Having difficulties with your experiment?
We are dedicated to your success. Get back on track. View our expert recommendations for commonly encountered problem scenarios.
View the relevant questions below:
Beginning your experiment?
Visit our
You may occasionally observe a small amount of aggregation of the Qdot™ nanocrystals during proper storage. To remove any aggregates that may have formed prior to use, we recommend centrifuging the vial at 2,000 x g for 1 min. Pipette only the supernatant and avoid the pellet. In our experience, pelleting any aggregates that may have formed typically results in a loss of less than 10% of the product.
Once your product undergoes aggregation, it cannot be dispersed back into solution. We recommend purchasing a new product.
Freezing will cause the product to aggregate. The Qdot™ nanocrystals cannot be dispersed into solution after aggregation.
Appropriate mounting media selection is very important to retain the fluorescence of Qdot™ nanocrystals. In our studies, Qdot™ nanocrystals work best with the following mountants:
Note: We do not recommend using ProLong™ mounting media with Qdot™ nanocrystals as it will quench their fluorescence.
Blinking is an inherent property of quantum dots; in fact, all single-luminescent molecules blink, including organic dyes. The brightness and photostability of Qdot™ nanocrystals makes the blinking more visibly apparent. Under higher energy excitation, Qdot™ nanocrystals blink even faster. Beta-mercaptoethanol can reduce blinking.
The precipitate in the organic ITK™ Qdot™ nanocrystals occurs with some frequency. The ITK™ Qdot™ nanocrystals sometimes include impurities that show as a white precipitate.
Spinning your ITK™ Qdot™ nanocrystals at ~3,000 rpm for 3_5 minutes should remove the white precipitate from the supernatant. Use the supernatant immediately.
Here are some suggestions:
Here are some suggestions: