Polymer Delivery Pump Cleaning Kit Dx
Polymer Delivery Pump Cleaning Kit Dx
Applied Biosystems™

Polymer Delivery Pump Cleaning Kit Dx

This is a pump cleaning kit for Applied Biosystems Dx genetic analyzers. It is highly recommended that a field service engineer perform any necessary pump cleaning.
Have Questions?
Número de catálogo 4461875
Precio (MXN)
-

This is a pump cleaning kit for Applied Biosystems Dx genetic analyzers. It is highly recommended that a field service engineer perform any necessary pump cleaning.

This is a pump cleaning kit for Applied Biosystems 3500 and SeqStudio Flex series genetic analyzers. It is highly recommended that a field service engineer perform any necessary pump cleaning.
Especificaciones
Para utilizar con3500 Series Genetic Analyzers, SeqStudio Flex Series Genetic Analyzers
Cantidad1 unidad
TipoPump Cleaning Kit
Unit SizeEach

Preguntas frecuentes

When I started my run on the Applied Biosystems 3500/3500xL Genetic Analyzer, an error occurred: "Unstable Electrophoresis current detected, check for air bubbles" but I don't see any bubbles. What else can cause this?

Some of the other causes of an Unstable Electrophoresis current detected error message are:

1.Leak on the system
2. Polymer that:
-Has expired
-Has been left on the instrument for more than the recommended time
-Is a mixture of expired polymer and non-expired polymer
3. Running Buffer that was:
-On the instrument longer than 2 weeks
-Not filled to the fill line or evaporated below the fill line
4. An arcing event that was not cleaned afterwards using the water wash wizard
5. Not performing regular maintenance on the instrument
6. Hardware issues

Inspect the system for leaks. If you do not see any leaks on the system, perform the wash the pump chamber and channels wizard using the conditioning pouch and place fresh, non-expired polymer and fresh Anode and Cathode buffer containers. If the problem persists, a service call may be required.

Find additional tips, troubleshooting help, and resources within our Capillary Electrophoresis Instruments Support Center.

Do I need to replace the buffer every 14 days if the Applied Biosystems 3500/3500xL Genetic Analyzer is not in use?

If the instrument is not in use, it is not necessary to replace the buffer every 14 days. However, the buffer level needs to remain at the fill line to prevent the capillary or array from drying out. Top off the buffer volume with water if necessary. Replace the anode and cathode buffer containers prior to starting a new run.

Find additional tips, troubleshooting help, and resources within our Capillary Electrophoresis Instruments Support Center.