanning: Double Monochromator, Fluorescence Intensity: Double Excitation and Emission Monochromator |
정확도 | 0.003 Abs or ±2%, at 200 to 399 nm (0 to 2 Abs) 0.003 Abs or ±1%, at 400 to 1000 nm (0 to 3 Abs) |
깊이(영국식 단위) | 23 in. |
깊이(미터법) | 58 cm |
동적 범위 | Fluorescence intensity: top reading > 6 decades Luminescence: > 7 decades |
높이(영국식 단위) | 20 in. |
높이(미터법) | 51 cm |
헤르츠 | 50/60 Hz |
선형성 | 96-well plate: ±2% (0 to 4 Abs at 450 nm) 384-well plate: ±2% (0 to 3 Abs at 450 nm) |
측정 유형 | Endpoint, Kinetic, Spectra, Multipoint, Kinetic Spectra |
No. of Dispensers | None |
정밀도 | SD < 0.001 Abs or CV < 0.5%, at 450 nm (0 to 3 Abs) |
민감도 | Fluorescence intensity (top reading): < 0.4 fmol fluorescein/well (black 384-well plate) Luminescence: < 7 amol ATP/well (white 384-well plate) |
소프트웨어 언어 버전 | English, German, French, Spanish, Portuguese, Italian, Japansese, Chinese and Russian |
전압 | 100/240 V |
파장 범위 | Absorbance: 200 to 1000 nm Fluorescence intensity: Excitation: 200 to 1000 nm, Emission: 270 to 840 nm Luminescence: 360 to 670 nm |
중량(영국식 단위) | 123 lb. |
중량(미터법) | 56 kg |
폭(영국식 단위) | 21 in. |
폭(미터법) | 53 cm |
Unit Size | Each |
호환성 | Robot |
검출기 유형 | Photomultiplier Tube (PMT) |
인큐베이터 온도 | 45°C |
인터페이스 | PC Software (Thermo Scientific SkanIt Software) |
광원 | Xenon Flash Lamp |
측정 속도 | 15 s for 96-well Plate, 45 s for 384-well Plate, 135 s for 1536-Well Plate (Minimum Times) |
측정 기술 | Absorbance, Fluorescence Intensity, Luminescence, Time-resolved Fluorescence |
플레이트 유형 | 6 to 1536-well Plates (Absorbance 6 to 384-well Plates) |
제품라인 | Varioskan |
Reading Options | Top |
판독 범위 | 0 to 6 Abs |
쉐이킹 | Orbital |
Software Type | SkanIt Software |
스펙트럼 스캔 속도 | <2.2 s/Well 400 to 500 nm, 1 Flash, 2 nm Step |
파장 선택 | Absorbance & Spectral Scanning: Double Monochromator, Fluorescence Intensity: Double Excitation and Emission Monochromator, Time-Resolved Fluorescence: Filters |
정확도 | 0.003 Abs or ±2%, at 200 to 399 nm (0 to 2 Abs) 0.003 Abs or ±1%, at 400 to 1000 nm (0 to 3 Abs) |
깊이(영국식 단위) | 23 in. |
깊이(미터법) | 58 cm |
동적 범위 | Fluorescence intensity: top reading > 6 decades Time-resolved fluorescence: > 6 decades Luminescence: > 7 decades |
높이(영국식 단위) | 20 in. |
높이(미터법) | 51 cm |
선형성 | 96-well plate: ±2% (0 to 4 Abs at 450 nm) 384-well plate: ±2% (0 to 3 Abs at 450 nm) |
측정 유형 | Endpoint, Kinetic, Spectra, Multipoint, Kinetic Spectra |
No. of Dispensers | None |
정밀도 | SD < 0.001 Abs or CV < 0.5%, at 450 nm (0 to 3 Abs) |
민감도 | Fluorescence intensity (top reading): < 0.4 fmol fluorescein/well (black 384-well plate) Time-resolved fluorescence: < 1 amol/well (white low volume 384 well plate) Luminescence: < 7 amol ATP/well (white 384-well plate) |
소프트웨어 언어 버전 | English, German, French, Spanish, Portuguese, Italian, Japansese, Chinese and Russian |
파장 범위 | Absorbance: 200 to 1000 nm Fluorescence intensity: Excitation: 200 to 1000 nm, Emission: 270 to 840 nm Timetoresolved fluorescence: Excitation 334 nm, Emission 400 to 700 nm Luminescence: 360 to 670 nm |
중량(영국식 단위) | 125 lb. |
중량(미터법) | 57 kg |
폭(영국식 단위) | 21 in. |
폭(미터법) | 53 cm |
Unit Size | Each |
호환성 | Robot |
검출기 유형 | Photomultiplier Tube (PMT) |
