Especificaciones
Relés de alarma/falloSignal horn: OUTO
Malfunction: OUT1
Alarm 1 (pre-Alarm): OUT2
Alarm 2: (main Alarm): OUT3
Measurement in progress: OUT4
Temperatura ambienteOperating: Min. +15°C; Max. +45°C
ConexionesUSB port, ethernet port, USB stick
Fondo (imperial)25.6 in. without packing
Fondo (métrico)65 cm without packing
DescripciónMonitor de yodo móvil
Eficacia del detector0.05 Cps/Bq
Entradas digitalesPush button “Ack alarm”: IN0
Signal “Air flow ok”: IN1
Signal “Heating ok”: IN2
Reserved: IN3
Tipo de pantallaWindows-based touch screen panel PC
EficaciaTyp. 0.05 cps/Bq131I
Requisitos eléctricosOperating voltage: 900 V
Condiciones ambientalesPermitted temp. gradient: Max: 15 K/h
Sampled air intake temperature: Min. -10°C, Max. +45°C with optional heating on
Barometric pressure: Min. 950, Max. 1050 hPa
Rel. humidity of sampled air at intake: Min. 20, Max. 80 % R.H. non-condensing
Tipo de filtroGlass fiber GF10, grade H12 for prevention of contamination and pollution of the iodine cartridge by aerosols
Air intake flange: DN 16; KF
Air intake filter diameter: 50 mm; fiber glass GF 10
Filter cartridge diameter: 58 mm; charcoal
Cartridge service life: several weeks depending on the iodine conc. of the ambient air
Cartridge activity load: 1 MBq
Altura (imperial)51.2 in. without packing
Entradas6 counters in total
4 counters are used for the three energy windows: Background energy window left; iodine energy window; background energy window right
Intervalo de mediciónMin. 100; Max. 105 Bq/m3 related to 60 minutes sampling time
Measuring Principle: Fixed filter monitor for the detection of the 364 keV line of 131I by a NaI detector. Compensation of higher energy gamma lines is realized with the help of online background subtraction by the trapezoid method. A built-in lutetium adapter is used for continuous peak-stabilization.
Consumo de energía300 VA
BombaRotary vane vacuum pump; oil-free; air-cooled
Air flow rate: typ. 4, Max. 5 m3/h
Supply voltage: 230V AC, single phase
Supply frequency: 50 Hz
Power consumption: 250 VA
Protection class: IP54
Noise level: 68 dB at 1 m distance
Maintence: 2000 operation hours
Resolución10% for 137 Cs
Sensibilidad10 Cps/μSvH-1 without compensation
Tipo de blindaje25 mm, lead in all directions
Omni-directional lead shielding made of ring elements above the filter drawer with top lid. A built-in lutetium adapter is used for continuous peak-stabilization is integrated in the lead shield.
Peso (imperial)187 lb. net, without packing
Anchura (imperial)18.9 in. without packing
Anchura (métrico)48 cm (Without Packing)
Límites de detecciónMinimum Detectable Activity Concentration (MDAC) provided according to ISO 11929
<1 Bq/m3 related to 131I with t1 = 5 min, t2 = 30 min at a constant dose rate level of 100 nSv/h
<40 Bq/m3 related to 131I with t1 = 5 min, t2 = 30 min at a constant dose rate level of 100 nSv/h (60Co)
Tipo de detectorNaI
Diameter (Metric) Detector Window50.8 mm
Para utilizar con (aplicación)Iodine Monitoring
Altura (métrico)130 cm (Without Packing)
Tipo de productoIodine Monitor
Tipo de radiación131L
Peso (métrico)85 kg (net, without packing)
Unit SizeEach