Plaques de 96 puits non stériles à fond plat transparent Immuno
Plaques de 96 puits non stériles à fond plat transparent Immuno
Le produit réel peut varier
Thermo Scientific™

Plaques de 96 puits non stériles à fond plat transparent Immuno

Choisissez vos surfaces de fixation pour des dosages colorimétriques fiables et reproductibles.
Have Questions?
Modifier l'affichagebuttonViewtableView
RéférenceTraitement de surfaceCertifiésVolume (métrique) de travailCode-barres
460984MaxiSorpOui350 μlCode-barres inclus
3355Immulon 1 BNo230 μlAucun code-barres
3855Immulon 4 HBXNo280 μlAucun code-barres
439454MaxiSorpOui350 μlAucun code-barres
442404MaxiSorpNo350 μlAucun code-barres
456529PolySorpNo350 μlAucun code-barres
456537MaxiSorpOui350 μlAucun code-barres
467320MediSorpOui350 μlAucun code-barres
467340MultiSorpNo350 μlAucun code-barres
475094PolySorpNo350 μlAucun code-barres
9502227Universal Binding (UB, liaison universelle)No330 μlAucun code-barres
95029330Meilleure capacité de liaisonOui330 μlAucun code-barres
Référence 460984
Prix (EUR)
754,00
Case of 180
Ajouter au panier
Traitement de surface:
MaxiSorp
Certifiés:
Oui
Volume (métrique) de travail:
350 μl
Code-barres:
Code-barres inclus
Prix (EUR)
754,00
Case of 180
Ajouter au panier
Les plaques Immuno à 96 puits à fond plat et transparentes non stériles Thermo Scientific™ permettent de choisir parmi les surfaces de fixation pour des résultats fiables et reproductibles dans les dosages colorimétriques. Elles incluent les plaques transparentes Thermo Scientific™ Nunc™, Immulon™ et Microtiter™ qui sont idéales pour les immunoessais quantitatifs et qualitatifs.

Plaques de 96 puits pour matières solides :

  • Polystyrène transparent
  • Choix de quatre surfaces pour une fixation optimale
  • Rangs et colonnes avec étiquetage alphanumérique

Surfaces :

  • Les plaques Thermo Scientific™ PolySorp™, Immulon™ 1B, et Universal Binding (UB) sont destinées à l’adsorption des molécules hydrophobes.
  • Les plaques Thermo Scientific™ Medisorp™ et Immulon 2HB sont légèrement hydrophiles et fixent divers types de biomolécules.
  • Les plaques Thermo Scientific™ MaxiSorp™, Immulon 4 HBX, et Enhanced Binding (EB) sont hydrophiles et idéales pour les tests sur anticorps en sandwich.
  • La plaque Thermo Scientific™ MultiSorp™ est très hydrophile et convient bien aux glycanes et aux protéines hydrosolubles.

Informations de commande :

Remarque : la disponibilité peut varier d’un pays à l’autre. Les références 467320, 456537, 460984 et 439454 sont livrées avec certificat.

Clause de non-responsabilité relative aux restrictions d’utilisation pour la recherche interne : Toute restriction applicable à l’acheteur et relative à l’utilisation du ou des produit(s) uniquement à des fins de recherche interne ne s’applique pas au(x) produit(s) présent(s) sur les pages Web associées à la présente déclaration.
Spécifications
Propriété de liaisonHydrophile
Type de liaisonAdsorption
CertifiésOui
Méthode de détectionColorimétrique
CouvercleSans couvercle
Nbre par carton180
Nb par paquet180
Conditionnement180 boîtes
Blocage de la plaqueAucun
Type de produitMicroplaque
Molécule cibleAnticorps (glycoprotéines)
Volume (métrique) de travail350 μl
Array8 x 12
Code-barresCode-barres inclus
CouleurTransparent
FormatÉtalon
MatériauPolystyrène
Nbre de puits96
StérilitéNon stérile
Traitement de surfaceMaxiSorp
Volume (métrique) du puits400 μl
Type de puitsRond
Unit SizeCase of 180

Foire aux questions (FAQ)

For Clear Flat-Bottom Immuno Nonsterile 96-Well Plates, is the surface treatment solely on the bottom of the well, or is it also applied to the sides?

For energy treated plates, the entire plate surface receives the energy treatment. The MaxiSorp, PolySorp, MediSorp and MultiSorp plates are energy treated plates. For traditionally treated plates, the well surface is treated up to the listed working volume.

What length of peptide is ideal for binding to the Thermo Scientific MaxiSorp surface? What are the detection limitations?

We have tested and found that a 3 amino acid peptide (Pro, Leu, Gly) cannot be detected when passively adsorbed on the MaxiSorp surface. However, this peptide can be detected when covalently immobilized using CovaLink NH Modules and CovaLink NH2 Modules and Plates. Using covalent immobilization of small peptide residues, you can typically obtain a better orientation of the molecule and reduced problems with antibody recognition of the peptide due to masking of the epitope. We have discovered that a 7 amino acid peptide from the MHC Class II antigen can be detected when adsorbed on the MaxiSorp surface. We state that the detection limitation using the MaxiSorp surface is between 3 and 7 amino acid residues.
One additional note is that detection is contingent upon the orientation of the peptide when immobilized. If the active site is inactivated or hidden at the site facing the solid phase, no detection signal is observed.

Can I bind either single- or double-stranded DNA to the Thermo Scientific MaxiSorp surface?

Single-stranded DNA can be adsorbed to our MaxiSorp surface using approximately 10 µg ssDNA per mL PBS, pH 8.2, although the stability is uncertain. Based on our experience, ssDNA immobilized on the MaxiSorp surface is so loosely bound that it can be removed by stringent washing.
Double-stranded DNA will not bind to the MaxiSorp surface. DNA, however, can be covalently bound to Nunc NucleoLink Strips.

Are your Thermo Scientific 8-Well Strip Caps for Immuno Standard Modules compatible with all Nunc MicroWell 96-Well Microplates and Nunc Immuno 96-Well Plates?

Our Thermo Scientific 8-Well Strip Caps for Immuno Standard Modules were designed to provide a positive seal for flat and round bottom wells of our Nunc MicroWell 96-Well Microplates and Nunc Immuno 96-Well Plates.
8-Well Strip Caps for Immuno Standard Modules are not compatible with C or V bottom wells of the Nunc Microwell 96-Well Microplates and Nunc Immuno 96-Well Plates.

For Nunc Immuno plates and modules, what are the advantages of one well geometry type over another? Which types should I use for various applications?

The following list describes the geometries of wells available for Nunc Immuno-plates and modules:
- Flat bottom (F): Allows maximum transmission of light. These plates can be read on a monochromatic reader.
- Round bottom (U): This geometry optimizes washing and coating. These plates can be read using a dual wavelength reader.
- "C" bottom (C): This design of the well is a combination of both flat and round bottoms. Basically, it is a flat bottomed well with curved edges at the bottom. These plates also can be read using a monochromatic reader combining optimal reading and washing.
- StarWell: These wells have a modified "C" shape geometry with eight fins strategically placed at the bottom. This feature increases surface area, allowing more molecules to become immobilized which reduces incubation times.