인큐베이터 온도 | 45°C |
인터페이스 | PC Software (Thermo Scientific SkanIt Software) |
광원 | Xenon Flash Lamp |
측정 속도 | 15 s for 96-well Plate, 45 s for 384-well Plate, 135 s for 1536-Well Plate (Minimum Times) |
측정 기술 | Absorbance, Fluorescence Intensity, Luminescence |
플레이트 유형 | 6 to 1536-well Plates (Absorbance 6 to 384-well Plates) |
제품라인 | Varioskan |
Reading Options | Top and Bottom |
판독 범위 | 0 to 6 Abs |
쉐이킹 | Orbital |
Software Type | SkanIt Software |
스펙트럼 스캔 속도 | <2.2 s/Well 400 to 500 nm, 1 Flash, 2 nm Step |
파장 선택 | Absorbance: Double Monochromator, Fluorescence Intensity: Double Excitation and Emission Monochromator |
정확도(광도) | 0.003 Abs or ±2% at 200 to 399 nm (0 to 2 Abs), 0.003 Abs or ±1%, at 400 to 1000 nm (0 to 3 Abs) |
CO2 농도 범위 | 0.1 to 15% |
깊이(영국식 단위) | 23 in. |
깊이(미터법) | 58 cm |
동적 범위 | Fluorescence intensity: top reading > 6 decades; Luminescence: > 7 decades |
높이(영국식 단위) | 20 in. |
높이(미터법) | 51 cm |
선형성 | 96-well plate: 0 to 4 Abs, ±2% at 450 nm, 384-well plate: 0 to 3 Abs ±2% at 450 nm |
측정 유형 | Endpoint, Kinetic, Spectra, Multipoint, Kinetic Spectra |
No. of Dispensers | None |
정밀도 | SD <0.001 Abs or CV <0.5%, at 450 nm (0 to 3 Abs) |
민감도 | Fluorescence intensity (top reading): < 0.4 fmol fluorescein/well (black 384-well plate); Luminescence: < 7 amol ATP/well (white 384-well plate) |
소프트웨어 언어 버전 | English, German, French, Spanish, Portuguese, Italian, Japansese, Chinese and Russian |
파장 범위 | Absorbance: 200 to 1000 nm, Fluorescence intensity: Excitation: 200 to 1000 nm, Emission: 270 to 840 nm, Luminescence: 360 to 670 nm |
중량(영국식 단위) | 125 lb. |
중량(미터법) | 57 kg |
폭(영국식 단위) | 21 in. |
폭(미터법) | 53 cm |
Unit Size | Each |
호환성 | Robot |
검출기 유형 | Photomultiplier Tube (PMT) |
인큐베이터 온도 | 45°C |
인터페이스 | PC Software (Thermo Scientific SkanIt Software) |
광원 | Xenon Flash Lamp |
측정 속도 | 15 s for 96-well Plate, 45 s for 384-well Plate, 135 s for 1536-Well Plate (Minimum Times) |
측정 기술 | Absorbance, Fluorescence Intensity, Luminescence |
플레이트 유형 | 6 to 1536-well Plates (Absorbance 6 to 384-well Plates) |
제품라인 | Varioskan |
Reading Options | Top and Bottom |
판독 범위 | 0 to 6 Abs |
쉐이킹 | Orbital |
Software Type | SkanIt Software |
스펙트럼 스캔 속도 | <2.2 s/Well 400 to 500 nm, 1 Flash, 2 nm Step |
파장 선택 | Absorbance: Double Monochromator, Fluorescence Intensity: Double Excitation and Emission Monochromator |
정확도(광도) | 0.003 Abs or ±2% at 200 to 399 nm (0 to 2 Abs), 0.003 Abs or ±1%, at 400 to 1000 nm (0 to 3 Abs) |
CO2 농도 범위 | 0.1 to 15% |
깊이(영국식 단위) | 23 in. |
깊이(미터법) | 58 cm |
동적 범위 | Fluorescence intensity: top reading > 6 decades; Luminescence: > 7 decades |
높이(영국식 단위) | 20 in. |
높이(미터법) | 51 cm |
선형성 | 96-well plate: 0 to 4 Abs, ±2% at 450 nm, 384-well plate: 0 to 3 Abs ±2% at 450 nm |
측정 유형 | Endpoint, Kinetic, Spectra, Multipoint, Kinetic Spectra |
No. of Dispensers | 1 |
O2 농도 범위 | 1 to 21% |
정밀도 | SD <0.001 Abs or CV <0.5%, at 450 nm (0 to 3 Abs) |
민감도 | Fluorescence intensity (top reading): < 0.4 fmol fluorescein/well (black 384-well plate); Luminescence: < 7 amol ATP/well (white 384-well plate) |
소프트웨어 언어 버전 | English, German, French, Spanish, Portuguese, Italian, Japansese, Chinese and Russian |
