KingFisher™ Flex 96 Deep Well head
KingFisher™ Flex 96 Deep Well head
Thermo Scientific™

KingFisher™ Flex 96 Deep Well head

Se trata de un cabezal magnético para la purificación de partículas magnéticas KingFisher™ Flex (n.º de cat. 5400630 y 5400640)Más información
Have Questions?
Número de catálogoCantidad
24074430Cada uno
Número de catálogo 24074430
Precio (MXN)
-
Cantidad:
Cada uno
Se trata de un cabezal magnético para la purificación de partículas magnéticas KingFisher™ Flex (n.º de cat. 5400630 y 5400640) Este cabezal magnético puede servir como cabezal de sustitución para el sistema de formato de 96 pocillos profundos o como cabezal suplementario para el sistema de formato de 24 pocillos profundos.
Para uso exclusivo en investigación. No apto para uso en procedimientos diagnósticos.
Especificaciones
Línea de productosKingFisher
CantidadCada uno
Condiciones de envíoTemperatura ambiente
Unit SizeEach

Preguntas frecuentes

Do you offer an alternative for the discontinued MagMAX Express-96 Deep Well Magnetic Head (Cat. No. 4388435)?

Unfortunately, we do not offer an alternative for the discontinued MagMAX Express-96 Deep Well Magnetic Head (Cat. No. 4388435).

I see that the KingFisher 24 Deep Well Head has a processing volume of 5 mL. Does that mean that I can process up to 5 mL of my sample?

No, the processing volume refers to the maximum volume that can be held in the 24 Deep Well plates. This incorporates the volume of binding buffers and beads, so when scaling up reaction volumes, it is important to make sure that the total volume does not exceed 5 mL.

My KingFisher Flex run was stopped due to an error message. Can I start the protocol from where it stopped?

No, once a protocol was stopped, it will restart from the beginning. The extraction procedure can either be finished manually by using a magnetic stand, or a modified version of the original KingFisher Flex script would have to be created that starts off at the point where the run was stopped.

I have a 12-plate protocol but there are only 8 slots on the KingFisher Flex instrument turntable. What should I do with the remaining 4 plates?

Set up the protocol on the BindIt software. After the first 8 plates are done, the instrument will pause and instruct you to switch out the plates.

Is it always necessary to use the transport locks on the KingFisher Flex instrument?

The transport locks are only necessary when relocating the instrument.