TSQ Quantis™ Triple-Quadrupol-Massenspektrometer
TSQ Quantis™ Triple-Quadrupol-Massenspektrometer
TSQ Quantis™ Triple-Quadrupol-Massenspektrometer
TSQ Quantis™ Triple-Quadrupol-Massenspektrometer
TSQ Quantis™ Triple-Quadrupol-Massenspektrometer
Thermo Scientific™

TSQ Quantis™ Triple-Quadrupol-Massenspektrometer

Robuste Targeted-Quantifizierungs-Workflows mit Best-in-Class-Empfindlichkeit für alle Analyten und für Experts und Non-Experts.

Have Questions?
KatalognummerTyp
TSQ02-10001Triple-Quadrupol-LC-MS
Katalognummer TSQ02-10001
Preis (EUR)
-
Typ:
Triple-Quadrupol-LC-MS

Sichere und effiziente Beantwortung kritischer analytischer Herausforderungen bei Targeted-Quantifizierungs-Workflows in Ihrem Labor. Mit seinen segmentierten Quadrupolen, der schnelleren Elektronik und dem verbesserten Detektionssystem bietet das Thermo Scientific™ TSQ Quantis™ Triple-Quadrupol-Massenspektrometer beispiellose Robustheit mit marktführender Empfindlichkeit für den täglichen Gebrauch und für jeden Anwender.

Das TSQ Quantis MS lässt sich nahtlos mit branchenführender Quantifizierungssoftware, HPLCs und UHPLCs sowie Probenvorbereitungslösungen kombinieren. Es bietet zudem verbesserte Effizienz bei Targeted-Quantifizierungs-Workflows für eine Vielzahl von Analyten in komplexen Matrices. Weitere Vorteile sind u. a.:

  • Überragende Empfindlichkeit, aus der überzeugende Daten für alle molekularen Spezies selbst in komplexen Matrices resultieren.
  • Hervorragende/hohe Robustheit, die erhöhte Produktivität ermöglicht.
  • Zuverlässigkeit und Reproduzierbarkeit verbessern die Qualität der Analysen – Lauf für Lauf, Probe für Probe.
  • Die Anbindung an anwendungsspezifische Software sorgt für mehr Produktivität in allen Anwendungsgebieten.
  • Simplizität und Anwenderfreundlichkeit ermöglichen Bedienern mit unterschiedlicher Erfahrung das Erfassen hochwertiger Ergebnisse bei erhöhter Ergebniszuverlässigkeit.
Specifications
Zur Verwendung mit (Anwendung)Pharmazeutisch, Umwelt, Lebensmittelsicherheit, klinische Forschung und forensische Toxikologie
Massenbereich5 bis 3000 m/z
TypTriple-Quadrupol-LC-MS
Unit SizeEach

Häufig gestellte Fragen (FAQ)

How do I clean/store my Triple Quadrupole mass spectrometry instruments after analysis?

For short periods (less than 24 hr) the system can just be put into standby. However, it is good practice to flush the system with 50:50 solvent/water (where the solvent is either methanol or acetonitrile) to remove buffers or additives and stop microbial growth.

For longer periods, flush the LC system with 50:50 solvent/water. Then flush with 100% solvent. Put the system into standby.

Find additional tips, troubleshooting help, and resources within our Mass Spectrometry Support Center.

How do I perform mass calibration on Triple Quadrupole mass spectrometry instruments and how often should I do it?

The mass calibration should ideally be performed every 3 months and definitely performed at least every 6 months. Refer to the Instrument Manual (https://tools.thermofisher.com/content/sfs/manuals/TSQ-Series-Hardware.pdf) for detailed instructions on how to perform mass calibration.

Find additional tips, troubleshooting help, and resources within our Mass Spectrometry Support Center.

How should I clean the Ion Transfer Tube on Triple Quadrupole mass spectrometry instruments?

The Ion Transfer Tube should be removed and sonicated in a 50:50 solution of methanol/water that contains 20% formic acid for 30 min. Rinse thoroughly with water. Then sonicate in methanol for 15 min. Dry with nitrogen gas.

The same procedure can be applied to the Ion Sweep Cone.

Refer to the Instrument Manual (https://tools.thermofisher.com/content/sfs/manuals/TSQ-Series-Hardware.pdf) for more detailed instructions.

Find additional tips, troubleshooting help, and resources within our Mass Spectrometry Support Center.

How should I adjust my ion source position on Triple Quadrupole mass spectrometry instruments?

Perform loop injections of your analyte using isocractic elution through a short column at the flow rate and approximate mobile phase composition. Use a concentration where the signal to noise is approximately 10:1. Record the raw data file and integrate the peaks, comparing peak areas at different source positions (i.e., left and right, up and down, fore and aft). Ensure that the background noise is not increasing more than the peak area.

Note: The needle position on the Thermo Fisher Scientific TSQ Altis MS, TSQ Quantis MS, and TSQ Fortis MS instruments is set and should not be adjusted.

Find additional tips, troubleshooting help, and resources within our Mass Spectrometry Support Center.

Do I need to warm/equilibrate Triple Quadrupole mass spectrometry instruments prior to starting a session of analysis?

It is advisable to turn both the LC system and the MS system on for at least 30 min prior to analysis. This time can usually be spent testing one or more solvents or matrix blanks.

Find additional tips, troubleshooting help, and resources within our Mass Spectrometry Support Center.