파장 범위 | Absorbance: 200 to 1000 nm, Fluorescence intensity: Excitation: 200 to 1000 nm, Emission: 270 to 840 nm, Luminescence: 360 to 670 nm |
중량(영국식 단위) | 125 lb. |
중량(미터법) | 57 kg |
폭(영국식 단위) | 21 in. |
폭(미터법) | 53 cm |
Unit Size | Each |
호환성 | Robot |
검출기 유형 | Photomultiplier Tube (PMT) |
인큐베이터 온도 | 45°C |
인터페이스 | PC Software (Thermo Scientific SkanIt Software) |
광원 | Xenon Flash Lamp |
측정 속도 | 15 s for 96-well Plate, 45 s for 384-well Plate, 135 s for 1536-Well Plate (Minimum Times) |
측정 기술 | Absorbance, Fluorescence Intensity |
플레이트 유형 | 6 to 1536-well Plates (Absorbance 6 to 384-well Plates) |
제품라인 | Varioskan |
Reading Options | Top and Bottom |
판독 범위 | 0 to 6 Abs |
쉐이킹 | Orbital |
Software Type | SkanIt Software |
스펙트럼 스캔 속도 | <2.2 s/Well 400 to 500 nm, 1 Flash, 2 nm Step |
파장 선택 | Absorbance & Spectral Scanning: Double Monochromator, Fluorescence Intensity: Double Excitation and Emission Monochromator |
정확도 | 0.003 Abs or ±2%, at 200 to 399 nm (0 to 2 Abs) 0.003 Abs or ±1%, at 400 to 1000 nm (0 to 3 Abs) |
깊이(영국식 단위) | 23 in. |
깊이(미터법) | 58 cm |
동적 범위 | Fluorescence intensity: top reading > 6 decades, bottom reading > 5.5 decades |
높이(영국식 단위) | 20 in. |
높이(미터법) | 51 cm |
헤르츠 | 50/60 Hz |
선형성 | 96-well plate: ±2% (0 to 4 Abs at 450 nm) 384-well plate: ±2% (0 to 3 Abs at 450 nm) |
측정 유형 | Endpoint, Kinetic, Spectra, Multipoint, Kinetic Spectra |
No. of Dispensers | None |
정밀도 | SD < 0.001 Abs or CV < 0.5%, at 450 nm (0 to 3 Abs) |
민감도 | Fluorescence intensity (top reading): < 0.4 fmol fluorescein/well (black 384-well plate) Fluorescence intensity (bottom reading): < 4 fmol fluorescein/well, (clear bottom black 384 square well plate) |
소프트웨어 언어 버전 | English, German, French, Spanish, Portuguese, Italian, Japansese, Chinese and Russian |
전압 | 100/240 V |
파장 범위 | Absorbance: 200 to 1000 nm Fluorescence intensity: Excitation: 200 to 1000 nm, Emission: 270 to 840 nm |
중량(영국식 단위) | 119 lb. |
중량(미터법) | 54 kg |
폭(영국식 단위) | 21 in. |
폭(미터법) | 53 cm |
Unit Size | Each |
호환성 | Robot |
검출기 유형 | Photomultiplier Tube (PMT) |
인큐베이터 온도 | 45°C |
인터페이스 | PC Software (Thermo Scientific SkanIt Software) |
광원 | Xenon Flash Lamp |
측정 속도 | 15 s for 96-well Plate, 45 s for 384-well Plate, 135 s for 1536-Well Plate (Minimum Times) |
측정 기술 | Absorbance, Fluorescence Intensity, Luminescence |
플레이트 유형 | 6 to 1536-well Plates (Absorbance 6 to 384-well Plates) |
제품라인 | Varioskan |
Reading Options | Top and Bottom |
판독 범위 | 0 to 6 Abs |
쉐이킹 | Orbital |
Software Type | SkanIt Software |
스펙트럼 스캔 속도 | <2.2 s/Well 400 to 500 nm, 1 Flash, 2 nm Step |
파장 선택 | Absorbance: Double Monochromator, Fluorescence Intensity: Double Excitation and Emission Monochromator |
정확도(광도) | 0.003 Abs or ±2% at 200 to 399 nm (0 to 2 Abs), 0.003 Abs or ±1%, at 400 to 1000 nm (0 to 3 Abs) |
CO2 농도 범위 | 0.1 to 15% |
깊이(영국식 단위) | 23 in. |
깊이(미터법) | 58 cm |
동적 범위 | Fluorescence intensity: top reading > 6 decades; Luminescence: > 7 decades |
높이(영국식 단위) | 20 in. |
높이(미터법) | 51 cm |
선형성 | 96-well plate: 0 to 4 Abs, ±2% at 450 nm, 384-well plate: 0 to 3 Abs ±2% at 450 nm |
측정 유형 | Endpoint, Kinetic, Spectra, Multipoint, Kinetic Spectra |
No. of Dispensers | 2 |
O2 농도 범위 | 1 to 21% |
정밀도 | SD <0.001 Abs or CV <0.5%, at 450 nm (0 to 3 Abs) |
민감도 | Fluorescence intensity (top reading): < 0.4 fmol fluorescein/well (black 384-well plate); Luminescence: < 7 amol ATP/well (white 384-well plate) |
소프트웨어 언어 버전 | English, German, French, Spanish, Portuguese, Italian, Japansese, Chinese and Russian |
파장 범위 | Absorbance: 200 to 1000 nm, Fluorescence intensity: Excitation: 200 to 1000 nm, Emission: 270 to 840 nm, Luminescence: 360 to 670 nm |
중량(영국식 단위) | 125 lb. |
중량(미터법) | 57 kg |
폭(영국식 단위) | 21 in. |
폭(미터법) | 53 cm |
Unit Size | Each |
호환성 | Robot |
검출기 유형 | Photomultiplier Tube (PMT) |
인큐베이터 온도 | 45°C |
인터페이스 | PC Software (Thermo Scientific SkanIt Software) |
광원 | Xenon Flash Lamp |
측정 속도 | 15 s for 96-well Plate, 45 s for 384-well Plate, 135 s for 1536-Well Plate (Minimum Times) |
측정 기술 | Absorbance, Fluorescence Intensity, Luminescence |
플레이트 유형 | 6 to 1536-well Plates (Absorbance 6 to 384-well Plates) |
제품라인 | Varioskan |
Reading Options | Top and Bottom |
판독 범위 | 0 to 6 Abs |
쉐이킹 | Orbital |
Software Type | SkanIt Software |
스펙트럼 스캔 속도 | <2.2 s/Well 400 to 500 nm, 1 Flash, 2 nm Step |
파장 선택 | Absorbance & Spectral Scanning: Double Monochromator, Fluorescence Intensity: Double Excitation and Emission Monochromator |
정확도 | 0.003 Abs or ±2%, at 200 to 399 nm (0 to 2 Abs) 0.003 Abs or ±1%, at 400 to 1000 nm (0 to 3 Abs) |
깊이(영국식 단위) | 23 in. |
깊이(미터법) | 58 cm |
동적 범위 | Fluorescence intensity: top reading > 6 decades, bottom reading > 5.5 decades Luminescence: > 7 decades |
높이(영국식 단위) | 20 in. |
높이(미터법) | 51 cm |
헤르츠 | 50/60 Hz |
선형성 | 96-well plate: ±2% (0 to 4 Abs at 450 nm) 384-well plate: ±2% (0 to 3 Abs at 450 nm) |
측정 유형 | Endpoint, Kinetic, Spectra, Multipoint, Kinetic Spectra |
No. of Dispensers | None |
정밀도 | SD < 0.001 Abs or CV < 0.5%, at 450 nm (0 to 3 Abs) |
민감도 | Fluorescence intensity (top reading): < 0.4 fmol fluorescein/well (black 384-well plate) Fluorescence intensity (bottom reading): < 4 fmol fluorescein/well, (clear bottom black 384 square well plate) Luminescence: < 7 amol ATP/well (white 384-well plate) |
소프트웨어 언어 버전 | English, German, French, Spanish, Portuguese, Italian, Japansese, Chinese and Russian |
전압 | 100/240 V |
파장 범위 | Absorbance: 200 to 1000 nm Fluorescence intensity: Excitation: 200 to 1000 nm, Emission: 270 to 840 nm |
중량(영국식 단위) | 123 lb. |
중량(미터법) | 56 kg |
폭(영국식 단위) | 21 in. |
폭(미터법) | 53 cm |
Unit Size | Each |
호환성 | Robot |
검출기 유형 | Photomultiplier Tube (PMT) |
인큐베이터 온도 | 45°C |
인터페이스 | PC Software (Thermo Scientific SkanIt Software) |
광원 | Xenon Flash Lamp |
측정 속도 | 15 s for 96-well Plate, 45 s for 384-well Plate, 135 s for 1536-Well Plate (Minimum Times) |
측정 기술 | Absorbance, Fluorescence Intensity, Luminescence, Time-resolved Fluorescence |
플레이트 유형 | 6 to 1536-well Plates (Absorbance 6 to 384-well Plates) |
제품라인 | Varioskan |
Reading Options | Top and Bottom |
판독 범위 | 0 to 6 Abs |
쉐이킹 | Orbital |
Software Type | SkanIt Software |
스펙트럼 스캔 속도 | <2.2 s/Well 400 to 500 nm, 1 Flash, 2 nm Step |
