JUN™ Dye Spectral Calibration Plate for Multiplex qPCR, Fast 96-well
Applied Biosystems™
JUN™ Dye Spectral Calibration Plate for Multiplex qPCR, Fast 96-well
This plate is used for JUN dye spectral calibration of your 7500 Fast, ViiA 7, QuantStudio 6 Flex, QuantStudio 7Weitere Informationen
Have Questions?
Katalognummer
Menge
A24735
1 plate
Katalognummer A24735
Preis (EUR)
215,00
Each
Zum Warenkorb hinzufügen
Menge:
1 plate
Preis (EUR)
215,00
Each
Zum Warenkorb hinzufügen
This plate is used for JUN dye spectral calibration of your 7500 Fast, ViiA 7, QuantStudio 6 Flex, QuantStudio 7 Flex, and QuantStudio 12K Flex real-time PCR systems with a Fast 96-well block. Calibration and verification should be run at least every six months. Please see your instrument maintenance guide for full instructions on use. The dye formulations will improve results with multiplexing by more accurately representing fluorescent spectra found in your real-time PCR experiments.
Nur für Forschungszwecke.Nicht zur Verwendung bei diagnostischen Verfahren.
Specifications
Kalibrierter FarbstoffJUN™, JUN™
KalibrierungsartSpektralkalibrierung
Zur Verwendung mit (Anwendung)Real Time PCR (qPCR)
Zur Verwendung mit (Geräte)7500 Fast System, QuantStudio™, Viia™ 7 System
Menge1 plate
ProdukttypPlatte zur Spektrenkalibrierung mit Farbstoff
Unit SizeEach
Inhalt und Lagerung
Bei -15 °C ± 25 °C lagern
Häufig gestellte Fragen (FAQ)
What are the excitation and emission profiles for ABY, JUN, Cyanine 5, and Cyanine 5.5 dyes?
ABY dye has an excitation wavelength of 568 nm and emission wavelength of 583 nm.
JUN dye has an excitation wavelength of 606 nm and emission wavelength of 618 nm.
Cynanine 5 dye has an excitation wavelength of 649 nm and emission wavelength of 667 nm.
Cyanine 5.5 dye has an excitation wavelength of 678 nm and emission wavelength of 695 nm.
Please refer to your instrument user guide for additional information on compatibility with different dye channels.