Search Thermo Fisher Scientific
Search Thermo Fisher Scientific
Planicidad de línea de base | ±0.002 A |
Compartimentos | 8-position cell changer (standard) Also accommodates (optional): 4-position cell changer (long path cells) Peltier thermostatted cell holder (20-60 °C) Sipper accessory Fiber optic probe coupler |
Conexiones | Single USB-A Duplex USB-A Export data to network or PC via USB, Ethernet, or Wi-Fi USB adaptor, available Print via USB, Ethernet or Wi-Fi USB adaptor available |
Descripción | GENESYS 180 UV-Vis Spectrophotometer with UK Power Cable |
Tipo de detector | Dual Silicon Photodiodes |
Pantalla | 7 in. High Definition Color Touchscreen, Tiltable, 800 x 1280 pixels |
Dimensiones (Largo x Ancho x Alto) | 35.5 x 38.5 x 19.5 cm |
Ancho de banda espectral | 2 nm |
Wavelength Range | 190 to 1100 nm |
Desviación | (At 500 nm after 1 hour warmup) <0.0005 A/hr. |
Requisitos eléctricos | External AC to DC converter. Voltage and Frequency (Hz) selected automatically, 100/240 V, 50/60 Hz |
Para utilizar con (aplicación) | Sample Concentration |
Teclado | Touchscreen |
Vida útil de la lámpara | > 5 years typical, 3 years guaranteed |
Ruido | (RMS at 500 nm 60 consecutive measurements) ≤0.00020A at 0A at 260 and 500 nm ≤0.00030A at 1A at 260 and 500 nm ≤0.00040A at 2A at 260 and 500 nm |
Diseño óptico | Double Beam |
Exactitud fotométrica | ±0.002 A at 0.5 A, ±0.004 A at 1.0 A, ±0.008 A at 2.0 A |
Pantalla fotométrica | –3A to +5A |
Intervalo fotométrico | –2A to +3.5A |
Repetibilidad fotométrica | ±0.001 A at 1 A measured at 1.0 A at 546 nm |
Impresora incluida | Snap-on printer available |
Velocidad de barrido | Slow, medium and fast - up to 1600 nm/min. |
Luz difusa | < 1.0%T 198 nm (KCl) , <0.05%T at 220 nm (NaI), <0.03%T at 340 nm (NaNO2) |
Garantía | 2 years standard with 1 year extension upon registration of the instrument with Thermo Fisher Scientific within 6 months of purchase |
Exactitud de la longitud de onda | ±0.5 nm |
Intervalo de datos de longitud de onda | 0.1 nm, 0.2 nm, 0.5 nm, 1 nm, 2 nm, 5 nm |
Repetibilidad de longitud de onda | ±0.2 nm |
Peso (métrico) | 7.5 kg |
Unit Size | Each |
Planicidad de línea de base | ± 0,002A |
Compartimentos | Cambiador de celdas de 8 posiciones (estándar) También admite (opcional): Cambiador de celdas de 4 posiciones (celdas de paso óptico) Portacubetas con termostato Peltier (20-60 ºC) Accesorio aspirador Acoplador de sonda de fibra óptica |
Conexiones | El puerto USB-A admite dispositivos de memoria flash para el almacenamiento de datos y métodos El puerto doble USB-A del lateral admite la conexión a un ordenador Windows™ que ejecute el software de control remoto opcional, al teclado o al ratón Exportación de datos a la red o PC mediante USB, Ethernet o Wi-Fi Impresión a través de USB, Ethernet o Wi-Fi |
Descripción | Espectrofotómetro UV-Vis GENESYS 180 |
Tipo de detector | Dos fotodiodos de silicio |
Pantalla | Pantalla táctil a color de 7 pulgadas, inclinable y de alta definición, 800 × 1280 píxeles |
Dimensiones (Largo x Ancho x Alto) | 35,5 x 38,5 x 19,5 cm |
Ancho de banda espectral | 2 nm |
Wavelength Range | 190 nm-1100 nm |
Desviación | (A 500 nm tras 1 hora de calentamiento) < 0,0005 A/h |
Requisitos eléctricos | Convertidor de CA a CC externo. Selección automática de tensión y frecuencia (Hz), 100-240 V, 50-60 Hz. |
Teclado | Pantalla táctil |
Vida útil de la lámpara | > 5 years typical, 3 years guaranteed |
Ruido | (RMS a 500 nm, 60 mediciones consecutivas) ≤ 0,00020 A a 0 A a 260 y 500 nm ≤ 0,00030 A a 1 A a 260 y 500 nm ≤ 0,00040 A a 2 A a 260 y 500 nm |