파장 선택 | Absorbance & Spectral Scanning: Double Monochromator, Fluorescence Intensity: Double Excitation and Emission Monochromator, Time-Resolved Fluorescence: Filters |
정확도(광도) | 0.003 Abs or ±2% at 200 to 399 nm (0 to 2 Abs), 0.003 Abs or ±1%, at 400 to 1000 nm (0 to 3 Abs) |
CO2 농도 범위 | 0.1 to 15% |
깊이(영국식 단위) | 23 in. |
깊이(미터법) | 58 cm |
동적 범위 | Fluorescence intensity: top reading > 6 decades; Luminescence: > 7 decades; Time-resolved fluorescence: > 6 decade |
높이(영국식 단위) | 20 in. |
높이(미터법) | 51 cm |
선형성 | 96-well plate: 0 to 4 Abs, ±2% at 450 nm, 384-well plate: 0 to 3 Abs ±2% at 450 nm |
측정 유형 | Endpoint, Kinetic, Spectra, Multipoint, Kinetic Spectra |
No. of Dispensers | None |
O2 농도 범위 | 1 to 21% |
정밀도 | SD <0.001 Abs or CV <0.5%, at 450 nm (0 to 3 Abs) |
민감도 | Fluorescence intensity (top reading): < 0.4 fmol fluorescein/well (black 384-well plate); Luminescence: < 7 amol ATP/well (white 384-well plate); Time-resolved fluorescence: < 1 amol/well (white low volume 384 well plate) |
소프트웨어 언어 버전 | English, German, French, Spanish, Portuguese, Italian, Japansese, Chinese and Russian |
파장 범위 | Absorbance: 200 to 1000 nm, Fluorescence intensity: Excitation: 200 to 1000 nm, Emission: 270 to 840 nm, Luminescence: 360 to 670 nm, Time-resolved fluorescence: Excitation 334 nm, Emission 400 to 700 nm |
중량(영국식 단위) | 125 lb. |
중량(미터법) | 57 kg |
폭(영국식 단위) | 21 in. |
폭(미터법) | 53 cm |
Unit Size | Each |
카탈로그 번호 | 사양 | 제품 사이즈 | 측정 기술 | Reading Options | No. of Dispensers | 제품 정가(KRW) | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
VLBLATD0 | Each | Absorbance, Fluorescence Intensity, Luminescence, Time-resolved Fluorescence, AlphaScreen | Top and Bottom | None | 견적 요청하기 | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
VL0000D0 | Each | Absorbance, Fluorescence Intensity | Top | None | 견적 요청하기 | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
VL0L00D0 | Each | Absorbance, Fluorescence Intensity, Luminescence | Top | None | 견적 요청하기 | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
VL0L0TD0 | Each | Absorbance, Fluorescence Intensity, Luminescence, Time-resolved Fluorescence | Top | None | 견적 요청하기 | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
VLBL00GD0 | Each | Absorbance, Fluorescence Intensity, Luminescence | Top and Bottom | None | 견적 요청하기 | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
VLBL00GD1 | Each | Absorbance, Fluorescence Intensity, Luminescence | Top and Bottom | 1 | 견적 요청하기 | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
VLB000D0 | Each | Absorbance, Fluorescence Intensity | Top and Bottom | None | 견적 요청하기 | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
VLBL00GD2 | Each | Absorbance, Fluorescence Intensity, Luminescence | Top and Bottom | 2 | 견적 요청하기 | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
VLBL00D0 | Each | Absorbance, Fluorescence Intensity, Luminescence | Top and Bottom | None | 견적 요청하기 | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
VLBL0TGD0 | Each | Absorbance, Fluorescence Intensity, Luminescence, Time-resolved Fluorescence | Top and Bottom | None | 견적 요청하기 | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|