Glass Syringe, 1.0 mL
Product Image
Applied Biosystems™

Glass Syringe, 1.0 mL

Diese Glasspritze eignet sich zum Befüllen von Genanalysatorkapillaren vom Typ 310 mit POP-4™ Performance Optimized Polymer, POP-6™ Performance Optimized PolymerWeitere Informationen
Have Questions?
KatalognummerMenge
43044711 syringe
Katalognummer 4304471
Preis (EUR)
678,00
Each
Zum Warenkorb hinzufügen
Menge:
1 syringe
Preis (EUR)
678,00
Each
Zum Warenkorb hinzufügen
Diese Glasspritze eignet sich zum Befüllen von Genanalysatorkapillaren vom Typ 310 mit POP-4™ Performance Optimized Polymer, POP-6™ Performance Optimized Polymer und GeneScan™ Polymer.

Zur Verwendung mit:
Genanalysatorkapillaren Typ 310, 47 cm (36 cm wtr), Laufpuffer 310, 10 X, BigDye™ Terminator v1.1 Zyklussequenzierungskit, BigDye™ Terminator v3.1 Zyklussequenzierungskit, Genanalysator-Probenröhrchen, 0,5 ml, Hi-Di™ Formamid, POP-4™ Polymer für 3100/3100-Avant™ Genanalysatoren, POP-4™ Polymer für den Genanalysator 310, POP-6™ Polymer für den Genanalysator 310, Septa für 0,5-ml-Probenröhrchen.

Nur für Forschungszwecke.Darf nicht für diagnostische Verfahren eingesetzt werden.
Nur für Forschungszwecke. Nicht zur Verwendung bei diagnostischen Verfahren.
Specifications
ProdukttypSpritze
Menge1 syringe
Zur Verwendung mit (Geräte)310 genetischer Analysator
MaterialGlas
Volumen (metrisch)1 mL
Unit SizeEach

Häufig gestellte Fragen (FAQ)

How often should I change my polymer and buffer? 

Polymer in the syringe should be changed every three days. Do not return unused polymer to the bottle.

Buffer and water should be changed every 48 hours or, if the run lasts for more than 48 hours, after the run is complete.

Find additional tips, troubleshooting help, and resources within our Capillary Electrophoresis Instruments Support Center.

How often do I really need to clean the pump block?

The pump block needs to be cleaned when:

Installing a syringe.

Switching applications (i.e., from sequencing to fragment analysis).

Shutting down the instrument for a long period of time.

If the instrument has run for more than 4 days without any of the above occurring.

Appearance of unusual spikes in the electropherogram.

If the syringe containing the polymer has been on the instrument for more than a week, it is important to clean the syringe and the block, and to change the polymer. At room temperature, enough urea decomposition can occur in one week to cause transient current increases during electrophoreses. If polymer is allowed to dry in the pump block channel, the pump block may be ruined.

Find additional tips, troubleshooting help, and resources within our Capillary Electrophoresis Instruments Support Center.

What happens to the 310 Genetic Analyzer if it runs out of polymer during the run?

A message will appear at the start of the run telling you if you have enough polymer to complete the run. If you should continue without adding more polymer, then the run will be cancelled at the injection where it ran out and an entry will be made into the run log that there is no polymer. The following injections will appear to run for 1–2 minutes each and will be recorded into the run log. No data will be collected for these injections.

Find additional tips, troubleshooting help, and resources within our Capillary Electrophoresis Instruments Support Center.