Diseño óptico | Doble haz |
Exactitud fotométrica | ± 0,002 A a 0,5 A, ± 0,004 A a 1,0 A, ± 0,008 A a 2,0 A |
Pantalla fotométrica | De -3A a +5A |
Intervalo fotométrico | De -2A a +3,5A |
Repetibilidad fotométrica | ± 0,001 A a 1 A medido a 1,0 A a 546 nm |
Impresora incluida | Impresora conectable disponible |
Velocidad de barrido | Lento, medio y rápido: hasta 1600 nm/min |
Luz difusa | < 1,0 T% 198 nm (KCl), < 0,05 T% a 220 nm (NaI), < 0,03 T% a 340 nm (NaNO2) |
Garantía | 2 años estándar con ampliación de un año tras el registro del instrumento con Thermo Fisher Scientific en un periodo máximo de 6 meses desde la compra |
Exactitud de la longitud de onda | ± 0,5 nm |
Intervalo de datos de longitud de onda | 0,1 nm; 0,2 nm; 0,5 nm; 1 nm; 2 nm; 5 nm |
Repetibilidad de longitud de onda | < ± 0,2 nm |
Peso (métrico) | 7,5 kg |
Unit Size | Each |
Planicidad de línea de base | ±0.002 A |
Compartimentos | 8-position cell changer (standard) Also accommodates (optional): 4-position cell changer (long path cells) Peltier thermostatted cell holder (20-60 °C) Sipper accessory Fiber optic probe coupler |
Conexiones | Single USB-A Duplex USB-A Export data to network or PC via USB, Ethernet, or Wi-Fi USB adaptor, available Print via USB, Ethernet or Wi-Fi USB adaptor available |
Descripción | GENESYS 180 UV-Vis Spectrophotometer with EU Power Cable |
Tipo de detector | Dual Silicon Photodiodes |
Pantalla | 7 in. High Definition Color Touchscreen, Tiltable, 800 x 1280 pixels |
Dimensiones (Largo x Ancho x Alto) | 35.5 x 38.5 x 19.5 cm |
Ancho de banda espectral | 2 nm |
Wavelength Range | 190 to 1100 nm |
Desviación | (At 500 nm after 1 hour warmup) <0.0005 A/hr. |
Requisitos eléctricos | External AC to DC converter. Voltage and Frequency (Hz) selected automatically, 100/240 V, 50/60 Hz |
Para utilizar con (aplicación) | Sample Concentration |
Teclado | Touchscreen |
Vida útil de la lámpara | > 5 years typical, 3 years guaranteed |
Ruido | (RMS at 500 nm 60 consecutive measurements) ≤0.00020A at 0A at 260 and 500 nm ≤0.00030A at 1A at 260 and 500 nm ≤0.00040A at 2A at 260 and 500 nm |
Diseño óptico | Double Beam |
Exactitud fotométrica | ±0.002 A at 0.5 A, ±0.004 A at 1.0 A, ±0.008 A at 2.0 A |
Pantalla fotométrica | –3A to +5A |
Intervalo fotométrico | –2A to +3.5A |
Repetibilidad fotométrica | ±0.001 A at 1 A measured at 1.0 A at 546 nm |
Impresora incluida | Snap-on printer available |
Velocidad de barrido | Slow, medium and fast - up to 1600 nm/min. |
Luz difusa | < 1.0%T 198 nm (KCl) , <0.05%T at 220 nm (NaI), <0.03%T at 340 nm (NaNO2) |
Garantía | 2 years standard with 1 year extension upon registration of the instrument with Thermo Fisher Scientific within 6 months of purchase |
Exactitud de la longitud de onda | ±0.5 nm |
Intervalo de datos de longitud de onda | 0.1 nm, 0.2 nm, 0.5 nm, 1 nm, 2 nm, 5 nm |
Repetibilidad de longitud de onda | ±0.2 nm |
Peso (métrico) | 7.5 kg |
Unit Size | Each |
Número de catálogo | Especificaciones | Formato | Descripción | Precio (EUR) | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
840-309400 | Each | GENESYS 180 UV-Vis Spectrophotometer with UK Power Cable | Solicite un presupuesto | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
840-309000 | Each | Espectrofotómetro UV-Vis GENESYS 180 | Solicite un presupuesto | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
840-309300 | Each | GENESYS 180 UV-Vis Spectrophotometer with EU Power Cable | Solicite un presupuesto | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|