Search Thermo Fisher Scientific
Search Thermo Fisher Scientific
Select an article below to learn how to care for and maintain your Thermo Scientific Nicolet Summit FTIR Spectrometer.
Your Thermo Scientific™ Nicolet™ Summit FTIR Spectrometer requires little upkeep and no daily maintenance. However, basic care will ensure continued peak performance. Follow the guidelines in this article. For best performance, leave the spectrometer powered on.
NOTICE Static electricity can permanently damage critical components in your spectrometer. To help prevent such damage, follow these recommendations:
The OMNIC Paradigm Software includes qualification and performance verification (PV) workflows for your Summit spectrometer. The qualification workflows are standard industry-wide qualification tests you most likely recognize (for example, European Pharmacopoeia, Japanese Pharmacopoeia, etc.). The PV workflow runs a series of standard tests to verify instrument operation and ensure the accuracy of your data. Any required reference standards are supplied inside the instrument and controlled by the OMNIC Paradigm Software.
We recommend that you run the PV workflow or your preferred qualification workflow at least once a week. For more information, see the article titled “Qualify Your Summit Spectrometer” in the online help.
The spectrometer and touchscreen can be cleaned only as recommended. See the article titled “Clean the Spectrometer and Touchscreen” in the online help for details.
The spectrometer’s optical components can be easily damaged by excessive moisture in the air. The spectrometer is sealed and its components are protected by two desiccant canisters that absorb moisture. A humidity indicator located here monitors the humidity level inside the spectrometer.
Figure 1. Location of Humidity Indicator
Check the humidity indicator at least once a month and change the desiccant cannisters when needed (see the table below), or purchase and install a purge kit. See the article titled “Install and Maintain a Purge Kit” for more information.
Table 1. Humidity Indicator States and Recommendations
Humidity Indicator State | Meaning | Action |
---|---|---|
Blue | Desiccant is fully charged | none |
Light blue | Desiccant is becoming saturated with moisture and no longer providing sufficient protection | Replace the desiccant |
Pink or white | Desiccant has expired | Replace the desiccant and the humidity indicator |
NOTICE We recommend that you keep the spectrometer sealed and desiccated, or sealed and purged with dry air or nitrogen, at all times. Equipment damage due to failure to maintain seal and desiccation and/or purge is not covered under the warranty. If you have questions about this requirement, please contact us.
If your Summit spectrometer is purged with nitrogen or dry air, check the purge filter at least once a month.
Replace the filter when it is yellow, or otherwise discolored, or if it is contaminated with debris or foreign particles. For more information, see the section “Checking and Changing the Purge Gas Filter” in the article titled “Install and Maintain a Purge Kit” in the online help.
Figure 2. The purge filter is located inside of the plastic bowl below the pressure gauge
Align your spectrometer’s optics if a qualification workflow, such as the Nicolet FTIR - PV test, indicates a failing test. Aligning the spectrometer optimizes the energy that reaches the detector and maximizes the detector signal. If a performance or qualification test fails, align the spectrometer’s optics, calibrate the laser frequency, and run the test again. If the test still fails, contact your local technical support representative for help.
Calibrate your laser frequency if a qualification workflow, such as the Nicolet FTIR - PV test, indicates a failing test. When a test fails, first align the spectrometer, then calibrate the laser frequency and run the test again. If the test still fails, contact your local technical support representative for help.
Before cleaning the Thermo Scientific™ Nicolet™ Summit Spectrometer, turn it off and disconnect the power supply.
Gently clean the exterior of the instrument with a soft, clean cloth lightly dampened with mild soap.
Dust can collect on the electronics enclosure on the back of the instrument and can potentially interfere with heat dissipation, which may reduce the lifetime of the electronic components.
To remove dust from the back of the instrument, use compressed air to blow the dust off. Do not use any liquid when removing dust from the back of the spectrometer.
CAUTION To avoid shock hazard, do not allow liquid to run into the power supply or the back of the instrument.
NOTICE Do not use harsh detergents, solvents, chemicals, or abrasives; these can damage the finish. Do not allow liquid to contact any windows in the sample compartment.
CAUTION Windows can be scratched and ruined very easily. Do not touch or attempt to clean them. Dust will not affect the signal, but fingerprints can degrade the performance of the instrument and permanently damage mirrors or windows. If you wish to remove dust from a mirror or window, blow it off with a gentle stream of clean, dry air or nitrogen. Never allow any liquid to come into contact with a window or optical component in the instrument.
To clean the back and sides of the touchscreen device, use a soft, clean cloth lightly dampened with mild soap.
To clean the screen, apply window or glass cleaner to a clean cloth or sponge, and gently wipe the screen with the cloth.
CAUTION Do not use alcohol (methyl, ethyl, or isopropyl), thinner, benzene, or other abrasive cleaners.
NOTICE Never apply the cleaner directly to the touchscreen.
If your instrument is not equipped with a purge kit, or if the purge is turned off, you must monitor the humidity level inside the spectrometer. Replace the desiccant when the paper humidity indicator turns pink (can be light pink or almost white). Replace the paper humidity indicator each time you replace the desiccant.
