This is a pump cleaning kit for Applied Biosystems Dx genetic analyzers. It is highly recommended that a field service engineer perform any necessary pump cleaning.
Have Questions?
Catalog Number
Quantity
4461875
1 Ea.
Catalog number 4461875
Price (TWD)
3,360.00
Each
Add to cart
Quantity:
1 Ea.
Price (TWD)
3,360.00
Each
Add to cart
This is a pump cleaning kit for Applied Biosystems Dx genetic analyzers. It is highly recommended that a field service engineer perform any necessary pump cleaning.
This is a pump cleaning kit for Applied Biosystems 3500 and SeqStudio Flex series genetic analyzers. It is highly recommended that a field service engineer perform any necessary pump cleaning.
For In Vitro Diagnostic Use Only
Specifications
For Use With3500 Series Genetic Analyzers, SeqStudio Flex Series Genetic Analyzers
Quantity1 Ea.
TypePump Cleaning Kit
Unit SizeEach
Contents & Storage
Each packet contains enough parts for at least a one time cleaning of the pump. Store at room temperature.
Frequently asked questions (FAQs)
When I started my run on the Applied Biosystems 3500/3500xL Genetic Analyzer, an error occurred: "Unstable Electrophoresis current detected, check for air bubbles" but I don't see any bubbles. What else can cause this?
Some of the other causes of an Unstable Electrophoresis current detected error message are:
1.Leak on the system
2. Polymer that:
-Has expired
-Has been left on the instrument for more than the recommended time
-Is a mixture of expired polymer and non-expired polymer
3. Running Buffer that was:
-On the instrument longer than 2 weeks
-Not filled to the fill line or evaporated below the fill line
4. An arcing event that was not cleaned afterwards using the water wash wizard
5. Not performing regular maintenance on the instrument
6. Hardware issues
Inspect the system for leaks. If you do not see any leaks on the system, perform the wash the pump chamber and channels wizard using the conditioning pouch and place fresh, non-expired polymer and fresh Anode and Cathode buffer containers. If the problem persists, a service call may be required.
Do I need to replace the buffer every 14 days if the Applied Biosystems 3500/3500xL Genetic Analyzer is not in use?
If the instrument is not in use, it is not necessary to replace the buffer every 14 days. However, the buffer level needs to remain at the fill line to prevent the capillary or array from drying out. Top off the buffer volume with water if necessary. Replace the anode and cathode buffer containers prior to starting a new run.