Analytical sensitivity
29 pg/mL
Assay range
97.66-100,000 pg/mL
Sample type/volume
Fecal Extract
50 µL
Plasma
1 µL
Saliva
10 µL
Serum
1 µL
Urine
10 µL
Hands-on time
1 hr 20 min
Time-to-result
2 hr 30 min
Homogenous (no wash)
No
Interassay CV
11.9%
Intraassay CV
8.6%
Instrument
Colorimetric Microplate Reader
Product size
96 Tests
Contents
Pre-coated 96 well plate
Standard
Assay Diluent concentrate
HRP-Conjugated Detection Antibody
Wash Buffer
Chromogen
Stop Solution
Adhesive Plate Covers
Shipping condition
Wet or Dry Ice
Storage
-20°C
Protein name
Cortisone
Protein aliases
C21H28O5
Species (tested)
Chemical
Assay kit format
Competitive ELISA
Detector antibody conjugate
HRP
Label or dye
HRP
The Cortisone ELISA quantitates Cortisone in serum, saliva, urine, plasm, or dried fecal extracts.
Principle of the method
The Cortisone Competitive ELISA research-use-only kit is designed to quantitatively measure cortisone in extracted dried fecal samples, urine, saliva, plasma, and serum samples. This kit measures total cortisone in serum and plasma and in extracted fecal samples. A cortisone standard is provided to generate a standard curve for the assay and all samples should be read off the standard curve. Standards or diluted samples are pipetted into a clear microtiter plate coated with an antibody to capture rabbit antibodies. A cortisone-peroxidase conjugate is added to the standards and samples in the wells. The binding reaction is initiated by the addition of a polyclonal antibody to cortisone to each well. After a two-hour incubation the plate is washed and TMB substrate is added. The substrate reacts with the bound cortisone-peroxidase conjugate to produce a colored product. After a short incubation, the reaction is stopped and the intensity of the generated color is detected in a microtiter plate reader capable of measuring at 450 nm.
Cortisone (Kendall's Compound E) is a glucocorticoid which is the product of cortisol (Compound F) deactivation by 11-beta-HSD2 (hydroxysteroid dehydrogenase), prohibiting receptor activation. The balance between cortisone and cortisol glucocorticoids helps to regulate the anti-inflammatory response. Cortisone is identical across all species.
Rigorous validation
Each manufactured lot of this ELISA kit is quality tested for criteria such as sensitivity, specificity, precision, and lot-to-lot consistency. See manual for more information on validation.
For Research Use Only. Not for use in diagnostic procedures. Not for resale without express authorization.
If an Invitrogen™ antibody doesn't perform as described on our website or datasheet,we'll replace the product at no cost to you, or provide you with a credit for a future purchase.*
Learn moreGet expert recommendations for common problems or connect directly with an on staff expert for technical assistance related to applications, equipment and general product use.