Replacing the desiccant requires you to remove the desiccant cartridge and replace the desiccant canisters and the O‑ring.
Before you begin, you will need the following:
WARNING Before opening the desiccant compartment, turn off the instrument power, unplug the power cord and remove all accessories and samples from the system. Entry of flammable liquids or gases into the desiccant compartment is an explosion hazard. If such entry occurs, contact us immediately and do not apply power to the instrument until the condition has been corrected.
NOTICE Make sure nothing falls into the instrument while the desiccant cover is removed.
WARNING Replace the desiccant only with replacement parts supplied by us.
Note You can use either desiccant cartridge (with or without the purge connector) to maintain the required humidity. The illustrations below show the nonpurged version of the desiccant cartridge. The instructions for replacing the desiccant canisters are the same for both cartridges.
![]() | ![]() |
Note Be careful to not drop the O-ring into the desiccant compartment.
WARNING The contents of the desiccant canisters (hydrated metal aluminosilicate molecular sieve) could be harmful if ingested. If you discard the saturated desiccant canisters, make sure they are properly disposed of.
![]() |
|
NOTICE Always wear lab gloves or finger cots, or use laboratory tissue, when handling a new humidity indicator. Oil or moisture from skin can discolor the indicator.
NOTICE Make sure that the indicator is pressed tightly against the window so that it does not come loose in the desiccant compartment when you reinstall the cartridge.
The IR source is easily accessible from the bottom of the spectrometer and can be replaced without having to remove the spectrometer’s cover.
Time needed: 25 minutes or less
CAUTION Before replacing the source, turn off the instrument. Replace the source only with replacement parts supplied by us.
CAUTION The source becomes extremely hot while you are using the instrument. Wait at least 10 minutes after turning the instrument off before you perform the next step.
![]() | ![]() |
NOTICE Never touch a source element with your bare fingers. Skin oils or other deposits on the element will shorten its life. Always use clean finger cots, gloves, or clean laboratory tissue when handling a source element.
WARNING To ensure a good ground connection and avoid shock hazard, do not use an outlet that is connected to a conduit ground. The ground must be a noncurrent-carrying wire connected to earth ground at the main distribution box.
Note Do not install any sampling accessing until after you have aligned the spectrometer and verified its performance
Install the touchscreen monitor to save space in your lab and to take full advantage of OMNIC Paradigm software for touchscreen for quickly executing workflows and carrying out routine procedures.
Time needed: 25 minutes or less
![]() 1. WiFi dongle port 2. USB port 3. Power port 4. HDMI port | ![]() The power cable is shown highlighted in yellow. Connect the power cable and lay it in the guides along the back of the spectrometer. |
![]() | ![]() |
WARNING To ensure a good ground connection and avoid shock hazard, do not use an outlet that is connected to a conduit ground. The ground must be a noncurrent-carrying wire connected to earth ground at the main distribution box.
Congratulations. Your touchscreen monitor is installed and ready for use. With the touchscreen monitor installed, you can open OMNIC Paradigm software and switch to the touchscreen interface for a convenient, streamlined experience.
The Thermo Scientific™ Nicolet™ Summit FTIR Spectrometer has windows on both sides of the sample compartment that seal the instrument from moisture and other contaminants but allow the infrared beam to enter and exit the sample compartment. The windows should always be installed, even if your system is purged with dry air or nitrogen.
WARNING Prevent fire and explosion. The infrared source inside the instrument is an ignition source. If you are using volatile solvents, provide a fume hood or other active venting system that is free of spark and other ignition sources and prevents flammable vapors from collecting in the atmosphere surrounding the instrument.
We install potassium bromide (KBr) windows at the factory and run all of our operating and performance tests with the windows installed. The KBr windows are hygroscopic but have a protective coating. On this spectrometer, they provide a working spectral range between 8,000 and 350 cm-1 with no significant optical absorptions lines.
If you purchased your spectrometer with zinc selenide (ZnSe) windows, the spectrometer is delivered with KBr windows installed. You must install the ZnSe windows yourself using the instructions below. ZnSe windows are non-hygroscopic, which makes them more suitable for extremely humid environments when the spectrometer is not purged. They have an anti-reflection coating to optimize performance and, on this spectrometer, provide a working spectral range between 7,800 and 500 cm-1.
For optimal spectrometer performance, the sample compartment windows must be clear (not cloudy) and clean (no dust or fingerprints). If your windows become contaminated, the spectrometer may fail performance and qualification tests. You can order new windows from us and install them yourself using the instructions below.
NOTICE
Time needed: 10 minutes
Tools needed:
Note If your instrument is purged, leave the purge on while you replace the windows to prevent ambient air from entering the spectrometer.
Note The window fittings are tight. Use one hand to stabilize the instrument and the other to remove the window.
NOTICE Use nitrile gloves to handle the window and handle it by the rims only. (Avoid touching the window surfaces even when wearing gloves.)
NOTICE The window should turn easily at the start; if it doesn’t, reverse the direction of rotation and restart the insertion to avoid stripping the threads.