Columna preempaquetada POROS™ GoPure™ 50 IP, 0,5 x 5 cm, 1 ml
Columna preempaquetada POROS™ GoPure™ 50 IP, 0,5 x 5 cm, 1 ml
Thermo Scientific™

Columna preempaquetada POROS™ GoPure™ 50 IP, 0,5 x 5 cm, 1 ml

Thermo Scientific™ POROS™ GoPure™ 0.5 cm pre-packed columns are intended for the screening of POROS ion exchange resins or for small scale purification.
Have Questions?
Número de catálogoCantidad
44813181 columna
Número de catálogo 4481318
Precio (MXN)
-
Cantidad:
1 columna

POROS GoPure 0.5 cm pre-packed columns are intended for the screening of POROS ion exchange resins or for small scale purification.

Estas columnas preempaquetadas están listas para conectarse y tienen las siguientes características:

  • Dimensiones de 0,5 x 5 cm, con un volumen nominal del lecho de 1 ml, lo que proporciona una plataforma de selección uniforme
  • Conexión directa a sistemas de cromatografía estándar
  • Empaquetada con resinas POROS de alto rendimiento que ofrecen una alta capacidad de unión dinámica y alta resolución en una amplia gama de caudales

La resina de esta columna POROS GoPure IP 50 50 μm tiene un grupo funcional de polietilenimina para cromatografía de intercambio aniónico fuerte.

  • Las columnas cómodas —y listas para usar simplifican y aceleran las evaluaciones de resina POROS
  • Conserva la muestra— el volumen nominal del lecho de 1 ml minimiza la necesidad de muestras para la selección de múltiples resinas
  • Escalabilidad simplificada— la consistencia y el rango de la familia POROS GoPure de columnas preempaquetadas simplifica el proceso de escalado y modelado de procesos

Las columnas preempaquetadas POROS GoPure de 0,5 cm se han desarrollado para satisfacer las necesidades de los científicos de desarrollo de procesos que requieren una plataforma de selección de resina POROS cómoda y uniforme. El volumen del lecho de 1 ml reduce los requisitos de muestras y el tiempo de residencia recomendado de ≥3 min (0,3 ml/min) en esta columna de 5 cm de longitud equivale a 400 cm/h en una columna de 20 cm de altura. Las especificaciones detalladas se pueden encontrar en las instrucciones de uso de POROS GoPure 0,5 cm a continuación.

Otras resinas de intercambio iónico POROS, así como resinas de intercambio catiónico, también están disponibles en este formato de columna. Todas las resinas de cromatografía de alta resolución POROS ™ tienen una estructura de poros única que proporciona un excelente transporte de masa para una alta capacidad y resolución en una amplia gama de caudales.

Para el escalado, las columnas preempaquetadas POROS GoPure están disponibles en una variedad de diámetros y en múltiples alturas de lecho que equivalen a un rango de volumen de columna de 5 ml a 6,3 litros.

Para uso exclusivo en investigación. No apto para uso en procedimientos diagnósticos.
Especificaciones
Tipo de columnaIntercambio aniónico
Para utilizar conFPLC
Compatibilidad de alto rendimientoProtocolos automatizados
Presión máxima30 bar
Cantidad1 columna
Tipo de muestraMuestra de purificación biofarmacéutica, sobrenadantes de cultivo celular, suero
Fase estacionariaPolietileneimina
Volumen (métrico)1 mL
Diámetro (métrico)0.5 cm
Longitud (métrico)5 cm
Tamaño de partículas50 μm
Línea de productosGoPure
Unit SizeEach
Contenido y almacenamiento
Shipped in 20% ethanol at 2°C to 30°C.

Preguntas frecuentes

Are POROS analytical columns compatible with HPLC systems from Waters or Agilent?

All of the pre-packed POROS columns, regardless of column size or resin, have the same 10-32 female fittings on the column. Columns are compatible with Waters and Agilent HPLC systems. You will need 1/16 inch male fittings for the HPLC to attach the column.

Find additional tips, troubleshooting help, and resources within our Protein Purification and Isolation Support Center.

For any POROS analytical column, is the column compatible with an AKTA or Avant purification system?

All of the pre-packed columns, regardless column size or resin, have the same 10-32 female fittings on the column. They are all compatible with the AKTA instrument platform, however due to the column volume of the 2.1 mmD x 30 mmL columns (0.1 mL), these will not perform well in AKTA or Avant purification systems due to the relative size of the hold- up volume in these systems. The smallest column (PEEK or SS) that could feasibly be used in these systems is the 4.6mmD x 50mmL (0.8 mL) format.

Find additional tips, troubleshooting help, and resources within our Protein Purification and Isolation Support Center.

Are the pre-packed POROS CaptureSelect or any POROS analytical columns compatible with an UPLC or is the use of these columns limited to HPLC?

UPLC is typically described as chromatography with sub-2 µm particles, which drives higher efficiency/better separation independent of operating flow rate. As particle size decreases, backpressure increases so the UPLC systems tend to have higher pressure ratings compared to HPLC systems. The flow rates on UPLC systems would be sufficient for POROS columns. Pump capabilities are typically 0.1-1.0 mL/min, so not as fast as a HPLC, but sufficient. Hold-up volumes on UPLC systems are lower than HPLCs, so this can help with sensitivity and efficiency. So from a pressure and flow standpoint there should be no challenge to operate a POROS CaptureSelect column on a UPLC system.

That being said, care should be taken running POROS columns on UPLC systems. Higher pressure safety interlock MUST be set at the rated column pressure, typically 180 bar. HPLC column hardware is not rated to operate at the static pressures possible and typical on UPLC systems. A clog or the like on a UPLC system could generate unsafe pressures on any HPLC-rated column before the default UPLC safety pressure interlock is tripped.

Find additional tips, troubleshooting help, and resources within our Protein Purification and Isolation Support Center.

What is the physical or mechanical stability of POROS resins?

The physical stability of POROS resins is high. Customers have agitated, with overhead mixing and or bottom gassing for long periods of time (5-10 hours) and have performed hundreds of diaphragm pump passes during stall packs. Keep in mind that agitation with any device that creates grinding and sheer, such as with a magnetic stir bar, is not recommended for any type of resin.

Find additional tips, troubleshooting help, and resources within our Protein Purification and Isolation Support Center.

What should I do if a POROS column dries out or air is pumped onto the column?

POROS resins rehydrate quite well. Here is the procedure that we recommend:

1. Flow 5% ethanol over the column at ?350 cm/hour for 3 CVs, followed by water or 0.1 M sodium chloride at 300 cm/hour for 3 CVs.
2. If baseline noise is still present, flow 20% ethanol over the column at 300 cm/hour for 3 CVs, followed by water or 0.1 M sodium chloride at 300 cm/hour for 3 CVs.
3. If you have established column qualification tests and specifications, perform the tests after rewetting.

Find additional tips, troubleshooting help, and resources within our Protein Purification and Isolation Support Center.