7900HT TaqMan™ RNase P-Kit zur Verifizierung von Geräten mit niedriger Dichte
Applied Biosystems™
7900HT TaqMan™ RNase P-Kit zur Verifizierung von Geräten mit niedriger Dichte
The 7900HT TaqMan® Low Density Array Instrument Verification RNase P kit is used to validate the performance of the AppliedWeitere Informationen
Have Questions?
Katalognummer
Menge
4351468
1 kit
Katalognummer 4351468
Preis (EUR)
3.710,00
Each
Zum Warenkorb hinzufügen
Menge:
1 kit
Preis (EUR)
3.710,00
Each
Zum Warenkorb hinzufügen
The 7900HT TaqMan® Low Density Array Instrument Verification RNase P kit is used to validate the performance of the Applied Biosystems® 7900HT Fast Real-Time PCR Systems with the TaqMan® Low Density Array Block Upgrade. The 7900HT TaqMan® Low Density Array Instrument Verification RNase P kit contains an empty TaqMan® Low Density Array and 8 vials of RNase P detection reagents to detect and quantitate genomic copies of the human RNase P gene, a single-copy gene encoding the RNA moiety of the RNase P enzyme. Each vial contains premixed reagents (1X TaqMan® Universal PCR Master Mix, RNase P primers, and FAM™ dye–labeled probe) and template. The 7900HT TaqMan® Low Density Array Instrument Verification RNase P kit must demonstrate the ability to distinguish between 5,000 and 10,000 genomic equivalents with a 99.7% confidence level for a subsequent sample run in a single well. To ensure accurate results, Applied Biosystems® recommends that the performance of your instrument be verified every six months.
Nur für Forschungszwecke. Nicht zur Verwendung bei diagnostischen Verfahren.
Zur Verwendung mit (Anwendung)Real Time PCR (qPCR)
Zur Verwendung mit (Geräte)7900 HT LDA-System
ProduktlinieTaqMan
Menge1 kit
ProdukttypInstrument Verification Kit
Unit SizeEach
Inhalt und Lagerung
Contains an empty TaqMan™ Low Density Array and 8 vials of RNase P detection reagents. Store at -15 to 25°C. Protect from light
Häufig gestellte Fragen (FAQ)
When do I need to calibrate the 7900HT Fast Real-Time PCR System?
Recommended Maintenance Schedule
Power on/off the computer controlling the instrument: Weekly
Check computer disk space. If necessary, archive or back up your experiment files and instrument settings: Weekly
Run disk cleanup and disk defragmentation: Every month
Background calibration: Every month
Perform an instrument self test: Every month
Pure dye calibrations: Every 6 months
RNaseP instrument verification: After the instrument has been moved, or as needed to verify instrument performance
Can I use my 7900HT System RNase P instrument verification plates with both the 7900HT System and the QuantStudio 6 and 7 Flex Real-Time PCR Systems?
Only the TaqMan RNase P Instrument Verification Plate, Fast 96-well (for 0.1 mL block) (Cat. No. 4351979) is compatible with both the 7900HT System and the QuantStudio 6 and 7 Flex Real-Time PCR Systems. The following RNase P instrument verification plates can be used on the 7900HT System, but not on the QuantStudio 6 and 7 Flex Real-Time PCR Systems:
• TaqMan RNase P Instrument Verification Plate, 96-well (Cat. No. 4310982) (for 96-well standard block)
• TaqMan RNase P Instrument Verification Plate for 7900HT System, 384-well (Cat. No. 4323306) (for 384-well block)
• 7900HT TaqMan Low Density Array Instrument Verification RNase P Kit (Cat. No. 4351468) (for TaqMan Array Card block)
The following RNase P instrument verification plates can be used on the QuantStudio 6 and 7 Flex Real-Time PCR Systems, but not on the 7900HT System:
• TaqMan RNase P Instrument Verification Plate for ViiA7 System, 96-well (Cat. No. 4432382) (for 96-well standard block)
• TaqMan RNase P Instrument Verification Plate for ViiA7 System, 384-well (Cat. No. 4455280) (for 384-well block)
• ViiA7 Array Card RNase P Verification Kit (Cat. No. 4432265) (for TaqMan Array Card block)
What is the difference between a Fast 96-well and standard 96-well block?
There are two main differences between a Fast 96-well and standard 96-well block for the Applied Biosystems real-time PCR systems:
• Reaction well volume: The Fast 96-well blocks have 0.1 mL reaction wells and are compatible with 0.1 mL plates and tubes. The standard 96-well blocks have 0.2 mL reaction wells and are compatible with 0.2 mL plates and tubes.
• Ramp rate: The fast 96-well blocks have a higher maximum ramp rate than standard blocks for the same real-time PCR system. However, both the fast 96-well and standard 96-well blocks are capable of running standard and fast chemistry.
Note: For newer instruments (QuantStudio 3 and 5 Real-Time PCR Systems and later) "0.1 mL" and "0.2 mL" are being used in place of "fast" and "standard" as designations for the two 96-well block formats.